The A8, A9 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A8, A9 is intended for use by licensed clinicians in the administration of general anesthesia, for patients requiring anesthesia within a health care facility, and can be used in adult, pediatric and neonate populations. High Flow Nasal Cannula (HFNC) is indicated for delivery of nasal high flow oxygen to spontaneously breathing adult patients. It can be used for pre-oxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms. It is not intended for apneic ventilation. HFNC is indicated for use in adults only.
Device Story
A8/A9 Anesthesia System: continuous flow inhalation gas anesthesia workstation; delivers anesthetic vapor; provides automatic/manual ventilation. Inputs: O2, CO2, N2O, anesthetic agents (Desflurane, Isoflurane, Halothane, Sevoflurane); patient physiological data via integrated Anesthetic Gas Module (AGM) and patient monitor. Operation: clinician-controlled via interface; electronic flowmeters; electronic vaporizer (A9); automatic ventilation modes (VCV, PCV, PCV-VG, SIMV-VC, SIMV-PC, SIMV-VG, PS, CPAP/PS, APRV). Outputs: anesthetic gas delivery, ventilation support, monitoring data (gas concentrations, airway pressure, flow, volume). Used in operating rooms by licensed clinicians. Benefits: precise anesthetic delivery, respiratory support, and integrated monitoring for patient safety during surgery.
Clinical Evidence
Bench testing only. No clinical data presented. Functional and system-level testing validated performance against specifications and consensus standards (ISO 10993, ISO 18562, ANSI/AAMI ES60601-1, IEC 60601-1-2, ISO 80601-2-13, ISO 80601-2-55, ASTM F1101-90).
Technological Characteristics
Anesthesia workstation; gas delivery via electronic flowmeters; electronic vaporizer (A9); sealed lead acid battery; integrated AGM; HFNC capability. Materials comply with ISO 10993 and ISO 18562. Connectivity: integrated with patient monitor (Passport 12M/17M). Standards: ISO 80601-2-13 (anesthetic workstation), ISO 80601-2-55 (gas monitors), IEC 60601-1-8 (alarms).
Indications for Use
Indicated for administration of general inhalation anesthesia and maintenance of ventilation in adult, pediatric, and neonate patients in healthcare facilities. HFNC indicated for delivery of nasal high flow oxygen to spontaneously breathing adult patients for pre-oxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation; not for apneic ventilation.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
March 26, 2021
Shenzhen Mindray Bio-Medical Electronics Co., LTD. Yanhong Bai Manager Regulatory Affairs Mindray Building, Keji 12th Road South Hi-tech Industrial Park, Nanshan Shenzhen, Guangdong 518057 China
Re: K201957
Trade/Device Name: A8, A9 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine For Anesthesia Or Analgesia Regulatory Class: Class II Product Code: BSZ, CCK, NHO, CBQ, NHP, CBS, CBR, CCL, KDP, NHQ Dated: February 24, 2021 Received: February 25, 2021
Dear Yanhong Bai:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201957
Device Name A8, A9 Anesthesia System
#### Indications for Use (Describe)
The A8, A9 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A8, A9 is intended for use by licensed clinicians in the administration of general anesthesia, for patients requiring anesthesia within a health care facility, and can be used in adult, pediatric and neonate populations.
High Flow Nasal Cannula (HFNC) is indicated for delivery of nasal high flow oxygen to spontaneously breathing adult patients. It can be used for pre-oxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms. It is not intended for apneic ventilation. HFNC is indicated for use in adults only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Mindray A8, A9 Anesthesia System is provided below.
# 1. SUBMITTER
| Applicant: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81888998<br>Fax: +86 755 26582680 |
|----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Contact Person: Yanhong Bai<br>Title: Manager Regulatory Affairs<br>Phone: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | March 25, 2021 |
#### 2. DEVICE
| Device Trade Name: | A8, A9 Anesthesia System |
|-----------------------|---------------------------------------------|
| Device Common Name: | Gas-Machine, Anesthesia |
| Classification Name: | 868.5160, Class II, Gas-Machine, Anesthesia |
| Regulatory Class: | Class II |
| Primary Product Code: | BSZ |
| Table 1: Secondary Product Codes | | | |
|----------------------------------|-------------------------|-----------------------------|-----------------------------|
| Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name |
| 868.1400, II | CCK | Carbon Dioxide Gas Analyzer | Carbon Dioxide Gas Analyzer |
| 868.1500, II | NHO/CB<br>Q/NHQ/<br>NHP | Enflurane gas analyzer | Enflurane gas analyzer |
#### Table 1· Product Code
{4}------------------------------------------------
| Regulation<br>Number/Class | Product<br>Code | Regulation description | Device Common Name |
|----------------------------|-----------------|----------------------------|----------------------------|
| 868.1620, II | CBS | Halothane Gas Analyzer | Halothane Gas Analyzer |
| 868.1700, II | CBR | Nitrous Oxide Gas Analyzer | Nitrous Oxide Gas Analyzer |
| 868.1720, II | CCL | Oxygen Gas Analyzer | Oxygen Gas Analyzer |
| 880.6740, II | KDP | Vacuum | Vacuum |
## 3. PREDICATE DEVICES
#### Primary predicate:
- K171292 A7 Anesthesia System, ShenZhen Mindray Bio-Medical Electronics Co., Ltd. .
