K201620 · 3B Medical, Inc. · BZD · Jan 29, 2021 · Anesthesiology
Device Facts
Record ID
K201620
Device Name
Luna® G3 BPAP 25A
Applicant
3B Medical, Inc.
Product Code
BZD · Anesthesiology
Decision Date
Jan 29, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5905
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Luna® G3 BPAP 25A is a Bi-level PAP (Bi-level Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA). The integrated humidifier is indicated for the humidification and warming of air from the flow generator device. These devices are intended for single-patient use by prescription in the home or hospital/institutional environment on adult patients. It is to be used on patients >66lbs/30kg for whom CPAP therapy has been prescribed. The system can deliver bi-level therapy or auto bi-level therapy.
Device Story
Microprocessor-controlled, blower-based system generating bi-level positive airway pressure (IPAP/EPAP) to treat obstructive sleep apnea; delivers fixed or auto-adjusting pressure (4-25 cmH2O). Inputs include patient respiratory signals; device processes these to detect respiratory events and adjust pressure. Output is pressurized air delivered to patient's oral/nasal airway. Used in home or hospital by adult patients; healthcare providers manage therapy via integrated cellular module (4G) and iCodeConnect software. Features integrated humidifier and optional heated tubing to reduce condensation. Benefits include effective OSA treatment, data-driven therapy management, and improved patient comfort via humidification and auto-adjusting pressure modes.
Clinical Evidence
Bench testing only. Performance validated via biocompatibility (ISO 10993, ISO 18562), electrical safety/EMC (IEC 60601 series), and mechanical testing (shock, vibration, drop, leakage). Software verification and validation conducted per FDA guidance.
Technological Characteristics
Microprocessor-controlled blower system; Class II, Type BF. Pressure range 4-25 cmH2O. Integrated humidifier; optional heated tubing. Connectivity via 4G cellular module for data upload. Standards: ISO 80601-2-70, ISO 80601-2-74, IEC 60601-1, IEC 60601-1-2, ISO 10993, ISO 18562. IP22 water ingress protection.
Indications for Use
Indicated for adult patients >30kg (66lbs) with Obstructive Sleep Apnea (OSA) prescribed CPAP therapy; for use in home or hospital/institutional environments.
Regulatory Classification
Identification
A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 29, 2021
3B Medical, Inc. Yasser Estafanous Director of RA/OA 203 Avenue A NW. Suite 300 Winter Haven, Florida 33881
Re: K201620
Trade/Device Name: Luna® G3 BPAP 25A Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous ventilator (IPPB) Regulatory Class: Class II Product Code: BZD Dated: December 17, 2020 Received: December 18, 2020
Dear Yasser Estafanous:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Rachana Visaria, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201620
Device Name Luna® G3 BPAP 25A
#### Indications for Use (Describe)
The Luna® G3 BPAP 25A is a Bi-level Positive Airway Pressure) device designed for the treatment of adult Obstructive Sleep Apnea (OSA). The integrated humidification and warming of air from the flow generator device. These devices are intended for single-patient use by prescription in the home or hospital/ institutional environment on adult patients. If is to be used on patients >661bs/30kg for whom CPAP therapy has been prescribed. The system can deliver bi-level therapy or auto bi-level therapy.
Type of Use (Select one or both, as applicable)
| <div> <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|-----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# 510(k) Summary
| Device Trade Name | Luna® G3 BPAP 25A |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Model | LG3700 |
| Common/Usual Name | BPAP System |
| Date Prepared | January 29, 2021 |
| Sponsor Identification | 3B Medical, Inc.<br>203 Avenue A NW, Suite 300, |
| Submission Correspondent | Yasser Estafanous<br>3B Medical, Inc.<br>203 Avenue A NW, Suite 300,<br>Winter Haven, FL 33881 |
| Phone | 863-226-6285 |
| Fax | 863-226-6284 |
| Email | yestafanous@3bproducts.com |
| Establishment Registration # | 3008566132 |
| | BMC Medical CO., LTD.<br>Room 901, Building 1, No.28 Pingguoyuan<br>Road, Shijingshan, Beijing 100041, CHINA |
| Classification | Class II Device (21 CFR 868.5905) |
| Classification Name | Noncontinuous ventilator |
| Classification Panel | Medical Device |
| Products Code | BZD |
| Medical Specialties | Anesthesiology |
| Predicate Device(s) | RESmart® BPAP 25A (K133769) |
| | DreamStation Auto BiPAP (K131982) |
| Reference Device | Luna® CPAP and Auto CPAP System<br>(K153387) |
| Reason for Submission: | New Device |
| Intended Use | The Luna® G3 BPAP 25A is a Bi-level PAP (Bi-<br>level Positive Airway Pressure) device designed<br>for the treatment of adult Obstructive Sleep Apnea<br>(OSA). The integrated humidifier is indicated for<br>the humidification and warming of air from the<br>flow generator device. These devices are intended<br>for single-patient use by prescription in the home<br>or hospital/institutional environment on adult<br>patients. It is to be used on patients >66lbs/30kg<br>for whom CPAP therapy has been prescribed. The<br>system can deliver bi-level therapy or auto bi-level<br>therapy. |
| Device Description | The Luna® G3 BPAP 25A is a microprocessor<br>controlled, blower-based system that generates bi-<br>level positive airway pressure (IPAP/EPAP) to<br>support treatment of obstructive sleep apnea. The<br>system provides fixed or auto-adjust pressure from |
{4}------------------------------------------------
4 to 25 cmH2O above the ambient atmospheric
pressure to a patient's oral/nasal airway.
