To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following: - Primary/Secondary Brain Tumors - Vascular Abnormalities/Malformations - Intraventricular Tumors/Cysts
Device Story
NICO BrainPath provides minimally invasive subcortical access to neurological tissues; creates atraumatic surgical corridor. Device consists of reusable aluminum obturators and single-use cyclic olefin copolymer (COC) sheaths. Surgeon assembles obturator and sheath in OR; inserts into brain; removes obturator to leave sheath in place, providing 11 mm or 13.5 mm surgical corridor for visualization and instrument access. Used in neurosurgery; compatible with third-party navigation and retractor arms. Modifications include concave proximal sheath ring to reduce optical glare and expanded size offerings (11 mm diameter; 95 mm length). Benefits include minimally invasive access to deep-seated brain pathologies, potentially reducing tissue trauma compared to traditional open surgery.
Clinical Evidence
Bench testing only. Testing included cytotoxicity, sensitization, irritation, simulated use with third-party instruments, packaging/shelf-life validation, cleaning validation (bioburden/endotoxin/protein), and sterility validation (autoclave/gas plasma). All tests passed.
Technological Characteristics
Materials: Aluminum (obturator), COC (sheath). Principle: Mechanical obturator-sheath assembly for surgical corridor creation. Dimensions: 11 mm and 13.5 mm inner diameters; 50-95 mm lengths. Connectivity: None. Sterilization: Gamma (single-use), Autoclave/Hydrogen Peroxide Gas Plasma (reusable).
Indications for Use
Indicated for patients requiring subcortical access for primary/secondary brain tumors, vascular abnormalities/malformations, or intraventricular tumors/cysts during brain and spinal surgery.
Regulatory Classification
Identification
A self-retaining retractor for neurosurgery is a self-locking device used to hold the edges of a wound open during neurosurgery.
{0}------------------------------------------------
Public Health Service
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 7, 2017
NICO Corporation Sean Spence Regulatory Affairs Manager 250 E 96th St. Suite 125 Indianapolis, Indiana 46240
Re: K172433
Trade/Device Name: NICO BrainPath Regulation Number: 21 CFR 882.4800 Regulation Name: Self-Retaining Retractor for Neurosurgery Regulatory Class: Class II Product Code: GZT Dated: August 10, 2017 Received: August 11, 2017
Dear Mr. Spence:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172433
Device Name NICO BrainPath
Indications for Use (Describe)
To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:
- Primary/Secondary Brain Tumors
- · Vascular Abnormalities/Malformations
· Intraventricular Tumors/Cysts"
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <input checked="true" type="checkbox"/> Recreational Use (Part 21 CFR 201.327(p)) </div></td></tr><tr><td><div> <input type="checkbox"/> Sunscreen Use (21 CFR 201.327(c)) </div></td></tr></table> | <div> <input checked="true" type="checkbox"/> Recreational Use (Part 21 CFR 201.327(p)) </div> | <div> <input type="checkbox"/> Sunscreen Use (21 CFR 201.327(c)) </div> |
|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Recreational Use (Part 21 CFR 201.327(p)) </div> | | | |
| <div> <input type="checkbox"/> Sunscreen Use (21 CFR 201.327(c)) </div> | | | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NICO Corporation. The letters "NICO" are in a bold, sans-serif font, with the "O" being a filled circle in a golden color. Below the letters is the word "CORPORATION" in a smaller, sans-serif font.
# NICO BrainPath®
21 CFR §807.92 Date Prepared: 06 September 2017
510(k) Number: K172433
| Submitter/Manufacturer: | NICO Corporation<br>250 E. 96th Street, Suite 125<br>Indianapolis, IN 46240 |
|-------------------------|-----------------------------------------------------------------------------|
| Contact Person: | Sean Spence, RAC<br>Regulatory Affairs Manager<br>Office: 317.660.7118 |
| Trade Name: | NICO® BrainPath® |
| Common/Usual Name: | Self-retaining retractor for neurosurgery |
| Classification / Panel: | 21 CFR §882.4800 / Neurology |
| Product Codes: | GZT |
| Predicate Device: | K150378, NICO BrainPath |
#### Device Description
The NICO BrainPath and accessories are designed to provide minimally invasive access to neurological tissues. The design specifically supports the creation of an atraumatic surgical corridor to access the brain. These BrainPath devices are part of what is being called the BrainPath Approach™, which integrates the expanding neurosurgical armamentarium of trajectory planning and navigation, optics, corridor resection and biopsy, and tissue preservation. To date, the BrainPath technology has been used to successfully access primary and secondary brain tumors, vascular abnormalities or malformations, and intraventricular tumors and cysts.
