Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box)
Applicant
Rebound Therapeutics
Product Code
GWG · Neurology
Decision Date
Nov 4, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
Device Story
System comprises sterile, single-use sheath with integrated LEDs/camera and obturator; non-sterile, reusable Image Control Box (ICB). Sheath provides surgical access and instrument channel; obturator facilitates insertion via conical tip. LEDs provide illumination; camera captures video. ICB powers components, relays video to external monitor, and allows user adjustments (zoom, contrast, brightness, orientation) via membrane keypad. Used in neurosurgery/endoscopic neurosurgery by physicians. Provides real-time visualization to assist in diagnostic/therapeutic procedures; aids in navigation and tissue removal; benefits patient by enabling minimally invasive access to surgical sites.
Clinical Evidence
No clinical data. Evidence provided via bench testing: biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1-1, 60601-1-2, 60601-2-18), software verification/validation, mechanical testing (dimensional, imaging, illumination, tensile strength), simulated use testing, particulate testing (USP 788), sterilization validation (ISO 11135-1), and shelf-life testing (ISTA 2A, ASTM F1980).
Indicated for patients undergoing neurosurgery, endoscopic neurosurgery, or pure neuroendoscopy (e.g., ventriculoscopy) requiring visualization, diagnostic, or therapeutic procedures including ventriculostomies, biopsies, and removal of cysts, tumors, or other obstructions.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 4, 2020
Rebound Therapeutics Naomi Gong VP of Regulatory Affairs 13900 Alton Parkway Suite 120 Irvine, California 92618
Re: K201840
Trade/Device Name: Aurora Surgiscope System (Surgiscope), Aurora Surgiscope System (Image Control Box) Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG, GZT Dated: July 1, 2020 Received: July 2, 2020
Dear Naomi Gong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201840
Device Name AURORA Surgiscope System
#### Indications for Use (Describe)
The AURORA Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary (K201840)
#### SUBMITTER
Rebound Therapeutics 13900 Alton Parkway, Suite 120 Irvine, CA 92618
| Contact Person: | Naomi Gong, RAC<br>Vice President of Regulatory Affairs |
|-----------------|---------------------------------------------------------|
| Telephone: | (949) 523-6510 |
| Email: | naomi.gong@integralife.com |
| Date Prepared: | November 4, 2020 |
DEVICE
| Name of Device: | AURORA Surgiscope System |
|--------------------|------------------------------------------------------------------|
| Regulation Number: | 21 CFR 882.1480, 21 CFR 882.4800 |
| Regulation Name: | Neurological endoscope, Self-retaining retractor for neurosurger |
| Regulatory Class: | II |
| Product Code: | GWG, GZT |
#### PREDICATE DEVICE
AURORA Surgiscope System, Rebound Therapeutics, K191861 NICO BrainPath, NICO Corp, K172433 (Reference device)
## DEVICE DESCRIPTION
The Aurora Surgiscope System consists of two components: (1) a sterile, single use, sheath with integrated illumination LEDs and camera, with an obturator, and (2) a non-sterile, reusable control unit, Image Control Box (ICB).
The sheath is intended to provide access to the surgical site by acting as the insertable portion of the device, as well as the instrument channel to accommodate other surgical tools. Depth markers are present along the length of the sheath for user reference.
At the proximal end of the sheath is the imager, which comprises the following components: LEDs (light emitting diodes), camera (and optical components), and focus knob.
- . The LEDs provide illumination to the surgical field by directing light down the sheath, along the instrument channel.
- The camera captures video image of the surgical field.
The proximal end of the sheath also contains a tab, which serves as the location for the fixation arm to hold/fix the device. To facilitate insertion of the surgical site, an obturator is provided with the device. During insertion, the obturator is fully inserted into the entire unit is advanced to the desired location. The distal end of the obturator is conical in shape to minimize tissue damage. In addition, the proximal handle of the obturator is designed to accommodate various stereotactic instruments for neuronavigation. Once inserted, the obturator is removed.
The ICB is a non-sterile device that provides three main functions in the AURORA Surgiscope System:
{4}------------------------------------------------
- . To power the Surgiscope LEDs and camera
- . To relay the video feed captured by the Surgiscope camera to a display monitor for real-time image visualization
- . To allow the user to make adjustments to the displayed video feed (e.g., contrast, brightness), as well as vary the LED light output.
The user interface is a membrane keypad with buttons located on the ICB that can be depressed for image adjustment, such as zoom, contrast, brightness, and orientation. The ICB is supplied with two cables: A power cable for connection to an AC wall outlet, and a display cable for connection to a high definition surgical monitor.
## INDICATIONS FOR USE
The Aurora Surgiscope System is intended for use in neurosurgery and endoscopic neurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.
### SUMMARY OF NON-CLINICAL TESTING
The following testing was conducted to demonstrate the safe and substantial equivalence to the predicate device.
- (1) Biocompatibility testing
The AURORA Surgiscope System is intended to come into direct contact with human tissue during use and was evaluated and tested per the requirements and recommendations of the FDA Guidance, "Use of ISO-10993-1, Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process." The battery of testing included the following tests:
-Cytotoxicity (MEM Elution) -Sensitization (Kligman Maximization) -Irritation (Intracutaneous Injection) -Systemic Toxicity (Systemic Injection) -Hemolysis (In direct) -Materials Mediated Pyrogenicity
The sheath with LEDs and camera portion is considered a tissue contacting and externally communicating for a duration of less than 24 hours. The ICB is not intended to have direct or indirect patient contact and therefore biocompatibility for it was not evaluated.
