PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, PICO Fluid Management Pack
K202157 · Smith & Nephew Medical Limited · OMP · Oct 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K202157
Device Name
PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, PICO Fluid Management Pack
Applicant
Smith & Nephew Medical Limited
Product Code
OMP · General, Plastic Surgery
Decision Date
Oct 30, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K202157 · Oct 30, 2020
PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, PICO Fluid Management Pack
Smith & Nephew Medical Limited
MAUDE database; Manufacturer complaint data
The sponsor conducted a systematic review of postmarket surveillance data (MAUDE database and internal complaint records) to evaluate the risk of magnetic interference with implantable medical devices.
Postmarket surveillance; MAUDE database; Complaint data; Magnetic interference
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Postmarket surveillance review of magnetic interference; Retrospective analysis of postmarket surveillance records
Users of PICO devices and patients with implantable medical devices (IMDs); Sample Size: Over 60,000 records
Not applicable for this study
Incidence of magnetic interference with implantable medical devices
Indications for Use
PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, and the PICO Fluid Management Pack are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include: Chronic Acute Traumatic Subacute and dehisced wounds Partial-thickness burns Ulcers (such as diabetic or pressure) Flaps and grafts Closed surgical incisions PICO 7, PICO 7Y, PICO 14 Single Use Negative Pressure Wound Therapy Systems and PICO Fluid Management Pack is suitable for use both in a hospital and homecare setting.
Device Story
Canister-free, single-use negative pressure wound therapy (NPWT) system; consists of absorbent dressing connected to small NPWT pump via tubing/port. Pump utilizes custom voice-coil mechanism controlled by microprocessor to maintain -80mmHg nominal pressure under dressing. Dressing manages exudate through absorption and evaporation. Used in hospital and homecare settings by clinicians or patients. Output is negative pressure application to wound site; healthcare providers monitor wound healing progress. Benefits include promotion of wound healing through exudate removal. Subject devices are updates to existing PICO family, specifically modifying magnet safety warnings and removing pump warning labels.
Clinical Evidence
Bench testing only. Evidence includes magnetic field strength testing demonstrating field strength is weaker than common consumer electronics (laptops, phones). Systematic literature review of implantable medical devices (IMDs) and magnetic interference, combined with analysis of over 60,000 MAUDE database records and internal complaint data, confirmed no reported instances of death or serious harm related to magnetic interference from PICO devices. Previous biocompatibility (ISO 10993), electrical safety (IEC 60601-1 series), and wound model performance data from cleared predicates remain applicable.
Technological Characteristics
Canister-free NPWT system; PU plastic pump casing; custom voice-coil pump; microprocessor-controlled; -80mmHg nominal pressure; ethylene oxide sterilization. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 10993.
Indications for Use
Indicated for patients requiring negative pressure wound therapy to promote healing via removal of low to moderate exudate and infectious materials. Applicable to chronic, acute, traumatic, subacute, dehisced wounds, partial-thickness burns, ulcers (diabetic/pressure), flaps, grafts, and closed surgical incisions.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
PICO 7 Single Use Negative Pressure Wound Therapy System (K180698)
Reference Devices
PICO 7Y Single Use Negative Pressure Wound Therapy System (K182323)
PICO 14 Single Use Negative Pressure Wound Therapy System (K191760)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
October 30, 2020
Smith & Nephew Medical Limited Steeve Lamvohee Regulatory Affairs Director 101 Hessle Road Hull, Yorkshire HU3 2BN United Kingdom
Re: K202157
Trade/Device Name: PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, PICO Fluid Management Pack Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II
Product Code: OMP Dated: July 31, 2020 Received: August 3, 2020
Dear Steeve Lamvohee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Kimberly M. Ferlin, Ph.D. Assistant Director (Acting) DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known) K202157
Device Name
PICO 7 Single Use Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
PICO 7 Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- · Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcers (such as diabetic or pressure)
- · Flaps and grafts
- · Closed surgical incisions
PICO 7 Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a hospital and homecare setting.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{3}------------------------------------------------
510(k) Number (if known) K202157
Device Name
PICO 7Y Single Use Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
PICO 7Y Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (Negative Pressure Wound Therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- · Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcers (such as diabetic or pressure)
- · Flaps and grafts
- · Closed surgical incisions
PICO 7Y Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a homecare setting.