The BrainPath consists of multiple-sized reusable and re-sterilizable obturators with coordinating single patient use disposable sheaths. The obturator and sheath are assembled in the operating room immediately prior to use. After placement, the obturator is removed leaving behind the sheath which provides a 13.5 mm or 11 mm surgical corridor. Table 1 outlines the available BrainPath configurations.
| Name | Sheath Inner-Diameter (mm) | Sheath Length (mm) | Obturator Tip (mm) |
|----------------------|----------------------------|--------------------|--------------------|
| 13.5 mm x 50 mm (ST) | | 50 | 7.5 |
| 13.5 mm x 50 mm | | 50 | |
| 13.5 mm x 60 mm | 13.5 | 60 | |
| 13.5 mm x 75 mm | | 75 | |
| 13.5 mm x 95 mm | | 95 | 15 |
| 11 mm x 50 mm | | 50 | |
| 11 mm x 60 mm | 11 | 60 | |
| 11 mm x 75 mm | | 75 | |
#### Table 1: BrainPath Configurations
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for NICO Corporation. The logo consists of the word "NICO" in a bold, sans-serif font, with the "O" being a circle filled with a golden color. Below the word "NICO" is the word "CORPORATION" in a smaller, sans-serif font.
The BrainPath accessories include a Manipulation Tool which is used for manipulating the position of the sheath after it has been placed, Shepherd's Hooks which are for attaching to various commercially available retractor arms, a 95 mm Stylet for use with the 95 mm obturator, and a Sterilization Tray for sterilization of the reusable components (i.e., obturators, manipulation tools, and 95 mm stylet).
| BrainPath Component | Sterile or Non-Sterile | Single-Use or Reusable |
|------------------------------|------------------------|------------------------|
| Sheath (all sizes) | Sterile | Single-Use |
| Obturators (all sizes) | Non-Sterile | Reusable |
| Manipulation Tool | Non-Sterile | Reusable |
| 95mm Stylet | Non-Sterile | Reusable |
| Sterilization Tray | Non-Sterile | Reusable |
| Shepherd's Hooks (all types) | Sterile | Single-Use |
#### Table 2: NICO BrainPath Components Supplied
## Intended Use
To provide for access and allow for visualization of the surgical field during brain and spinal surgery. Indications may include subcortical access to diseases such as the following:
- Primary/Secondary brain tumors -
- Vascular abnormalities/malformations -
- -Intraventricular tumors/cysts
#### Comparison to Predicate
The subject BrainPath iteration is nearly identical to the predicate NICO BrainPath (K150378). Modifications include a concave proximal sheath ring geometry to reduce optical glare, and a product line extension to include a 13.5 mm x 95 mm and 11 mm set of offerings. The 95 mm length Obturator also necessitated the creation of a new 95 mm stylet accessory. Table 3 provides additional details on how the subject and predicate devices compare.
