The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.
Device Story
Continuous flow inhalation gas anesthesia system; delivers anesthetic vapor; provides automatic/manual ventilation. Inputs: O2, N2O, anesthetic agents (Desflurane, Isoflurane, Enflurane, Sevoflurane, Halothane). Monitors: CO2, O2, N2O, agent concentration. Operation: Licensed clinicians in health care facilities. Features: Fresh Flow Optimizer, lung recruitment function, heated breathing circuit, spirometry loops. Output: Anesthetic delivery, ventilation support, real-time gas/pressure monitoring. Clinical decision-making: Clinicians use real-time data to adjust anesthesia depth and ventilation parameters. Benefits: Precise anesthetic delivery and respiratory support for diverse patient populations.
Clinical Evidence
Bench testing only. Testing included software unit/integration/system testing, performance testing of gas modules, fresh flow optimizer, suction, heating, power, thermal, and cleaning/disinfection. Biocompatibility testing (VOCs, particulates, cytotoxicity, sensitization, irritation, E&L, inorganic gases) and human factors validation were performed. Device conforms to AAMI/ANSI ES60601-1, IEC 60601-1-2, ISO 80601-2-13, ISO 80601-2-55, ASTM F1101-90, and ISO 10079-3.
Technological Characteristics
Continuous flow gas anesthesia system. Materials: Airway pressure gauge housing material changed. Sensing: O2, CO2, N2O, and anesthetic agent sensors. Connectivity: Data integration with Passport 12M/17M monitors. Control: DSP hardware platform for ventilator control. Modes: VCV, PCV, PCV-VG, SIMV-VC, SIMV-PC, PS, CPAP/PS, APRV. Standards: ISO 80601-2-13, ISO 80601-2-55, IEC 60601-1-2.
Indications for Use
Indicated for adult and pediatric (including neonate, infant, child, adolescent) patients requiring general inhalation anesthesia and ventilation maintenance within a health care facility.
Regulatory Classification
Identification
A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.
January 12, 2018
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Yanhong Bai Manager Regulatory Affairs, Technical Regulation Department Mindray Building, Keji 12th Road South Hi-tech Industrial Park. Shenzhen, 518057 China
Re: K171292
Trade/Device Name: A7 Anesthesia System Regulation Number: 21 CFR 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: Class II Product Code: BSZ, CCK, NHO, CBS, CBQ, CBR, CCL, NHQ, NHP, KDP Dated: December 8, 2017 Received: December 11, 2017
Dear Yanhong Bai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/SafetyReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed by Tara A. Ryan -S
DN: c=US, o=U.S. Government, ou=HHS,
Tara A. Ryan -s ou=FDA, ou People, cn=Tara A. Ryan-5,
0.9.2342.19200300.100.1.1=1300030749
Date: 2018.01.12 11:26:05-05'00'
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K171292
Device Name A7 Anesthesia System
Indications for Use (Describe)
The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:12px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the A7 Anesthesia System is provided below:
| Device Common Name: | Gas-Machine, Anesthesia |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Proprietary Name: | A7 Anesthesia System |
| Submitter: | SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>Shenzhen 518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680 |
| Contact: | Yanhong Bai<br>Manager Regulatory Affairs<br>SHENZHEN MINDRAY BIO-MEDICAL<br>ELECTRONICS CO., LTD.<br>Mindray Building, Keji 12th Road South<br>High-tech Industrial Park, Nanshan<br>518057, P.R. China<br>Tel: +86 755 81885635<br>Fax: +86 755 26582680<br>E-mail: baiyanhong@mindray.com |
| Date Prepared: | December 6th, 2017 |
| Panel: | Anesthesiology |
| Classification Regulation and Product Code: | |
| Primary: | |
| 868.5160- BSZ Anesthesia Gas Machine Class II | |
| Secondary: | |
| 868.1400 – CCK - Carbon Dioxide Gas Analyzer | |
| 868.1500 – NHO/CBQ/NHQ/NHP - Enflurane gas analyzer | |
| 868.1620 - CBS - Halothane Gas Analyzer | |
| 868.1700 – CBR - Nitrous Oxide Gas Analyzer | |
| 868.1720 – CCL- Oxygen Gas Analyzer | |
| 880.6740 – KDP- Vacuum Regulator | |
{4}------------------------------------------------
| 510(k) number | Trade or proprietor or model<br>name | Manufacturer |
|---------------|--------------------------------------|--------------------------------------------------------|
| K151954 | A7 Anesthesia System | Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd. |
#### Predicate Device:
#### Reference Devices:
| 510(k) number | Trade or proprietor or model<br>name | Manufacturer |
|---------------|--------------------------------------|--------------------------------------------------------|
| K123211 | A5 Anesthesia Delivery System | Shenzhen Mindray Bio-<br>Medical Electronics Co., Ltd. |
| K083050 | Evita XL Ventilator | Dräger Medical AG & Co. KG |
| K110213 | GE Datex-OhmedaAisys | Datex-Ohmeda Inc. |
| K132530 | GE Datex-OhmedaAisys CS2 | Datex-Ohmeda Inc. |
### Indications for Use:
The A7 Anesthesia System is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation.
