Dual Cure Resin Cement

K201790 · Rizhao Huge Biomaterials Company, Ltd. · EMA · Sep 29, 2020 · Dental

Device Facts

Record IDK201790
Device NameDual Cure Resin Cement
ApplicantRizhao Huge Biomaterials Company, Ltd.
Product CodeEMA · Dental
Decision DateSep 29, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3275
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Dual Cure Resin Cement is a dental luting system designed for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, metal-ceramic, all-ceramic and/or porcelain, composites, and their combinations.

Device Story

Dual Cure Resin Cement is a dental luting system used by clinicians for permanent cementation of indirect restorations. The device consists of a base paste and a catalyst paste containing methacrylate resins, inorganic fillers, pigments, and photo/chemical initiators. It is supplied in either two separate syringes for manual mixing or a double-barrel syringe for automixing. The material is dual-curing, utilizing both light-activation and chemical polymerization to set. Once mixed in a 1:1 ratio, the cement is applied to the restoration, which is then seated in the patient's mouth. The cement is radiopaque, allowing for radiographic verification of excess material. By providing a strong, durable bond between the restoration and the tooth structure, the device restores function and aesthetics to the patient's dentition.

Clinical Evidence

Bench testing only. No clinical data. Performance verified against ISO 4049:2009 (film thickness, working/setting time, flexural strength, water sorption/solubility, radio-opacity) and ISO 29022:2013 (shear bond strength). All tests met pass/fail criteria.

Technological Characteristics

Methacrylate-based resin matrix with silanated barium glass fillers (0.01-3 microns, 40% volume). Dual-cure (photo-initiation and chemical polymerization). Radiopaque. Delivery via hand-mix (separate syringes) or automix (double-barrel syringe). Conforms to ISO 4049:2009 and ISO 29022:2013.

Indications for Use

Indicated for cementation of dental restorations (crowns, bridges, inlays/onlays, veneers, posts) made of metal/alloys, metal-ceramic, all-ceramic/porcelain, or composites in patients requiring dental luting.

