K201594 · High Technology Products S.L.U · GEX · Dec 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201594
Device Name
Primelase Excellence
Applicant
High Technology Products S.L.U
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.
Device Story
Diode laser system (755nm, 810nm, 1060nm, or 810-1060nm) for dermatologic/surgical procedures. Central unit controls power/frequency; external applicators emit laser pulses via sapphire tip. Sapphire tip provides contact cooling to anesthetize tissue/protect epidermis. Operated by clinicians in clinical settings. User selects parameters via LCD touchscreen; activates pulses via applicator button. Static and dynamic modes available. Output is infrared laser energy for hair reduction, lesion treatment, or skin rejuvenation. Benefits include permanent hair reduction and treatment of vascular/pigmented lesions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), laser safety (IEC 60825-1), and biocompatibility (ISO 10993).
Technological Characteristics
AlGaAs laser diode array; Class IV laser; 755/810/1060nm wavelengths; sapphire contact cooling (5°C); LCD touchscreen interface; finger switch control; single-phase power (100-240V). Standards: IEC 60601-1, IEC 60601-1-2, IEC 62304, IEC 60601-2-22, IEC 60601-1-6, IEC 60825-1, ISO 14971, ISO 10993-5/10.
Indications for Use
Indicated for hair removal (permanent reduction in hair regrowth), treatment of Pseudofolliculitis barbae (PFB), benign pigmented lesions, benign vascular lesions (including angiomas, hemangiomas, telangiectasia, leg veins), and wrinkles. Applicable to all Fitzpatrick skin types I-VI. Prescription use only.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
December 17, 2020
High Technology Products SLU % Vardhini Kirthivas Vice President - Regulatory Services Freyr Global Regulatory Solutions & Services Level 4 Building No. H-08 Phoenix SEZ Phase 2 Gachibowli, Hyderabad, Telangana 500081 India
Re: K201594
Trade/Device Name: Primelase Excellence Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: June 8, 2020 Received: June 12, 2020
Dear Vardhini Kirthivas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201594
Device Name Primelase Excellence
### Indications for Use (Describe) Intended Use: PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.
Indications for Use:
PRIMELASE Excellence System with 810nm Laser applicator is intended for:
· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- · Treatment of Pseudofolliculitis barbae (PFB).
- · Use on all skin types (Fitzpatrick I-VI).
- · Treatment of benign pigmented lesions
PRIMELASE Excellence System with 755nm Laser applicator is intended for:
· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- · Treatment of Pseudofolliculitis barbae (PFB).
- · Use on all skin types (Fitzpatrick I-VI).
- · Treatment of benign pigmented lesions
PRIMELASE Excellence System with 1060nm Laser applicator is intended for:
· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB).
- · Use on all skin types (Fitzpatrick I-VI).
· Treatment of benign vascular lesions, including angiomas, telangiectasia, and other benign vascular lesions and leg veins.
- Treatment of wrinkles
PRIMELASE Excellence System with 810-1060nm Laser applicator is intended for:
· Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB).
· Use on all skin types (Fitzpatrick I-VI).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
High Technology Products SLU, Spain
Traditional 510(k) Primelase Excellence
# 5. 510(k) Summary
| 5.1. Submitter Information:<br>Application Correspondent: | Vardhini Kirthivas<br>Freyr Global Regulatory Solutions and Services<br>Level 4, Building No. H-08, Phoenix SEZ,<br>Phase 2, Gachibowli, Hyderabad,<br>Telangana, India, 500081 |
|-----------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | +91 - 9940445965 |
| E-mail: | vardhini.kirthivas@FreyrSolutions.com |
| Legal Manufacturer: | High Technology Products, S.L.U<br>Passatge Masoliver, 24-28<br>08005 Barcelona, Catalonia Spain |
| Phone: | +34 93 458 85 66 |
| Contact Person: | Mr. Sergi Lozano |
| E-mail: | slozano@cocoonmedical.com |
| Date Prepared: | 14 December 2020 |
#### 5.2. Device Identification:
| Device Trade Name: | Primelase Excellence |
|----------------------|-----------------------------------|
| Device Common Name: | PRIMELASE Excellence System |
| Classification Name: | Powered Laser Surgical Instrument |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Product Code: | GEX |
### 5.3. Predicate Devices:
| Device Name | 510(k) Number |
|------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Primelase Excellence | K191321 |
| The Modified Alma Lasers XL TM Family<br>of Multi-Application and Multi-<br>Technology Platforms [SopranoXL<br>SopranoXL and SopranoICE] | K172193 |
{5}------------------------------------------------
| Device Name | 510(k) Number |
|----------------------------------------------------------------------------------------|---------------|
| Lightsheer Desire; Lightsheer Desire<br>Light; Lightsheer Duet; Lightsheer<br>Infinity | K170179 |
#### 5.4. Device Description
PRIMELASE Excellence is a diode laser device which consists of a central unit and external applicators.
The PRIMELASE equipment emits laser radiation (beam infrared light with a wavelength range of 755 nm to 1060 nm, typically and most used 810 nm.), pulsed through the laser aperture situated at the tip of the applicator.
The device applicator contains the diode which emits the laser energy whereas the power delivered, and the working frequency being controlled by the machine's central unit.
The emission of energy is activated in the form of continuous pulses when pressing the applicator button. The applicator sapphire tip is cooled to a constant temperature to cool the skin, so that it partially anaesthetizes the tissue reducing the risk of damage to the epidermis during treatment.