Comparison of Technological Characteristics with the Predicate Devices.
The Luna® G3 BPAP 25A utilizes the same blower and the same algorithm as the primary predicate device RESmart® BPAP 25A (K133769) for respiratory event detection and therapy for sleep disordered breathing events. The subject device and the primary predicate device share the same intended use, same operating principal, and are manufactured and packaged with similar processes. The basic functionality and performance characteristics of the subject device are the same as
{5}------------------------------------------------
the primary predicate device. Compared with the primary predicate device, the subject device includes the following main modifications:
- The humidifier is integrated into the BPAP device, which makes the physical size of the subject device smaller and does not affect the safety or effectiveness of the subject device.
- An optional heated tubing (LH1) is intended to provide warmed and/or ● humidified breathing gases before entering the patient's airway. The purpose of the heated tubing is to maintain or raise the gas temperature to or above the dew point (of the air exiting the humidifier) reducing or eliminating water condensation and/or pooling of water in the breathing circuit. The biocompatibility of the heated tubing has been tested, and there is no risks of safety or effectiveness.
- The subject device includes an integrated cellular module, which enables the . device to upload therapy data to the software iCodeConnect (K160127) managed by the healthcare provider, and the therapy parameters of the device can be adjusted by the healthcare provider through the cellular module. The EMC and wireless coexistence of the cellular module and the cybersecurity of its software are tested and assessed, and there is no risks of safety or effectiveness.
- The subject device provides an AutoCPAP mode in addition to the three therapy modes of the primary predicate device RESmart® BPAP 25A (K133769). The AutoCPAP mode delivers CPAP therapy and provides an air pressure no less than the prescribed one based on the patient's needs. The auto adjusting CPAP algorithm of the AutoCPAP mode is identical to that of the reference device Luna® CPAP and Auto CPAP System (K153387). It did not raise new safety or effectiveness questions.
The predicate device DreamStation Auto BiPAP (K131982) comprises an optional heated tubing and an optional cellular modem. The basic functionality and performance characteristics of the two components of the subject device Luna® G3 BPAP 25A are the same as those of those of of the predicate device
{6}------------------------------------------------
DreamStation Auto BiPAP (K131982). Testings and assessments were conducted on the subject device and they did not raise new safety or effectiveness questions.
The substantial equivalence comparison is provided below.
