K123397 · Ivoclar Vivadent, AG · EBF · Jan 31, 2013 · Dental
Device Facts
Record ID
K123397
Device Name
MULTILINK AUTOMIX
Applicant
Ivoclar Vivadent, AG
Product Code
EBF · Dental
Decision Date
Jan 31, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Indications for Use
Multilink Automix is used for the permanent cementation of indirect restorations where a strong bond is desired: - Inlays, onlays, crowns, bridges and root posts made of: – - Metal and metal ceramics - All-ceramics, in particular opaque zirconium oxide ceramics - Composites and fibre-reinforced composites
Device Story
Multilink Automix is a self-curing and self-etching luting composite system for permanent cementation of indirect dental restorations. Used in dental clinics by professionals; applied with Multilink Primer; Monobond Plus recommended as coupling agent for bonding to alloys, zirconium, aluminum oxide, and silicate ceramics. Device provides mechanical fixation and chemical bonding to tooth structure and restorative materials. Benefits include strong, durable retention of indirect restorations. Performance is verified through physical property testing (flexural strength, bond strength, radiopacity) to ensure clinical suitability.
Clinical Evidence
Bench testing only. Evaluated per ISO 4049:2000 for water absorption, water solubility, radiopacity, flexural strength, modulus of elasticity, compressive strength, and shear bond strength. Biocompatibility evaluated per ISO 10993. Results demonstrate performance comparable to the predicate device.
Technological Characteristics
Self-curing/self-etching luting composite system. Materials tested per ISO 4049:2000. Properties include radiopacity (350% Al), flexural strength (70-110 MPa), and compressive strength (240-280 MPa). Delivery via automix system. Biocompatibility per ISO 10993.
Indications for Use
Indicated for permanent cementation of indirect restorations including inlays, onlays, crowns, bridges, and root posts composed of metal, metal-ceramics, all-ceramics (specifically opaque zirconium oxide), composites, and fiber-reinforced composites.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
**MULTILINK AUTOMIX**
JAN 3 1 2013
K | 23397
Donna Marie Hartnett Contact:
Ivoclar Vivadent, Inc. 175 Pineview Drive, Amherst, NY 14228 Company: (716) 691-0010
December 4, 2012 Date Prepared:
Multilink Automix Proprietary Name:
Material, Tooth Shade, Resin (872.3690) Classification Name:
Predicate Devices: Multilink (K032470)
Multilink Automix is a self-curing and self-etching luting Device Description: composite system for the adhesive luting of indirect restorations made of metalceramic all-ceramic and composite. Multilink Automix is applied together with the selfetching and self-Ocuring Multilink Primer. Monobond Plus is recommended as a coupling agent to achieve a strong bond to precious and non-precious alloys, as well as to all ceramics made of zirconium and aluminum oxide and silicate ceramics.
The predicate device to which Multilink Automix has been compared is Multilink (K032470). For this application. Multilink Automix has been compared to its predicate with regard to chemical composition, performance data and indications for use. The comparison shows that Multilink Automix is substantially equivalent to the predicate device.
## Intended Use:
Multilink Automix is used for the permanent cementation of indirect restorations where a strong bond is desired:
- Inlays, onlays, crowns, bridges and root posts made of: –
- o Metal and metal ceramics
- o All-ceramics, in particular opaque zirconium oxide ceramics
- o Composites and fibre-reinforced composites
Technological Characteristics: The device design, i.e. delivery form, and intended use of Multilink Automix and the predicate device are the same. The composition of the subject device has been modified from the predicate, however, there are no ingredients in the subject device which pose any new issues of safety and effectiveness.
Testing Summary: The device was tested in accordance with ISO 4049:2000 for Polymer based dental restorative materials for Water Absorption, Water solubility, radiopacity, Flexural Strength, Modulus of Elasticity, Compressive Strength and Shear Bond Strength and the results from testing demonstrates that Multilink Automix is substantially equivalent to the predicate device. Biocompatibility testing and evaluation was also carried out according to ISO 10993.
510K Summary 5-1
{1}------------------------------------------------
## 510(K) SUMMARY MULTILINK AUTOMIX
The following table shows the performance data of Multilink Automix and its predicate device: . .
・
.
.
| Property | UOM | Multilink<br>(K032470) | Multilink Automix | Multilink Automix |
|--------------------------------------------------------------------------|--------|---------------------------|---------------------------|---------------------------|
| | | | Self Curing | Dual Curing |
| Working time (37C) | Min | 3-4 | 3-4 | -- |
| Setting time | Min | 7-9 | 6-7 | -- |
| Film Thickness | μm | <20 | <50 | <50 |
| Water Absorption | µg/mm³ | <25.0 | <40 | <40 |
| Water solubility | µg/mm³ | <3.0 | <7.5 | <7.5 |
| Radiopacity | %Al | 350 | 350 | 350 |
| Flexural Strength | MPa | 90±10 | 70±20 | 110±10 |
| Modulus of Elasticity | MPa | 5000±1000 | 3250±400 | 6000±400 |
| Compressive<br>Strength | MPa | 250±20 | 240±20 | 280±20 |
| Transparency<br>Base (trans and Cat)<br>Yellow and Cat<br>Opaque and Cat | % | 12±1.5<br>10±1.5<br>2±0.5 | 12±1.5<br>10±1.5<br>2±0.5 | 12±1.5<br>10±1.5<br>2±0.5 |
| Vickers Hardness | MPa | 370 | -- | 440±30 |
| Shear Bond Strength<br>Dentin / 24h<br>Enamel/ 24hh | MPa | 21±2<br>23+4 | 17±5<br>18+3 | 21±2<br>23+4 |
.
:
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or feathers.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 31, 2013
Ivoclar Vivadent. AG C/O Ms. Donna Marie Hartnett Director of Quality Assurance / Regulatory Affairs Ivoclar Vivadent, Incorporated 175 Pineview Drive AMHEREST NY 14228
Re: K123397
Trade/Device Name: Multilink Automix Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: October 31, 2012 Received: November 5, 2012
Dear Ms. Hartnett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Hartnett
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 80}), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
hor
Ret
Harshita.D.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
510(k) Number (if known): K123397
Device Name: Multilink Automix
Indications For Use:
-
Multilink Automix is used for the permanent cementation of indirect restorations where a strong bond is desired:
Inlays, onlays, crowns, bridges and root posts made of:
- Metal and metal ceramics 0
- All-ceramics, in particular opaque zirconium oxide ceramics 0
- Composites and fibre-reinforced composites 0
Prescription Use 区 AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
2013.01.30 Susan Runner DDS, MA -15:42:28 -05'00'
(Division Sign-Off)
Division of Anasthesiology, ଭୌଗିନ୍ୟାପଣ୍ଡେଲ୍ସନ୍ of CDRH, Office of Device Evaluation (ODE) Infection Control, Dental Devices
| 510/k) Number: | n of | 20 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
|----------------|------|----|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | |
Page 1 of __ 1__
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.