| Type of Use (Select one or both, as applicable) |
|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{4}------------------------------------------------
510(k) Number (if known) K202157
#### Device Name
PICO 14 Single Use Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
PICO 14 Single Use Negative Pressure Wound Therapy System is indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- · Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcers (such as diabetic or pressure)
- · Flaps and grafts
- · Closed surgical incisions
PICO 14 Single Use Negative Pressure Wound Therapy Systems are suitable for use both in a homecare setting.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| <div> <span> <b> </b> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{5}------------------------------------------------
510(k) Number (if known)
#### K202157
Device Name PICO Fluid Management Pack
#### Indications for Use (Describe)
PICO single use negative pressure systems are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as they may promote wound healing via removal of low to moderate levels of exudate and infectious materials. Appropriate wound types include:
- Chronic
- Acute
- · Traumatic
- Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcers (such as diabetic or pressure)
- · Flaps and grafts
- · Closed surgical incisions
PICO single use negative pressure systems are suitable for use both in a hospital and homecare setting.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------|--|
| <span> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
{6}------------------------------------------------
PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System, PICO 14 Single Use Negative Pressure Wound Therapy System, and the PICO Fluid Management Pack.
| I. Submitter | |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Owner<br>Name/Address: | Smith &Nephew Medical Limited<br>101 Hessle Road<br>Hull<br>HU3 2BN<br>United Kingdom |
| Establishment<br>Registration<br>Number: | 8043484 |
| Contact Person: | Dr Steeve Lamvohee, Regulatory Affairs Director |
| Phone Number: | +44 7583 048727 |
| Date Prepared: | 08 Sep 2020 |
| II. Device | |
| Trade Names: | PICO 7 Single Use Negative Pressure Wound Therapy System<br>PICO 7Y Single Use Negative Pressure Wound Therapy System<br>PICO 14 Single Use Negative Pressure Wound Therapy System<br>PICO Fluid Management Pack |
| Common or Usual<br>Name: | Negative Pressure Wound Therapy powered suction pump |
| Classification Name: | Powered suction pump (21 CFR 878.4780) |
| Regulatory Class: | Class II |
| Product Code: | OMP |
| 510(k) Number: | K202157 |
| III. Predicate Device | |
| 510(k) Number: | K180698 |
| Device name: | PICO 7 Single Use Negative Pressure Wound Therapy System |
| Clearance Date: | 21 August 2018 |
| Recall Information: | The predicate has not been the subject of any recall |
{7}------------------------------------------------
Use Negative Pressure Wound Therapy Systems
Traditional 510(k) Premarket Notification K202157
# IV. Device Description
All of the PICO devices (PICO 7 (K180698), PICO 7Y (K182323), and PICO 14 (K191760) are canister-free single-use Negative Pressure Wound Therapy (NPWT) Systems and use an absorbent dressing connected to a small NPWT pump by a tubing and port. The dressing manages wound exudate by a combination of absorption and evaporation. The pump provides for a -80mmHg nominal pressure under the dressing, applying Negative Pressure Wound Therapy (NPWT) to the wound. PICO Fluid Management Pack are packs of 5 individually packaged PICO Dressings designed for use with PICO devices.
# V. Indications for Use
PICO 7 Single Use Negative Pressure Wound Therapy System, PICO 7Y Single Use Negative Pressure Wound Therapy System. PICO 14 Single Use Negative Pressure Wound Therapy System. and the PICO Fluid Management Pack are indicated for patients who would benefit from a suction device (negative pressure wound therapy) as it may promote wound healing via removal of low to moderate levels of exudate and infectious materials.
Appropriate wound types include:
- Chronic
- · Acute
- · Traumatic
- Subacute and dehisced wounds
- · Partial-thickness burns
- · Ulcers (such as diabetic or pressure)
- · Flaps and grafts
- Closed surgical incisions
PICO 7, PICO 7 Y and PICO 14 Single Use Negative Pressure Wound Therapy Systems are suitable for use in both a hospital and homecare setting.
