Storz Medical Duolith SD1 T-Top & Tower System with C-Actor Sepia Handpiece
K202112 · Storz Medical AG · PZL · Feb 21, 2021 · General, Plastic Surgery
Device Facts
Record ID
K202112
Device Name
Storz Medical Duolith SD1 T-Top & Tower System with C-Actor Sepia Handpiece
Applicant
Storz Medical AG
Product Code
PZL · General, Plastic Surgery
Decision Date
Feb 21, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4685
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K202112 · Feb 21, 2021
Storz Medical Duolith SD1 T-Top & Tower System with C-Actor Sepia Handpiece
Storz Medical AG
Published clinical literature
A review of five independent studies involving 89 patients with chronic soft tissue wounds (including diabetes mellitus) was used to demonstrate clinical effectiveness in wound healing and reduction of wound area.
Literature review; Chronic wounds; Diabetic foot ulcers; Clinical effectiveness
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Review of five independent studies; Retrospective/Observational clinical literature review; Follow-up/Duration: 13-20 weeks
89 patients with chronic soft tissue wounds, including diabetes mellitus; Sample Size: 89
Not applicable for this study
Complete wound healing (54-70% at 13-20 weeks); reduction of wound area (35-80% at 7-20 weeks)
Indications for Use
The DUOLITH SD1 C-ACTOR Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the exidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Device Story
Device uses electromagnetic shock wave generation for chronic wound treatment. Control unit discharges high-voltage capacitor into handpiece coil; transient magnetic field induces eddy currents in metallic membrane; membrane repulsion creates pressure wave in water; parabolic reflector focuses wave onto target tissue. Operated by clinicians in clinical settings. System includes microprocessor control of parameters (intensity, frequency, pulse count). Output is focused acoustic pressure shockwaves. Clinicians use device to promote wound healing in conjunction with standard care. Benefits include reduction of wound area and support for complete healing in chronic diabetic ulcers.
Clinical Evidence
Clinical effectiveness supported by literature review of 89 chronic soft tissue wounds (including diabetes mellitus). Five independent studies reported 54-70% complete wound healing at 13-20 weeks and 35-80% reduction in wound area at 7-20 weeks. Nonclinical testing confirmed design specifications, biocompatibility (ISO 10993-1), and electrical safety (IEC 60601 series).
Technological Characteristics
Electromagnetic shock wave generation; high-voltage capacitor/coil system; parabolic reflector. Dimensions: 454 x 187 x 460 mm. Energy flux density: 0.03 - 1.24 mJ/mm2. Pulse repeat rate: 1-8 Hz. Microprocessor-controlled. Biocompatibility per ISO 10993-1. Electrical safety per IEC 60601-1, IEC 60601-2-36. EMC per IEC 60601-1-2 and AIM 7351731.
Indications for Use
Indicated for adult patients (22+ years) with chronic, full-thickness diabetic foot ulcers (>30 days duration) measuring ≤16 cm2, extending through epidermis, dermis, tendon, or capsule, without bone exposure. Used with standard diabetic ulcer care.
Regulatory Classification
Identification
An extracorporeal shock wave device for treatment of chronic wounds is a prescription device that focuses acoustic shock waves onto the dermal tissue. The shock waves are generated inside the device and transferred to the body using an acoustic interface.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Non-clinical performance testing must be conducted to demonstrate that the system produces anticipated and reproducible acoustic pressure shock waves.
(2) The patient-contacting components of the device must be demonstrated to be biocompatible.
(3) Performance data must demonstrate that the reusable components of the device can be reprocessed for subsequent use.
(4) Performance data must be provided to demonstrate the electromagnetic compatibility and electrical safety of the device.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Performance data must support the use life of the system by demonstrating continued system functionality over the labeled use life.
(7) Physician labeling must include:
(i) Information on how the device operates and the typical course of treatment;
(ii) A detailed summary of the device's technical parameters;
(iii) Validated methods and instructions for reprocessing of any reusable components; and
(iv) Instructions for preventing hearing loss by use of hearing protection.
(8) Patient labeling must include:
(i) Relevant contraindications, warnings, precautions, adverse effects, and complications;
(ii) Information on how the device operates and the typical course of treatment;
(iii) The probable risks and benefits associated with the use of the device;
(iv) Post-procedure care instructions; and
(v) Alternative treatments.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA acronym is in a blue square, followed by the words "U.S. Food & Drug Administration" in blue text.
