K160176 · Covidien · GDW · Apr 26, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160176
Device Name
Signia Stapler
Applicant
Covidien
Product Code
GDW · General, Plastic Surgery
Decision Date
Apr 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Signia™ stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas. The Signia™ stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue. The Signia™ stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of the pancreas. The Signia™ stapler. when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloads preloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
Device Story
Battery-powered, microprocessor-controlled surgical stapler; provides powered maneuverability and firing of compatible stapling reloads. System includes power handle, power shell, linear adapter, insertion guide, manual retraction tool, and charger. User controls device via handle buttons for stapling activation (clamping/firing), articulation, and rotation. Handle provides feedback via OLED display, LED indicator, and audible annunciator. Used in OR by qualified medical professionals. Motors with built-in encoders provide shaft position data; software manages motor speed and position based on encoder pulses. Device facilitates tissue resection, transection, and anastomosis; benefits include powered control during complex surgical procedures.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing and animal studies. Testing included lifecycle reliability, cleaning validation, aseptic barrier, functional performance, biocompatibility, electrical safety, and EMC/EMI.
Technological Characteristics
Battery-powered, microprocessor-controlled handheld surgical stapler. Features OLED display, LED indicators, and audible annunciators. Motors with integrated encoders for position sensing. Compatible with various Endo GIA and Tri-Staple reloads. Biocompatibility per ISO 10993-1; electrical safety per IEC 60601-1; EMC/EMI per IEC 60601-1-2.
Indications for Use
Indicated for patients undergoing abdominal, gynecological, pediatric, or thoracic surgery requiring tissue resection, transection, or anastomosis; includes liver, hepatic vasculature, biliary structures, and pancreas procedures. Also indicated for blunt tissue dissection and deep pelvic procedures (e.g., low anterior resection).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
iDrive™ Ultra Powered Handle with Endo GIA™ Adapter (K121510)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, layered on top of each other. These profiles are connected to three flowing shapes below, resembling fabric or ribbons. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 26, 2016
Covidien Mr. Frank Gianelli Sr. Product Specialist Regulatory Affairs 60 Middletown Avenue North Haven, Connecticut 06471
Re: K160176
Trade/Device Name: Signia Stapler Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: January 22, 2016 Received: January 27, 2016
Dear Mr. Gianelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Food and Drug Administration | | Form Approved: OMB No. 0910-0120 |
| Indications for Use | | Expiration Date: January 31, 2017 |
| See PRA Statement below. | | |
| 510(k) Number (if known) | K160176 | |
| Device Name | Signia™ Stapler | |
| Indications for Use (Describe) | The Signia™ stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™<br>Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges,<br>has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of<br>anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and<br>for transection and resection of the pancreas. | |
| | The Signia™ stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use<br>reloads, can be used to blunt dissect or separate target tissue from other certain tissue. | |
| | The Signia™ stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has<br>applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection<br>and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection.<br>It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for<br>transection and resection of the pancreas. | |
| | The Signia™ stapler. when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded<br>with polyglycolic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloads preloaded with<br>polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for<br>resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance,<br>hepatic vasculature and biliary structures, and for transection and resection of the pancreas. | |
| Type of Use (Select one or both, as applicable) | <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| | *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: | |
| | Department of Health and Human Services | |
| | Food and Drug Administration | |
| | Office of Chief Information Officer | |
| | Paperwork Reduction Act (PRA) Staff | |
| | PRAStaff@fda.hhs.gov | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of | |
| | information unless it displays a currently valid OMB number." | |
| FORM FDA 3881 (8/14) | Page 1 of 1 | <i>PSC Publishing Services (301) 443-6740</i> |
| Covidien<br>Traditional 510(k) | Signia™ Stapler | January 22, 2016<br>Page 56 |
址
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
### Date Prepared:
January 22, 2016
#### Submitter:
Covidien 60 Middletown Avenue North Haven, CT 06473
#### Contact:
Frank Gianelli Senior Requlatory Affairs Product Specialist Telephone: (203) 492-5352 Fax: (203) 492-5029 frank.j.gianelli@medtronic.com Email:
#### Name of Device:
| Trade/Proprietary Name: | Signia™ Stapler |
|-------------------------|-------------------------------------------|
| Common Name: | Surgical Stapler with Implantable Staples |
| Classification Name: | Staples, Implantable |
| | a. Panel no and product code: 79 GDW |
| | b. Regulation no: 21 CFR 878.4750 |
#### Predicate Device:
| Trade/Proprietary Name: | iDrive™ Ultra Powered Handle with Endo GIA™ Adapter |
|-------------------------|-----------------------------------------------------|
| Common Name: | Surgical Stapler with Implantable Staples |
| Classification Name: | Staples, Implantable, (79 GDW, 21 CFR 878.4750) |
| 510(k) Number: | K121510 |
| Manufacturer: | Covidien |
## Device Description:
The Signia™ Stapler is a battery powered microprocessor controlled surgical stapler that provides push-button powered maneuverability and firing of compatible Covidien stapling reloads. The Signia™ Stapler is composed of the Signia™ Power Handle, Signia™ Power Shell, and Signia™ Linear Adapter. System accessories include the Signia™ Reusable Insertion Guide, Signia™ Manual Retraction Tool, and Signia™ Single Bay Charger.