#### Secondary predicate:
- . K192972 - BeneVision N Series Patient Monitors (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1), ShenZhen Mindray Bio-Medical Electronics Co., LTD. (Supports Anesthetic gas module and accessories)
## 4. REFERENCE DEVICES
Per the FDA Guidance, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued July 28, 2104, the following reference devices are provided to support substantial equivalence:
- K160665 - Flow-i Anesthesia System C20, Flow-i Anesthesia System C30, Flow-i Anesthesia System C40, Maquet Critical Care AB (Supports electronic vaporizers, sealed lead acid battery, electronic flowmeter total flow range)
- . K151570 - Carestation 620/650/650C. Datex-Ohmeda, Inc. (Supports apnea pressure and trig window, sealed lead acid battery, APL Valve Adjustable Range)
- K083050 Evita XL Ventilator, Drager Medical AG & Co. KG (Supports tidal volume, ● Inspiratory pressure, support pressure, Plimit, PEEP, Phigh, Plow, RR, PEEP on Exit)
- K180295- HAMILTON-G5, HAMILTON MEDICAL AG (Supports HFNC) ●
## 5. DEVICE DESCRIPTION
The A8, A9 Anesthesia System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor and provides for automatic and manual modes of ventilation. The A8, A9 Anesthesia System incorporates O2, CO2, N2O and Agent concentration monitoring (Desflurane, Isoflurane, Halothane, and Sevoflurane). The A8, A9 Anesthesia System is a modified version the previously cleared Mindray A7 Anesthesia System cleared in K171292.
{5}------------------------------------------------
## 6. INTENDED USE/INDICATIONS FOR USE
The A8, A9 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A8, A9 is intended for use by licensed clinicians in the administration of general anesthesia, for patients requiring anesthesia within a health care facility, and can be used in adult, pediatric and neonate populations.
High Flow Nasal Cannula (HFNC) is indicated for delivery of nasal high flow oxygen to spontaneously breathing adult patients. It can be used for pre-oxygenation and short-term supplemental oxygenation (up to 10 minutes) during intubation in operating rooms. It is not intended for apneic ventilation. HFNC is indicated for use in adults only.
# 7. SUBSTANTIAL EQUIVALENCE
#### Comparison of Indications
Comparing with the primary predicate A7 Anesthesia System (K171292), the indications for use for the subject device A8, A9 Anesthesia System includes the indications for use for the A7 and additionally adds the indications for use for HFNC in adult patients. As the definition of pediatric population subgroups includes the neonate population according to Table 1 of the FDA Guidance, "Guidance for Industry and FDA Staff: Premarket Assessment of Pediatric Medical Devices" issued May 14, 2004, all the populations of the subject device A8, A9 Anesthesia System were cleared in the primary predicate A7 Anesthesia System (K171292).
The BeneVision N Series Patient Monitor modules (Including BeneVision N12, BeneVision N15, BeneVision N17, BeneVision N19, BeneVision N22, BeneVision N1), ShenZhen Mindray Bio-Medical Electronics Co., LTD cleared under K192972 are incorporated into this workstation with no changes to the modules. This predicate serves as a reference device for the Anesthetic Gas Module. A more detailed comparison of the features is included in the sections below.
The indications for the BeneVision N Series Patient Monitors modules are the same since the modules are incorporated with no changes in indications into the anesthesia workstation.
As a conclusion, the indications for use of the subject device A8, A9 Anesthesia System is the same as the primary predicate A7 Anesthesia System as cleared in K171292 and the BeneVision N Series Patient Monitors cleared in K192972.
## Comparison of Technological Characteristics
The table below compares the key technological feature of the subject devices to the primary predicate device (A7 Anesthesia System (K151954)). The features in gray are features which are different between the predicate devices and the subject devices.