February 21, 2021
Storz Medical AG % Michael Dayton Principal Consultant Biomed Research. Inc. 3959 Van Dyke Road, Suite 245 Lutz, Florida 33558
Re: K202112
Trade/Device Name: Storz Medical DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece Regulation Number: 21 CFR 878.4685 Regulation Name: Extracorporeal Shock Wave Device for Treatment of Chronic Wounds Regulatory Class: Class II Product Code: PZL Dated: January 20, 2021 Received: January 22, 2021
Dear Michael Dayton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K202112
Device Name
DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece
Indications for Use (Describe)
The DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm2, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 14px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 14px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### 1. SPONSOR
Storz Medical AG Lohstampfestr. 8 CH-8274 Tägerwilen Switzerland
| Contact Person: | Pavel Novak, Ph.D. |
|-----------------|---------------------|
| Telephone: | 41 (0) 71 677 45 13 |
| FAX: | 41 (0) 71 677 45 05 |
Date Prepared: February 19, 2021
#### 2. DEVICE NAME
| Proprietary Name: | DUOLITH SD1 T-Top & Tower System with C-ACTOR Sepia Handpiece |
|-----------------------|------------------------------------------------------------------|
| Regulation Name: | Extracorporeal Shock Wave Device for Treatment of Chronic Wounds |
| Classification Panel: | General and Plastic Surgery Devices |
| Regulatory Number: | 21 CFR 878.4685 |
| Product Code: | PZL |
| Device Class: | II |
#### 3. PREDICATE DEVICES
Equivalence is claimed to the predicate devices: DERMAPACE device (DEN160037; primary) and the ORTHOGOLD device (K191961).
#### 4. DEVICE DESCRIPTION
The DUOLITH SD1 with the C-ACTOR Sepia Handpiece uses an electromagnetically generated shock wave produced within the hand-held applicator. The shock wave is generated by discharging a high voltage capacitor located in the Control Unit into a cylindrically shaped coil system in the C-ACTOR Handpiece which is surrounded by a cylindrical metallic membrane. The transient magnetic field produced by the coil induces eddy currents in the metal membrane, causing it to repel from the coil, producing a pressure wave. The membrane is immersed in water and the pressure wave produced by the membrane propagates through the water to a concentric parabolic reflector, where it is reflected to a focal point outside of the Handpiece in front of the reflector. The DUOLITH SD1 with C-ACTOR incorporates micro-processor control of the operating parameters.
{4}------------------------------------------------
#### ડાં INTENDED USE
The DUOLITH SD1 C-ACTOR Handpiece is indicated to provide acoustic pressure shockwaves in the treatment of chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the exidermis, dermis, tendon, or capsule, but without bone exposure. The DUOLITH SD1 with C-ACTOR is indicated for adult (22 years and older), diabetic patients presenting with diabetic foot ulcers greater than 30 days in duration and is indicated for use in conjunction with standard diabetic ulcer care.
#### 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE
The technological characteristics of the DUOLITH SD1 and the predicate devices are the same with the exception of acoustic wave generation. The predicate devices utilize a high voltage spark plug to initiate a shock wave, while the Duolith utilizes a high voltage electrical coil. Both methods generate a short, high amplitude pulse (shock wave) of about 5 micro-seconds duration and results in similar waveforms and output energies.
#### 7. PERFORMANCE TESTING
## Nonclinical Performance
Nonclinical verification and validation testing was performed and demonstrated that the DUOLITH SD1 with C-ACTOR meets the design specifications and is safe and effective for its intended use. All tests required by the verification and validation plan were completed and passed. The testing demonstrates that the DUOLITH SD1 with C-ACTOR is substantially equivalent to the predicate devices.
The DUOLITH SD1 software was validated and demonstrated to be of a Moderate level of concern; while hazard analysis / risk management was performed and demonstrated that all risks are mitigated to an acceptable level. There are no direct body contacting components since the treatment handpiece is separated from direct patient contact by a sterile barrier during treatment. However the treatment head component was nevertheless tested for biocompatibility (contact duration <24 hours) and found to conform to elements of ISO 10993-1:2009. The DUOLITH SD1 was tested and demonstrated to conform to the general safety requirements of IEC 60601-1:2012; IEC 60601-6:2013; and electromagnetic compatibility requirements of IEC 60601-1-2:2014; and the RFID electromagnetic immunity requirements of AMI 7351731.
No performance standards applicable to this device have been adopted under Section 514 of the Act. The DUOLITH SD1 System with C-ACTOR complies with the applicable requirements of the following international consensus standards:
- · ISO 14971: 2007: Medical devices: Application of risk management to medical devices.
{5}------------------------------------------------
- · ISO 10993-1:2009 (4th Ed.): Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process.
- · ISO 61846:1998 Ultrasonics Pressure pulse lithotripters Characteristics of fields.
- · IEC 60601-1:2012, (3rd Ed.): Medical electrical equipment Part 1: General requirements for basic safety and essential performance.
