The device is intended for patients suffering from eye disease which are diagnosed to have a condition or conditions which may benefit from ophthalmic cryotherapy.
Device Story
Disposable, handheld ophthalmic cryotherapy instrument. Operates via Joule-Thomson principle; pressurized N2O cryogen contained within internal canister circulates to stainless steel tip. Rapid gas expansion cools tip to create ice ball for tissue destruction or adhesion. Used by surgeons in clinical settings. Self-contained design eliminates need for external consoles, footswitches, or gas cylinders required by predicate systems. Surgeon activates cooling via button on device. Output is localized freezing of ocular tissue; facilitates surgical procedures like retinal reattachment or glaucoma management. Benefits include portability, ease of use, and elimination of reusable probe reprocessing.
Clinical Evidence
No clinical studies were performed. Substantial equivalence supported by non-clinical bench testing, including evaluation of activation mechanism, external tip temperature, ice ball thickness, freezing/defrost performance, dimensional analysis, sterility, shelf life, traction, biocompatibility, ventilation safety, and usability.
Technological Characteristics
Disposable handheld cryophthalmic unit. Materials: stainless steel tip. Energy source: internal pressurized N2O gas canister. Principle: Joule-Thomson effect for tip cooling. Form factor: self-contained, single-use device. No electrical power source required for activation. Sterilization: sterile barrier packaging.
Indications for Use
Indicated for ophthalmic surgery including retinal detachment cryopexy, refractory glaucoma cyclodestructive procedures, vitreous cavity fragment extraction, cataract extraction, trichiasis lash follicle cryodestruction, and retinopathy of prematurity (ROP) treatment.
Regulatory Classification
Identification
A cryophthalmic unit is a device that is a probe with a small tip that becomes extremely cold through the controlled use of a refrigerant or gas. The device may be AC-powered. The device is intended to remove cataracts by the formation of an adherent ice ball in the lens, to freeze the eye and adjunct parts for surgical removal of scars, and to freeze tumors.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
December 30, 2020
Vitreq B.V. % Debora Stapleton Regulatory Affairs Consultant Dynamic Strategies Inc. 25 Granite Street Medway, Massachusetts 02053
Re: K202038
Trade/Device Name: CryoTreQ Regulation Number: 21 CFR 886.4170 Regulation Name: Cryophthalmic unit Regulatory Class: Class II Product Code: HPS Dated: November 23, 2020 Received: November 24, 2020
Dear Ms. Stapleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for LT Charles Chiang Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K202038
Device Name CryoTreQ
#### Indications for Use (Describe)
The VitreQ CryoTreQ is indicated for use in ophthalmic surgery for retinal detachment, cyclodestructive procedures in refractory glaucoma, extraction of fragments within the vitreous cavity, cataract extraction, cryodestruction of lash follicles for trichiasis and the treatment of retinopathy of prematurity (ROP).
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) |
|----------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Submitter:
## Submitter's Information
Vitreq BV Seggelant-Noord 2 Vierpolders Zuid-Holland, NL 3237 MG
Contact Person: Debora Stapleton Telephone Number: 774-277-2320 Date Prepared: 12/30/2020
## Name of the Device
Trade or Proprietary Device Name: CryoTreQ Common Name: Cryophthalmic Unit Classification Name: Cryophthalmic Unit Classification Regulation: 21 CFR 886.4170 Product Code: HPS, Class II
## Legally Marketed Predicate
Primary Predicate Device Name: D.O.R.C. Cryo Unit, 510(k) Number K940373 Secondary Predicate Device Name: Keeler Cryomatic MKII Cryosurgical System, 510(k) Number K131787 Reference Predicate Device Name: PHAKOS Disposable Retinal Cryo Probe, 510(k) Number K162756
# Description of Subject Device
The VitreQ CryoTreQ is a disposable, handheld instrument intended for use in ophthalmic cryotherapy. The device tip is made of stainless steel. The device utilizes a pressurized cryogen which is circulated to the tip. The cryogenic material remains enclosed within the tip during application, serving to cool the metal tip during therapy. Rapid gas expansion in the tip causes freezing to the Joule-Thompson principle.
