Non-invasive measurement of systolic pressure (SYS), diastolic pressure (DIA), heart rate (HR), heart rate variability (HRV), and irregular heartbeats (IrrHB) for professionals at office or patients at home. The device is not intended for ambulatory use. The device has not been tested and it is not intended for pediatric use. The device provides a HRV parameter which is only mathematical analysis and is not intended to produce any interpretation of those measurements or any kind of diagnosis.
Device Story
Wrist-worn monitor; uses oscillometric method for blood pressure (SYS/DIA) and heart rate; piezo-electrical sensors in cuff capture radial arterial waveforms. Device performs 5-minute continuous waveform recording to detect irregular heartbeats; calculates HRV via time- and frequency-domain statistical analysis of pulse peak intervals. Operates via Bluetooth connection to smartphone/tablet app; app displays live test data, post-test reports, and plots; supports cloud data upload. Used by professionals or patients at home; not for ambulatory use. Output provides raw measurements and rankings based on standard values; no clinical interpretation or diagnosis provided. Benefits include non-invasive monitoring of cardiovascular parameters.
Clinical Evidence
Small-scale, open, random, sequential comparative study (n=10; 8 men, 2 women; age 18-86) comparing HRVWatch-Mini to HRVWatch-Full-Size (K123130). Endpoints included BP, HR, HRV, and irregular heartbeat detection. Results demonstrated technical design, function, and data accuracy equivalence. Bench testing confirmed compliance with ANSI/AAMI SP10-1992, IEC 60601-1, and ISO 13485 standards.
Technological Characteristics
Wrist-worn oscillometric monitor; piezo-electrical sensors for radial pulse waveform acquisition. Bluetooth connectivity to mobile app (Android/iOS). Complies with ANSI/AAMI SP10-1992, ANSI/AAMI ES60601-1, IEC 60601-1-2, and ETSI standards. Software-controlled signal acquisition and analysis.
Indications for Use
Indicated for non-invasive measurement of systolic/diastolic blood pressure, heart rate, heart rate variability, and irregular heartbeats in adults. For use by professionals in office settings or patients at home. Not for ambulatory or pediatric use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font.
June 7, 2021
Taiwan Scientific Corporation D.C. Sun General Manager 10F, 88-4, Ming-Chiuan Rd., Shin-Dian District New Taipei, 23141 Taiwan
Re: K201914
Trade/Device Name: HRVWatch-Mini Wrist Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: March 31, 2021 Received: May 3, 2021
Dear D.C. Sun:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201914
Device Name HRVWatch-Mini Wrist Monitor
#### Indications for Use (Describe)
Indications for Use: Non-invasive measurement of systolic pressure (SYS), diastolic pressure (DIA), heart rate (HR), heart rate variability (HRV), and irregular heartbeats (IrrHB) for professionals at home. The device is not intended for ambulatory use. The device has not been tested and it is not intended for pediatric use. The device provides a HRV parameter which is only mathematical analysis and is not intended to produce any interpretation of those measurements or any kind of diagnosis.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) Summary
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of Safe Medical Devices Act (SMDA) and 21 CFR 807.92.
## 5.1 510(k) Number: k201914
## 5.2 Applicant Information:
Date Prepared: May 20, 2020 Name: Taiwan Scientific Corporation Address: 10 F, 88-4, Ming-Chiuan Rd., Shin-Dian District, New Taipei 23141, Taiwan Phone: +886-2-8667-3362 Fax: +886-2-8667-3387 Contact Person: Dehchuan Sun Phone Number: Office: +886-2-8667-3362 / Cell +886-953-933-962 Fax Number: +886-2-8667-3387 E-mail: sundc@taiwanscientific.com.tw; sundc(@hotmail.com
## 5.3 Device Information:
Classification: DXN /Class II/870.1130 Trade Name: HRVWatch-Mini Wrist Monitor Common Name: Non-invasive blood pressure meter system Classification Name: Noninvasive blood pressure measurement system
## 5.4 Predicate Devices:
(1) HRVWatch Wrist Monitor K123130 (primary predicate device) (2) Full Automatic (NIBP) Blood Pressure Monitor HL158UA K162338 (3) Alivecor Heart Monitor K142743
## 5.5 Device Description:
The non-invasive blood pressure measuring system HRVWatch-Mini Wrist Monitor is a smaller size variation of the HRVWatch Wrist Monitor (K123130) previously cleared by FDA in 2013. Identical to its larger-size predecessor in terms of device main functions, the Mini monitor measures blood pressures (BP) (systolic pressure SYS and diastolic pressure DIA) and heart rate (HR), using the oscillometric method commonly employed by electronic blood pressure meters. In addition, the device utilizes piezo-electrical sensors embedded in the cuff (wrist band) to obtain detailed radial arterial waveforms. The 5-minute continuous waveform recording allows the device to detect and report irregular heartbeats. Heart rate variability (HRV) parameters are also obtained by time- and frequency-domain statistical analyses of pulse peak intervals. The main body of the device with its biosensors and electrical circuit boards has the full function of signal acquisition, analysis, and data storage. The testing by the device is controlled by a software programs (app) installed in a smart phone or tablet (the collateral device) via two-way blue-tooth communication. The display of the collateral device (phone or tablet) is used to show live test data plots during test, or post-test data reports and plots. Test data can be uploaded to
{4}------------------------------------------------
cloud using the collateral device's internet connectivity. HRVWatch-Mini Wrist Monitor is intended to provide only patient parameter measurements and associated rankings based upon established standard values, and is not intended to produce any clinical interpretation of those measurements or any kind of diagnosis.