{7}------------------------------------------------
| | Subject Device | Primary Predicate Device | Predicate Device | Reference Device | | |
|-----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Luna® G3 BPAP 25A<br>(K201620) | RESmart® BPAP 25A<br>(K133769) | DreamStation Auto BiP<br>AP (K131982) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Comparison | |
| Classification | | | | | | |
| Device<br>Classification | Class II Device | Class II Device | Class II Device | Class II Device | Identical to primary<br>predicate | |
| Product Code | BZD | BZD | BZD | BZD | Identical to primary<br>predicate | |
| Classification<br>Panel | Anesthesiology | Anesthesiology | Anesthesiology | Anesthesiology | Identical to primary<br>predicate | |
| Regulation<br>Number | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905 | 21 CFR 868.5905 | Identical to primary<br>predicate | |
| Regulation<br>Name | Noncontinuous ventilator | Noncontinuous ventilator | Noncontinuous ventilator | Noncontinuous ventilator | Identical to primary<br>predicate | |
| Intended Use and Indications for Use | | | | | | |
| | Subject Device | Primary Predicate<br>Device | Predicate Device | Reference Device | | |
| | Luna® G3 BPAP 25A<br>(K201620) | RESmart® BPAP 25A<br>(K133769) | DreamStation Auto BiP<br>AP (K131982) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Comparison | |
| Indications<br>For Use | The Luna® G3 BPAP<br>25A is a Bi-level PAP<br>(Bi-level Positive Airway<br>Pressure) device designed<br>for the treatment of adult<br>Obstructive Sleep Apnea<br>(OSA). The integrated<br>humidifier is indicated for<br>the humidification and<br>warming of air from the<br>flow generator device.<br>These devices are<br>intended for single-<br>patient use by<br>prescription in the home<br>or hospital/institutional<br>environment on adult<br>patients. It is to be used<br>on patients >66lbs/30kg<br>for whom CPAP therapy<br>has been prescribed. The<br>system can deliver bi-<br>level therapy or auto bi-<br>level therapy. | The RESmart® BPAP<br>25A is a Bi-level PAP<br>(Bi-level Positive<br>Airway Pressure) device<br>designed for the<br>treatment of adult<br>Obstructive Sleep Apnea<br>(OSA). The optional<br>integrated humidifier is<br>indicated for the<br>humidification and<br>warming of air from the<br>flow generator device.<br>These devices are<br>intended for single<br>patient use by<br>prescription in the home<br>or hospital/institutional<br>environment on adult<br>patients. It is to be used<br>on patients >66lbs/30kg<br>for whom CPAP therapy<br>has been prescribed. The<br>system can deliver bi-<br>level therapy or auto bi-<br>level therapy. | The DreamStation Auto<br>BiPAP device delivers<br>positive airway pressure<br>therapy for the treatment<br>of Obstructive Sleep<br>Apnea (OSA) in<br>spontaneously breathing<br>patients weighing over<br>30kg (66 lbs). It is for use<br>in the home or<br>hospital/institutional<br>environment. | The Luna® CPAP and<br>Auto CPAP Systems are<br>intended to deliver<br>positive pressure for the<br>treatment of Obstructive<br>Sleep Apnea (OSA). The<br>optional integrated heated<br>humidifier is indicated<br>for the humidification<br>and warming of air from<br>the flow generator. These<br>devices are intended for<br>single patient use by<br>prescription in the home<br>or hospital/institutional<br>environment on adult<br>patients. | Identical to primary<br>predicate | |
| | Subject Device | Primary Predicate<br>Device | Predicate Device | Reference Device | | |
| | Luna® G3 BPAP 25A<br>(K201620) | RESmart® BPAP 25A<br>(K133769) | DreamStation Auto BiP<br>AP (K131982) | Luna® CPAP and Auto<br>CPAP System<br>(K153387) | Comparison | |
| Therapy<br>Delivered | S,<br>Auto S,<br>CPAP,<br>AutoCPAP | S,<br>Auto S,<br>CPAP | S,<br>Auto S,<br>CPAP | CPAP,<br>AutoCPAP | Similar to primary<br>predicate. Adding the<br>AutoCPAP mode of the<br>reference device does<br>not affect safety or<br>effectiveness. | |
| Pressure Range | For CPAP and<br>AutoCPAP mode: 4.0–<br>20.0 cmH2O<br>For Auto S and S mode:<br>4.0-25.0 cmH2O | For CPAP mode: 4.0–<br>20.0 cmH2O<br>For Auto S and S mode:<br>4.0-25.0 cmH2O | For CPAP mode: 4.0–<br>20.0 cmH2O<br>For Auto Bi-level mode:<br>4.0-25.0 cmH2O | 4.0–20.0 cmH2O | Similar to primary<br>predicate. Pressure<br>range of the AutoCPAP<br>mode is added, which<br>does not affect safety or<br>effectiveness. | |
| Pressure<br>Regulation | $\pm$ 0.5 cmH2O | $\pm$ 0.5 cmH2O | $\pm$ 0.5 cmH2O | $\pm$ 0.5 cmH2O | Identical to primary<br>predicate | |
| Pressure<br>Display<br>Accuracy (hPa) | $\pm$ (0.8cmH2O+4%) | $\pm$ 0.5 cmH2O | $\pm$ (0.3cmH2O+3.7%) | $\pm$ (0.5 cmH2O+4%) | Similar to primary<br>predicate. Change does<br>not impact safety or<br>effectiveness. | |
| Algorithm | | | | | | |
| Automatic<br>adjusting<br>CPAP<br>algorithm | Yes | No | No | Yes | Identical to reference<br>device | |
| Ramp<br>(minutes) | 0-60 | 0-60 | 0-60 | 0-60 | Identical to primary<br>predicate | |
| Expiratory<br>Pressure Relief | Reslex® function<br>Level 1-3 | Reslex® function<br>Level 1-3 | Flex® function<br>Level 1-3 | Reslex® function<br>Level 1-3…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.