The A7 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care facility, and can be used in adult and pediatric (including neonate, infant, child and adolescent) populations.
## Device Description:
The A7 Anesthesia System is a continuous flow inhalation gas anesthesia system that delivers anesthetic vapor and provides for automatic and manual modes of ventilation. The A7 incorporates 02, CO2, N2O and Agent concentration monitoring (Desflurane, Isoflurane, Enflurane, Sevoflurane and Halothane).
### Nonclinical testing and Performance:
The following areas have been tested:
- . Software
- o Unit testing
- Integration testing o
- o System testing
- Performance ●
- o AG Module
- Fresh Flow Optimizer O
- Sample Gas Return and CO2-absorbent O
{5}------------------------------------------------
- Suction o
- Heating Module O
- Power Supply O
- Thermal O
- Cleaning and Disinfection o
- Waveform Comparison o
- Biocompatibility ●
- Volatile Organic Compounds o
- Particulate Testing O
- Cytotoxicity O
- Sensitization O
- Irritation / intracutaneous reactivity O
- Extractables and leachables (E&L) testing O
- Inorganic gases (03, 02, CO2, NO2) testing O
- Human Factors Validation Testing ●
- . Testing as per consensus standards:
- o AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, ,C1:2009/(R)2012 and A2:2010/(R)2012 Medical electrical equipment - Part 1: General Requirements for basic safety and essential performance
- IEC 60601-1-2 Edition 3: 2007 Medical electrical equipment Part 1-2: General o requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- ISO 80601-2-13 First edition 2011-08-11 Medical electrical equipment Part 2- O 13: Particular requirements for basic safety and essential performance of an anesthetic workstation
- ISO 80601-2-55 First edition 2011-12-15 Medical electrical equipment -Part 2-55: o Particular requirements for the basic safety and essential performance of respiratory gas monitors
- ASTM F1101-90 (Reapproved 2003) Standard Specification for Ventilators O Intended for Use During Anesthesia
- ISO 10079-3:2014 Medical suction equipment Part 3:Suction equipment o powered from a vacuum or positive pressure gas source
- AIM 7351731 medical electrical equipment and system electromagnetic O immunity test for exposure to radio frequency identification readers - an aim standard. (General II (ES/EMC))
{6}------------------------------------------------
The functional and system level testing showed that the device continues to meet specifications and the performance of the device is equivalent to the predicate.
## Substantial Equivalence:
### Comparison of Indications:
Both the predicate device and the subject A7 Anesthesia System are gas anesthesia machine used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The indications for use of the subject device (A7 Anesthesia System) have been modified to add the definition of the pediatric population subgroup. All of the pediatric population subgroup was cleared in the predicate device A7 Anesthesia System (K151954). In conclusion, the minor changes to the indications for use do not change the intended use of the A7 Anesthesia System.
{7}------------------------------------------------
# Comparison of Technological Characteristics:
The table below compares the key technological feature of the subject device to the predicate device (A7 Anesthesia System (K151954)).