{6}------------------------------------------------
| Technical<br>Characteristics | Subject device<br>A8 Anesthesia System<br>Shenzhen Mindray<br>Bio-Medical<br>Electronics Co., Ltd.<br>(Subject device) | Subject device<br>A9 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd.<br>(Subject device) | Primary predicate<br>A7 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co.,<br>Ltd.<br>(K171292) |
|------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Connection Type | Double, Selectatec®,<br>with interlocking<br>function | Double, Insertion type<br>connection, with interlocking<br>function | Two or three, Selectatec®,<br>with interlocking function |
| | - | Working with an electronic based<br>injector | - |
| Electronic<br>Vaporizer | - | Applicable anesthetic agent:<br>Isoflurane<br>Sevoflurane<br>Desflurane | - |
| | - | Filling method:<br>Isoflurane (Safety filling adapter)<br>Sevoflurane<br>(Safety filling<br>adapter)<br>Sevoflurane (Quik-fil)<br>Desflurane (Saf-T-Fill) | - |
| | - | Setting range:<br>Isoflurane: 0.0% to 5.0%<br>Sevoflurane: 0.0% to 8.0%<br>Desflurane: 0.0% to 18.0% | - |
| | - | Accuracy: ±15% of the setting<br>value or ±5% of the maximum<br>setting value, whichever is<br>greater | - |
| Anesthetic agent -<br>Sevoflurane | - | Yes | Yes |
| Anesthetic agent -<br>Isoflurane | Yes | | Yes |
| Anesthetic agent -<br>Desflurane | Yes | | Yes |
| Anesthetic agent -<br>Halothane | Yes | - | Yes |
| <b>Working Mode</b> | - | - | - |
| Standby mode | - | Yes | Yes |
| Manual Ventilation<br>mode | - | Yes | Yes |
| Automatic<br>Ventilator mode | - | Yes | Yes |
| Technical<br>Characteristics | Subject device<br>A8 Anesthesia System<br>Shenzhen Mindray<br>Bio-Medical<br>Electronics Co., Ltd.<br>(Subject device) | Subject device<br>A9 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd.<br>(Subject device) | Primary predicate<br>A7 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co.,<br>Ltd.<br>(K171292) |
| Lung Recruitment<br>Ventilation | Yes | Yes | Yes |
| Cardiac Bypass<br>mode | Yes | Yes | Yes |
| ACGO mode | Yes | Yes | Yes |
| Flow Pause | Yes | Yes | Yes |
| Monitor mode | Yes | Yes | Yes |
| Automatic Ventilation mode | | | |
| VCV | Yes | Yes | Yes |
| PCV | Yes | Yes | Yes |
| PCV-VG | Yes | Yes | Yes |
| SIMV-VC | Yes | Yes | Yes |
| SIMV-PC | Yes | Yes | Yes |
| SIMV-VG | Yes | Yes | Yes |
| PS | Yes | Yes | Yes |
| CPAP/PS | Yes | Yes | Yes |
| APRV | Yes | Yes | Yes |
| Specifications - Anesthetic Ventilator setting parameter | | | |
| Tidal Volume<br>range | 10 to 2000mL (VCV, SIMV-VC)<br>5 to 2000mL (PCV-VG, SIMV-VG) | | 20 to 1500mL |
| Inspiratory<br>Pressure range | 5 to 90cmH2O | | 5 to 70cmH2O |
| Support Pressure<br>range | 0, 3 to 60cmH2O | | OFF, 3 to 50cmH2O |
| Apnea Pressure<br>range | 3 to 60cmH2O | | 3 to 50cmH2O |
| Plimit range | 5 to 100cmH2O | | 10 to 100cmH2O |
| PEEP range | 0 to 50cmH2O | | OFF, 3 to 30cmH2O |
| Tslope range | | 0.0 to 2.0s | 0.0 to 2.0s |
| RR range | 4 to 100bpm | | 4 to 100bpm |
| Technical<br>Characteristics | Subject device<br>A8 Anesthesia System<br>Shenzhen Mindray<br>Bio-Medical<br>Electronics Co., Ltd.<br>(Subject device) | Subject device<br>A9 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd.<br>(Subject device) | Primary predicate<br>A7 Anesthesia System<br>Shenzhen Mindray Bio-<br>Medical Electronics Co.,<br>Ltd.<br>(K171292) |
| Min RR range | 2 to 60bpm | 2 to 60bpm | 2 to 60bpm |
| I:E range | 4:1 to 1:8 | 4:1 to 1:8 | 4:1 to 1:8 |
| Apnea I:E range | 4:1 to 1:8 | 4:1 to 1:8 | 4:1 to 1:8 |
| Apnea Tinsp range | 0.2 to 10s | 0.2 to 10s | 0.2 to 10s |
| Tinsp range | 0.2 to 10s | 0.2 to 10s | 0.2 to 10s |
| Tpause range | OFF, 5% to 60% of Tinsp | OFF, 5% to 60% of Tinsp | OFF, 5% to 60% of Tinsp |
| Trig Window range | 5% to 80% of Texp | 5% to 80% of Texp | 25% of Texp |
| F-Trig range | 0.2 to 15L/min | 0.2 to 15L/min | 0.2 to 15L/min |
| P-Trig range | -20 to -1cmH2O | -20 to -1cmH2O | - |
| Exp% range | 5% to 80% | 5% to 80% | 25% |
| Phigh range | 3 to 90cmH2O | 3 to 90cmH2O…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.