The User controls the Signia™ Stapler via button presses on the Handle and receives feedback status and other information from the Handle via its OLED Display, LED indicator and audible annunciator. Pushbutton / Toggle switches on the Handle are provided for:
- Stapling activation: Advancing the Firing Rod / Clamping / Unclamping .
- Left and Right Articulation .
- Clockwise and Counter Clockwise Rotation .
When the appropriate buttons are pressed, the Signia™ Stapler software initiates stapling (firing), articulation or rotation operation by turning on the appropriate motor. The motors have built-in encoders which provide shaft position information. The software controls motor speed and position using encoder pulses. Since it is possible to press multiple keys simultaneously, software must be able to process them and take appropriate actions.
The Signia™ Stapier is compatible with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single use reloads, and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges.
{4}------------------------------------------------
The Signia™ Stapler when used with the abovementioned family of Endo GIA™ reloads is a surgical device for stapling and cutting tissues. The Signia™ Stapler is intended to be used by medical professionals qualified in the transportation, cleaning, sterilization, and use of surgical devices.
#### Intended Use:
The Signia™ Stapler, when used with Endo GIA™ single-use reloads, Endo GIA™ single-use reloads with Tri-Staple™ Technology, Tri-Staple™ 2.0 single-use reloads and Signia™ loading units with Tri-Staple™ 2.0 single-use cartridges, has applications in abdominal, gynecological, pediatric, and thoracic surgery for resection, transection, and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature, and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ curved tip single use reloads or Tri-Staple™ 2.0 curved tip single-use reloads, can be used to blunt dissect or separate target tissue from other certain tissue.
The Signia™ Stapler, when used with Endo GIA™ single use Radial Reloads with Tri-Staple™ Technology, has applications in open or minimally invasive general abdominal, gynecologic, pediatric and thoracic surgery for resection and transection of tissue and creation of anastomosis, as well as application deep in the pelvis, i.e., low anterior resection. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures and for transection and resection of the pancreas.
The Signia™ Stapler, when used with Endo GIA™ single use reinforced reloads with Tri-Staple™ Technology preloaded with polyglycolic acid staple line reinforcement or Tri-Staple™ 2.0 single use reinforced reloaded with polyglycolic acid staple line reinforcement, has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transection and resection of the pancreas.
- Note: The Signia™ Stapler is intended for use with Endo GIA™ Reloads, Endo GIA™ Reloads with Tri-Staple™ Technology, Signia™ Loading Units with Tri-Staple™ 2.0 cartridges and Tri-Staple™ 2.0 Reloads and does not carry a separate indication from these stapling devices.
## Technological and Performance Characteristics:
The Signia™ Stapler is substantially equivalent to the predicate iDrive Ultra™ Powered Handle with Endo GIA™ Adapter in regard to the stapling technologies employed.
Qualitative and quantitative data were obtained and used to compare the Signia™ Stapler to the predicate iDrive Ultra™ Powered Handle with Endo GIA™ Adapter.
The design differences were found to not affect safety or performance through applicable design verification activities that showed conformance to applicable technical design specifications and performance requirements, applicable medical device performance standards, and other nonclinical testing.
# Tests performed to evaluate and compare technological and performance characteristics:
- 1. Lifecvcle Reliability Tests
- 2. Cleaning Validation Tests
- 3. Aseptic Barrier Tests
- 4. Functional Performance Test Bench and Animal
- 5. Usability Tests
- Biocompatibility tests in accordance with ISO Standard 10993-1 6.
{5}------------------------------------------------
- 7. Electrical Safety Tests per IEC 60601-1
- 8. EMC/EMI Tests per IEC 60601-1-2
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
# Conclusion:
Through the comparison of technological and performance characteristics and the results of rinough the Significancer vas found to be substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.