# Device Comparison Table
| | Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System<br>(subject device) | Predicate Device<br>A7 Anesthesia System<br>(K151954) |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| | Vaporizers | Two or Three, variable bypass | Two or three, variable bypass |
| | Agent - Sevoflurane | Yes | Yes |
| | Agent - Isoflurane | Yes | Yes |
| | Agent - Desflurane | Yes | Yes |
| | Agent - Halothane | Yes | Yes |
| | Agent - Enflurane | Yes | Yes |
| | Automatic Ventilator | Yes | Yes |
| | Bellows | Yes | Yes |
| | Bellows Volume | 1500mL | 1500mL |
| Ventilation Modes | | | |
| | VCV | Yes | Yes |
| | PCV | Yes | Yes |
| | PCV – VG | Yes | Yes |
| | SIMV - VC | Yes | Yes |
| | SIMV - PC | Yes | Yes |
| | PS | Yes | Yes |
| Specifications | | | |
| | Tidal Volume | Yes | Yes |
| | Range, ml | 20 to 1500 | 20 to 1500 |
| | Minute Volume | Yes | Yes |
| | Rate, bpm | 4 to 100 bpm | 4 to 100 bpm |
| | I:E Ratio | 4:1 to 1:8 with 0.5 increment | 4:1 to 1:8 with 0.5 increment |
| | Inspiratory Pause | Off, 5 to 60% of insp. Period | Off, 5 to 60% of insp. Period |
| | Air Flow Range | 0 to 15 L/min | 0 to 15 L/min |
| Fresh Gas | N2O Flow Range | 0 to 12 L/min | 0 to 12 L/min |
| | O2 Flow Range | 0 to 15 L/min | 0 to 15 L/min |
| Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System<br>(subject device) | Predicate Device<br>A7 Anesthesia System<br>(K151954) | |
| Individual Gas Flow<br>Accuracy | ±50 ml/min or ±5% of setting<br>value, whichever is greater | ±50 ml/min or ±5% of setting<br>value, whichever is greater | |
| Pressure Limit, cm H₂O | 0 to 100cm H₂O | 0 to 100cm H₂O | |
| PEEP, cm H₂O | Off, 3 to 30, 1 cm H₂O increment | Off, 3 to 30, 1 cm H₂O increment | |
| System Checks | Auto at start | Auto at start | |
| Airway Pressure Measured at | Inspiratory | Inspiratory | |
| High/Low Airway Pressure Alarm | Yes | Yes | |
| Pressure Limiting Alarm | Yes | Yes | |
| Sub Atmospheric Pressure Alarm | Yes | Yes | |
| Continuous Press Alarm | Yes | Yes | |
| Apnea >2 Minute Alarm | Yes | Yes | |
| Apnea Alarm | Yes | Yes | |
| High/Low Minute Volume Alarm | Yes | Yes | |
| High/Low O₂ Concentration Alarm | Yes | Yes | |
| Heated Breathing Circuit | Yes | Yes | |
| Spirometry: Pressure-Volume and<br>Flow-Volume loops | Yes | Yes | |
| Anesthetic Gas Module Sampling Rate | Adult/pediatric: 120, 150,<br>200mL/min<br>Neonate: 70, 90, 120mL/min | High volume: 120, 150,<br>200mL/min<br>Low volume: 70, 90, 120mL/min | |
| Anesthetic Gas Module Sampling<br>Delay Time: | <4 seconds | <4 seconds | |
| Anesthetic Gas Module Refresh Rate: | 1 second | 1 second | |
| Anesthetic Gas Module Warm-up<br>Time: | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status | 45 seconds to warm-up status<br>10 minutes to ready-to-measure<br>status | |
| Anesthetic Gas Module Accuracy<br>CO₂: | 0 to 1%: +/-.1%<br>1 to 5%: +/-.2%<br>5 to 7%: +/-.3%<br>7 to10%: +/-.5%<br>>10%: unspecified | 0 to 1%: +/-.1%<br>1 to 5%: +/-.2%<br>5 to 7%: +/-.3%<br>7 to10%: +/-.5%<br>>10%: unspecified | |
| Anesthetic Gas Module Accuracy<br>N₂O: | 0 to 20%: +/-2%<br>20 to 100%: +/-3% | 0 to 20%: +/-2%<br>20 to 100%: +/-3% | |
| Technical<br>Characteristics | Subject Device<br>A7 Anesthesia System<br>(subject device) | Predicate Device<br>A7 Anesthesia System<br>(K151954) | |
