K201672 · Covidien · GDW · Jul 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201672
Device Name
Signia Stapler (with new software)
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Jul 17, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas. The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2,0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue. The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection of the pancreas. The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyclic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloads preloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection of tissue and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
Device Story
Battery-powered, microprocessor-controlled surgical stapler; provides powered maneuverability and firing of compatible stapling reloads. Input: force measurements during tissue clamping and firing. Processing: Adaptive Stapling Algorithm (ASA) adjusts firing speed based on measured force to maintain predetermined safety limits and optimize staple formation. Output: real-time force gauge display, LED indicators, OLED display, and audible tones. Used in OR by qualified medical professionals. Modification: software update extends real-time force gauge display to all compatible reloads (previously limited to intelligent reloads with ID chips). Device assists surgeons in achieving consistent staple formation; benefits include optimized tissue handling and safety via automated force management.
Clinical Evidence
No clinical study performed. Substantial equivalence demonstrated via non-clinical performance data and software verification testing.
Technological Characteristics
Battery-powered, microprocessor-controlled surgical stapler. Components: Power Handle, Power Shell, Linear Adapter. Features: LED indicators, OLED display, audible tones, push-button controls for firing, articulation, and rotation. Software includes Adaptive Stapling Algorithm (ASA) for tissue sensing and force management. Connectivity: None (standalone).
Indications for Use
Indicated for patients undergoing abdominal, gynecological, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis; includes liver, hepatic vasculature, biliary structures, and pancreas procedures; also indicated for blunt tissue dissection and deep pelvic procedures (e.g., low anterior resection).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 17, 2020
Covidien Katherine Choi Principal Regulatory Affairs Specialist 60 Middletown Avenue North Haven, Connecticut 06473
Re: K201672
Trade/Device Name: Signia Stapler (with new software) Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW Dated: June 18, 2020 Received: June 19, 2020
## Dear Katherine Choi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
#### 510(k) Number (if known)
#### K201672
Device Name SigniaTM Stapler (with new software)
#### Indications for Use (Describe)
The Siguia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIAM single-use reloads with Tri-Staple™ Technology. Tri-Staple™ 2.0 single-use reloads and Sigma™ loading with Tri-Staple™ 2.0 single-use cartidges. has applications in abdominal, gynecological, pediatic, and thoracic surgery for resection, and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature, and biliary stuctures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The Signia TM Stabler, when used with single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement or Tri Staple™ 2.0 single use reinforced reloads preloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
#### DATE PREPARED:
June 18, 2020
#### SUBMITTER:
Covidien 60 Middletown Avenue North Haven, CT 06473 USA
#### CONTACT PERSON:
Katherine Y. Choi, RAC Principal Requlatory Affairs Specialist Telephone: (203) 492-8412 Fax: (203) 492-5029
## IDENTIFICATION OF DEVICE:
| Proprietary/Trade Name: | SigniaTM Stapler with new software |
|-------------------------|------------------------------------|
| Classification Name: | Staples, Implantable |
| Regulation Number: | 21 CFR 878.4750 |
| Product Code: | GDW |
| Device Class: | Class II |
| Review Panel: | General and Plastic Surgery |
| Common Name: | Surgical Stapler |
## PREDICATE DEVICE:
| Proprietary/Trade Name: Signia™ Stapler | |
|-----------------------------------------|-----------------------------|
| 510(k) Number: | K160176 (April 26, 2016) |
| Classification Name: | Staples, Implantable |
| Regulation Number: | 21 CFR 878.4750 |
| Product Code: | GDW |
| Device Class: | Class II |
| Review Panel: | General and Plastic Surgery |
| Common Name: | Surgical Stapler |
## DEVICE DESCRIPTION:
The Signia™ Stapler is a battery powered microprocessor controlled surgical stapler that provides pushbutton powered maneuverability and firing of compatible Covidien stapling reloads. The Signia™ Stapler when used with a compatible reload is a surgical device for stapling and cutting tissues. The Signia™ Stapler is intended to be used by medical professionals qualified in the transportation, preparation, cleaning, sterilization, and use of surgical devices. All of which stay unchanged when compared to the predicate device.
The Signia™ Stapler is composed of the Signia™ Power Handle, Signia™ Power Shell, and Signia™ Linear Adapter. System accessories include the Signia™ Reusable Insertion Guide, Signia™ Manual Retraction Tool, Signia™ Single Bay Charger, Signia™ Sterilization Tray (optional), and Signia™ Four-Bay Smart Charger (optional).
The design modification is to introduce the new software version for the Signia™ Power Handle. The new software will extend the real time force gauge display on the Signia™ Power Handle
{4}------------------------------------------------
for the non-intelligent reloads that are already in the market without an ID chip. Currently, the force gauge is displayed only when an intelligent reload with an ID chip is attached to the Signia™ Stapler. By installing the new software, the force gauge will be displayed consistently for all compatible Covidien reloads. This is a user interface/display change but does not impact how the handle operates.
#### INTENDED USE/INDICATIONS FOR USE:
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GJA™ curved tip single use reloads or Tri-Staple™ 2,0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection of liver substance, hepatic vasculature and biliary structures and for transection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyclic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloads preloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection of tissue and creation of anastomosis. It may be used for transection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:
The Signia™ Stapler with new software does not change the fundamental operating principle and mechanism of action when compared to the predicate device. The Signia™ Stapler remains as a powered surgical stapler operated by the built-in battery and microprocessor with software. The user controls the Signia™ Stapler via button presses on the Handle and receives feedback status and other information via its LED indicators, OLED display, and audible tones. Push buttons on the handle are provided for stabling stapling and activation a (firing/clamping/unclamping), articulation (left/right), and rotation (clockwise/counterclockwise).
The Signia™ Stapler includes adaptive stapling algorithm (ASA) in its software, which is often referred as adaptive firing technology or tissue sensing technology. This feature measures force when clamping and firing stables during use and will prevent the force from exceeding predetermined safety limits. As it enters three specially developed, predetermined force zones, the Signia handle will adjust its firing speed in order to maintain forces within the lowest zone possible and optimize staple formation. Based on the initial force measured during clamping, the device will program the initial appropriate speed selection. There are three speed settings set from clamp force: Zone 2 Medium, and Zone 3 Slow.
## SUBSTANTIAL EQUIVALENCE:
The Signia™ Stapler with the new software is substantially equivalent to the legally marketed Signia™ Stapler (K160176) since extending the existing real time force gauge display to cover all Covidien reloads through software change does not alter the intended use, indications, or user environments of the device. Applicable design control activities to ensure the Signia™ Stapler with new software functions as
{5}------------------------------------------------
intended have been completed without raising different types of questions in terms of safety and effectiveness when compared to the predicate device.
## SUMMARY OF STUDIES:
Non-clinical performance data such as software verification and performance testing have demonstrated substantial equivalence to the predicate device.
Clinical performance data – No clinical study has been performed. The substantial equivalence has been demonstrated by non-clinical studies.
#### CONCLUSION:
Based upon the supporting data summarized above, we concluded that Signia™ Stapler with the new software is as safe and effective as the legally marketed K160176 and does not raise different questions of safety and effectiveness than the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.