Regulatory Classification

Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services logo. The text on the right is in a blue sans-serif font, with "FDA" in a larger font size than the rest of the text. September 29, 2020 Maggie Zheng, Regulatory Affairs Manager Rizhao HuGe Biomaterials Company, Ltd. No.2 North Zhaoyang Road, District of Donggang Rizhao City. CHINA 276800 Shandong Province ### Re: K201790 Trade/Device Name: Dual Cure Resin Cement Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: Class II Product Code: EMA, EBF Dated: July 17, 2020 Received: July 27, 2020 ### Dear Maggie Zheng: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting- combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use 510(k) Number (if known) K201790 Device Name Dual Cure Resin Cement ### Indications for Use (Describe) Dual Cure Resin Cement is a dental luting system designed for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, metalceramic, all-ceramic and/or porcelain, composites, and their combinations. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: underline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) | |-----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | <div style="display:inline-block; border: 1px solid black; width: 13px; height: 13px;">\u2713</div> | <div style="display:inline-block; border: 1px solid black; width: 13px; height: 13px;"></div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. {3}------------------------------------------------ ## K201790 ## 510 (k) Summary This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92. - 1. Date Summary Prepared: September 29, 2020 ### 2. Submitter Information: | Owner's Name | Rizhao HuGe Biomaterials Company, Ltd. | |----------------|-------------------------------------------------------------------------------------------------| | Address | No.2 North Zhaoyang Road, District of Donggang, Rizhao City,<br>Shandong Province, China 276800 | | Telephone | 0086 633 2277268 | | Fax | 0086 633 2277298 | | Contact Person | Ms. Maggie Zheng | | Contact Title | Regulatory Affairs Manager | | E-mail | zhengxy@hugedent.com | ### 3. Device Name Trade name: Dual Cure Resin Cement Common name: Dual Cure Resin Cement Classification name: Dental Cement (21 CFR 872.3275) Regulatory Class: II Product Code: EMA, EBF ### 4. Predicate Device Information | Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate | |---------------------|-------------------|-------------|--------------|-----------| | IVOCLAR VIVADENT AG | Multilink Automix | K123397 | EBF | Primary | This predicate device has not been subject to a design-related recall. No reference devices were used in this submission. ### 5. Description of Device {4}------------------------------------------------ Dual Cure Resin Cement is a versatile and ultimate dental dual cure (chemical and/or light cure) luting cement. This multi-purpose device is designed for cementation of all sorts of dental restorations made from materials of metal-ceramic, all-ceramic and/or porcelain, composites, and their combinations. Dual Cure Resin Cement contains paste-paste of Base and Catalyst comprising primarily of dental methacrylate resins and inorganic fillers (with particle size range from about 0.01 - 3 microns and a filler volume of about 40%), inorganic pigments, photo-initiation and chemical curing systems. The mixing ratio, based on volume, is 1 part base paste : 1 part catalyst paste. Dual Cure Resin Cement is delivered in either two seperated single syringe for Hand-mix and double-barrel syringes for Automix and comes in seven shades: TR (Translucent), A0 (Light+), A 1 (Light), A2 (Medium), A3 (Dark), White (White Opaque) and Opaque (Universal Opaque). Dual Cure Resin Cement is radiopaque, allowing for easy identification on radiographs. ### 6. Indications for Use Dual Cure Resin Cement is a dental luting system designed for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, metal-ceramic and/or porcelain, composites, and their combinations. ### 7. Summary of Physical and Chemical Properties Tests Dual Cure Resin Cement is classified as Type 2 Class 3 in accordance with ISO 4049, which belongs to luting materials that are cured by the application of external energy and also have a self-curing mechanism present ("dual cure" materials). Test standards and methods based on ISO 4049:2009 and ISO 29022:2013: | ISO 4049: 2009 Dentistry - Polymer-Based Restorative Materials | |-----------------------------------------------------------------------------| | ISO 29022:2013 Dentistry - Adhesive - Notched-edge shear bond strength test | | Table 4: Summary of Physical and Chemical Properties Test | | | |-------------------------------------------------------------|----------------------------------------------------------------------------------------|--------------| | Items per<br>ISO 4049: 2009/<br>ISO 29022:2013 | Pass/fail criteria | Conclusion | | ISO 4049: 2009<br>5.2.2 Film thickness, luting<br>materials | The film thickness of luting materials in any event shall be<br>no greater than 50 µm. | Satisfactory | {5}------------------------------------------------ # IHH | ISO 4049: 2009<br>5.2.4 Working time, Class 1<br>and Class 3 luting materials | The material shall be capable of forming a thin layer;<br>during its formation there shall be no detectable change in<br>its homogeneity, shall be not less than 60 s. | Satisfactory | |-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------| | ISO 4049: 2009<br>5.2.6 Setting time, Class 3<br>materials | The setting time shall be not more than 10 min. | Satisfactory | | ISO 4049: 2009<br>5.2.9 Flexural strength | The flexural strength shall be equal to or greater than 50<br>MPa. | Satisfactory | | ISO 4049: 2009<br>5.2.10 Water sorption and<br>solubility | The water sorption shall be ≤ 40 µg/mm3.<br>The solubility of shall be ≤ 7.5 µg/mm3. | Satisfactory | | ISO 4049: 2009<br>5.5 Radio-opacity | The radio-opacity shall be equal to or greater than that of<br>the same thickness of aluminium (1 mm of material). | Satisfactory | | Shear bond strength<br>(The test method equal to the<br>ISO 29022:2013) | Shear bond strength for enamel and dentin≥8 MPa<br>(Internal standard) | Satisfactory | ### 8. Technological Characteristics All components of the subject device are based upon industry well-known chemistry. The curing mechanism of the subject device and predicate devices are all of polymerization of uncured methacrylate ester monomers. The reaction is caused by photo initiator and chemical polymerization initiator systems. The following table shows the significant technological characteristics for the subject device and indicates the following similarities and differences with the predicate devices: | Table 5: Technological Characteristics Comparison Table | | | |---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Technological<br>Characteristics | Subject device<br>(Dual Cure Resin Cement) | Primary predicate | | Monomer matrix | Methacrylate based | Methacrylate based | | Primary Filler(s) | Silanated Barium glass filler | Silanated Barium glass filler | | Physical Form | Pastes of Catalyst and Base | Pastes of Catalyst and Base | | Indications of Use | Dual Cure Resin Cement is a dental luting<br>system designed for cementation of all<br>sorts of dental restorations including<br>crowns, bridges, inlays/onlays, veneers,<br>dental posts and other restorations made<br>from materials of metals/alloys,<br>metal-ceramic, all-ceramic and/or<br>porcelain, composites, and their<br>combinations. | Multilink Automix is used for the permanent<br>cementation of indirect restorations where a<br>strong bond us desired:<br>- Inlays, onlays, crowns, bridges and root<br>posts made of:<br>● Metal and metal ceramics<br>● All-ceramics, in particular opaque<br>zirconium oxide ceramics<br>● Composites and fiber-reinforced<br>composites | | Prescription/over-<br>the-counter use | Prescription | Prescription | {6}------------------------------------------------ | Table 5: Technological Characteristics Comparison Table | | | |---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------| | Technological<br>Characteristics | Subject device<br>(Dual Cure Resin Cement) | Primary predicate | | Curing method | Dual cure | Dual cure | | Delivery form | Hand-mix / Automix | Automix | | Radiographic<br>Appearance | Radiopaque | Radiopaque | | Physical Properties | The subject device and the predicate devices have substantially equivalent physical<br>properties as they all conform to the specifications set by ISO 4049 and ISO 29022. | | The subject device has the same technological characteristics as the predicate devices with the exception of the additional hand-mixing option. As compared to the predicated devices, the subject device is also available in Hand-mix, which is having the Catalyst and Base resin cement pastes packaged into physically separated single syringes instead of being packaged into a double barrel syringe as for the Automix. Regardless of which mode of operation by Hand-mix or Automix, the base and catalyst are mixed in a 1:1 ratio to form dual-curing cement, and it will not pose any new issues of the safety and effectiveness. Therefore, the minor differences in the delivery system affect neither the intended use nor substantial equivalence. ### 9. Summary of Biocompatibility The new device, Dual Cure Resin Cement, is substantially equivalent to the predicate devices that have been legally marketed for decades and with no clinical adverse events. The formulation of new device does not contain any non-conventional chemicals compared to the legally marketed predicate device. We selected our Dual Cure Resin Cement [Hand-mix, shade: Opague (Universal Opaque)] as the representative model in biocompatibility tests because it is the worst case scenario. In addition, the Hand-mix and Automix of our Dual Cure Resin Cement have the same chemical compositions, raw material suppliers, curing method and other technological characteristic. Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Subchronic Toxicity and Genotoxicity. ### 10. Clinical Performance Data Not applicable. Clinical performance testing has not been performed for the subject device. ### 11. Conclusions Based on the indications for use, technological characteristics, performance testing and {7}------------------------------------------------ # HUGE comparison to predicate devices, the subject device has been shown to be safe and effective for its intended use and the minor differences in indications for use fall within the intended use of the predicate devices affecting neither the general intended use nor substantial equivalence. Rizhao HuGe Biomaterials Company, Ltd. concludes that the subject device is substantially equivalent to the predicate devices described herein.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...