# VI. Comparison of Technological Characteristics with the Predicate Device
The technological principle for delivering the negative wound therapy for the subject devices and predicate device are identical. The differences between the subject devices and the predicate device are:
- · Updated text for Magnet Warning
- · Removal of Pump warning label
{8}------------------------------------------------
# Smith Nephew
Traditional 510(k) Premarket Notification K202157
| Items | Subject Devices:<br>PICO 7 Single Use Negative<br>Pressure Wound Therapy<br>System, PICO 7Y Single Use<br>Negative Pressure Wound<br>Therapy System, PICO 14 Single<br>Use Negative Pressure Wound<br>Therapy System, PICO Fluid<br>Management Pack | Predicate Device:<br>PICO 7 Single Use<br>Negative Pressure<br>Wound Therapy System | Comparison |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications For<br>Use | PICO 7 Single Use Negative<br>Pressure Wound Therapy<br>System, PICO 7Y Single Use<br>Negative Pressure Wound<br>Therapy System, PICO 14 Single<br>Use Negative Pressure Wound<br>Therapy System, and PICO Fluid<br>Management Pack are<br>indicated for patients who would<br>benefit from a suction device<br>(NPWT) as it may promote<br>wound healing via removal of<br>low to moderate levels of<br>exudate and infectious materials.<br>Appropriate wound types<br>include:<br>Chronic<br>Acute<br>Traumatic<br>Subacute and dehisced wounds<br>Partial-thickness burns<br>Ulcers (such as diabetic or<br>pressure)<br>Flaps and grafts<br>Closed surgical incisions<br>PICO 7, PICO 7Y, PICO 14<br>Single Use Negative Pressure<br>Wound Therapy Systems and<br>PICO Fluid Management Pack is<br>suitable for use both in a hospital<br>and homecare setting. | PICO 7 Single Use<br>Negative Pressure<br>Wound Therapy System<br>is indicated for patients<br>who would benefit from<br>a suction device<br>(NPWT) as it may<br>promote wound healing<br>via removal of low to<br>moderate levels of<br>exudate and infectious<br>materials. Appropriate<br>wound types include:<br>Chronic<br>Acute<br>Traumatic<br>Subacute and dehisced<br>wounds<br>Partial-thickness burns<br>• Ulcers (such as diabetic<br>or pressure)<br>Flaps and grafts<br>Closed surgical incisions<br>PICO 7 Single Use<br>Negative Pressure<br>Wound Therapy System<br>is suitable for use both in<br>a hospital and homecare<br>setting. | Same |
| Environment of<br>use | Hospital, home | Hospital, home | Same |
| Items | Subject Devices:<br>PICO 7 Single Use Negative<br>Pressure Wound Therapy<br>System, PICO 7Y Single Use<br>Negative Pressure Wound<br>Therapy System, PICO 14 Single<br>Use Negative Pressure Wound<br>Therapy System, PICO Fluid<br>Management Pack | Predicate Device:<br>PICO 7 Single Use<br>Negative Pressure<br>Wound Therapy System | Comparison |
| Patient<br>Population | Patients who would benefit from<br>a suction device (negative<br>pressure wound therapy) | Patients who would<br>benefit from a suction<br>device (negative<br>pressure wound therapy) | Same |
| Materials | PU Plastic pump casing | PU Plastic pump casing | Same |
| Single-use or<br>Reusable | Single use | Single use | Same |
| Method of<br>Sterilization | Pump, dressing and fixation<br>strips sterilized by ethylene oxide | Pump, dressing and<br>fixation strips sterilized<br>by ethylene oxide | Same |
| Biocompatibility | Dressing complies with ISO<br>10993 | Dressing complies with<br>ISO 10993 | Same |
| Type of Pump | Custom designed "voice-coil"<br>pump controlled by<br>microprocessor | Custom designed<br>"voice-coil" pump<br>controlled by<br>microprocessor | Same |
| Electrical Safety<br>Testing | Complies with IEC 60601-1, IEC<br>60601-1-2, IEC 60601-1-11 | Complies with IEC<br>60601-1, IEC 60601-1-<br>2, IEC 60601-1-11 | Same |
| Instructions for<br>Use | Addition of the word medical<br>• If you have an electronic<br>medical device and are helping<br>take care of somebody else using<br>the PICO 7 system.<br>As requested by the FDA,<br>addition of "MR Unsafe. You<br>must remove the PICO 7 pump<br>from the dressing before entering<br>the MRI suite. Do not bring<br>PICO 7 into the MRI scan room.<br>The device presents a projectile<br>hazard." | • If you have an<br>electronic device and are<br>helping take care of<br>somebody else using the<br>PICO 7 system<br>N/A | Magnet warning<br>box in HCP and<br>Patient IFU<br>changed<br>Addition of<br>warning to HCP<br>and Patient IFU |
| | Texts amended in "Important<br>Information Section" and<br>"Glossary of Symbols" | The PICO [#] pumps<br>contain a MAGNET.<br>Keep the PICO [#] | Important<br>Information and<br>Glossary of |