- · IEC 60601-2-36: 2014 (2nd Ed.): Medical electrical equipment Part 2-36: Particular requirements for the basic safety and essential performance of equipment for extracorporeally induced lithotripsy.
- · IEC 60601-1-2: 2014 (4th Ed.): Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- · IEC 60601-1-6: 2013 (3.1 Ed.): Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- · AIM 7351731:2017 (Rev. 2.0): Medical Electrical Equipment and System Electromagnetic Immunity Test for Exposure to Radio Frequency Identification Readers.
- · IEC 62304:2015 (1.1 Ed.) Medical Device Software Software Life Cycle Process.
- · ISO 13485:2016: Medical devices Quality management systems Requirements for regulatory purposes.
## Clinical Performance
A review of published literature with the subject device demonstrated successful clinical use with the Duolith SD1 technology in both men and women with 89 chronic soft tissue wounds having a range of etiologies including diabetes mellitus. The reports from five independent studies show significant results in complete wound healing (54-70% at 13-20 weeks) and reduction of wound area (35-80% at 7-20 weeks). The data from these studies establish the effectiveness of the Duolith SD1 to treat chronic, full-thickness diabetic foot ulcers with wound areas measuring no larger than 16 cm, which extend through the epidermis, dermis, tendon, or capsule, but without bone exposure.
The conclusions drawn from the nonclinical and clinical tests demonstrate that the Duolith SD1 is as safe, as effective, and performs as well as the legally marketed predicate devices.
#### 8. SUBSTANTIAL EQUIVALENCE COMPARISON
| Product Comparison | Predicate Device<br>dermaPACE | Predicate Device<br>OrthoGold 100 | Subject Device<br>DUOLITH SD1 w/<br>C-ACTOR |
|------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 510(k) Number | DEN160037 | K191961 | K202112 |
| Product Class | II | - Same - | - Same - |
| Classification Panel | General and Plastic Surgery Devices | - Same - | - Same - |
| Product Code and<br>Regulation | 21 CFR 878.4685<br>PZL - Extracorporeal Shock Wave Device for<br>Treatment of Chronic Wounds | Same | - Same - |
| Indications for Use | To provide acoustic pressure shockwaves in the<br>treatment of chronic, full-thickness diabetic foot<br>ulcers with wound areas measuring no larger<br>than 16 cm2, which extend through the<br> | - Same - | - Same - |
| Modes of action | Focused pressure pulses | Unfocused pressure<br>pulses | Focused pressure<br>pulses |
| Mechanism of action | Extracorporeally induced pressure pulses | - Same - | - Same - |
| Operating mode | Continuous | - Same - | - Same - |
| Electrical Protection | Class I, B | - Same - | - Same - |
| User Interface | LCD with function keys | Touch Screen | Touch Screen |
| Firmware Controlled | Yes | Yes | Yes |
| Type of Energy | Pressure pulses | - Same - | - Same - |
| Energy Source | 100 - 240 V AC, 50-60 Hz | - Same - | - Same - |
| Type of Acoustic<br>Wave Generation | Electrohydraulic, pressure wave under water<br>caused by discharge of high voltage<br>condensers | Electrohydraulic,<br>pressure wave under<br>water caused by<br>discharge of high<br>voltage condensers | Electromagnetic,<br>pressure wave under<br>water caused by<br>discharge of high<br>voltage condensers |
| Selection of parameters<br>(Intensity, Frequency,<br>Number of Pulses) | Yes | Yes | Yes |
| Size of treatment<br>applicator heads | ø70 x 160 mm | ø70 x 230 mm | ø62 x 179 mm |
| Intensity Settings<br>Range (Min-max:<br>Energy Flux Density) | 1 to 6<br>(0.23 - 0.33mJ/mm2) | 1 to 16<br>(0.01 - 0.19mJ/mm2) | 0.03 - 1.24mJ/mm2 |
| Focus depth | 1.5 - 10.5mm | 0 - 38mm<br>"unfocused" | 0 - 30mm |
| Pulse repeat rate (1/s) | 1 - 4 Hz | 1 - 8 Hz | 1 - 8 Hz |
| Number of pulses<br>(min and max) | 31.25 - 500 | 100 - 1000 | 100 - 500 |
| System Dimensions<br>(W x H x D) | 473 x 219 x 427 mm | 400 x 218 x 459 mm | 454 x 187 x 460 mm |
| Operating<br>Temperature | N / A | 10° - 30°C | 10° - 30°C |
| Operating Relative<br>Humidity | N / A | 3 - 85% | 5 - 55% |
| Treatment Duration<br>(Typical) | 5 — 10 min | 10 - 20 min | 10 - 20 min |
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.