## Intended Use
The device is intended for patients suffering from eye disease which are diagnosed to have a condition or conditions which may benefit from ophthalmic cryotherapy.
## Indications for Use
The VitreQ CrvoTreQ is indicated for use in ophthalmic surgery such as cryopexy for retinal detachment, cyclodestructive procedures in refractory glaucoma, extraction of fragments within the vitreous cavity, cataract extraction, cryodestruction of lash follicles for trichiasis and the treatment of retinopathy of prematurity (ROP).
## Summary of Technological Characteristics
The CryoTreQ has materials, basic design, and cryogen type which are the same as for the predicate. The operational mechanism is the same; cryogenic material moves toward a closed metal tip and remains within the metal cavity during application, serving as a means to cool the tip.
{4}------------------------------------------------
## Comparison to Predicate - Technological Differences
The differences between the CryoTreQ and predicate cryotherapeutic devices are that the CryoTreQ is a fully disposable unit and the cryogenic function does not require connection to other devices. The predicate devices utilize connection of a handpiece (or probe), a footswitch (or footpedal) to an electronic console for activation, and a compressed gas cylinder for cryogenic function. The CryoTreQ is self contained. The cryogen-containing canister is supplied within the instrument for single procedure use.
| Characteristic | VitreQ<br>CryoTreQ | D.O.R.C. Cryo<br>Unit | Keeler<br>Cryomatic<br>MKII<br>Cryosurgical<br>System | PHAKOS<br>Disposable<br>Retinal Cryo<br>Probe |
|-----------------------------------------------|----------------------------------------|----------------------------------------|-------------------------------------------------------------------------------------------------------------|-----------------------------------------------|
| 510(k) Number<br>Product Code(s) | K202038 | K940373 | K131787 | K162756 |
| Intended Use | Ophthalmic<br>cryotherapy | Ophthalmic<br>cryotherapy | Ophthalmic<br>cryotherapy | Ophthalmic<br>cryotherapy |
| Operation<br>Mechanism | Cooled metal<br>contact | Cooled metal<br>contact | Cooled metal<br>contact | Cooled metal<br>contact |
| Non-clinical<br>Freeze and<br>Defrost Testing | Pre-specified<br>criteria were<br>met | Not evaluated | Comparisons to CryoTreQ – external temperature achieved, ice ball formation, defrost duration, and traction | |
| Activation<br>Method | Button<br>activation | Footswitch<br>activation | Footswitch<br>activation | Footswitch<br>activation |
| Electrical Source<br>for Activation | None | Battery | Mains | Mains |
| Cryogen Supply<br>Format | Internal<br>compressed<br>gas canister | External<br>compressed<br>gas cylinder | External<br>compressed<br>gas cylinder | External<br>compressed<br>gas cylinder |
| Cryogenic<br>Medium | N2O gas | CO2 or N2O<br>gas | CO2 or N2O<br>gas | CO2 or N2O<br>gas |
| Disposable or<br>Reusable | Disposable<br>unit | Reusable<br>probes | Disposable<br>and reusable<br>probes | Disposable<br>probe |
Table Comparison of Features of Vitre® CryoTreQ and Predicate Devices
## Summary of Clinical Testing
No clinical studies were performed.
## Summary of Non-Clinical Testing
Testing performed to evaluate the device included evaluations of the device activation mechanism, external temperature achieved, ice ball layer thickness formation, freezing events, defrost performance, dimensional analyses, device sterility, shelf life, evaluation of traction, bio-compatibility, ventilation safety, usability, transport stability, and sterile barrier integrity.
The body of testing performed on CryoTreQ devices demonstrated that they perform as well as the identified legally marketed devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.