| | Predicate Device<br>NICO BrainPath | Subject Device<br>NICO Brain Port |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Device Name | BrainPath | Identical |
| Intended Use | To provide for access and allow for visualization of the surgical field during brain and spinal surgery.<br>Indications may include subcortical access to diseases such as the following:<br>- Primary/Secondary brain tumors<br>- Vascular abnormalities/malformations<br>- Intraventricular tumors/cysts | Identical |
| Principle of<br>Operation | Consists of an "obturator-like" component and a “sheath-like” component which are assembled, inserted, and disassembled to provide corridor access | Identical |
| Operation and<br>placement | Handheld and can be assisted by third-party navigation (if desired) | Identical |
| | Predicate Device<br>NICO BrainPath | Subject Device<br>NICO Brain Port |
| Shipping<br>Configuration | Obturator and sheath packaged and shipped<br>separately and paired during surgical case | Identical |
| Reusable or Single<br>Patient Use | Single Patient Use and Reusable | Identical |
| Method of<br>Sterilization / SAL | Disposables: Gamma (sheath/hooks) / 10-6<br>Reusables: Autoclave/hydrogen peroxide gas plasma | Identical |
| Biocompatibility | Demonstrated based on externally<br>communicating device in direct contact with<br>tissue/bone/dentin for a limited duration | Identical |
| Materials of<br>Manufacture | Obturator: Aluminum<br>Sheath: COC | Identical |
| Cross Sectional<br>analysis of<br>Obturator/ Sheath | Obturator/Sheath combination has a circular<br>cross section | Identical |
| Depth markings | Incremental depth markings on both sheath and<br>obturator. | Identical |
| Sheath inner<br>diameter (mm)<br>options | 13.5 mm | 13.5 mm and 11 mm |
| Sheath Lengths | 13.5 mm diameter: 50, 60, and 75 mm | 13.5 mm dia.: 50, 60, 75, and 95 mm<br>11 mm dia.: 50, 60, 75 mm |
| General Shape of<br>Obturator Tip | Distal end of obturator has a conical shape with a<br>rounded tip and no opening. | Identical |
| Obturator Tip | For the blue, standard tip obturators, the tip<br>extends 15 mm beyond the sheath. For the Gold-<br>ST, shallow-tip obturator, the tip extends 7.5 mm<br>beyond the sheath. | Identical, new offerings all have 15 mm tips |
| Third-Party<br>instrumentation | Obturator component interfaces with third-party<br>instruments. | Identical |
| Surface of Sheath | Inner diameter and horizontal proximal surface<br>of the knurled ring are textured. | Identical |
| Proximal End of<br>Sheath | Knurled ring and proximal portion of the sheath<br>has holes, slots, and tabs. | Updated knurled ring with concave geometry<br>to reduce optical glare, still includes holes,<br>slots and tabs. |
| Manipulation Tool | "Manipulation Tool” is textured to prevent<br>glare/shine from surgical lighting. | Identical |
| Shepherd's Hooks | Three Shepherd's Hooks available to facilitate<br>interfacing with common third-party retractors. | Identical |
| 95mm Stylet | None | For use with 13.5 mm x 95 mm Obturator as a |
#### Table 3: Comparison Table
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for NICO Corporation. The word "NICO" is written in large, bold, black letters. The "O" in NICO is a golden color. Underneath the word "NICO" is the word "CORPORATION" written in smaller, black letters.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for NICO Corporation. The letters "N," "I," and "C" are in black, and the "O" is in gold. Underneath the letters, the word "CORPORATION" is written in smaller, black letters.
## Non-Clinical Testing
The following tests were confirmed or repeated to demonstrate that the subject device modifications met applicable design and performance requirements, and supports a determination of substantial equivalence:
#### Table 4: Non-Clinical Testing
| Testing | Device(s) Supported | Result/Conclusion |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|-------------------|
| Cytotoxicity - MEM Elution: 72 hour incubation | All | Non-cytotoxic |
| Sensitization - Maximization (2 extracts) | | Non-sensitizer |
| Irritation - Intracutaneous Reactivity (2 extracts) | All | Non-irritant |
| Simulated Use to demonstrate the BrainPath has the ability to<br>interface with third-party Instruments and meets design input<br>requirements | All | Pass |
| Packaging & Shelf Life – shipping/distribution simulation,<br>environmental conditions, aging, visual packaging inspection,<br>bubble and seal strength packaging testing, and functional<br>testing following aging, environmental and shipping simulation | All | Pass |
| Specification Review | All | Pass |
| Cleaning Validation (Reusable Devices) – Establishment of<br>cleaning validation per miles soil test using bioburden<br>endotoxin and protein testing | Obturator, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray | Pass |
| Sterility Validation (Reusable Devices) – Steam autoclaving,<br>IUSS, and hydrogen peroxide gas plasma | Obturators, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray | Pass |
| Sterility Validation (Single-Use) – B&F testing, VDmax for<br>SAL 10-6, along with routine Endotoxin testing | BrainPath Sheath and<br>Shepherd's Hooks | Pass |
| Sterilization Tray Drop Test | Obturators, Manipulation<br>Tools, Stylet, and<br>Sterilization Tray | Pass |
### Conclusion
Risk assessment, biocompatibility, non-clinical testing, design validation, and compliance with recognized standards have demonstrated that the subject NICO BrainPath modifications do not raise different questions of safety or effectiveness when compared to the predicate. Therefore, the results of these analyses provide reasonable assurance that the NICO BrainPath has a similar safety and effectiveness profile as the predicate device and supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.