- (2) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the AURORA Surgiscope System consisting of both components, sheath with LEDs and camera, and the ICB. The system complied with the IEC 60601-1-1, IEC 60601-1-2, and IEC 60601-2-18.
- (3) Software verification and validation testing
Software verification and validation testing were conducted and documentation was provided as recommended by the FDA Guidance. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". The software for this device was considered as a major level of concern, since a failure or latent flow in the software could directly result in a serious injury or death to the patient or operator.
{5}------------------------------------------------
- (4) Mechanical and other testing
- Bench testing dimensional, imaging (visualization, resolution, etc.), illumination, tensile . strength between components
- Simulated use testing – clinician evaluation
- . Compatibility - instrument use in instrument channel, use with other devices (i.e., bipolar electrosurgical)
- Particulate Testing per USP <788> .
- Sterilization per ISO 11135-1 to validate a SAL of 10-6 .
- . Packaging and Shelf-life per ISTA 2A and ASTM F1980 _
## COMPARISON OF TECHNICAL CHARACTERISTICS TO PREDICATE DEVICE
| | PREDICATE Device | SUBJECT Device |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Aurora Surgiscope System (K191861) | Aurora Surgiscope System |
| INDICATIONS<br>FOR USE | The Aurora Surgiscope System is intended for use in<br>endoscopic neurosurgery and pure neuroendoscopy<br>(i.e. ventriculoscopy) for visualization, diagnostic<br>and/or therapeutic procedures such as<br>ventriculostomies, biopsies and removal of cysts,<br>tumors and other obstructions. | The Aurora Surgiscope System is intended for use in<br>neurosurgery, and endoscopic neurosurgery and<br>pure neuroendoscopy (i.e. ventriculoscopy) for<br>visualization, diagnostic and/or therapeutic<br>procedures such as ventriculostomies, biopsies and<br>removal of cysts, tumors and other obstructions. |
| | | |
| Sheath | ■ OD ≤ 11.5mm, ID = 8 mm<br>■ Lengths = 70, 100, 130 mm<br>■ Incremental depth markings<br>■ Single working channel<br>■ With obturator (conical shape, rounded tip) | ■ OD = 16.2 mm, ID = 11 mm<br>■ Lengths = 60 and 80 mm<br>■ Same<br>■ Same<br>■ Same |
| Materials | Sheath: ABS, PET, adhesives, white ink depth marker<br>Obturator: Polycarbonate, 304 SS, ABS, adhesives | Sheath: Aluminum, adhesives<br>Obturator: Same |
| Imager | The device incorporates an imaging system with<br>camera and optics (lens, prism):<br>Direction of View = 0°<br>Depth of field = 0 to 3 cm | Same |
| Light Source<br>Camera | 6 LED incorporated into inner dia. of sheath near<br>distal end - fixed intensity level<br>Integrated CMOS camera and electronics | 4 LEDs incorporated into Imager (proximal end of<br>sheath) - user can vary intensity level<br>Same |
| Image Control Box<br>(control unit) | ■ Electronics (Circuit boards, CPU with software<br>control)<br>■ Keypad buttons for image adjustment by user | ■ Equivalent<br>■ Same |
| Other (accessories) | Power supply and cable<br>Display cable | Same |
| Display Monitor | Not supplied | Same |
| | | |
| Access to the<br>surgical site | Sheath and Obturator used to access the surgical site.<br>The Obturator extends 10mm beyond distal end of<br>sheath | Same |
| Single working<br>channel | Removal of obturator reveals working channel which<br>provides visual access for camera and working access<br>for instruments including suction, bipolar, and<br>irrigation. | Same |
| Image acquisition | Image acquisition is achieved through an integrated<br>camera | Same |
| Image processing | Image is digitally processed by control unit | Same |
| Image display | External display monitor connection | Same |
| Illumination | Illumination is achieved via direct LED light sources<br>incorporated into the device. | Same |
| Visualization | CMOS, color, video, camera (with software)<br>incorporated at proximal end of the device and<br>controlled via control unit. | Equivalent |
{6}------------------------------------------------
| | PREDICATE Device | SUBJECT Device |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|
| | Aurora Surgiscope System (K191861) | Aurora Surgiscope System |
| Biocompatibility | Surgiscope: Demonstrated based on externally<br>communicating device in direct contact with<br>tissue/bone/dentin for a limited duration | Same |
| Use and how<br>supplied | Surgiscope: single use, sterile<br>ICB (control unit): reusable, non-sterile | Same |
| Sterilization<br>Method | Surgiscope: Ethylene oxide gas | Same |
The NICO BrainPath device (K172433) is included as a reference device because its sheath diameter is comparable and its sheath/obturator has the same intended use to access the surgical field during neurosurgery.
## CONCLUSION
Based upon the performance data provided in this submission and comparing indications for use, design, materials, principle of operation and overall technological characteristics, the AURORA Surgiscope System has been determined to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.