| Anesthetic Gas Module Accuracy<br>Desflurane: | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 10%: +/-.4%<br>10 to 15%: +/-.6%<br>15 to 18%: +/-1%<br>>18%: unspecified | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 10%: +/-.4%<br>10 to 15%: +/-.6%<br>15 to 18%: +/-1%<br>>18%: unspecified | |
| Anesthetic Gas Module Accuracy<br>Sevoflurane: | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 8%: +/-.4%<br>>8%: unspecified | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>5 to 8%: +/-.4%<br>>8%: unspecified | |
| Anesthetic Gas Module Accuracy<br>Enflurane/Isoflurane/<br>Halothane: | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>>5%: unspecified | 0 to 1%: +/-.15%<br>1 to 5%: +/-.2%<br>>5%: unspecified | |
| Anesthetic Gas Module Accuracy O2: | 0 to 25%: +/-1%<br>25 to 80%: +/-2%<br>80 to 100%: +/-3% | 0 to 25%: +/-1%<br>25 to 80%: +/-2%<br>80 to 100%: +/-3% | |
| Anesthetic Gas Module Accuracy<br>awRR: | 2 to 60rpm: +/-1rpm<br>>60rpm: unspecified | 2 to 60rpm: +/-1rpm<br>>60rpm: unspecified | |
| Anesthetic Gas Module<br>Measurement<br>Rise Time: | CO2: <250ms<br>N2O: <250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: <350ms | CO2: <250ms<br>N2O: <250ms<br>O2: ≤500ms<br>Hal/Iso/Sev/Des: ≤300ms<br>Enf: <350ms | |
| Measurement Range CO2 | 0 to 30% | 0 to 30% | |
| Measurement Range N2O | 0 to 100% | 0 to 100% | |
| Measurement Range Des | 0 to 30% | 0 to 30% | |
| Measurement Range Sev | 0 to 30% | 0 to 30% | |
| Measurement Range Enf/Iso/Hal | 0 to 30% | 0 to 30% | |
| Measurement Range O2 | 0 to 100% | 0 to 100% | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
There are some differences between the subject A7 Anesthesia System and the predicate device A7 Anesthesia System (K151954). Per the FDA Guidance, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" issued July 28, 2014, the following devices are provided to support substantial equivalence.
| Differences | Reference Device | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|----------------------------------------------------------|
| Add Three Ventilation<br>Modes | SIMV-VG | Evita XL Ventilator (K083050) |
| | CPAP/PS | Aisys anesthesia system (K110213) |
| | APRV | Evita XL Ventilator (K083050) |
| Lung Recruitment Ventilation Function | | GE Datex-Ohmeda Aisys CS2<br>Anesthesia System (K132530) |
| Added O2 Sensor | | A5 Anesthesia Delivery System<br>(K123211) |
| Redesign the Existing Ventilator Control Board Using<br>DSP Hardware Platform | | A7 Anesthesia System (K151954) |
| Change about the Anesthetic Gas Module and<br>Accessories | | A7 Anesthesia System (K151954) |
| Get Data from Anesthetic Gas Module plugged into<br>the Passport 12M/17M patient monitor (K170876) to<br>support optimizer function and agent usage calculation<br>function | | A7 Anesthesia System (K151954) |
| Increase the Maximum Inspiratory Flow | | Evita XL Ventilator (K083050) |
| Change the Material of the Airway Pressure Gauge<br>Housing | | A7 Anesthesia System (K151954) |
### Substantial Equivalence Conclusion:
Based on the detailed comparison of specifications for each of the modifications to the previously cleared A7 Anesthesia System (K151954) and cited reference devices, the performance testing results and conformance with applicable standards show that the A7 Anesthesia System can be found substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.