#### 5.6 Indications for Use:
Indications for Use: Non-invasive measurement of systolic pressure (SYS), diastolic pressure (DIA), heart rate (HR), heart rate variability (HRV), and irregular heartbeats (IrrHB) for professionals at office or patients at home. The device is not intended for ambulatory use. The device has not been tested and it is not intended for pediatric use. The device provides a HRV parameter which is only mathematical analysis and is not intended to produce any interpretation of those measurements or any kind of diagnosis.
#### 5.7 Comparison to Predicate Device(s):
The subject device HRVWatch-Mini wrist monitor is shown to be substantially equivalent to:
- (1) HRVWatch Wrist Monitor K123130 (Prescription use; primary predicate device) for the measurement of BP, HR, HRV, and IrrHB
- (2) Full Automatic (NIBP) Blood Pressure Monitor HL158UA K162338 (OTC use) for the measurement of BP, HR, and IrrHB
- (3) Alivecor Heart Monitor K142743 (Prescription and OTC uses) for the measurement of IrrHB
In this submission, the subject device is shown to be substantially equivalent to the primary predicate device HRVWatch Wrist Monitor K123130 for the measurement of BP, HR, HRV, and IrrHB in the application of prescription use. HRVWatch-Mini is a size-reduced device designed from HRVWatch K123130, with identical key parts (the BP pressure IC sensor, the PZT biosensor module, and the analytical firmware, etc.). HRVWatch K123130 connects to a Windows-based PC for data display, print, and upload, while the subject device is operated in connection with a smart phone or tablet (Android and iOS) via blue-tooth transmission. The subject device is shown to be substantially equivalent to the predicate device "Full Automatic (NIBP) Blood Pressure Monitor HL158UA K162338" for the measurement of BP, HR, and IrrHB in the application of OTC use. K162338 does not measure HRV while the subject device does. However, the additional functionality of HRV measurement in the subject device does not pose any new safety concern.
The subject device is shown to be substantially equivalent to the predicate device "Alivecor Heart Monitor K142743" for the measurement of IrrHB in the application of prescription and OTC use. K142743 is designed for the recording, detection and data transmission of irregular heartbeat with a single-lead ECG, while the subject device performs the similar functions based upon the radial pulse waves.
From the documentation and test results included this submission, the subject device is substantially equivalent to the predicate devices list above for the measurement of BP, HR, HRV, and IrrHB in the application of prescription and over-the-counter (OTC) use.
#### 5.8 Performance Summary (Bench and Clinical):
HRVWatch-Mini Wrist Monitor has been tested by both in-vitro (bench) and human clinical studies. The test results show that HRVWatch-Mini Wrist Monitor is in compliance with the following international standards, FDA Guidance documents, and literature articles:
{5}------------------------------------------------
- (i) FDA Guidance "Noninvasive blood pressure meter submission guide" (March 10, 1997)
- (ii) ANSI/AAMI SP10-1992 Electronic or Automatic Sphygmomanometers
- (iii) ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 - Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- (iv) IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- (v) ETSI EN 301 489-1 and ETSI EN300328 for blue-tooth emission compatibility
- (vi) ISO 13485:2016 for Medical Devices
- (vii) EC Declaration of Conformity (Medical Device Directive 93/42/EEC)
- (vii) 1996 Heart Rate Variability publication in the journal of Circulation (Circulation 1996, 93, 1043-1065; regarded widely as the HRV International Standard)
# 5.9 Discussion of Clinical Test
(i) The submission subject device HRVWatch-Mini is a sized reduced model from the FDA cleared device HRVWatch Wrist Monitor K123130 (full size)
(ii) To confirm the design change, a small scale, open, random, sequential, and comparative study was conducted between HRVWatch-Mini and HRVWatch-Full-Size wrist monitors (iii) A total of 10 subjects were recruited from company employees and family members for the human clinical study. Among the participants, 8 were men and 2 were women. Age range was from 18 to 86 years-old, height from 150 (4 foot 11 inch) to186 cm (6 foot 1 inch), and body weight from 45 (99.2 lb) to 75 Kg (165.3 lb). All participants lived in Taiwan and healthy. (iv) Tests were conducted comparing two HRVWatch devices for measurement of BP, heart rate,
- heart rate variability (HRV), and irregular heartbeat (IrrHB)
- (v) The clinical test results showed that HRVWatch-Mini was substantially equivalent to the predicate device HRVWatch-full-size for technical designs, functions, and data accuracy.
## 5.10 Conclusions:
After bench and clinical performance tests, HRVWatch-Mini was shown to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.