| Items | Subject Devices:<br>PICO 7 Single Use Negative<br>Pressure Wound Therapy<br>System, PICO 7Y Single Use<br>Negative Pressure Wound<br>Therapy System, PICO 14 Single<br>Use Negative Pressure Wound<br>Therapy System, PICO Fluid<br>Management Pack | Predicate Device:<br>PICO 7 Single Use<br>Negative Pressure<br>Wound Therapy System | Comparison |
| | The PICO [#] pumps contain a<br>MAGNET. Keep the PICO [#]<br>pumps at least 4 inches (10 cm)<br>away from other medical devices<br>at all times. As with all electrical<br>medical equipment, failure to<br>maintain appropriate distance<br>may disrupt the operation of<br>nearby medical devices. | pumps at least 4 inches<br>(10 cm) away from other<br>medical devices at all<br>times. Failure to do so<br>can cause the other<br>medical device to fail<br>which can result in<br>serious harm including<br>death. | Symbols in HCP<br>and Patient IFU<br>changed.<br>Texts removed<br>"Failure to do so<br>can cause the<br>other medical<br>device to fail<br>which can result<br>in serious harm<br>including death."<br>Texts added "As<br>with all electrical<br>medical<br>equipment,<br>failure to<br>maintain<br>appropriate<br>distance may<br>disrupt the<br>operation of<br>nearby medical<br>devices." |
| Warning Label<br>on front of Pump | No warning label on the front of<br>the pump. | <br> | Warning Label<br>Removed |
| | | Image: PICO 7 pump | |
{9}------------------------------------------------
# Smith Nephew
Traditional 510(k) Premarket Notification K202157
{10}------------------------------------------------
Traditional 510(k) Premarket Notification K202157
# VII. Performance Data
Performance data provided in previously cleared 510(k)s for PICO 7, PICO 7Y and PICO 14 Single Use Negative Pressure Wound Therapy Systems (including PICO Fluid Management Pack) continue to support substantial equivalence. Additional data is provided to support the proposed change including testing of magnetic field, an analysis of postmarket surveillance, and a risk
{11}------------------------------------------------
Traditional 510(k) Premarket Notification K202157
analysis. In the cleared 510(k)s for PICO 7, PICO 7Y and PICO 14 Single Use Negative Pressure Wound Therapy Systems (including PICO Fluid Management Pack), the device met requirements for:
Biocompatibility:
- FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, "Biological ● Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995,
- International Standard ISO 10993-1 "Biological Evaluation of Medical Devices Part 1: ● Evaluation and Testing Within a Risk Management Process," as recognized by FDA.
Testing included the following:
- Cytotoxicity
- Sensitization
- . Irritation
- Systemic toxicity
- Pyrogen Testing
# Electrical Safety and Electromagnetic Compatibility:
- IEC 60601-1 ●
- IEC 60601-1-11
- IEC 60601-1-6 ●
- IEC 60601-1-2
# Wound model testing:
- Wound exudate ●
- Wound size
- NPWT Wound flow rate ●
- Testing duration
- Pressure sampling rates ●
## Magnetic Field Testing:
Bench testing demonstrated that the average magnetic field strength of PICO Family pump is weaker than multiple everyday devices (e.g. laptops, phones, tablets, and headphones). These products are abundant in today's society and do not pose an unacceptable risk to patients with implantable medical devices (IMDs). The PICO pump does not significantly change the magnetic environment an IMD user may encounter.
## Post Market and Risk Analysis:
A systematic literature review was conducted of IMDs and magnetic interference and this demonstrated that although the risk of magnetic interference with implantable medical devices exists, the risk is not as high as had been originally assessed. Magnetic fields are present in the everyday environment which are comparable with those presented by the PICO devices, however, there are no risks that have been reported in the clinical literature that relate to actual death or serious harm. A comprehensive review of over 60,000 records from the MAUDE database as well as the Smith +Nephew PICO devices complaint data confirmed the literature review outcome.
# VIII. Conclusion
Data to confirm substantial equivalence to the predicate were provided in the FDA cleared 510(k) for PICO Family of Single Use Negative Pressure Wound Therapy Systems. The proposed labeling changes are supported by updated IFUs, evaluation of risk and updated risk analysis, and magnetic strength testing. The subject devices are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.