#### Intended Use & Indications for Use 5.5.
Intended Use
PRIMELASE Excellence is intended for use in dermatologic and general surgical procedures.
# Indications for Use
PRIMELASE Excellence System with 810nm Laser applicator is intended for:
- Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB).
- Use on all skin types (Fitzpatrick I-VI). ●
- . Treatment of benign pigmented lesions
# PRIMELASE Excellence System with 755nm Laser applicator is intended for:
- Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
{6}------------------------------------------------
- Treatment of Pseudofolliculitis barbae (PFB).
- o Use on all skin types (Fitzpatrick I-VI).
- Treatment of benign pigmented lesions o
PRIMELASE Excellence System with 1060nm Laser applicator is intended for:
- Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB). ●
- Use on all skin types (Fitzpatrick I-VI). ●
- Treatment of benign vascular lesions, including angiomas, hemangiomas, ● telangiectasia, and other benign vascular lesions and leg veins.
- Treatment of wrinkles .
PRIMELASE Excellence System with 810-1060mm Laser applicator is intended for:
- Hair Removal with Static and Dynamic modes intended for permanent reduction in hair regrowth, defined as a long term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
- Treatment of Pseudofolliculitis barbae (PFB). 0
- Use on all skin types (Fitzpatrick I-VI). ●
### 5.6. Comparison of Technological Characteristics
The fundamental scientific technology, materials of construction and mechanism of operation are identical between the subject PRIMELASE equipment and the predicate devices. Table 2 summarizes the comparison of technological characteristics between the subject and predicate devices.
| S.No | Parameters | Soprano ICE<br>(K172193) | Lightsheer<br>Infinity<br>(K170179) | Primelase<br>Excellence<br>(K191321) | Primelase<br>Excellence<br>(Subject<br>device) |
|------|----------------------|--------------------------|-------------------------------------|------------------------------------------|------------------------------------------------|
| 1. | Manufacturer | Alma Lasers,<br>Ltd, | Lumenis Ltd | High<br>Technology<br>Products<br>S.L.U. | High<br>Technology<br>Products,<br>S.L.U. |
| 2. | Product Code | GEX | GEX | GEX | GEX |
| 3. | Regulation<br>Number | 21 CFR<br>878.4810 | 21 CFR<br>878.4810 | 21 CFR<br>878.4810 | 21 CFR<br>878.4810 |
| Table 2 - Substantial Equivalence Table | | |
|-----------------------------------------|--|--|
| | | |
{7}------------------------------------------------
High Technology Products SLU, Spain
| S.No | Parameters | Soprano ICE<br>(K172193) | Lightsheer<br>Infinity<br>(K170179) | Primelase<br>Excellence<br>(K191321) | Primelase<br>Excellence<br>(Subject<br>device) |
|------|---------------------------|------------------------------------------------------------------------|-------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| 4. | Principle of<br>Operation | AlGaAs<br>Laser diode<br>array | AlGaAs<br>Laser diode<br>array | AlGaAs Laser<br>diode<br>array | AlGaAs Laser<br>diode<br>array |
| 5. | Laser<br>Classification | Class IV | Class IV | Class IV | Class IV |
| 6. | User<br>Interface | LCD touch<br>screen | LCD touch<br>screen | LCD touch<br>screen | LCD touch<br>screen |
| 7. | Pulsing<br>Control | Finger switch | Finger switch | Finger switch | Finger switch |
| 8. | Configuration | Main unit,<br>Handpiece<br>and Foot<br>control | Main unit<br>and<br>Handpiece | Main unit,<br>Handpiece<br>and Foot<br>control<br>(optional) | Main unit,<br>Handpiece<br>and Foot<br>control<br>(optional) |
| 9. | Laser<br>Wavelength | 755 nm<br>810 nm<br>1064nm | 805 nm<br>1060 nm | 755nm<br>810nm<br>810-1060nm | 755nm<br>810nm<br>1060nm<br>810-1060nm |
| 10. | Spot Size<br>(CM × CM) | 12x10,<br>15x10,<br>20x10,<br>10x10 | 9x9, 27x9,<br>22x35 | 20x9, 30x9,<br>30x17 | 20x9, 30x9,<br>30x17<br>10x10 |
| 11. | Laser Contact | Sapphire<br>window | Pure<br>sapphire,<br>A1203 | Pure sapphire,<br>A12O3 | Pure sapphire,<br>A1203 |
| 12. | Frequency | Up to 3Hz<br>(HR)<br>5 - 10Hz<br>(SHR) | Up to 3Hz | Up to 3Hz<br>(static)<br>5 — 10Hz<br>(dynamic) | Up to 3Hz<br>(static)<br>5 — 10Hz<br>(dynamic) |
| 13. | Pulse<br>duration | 3.3-200ms | 5-400ms | 3-400ms | 3-400ms |
| 14. | Fluence | 120J/cm² | 100J/cm² | 80J/cm² | 120J/cm² |
| 15. | Tissue<br>cooling | Contact<br>continuous,<br>Thermo-<br>electrical | Chilltip<br>contact<br>cooling | Contact<br>continuous,<br>Thermo-<br>electrical | Contact<br>continuous,<br>Thermo-<br>electrical |
| 16. | Cooling<br>Temperature | 4 °C | 2 °C -12 °C | 5 °C | 5 °C |
| 17. | Power supply | 120VAC,<br>11A, 50/60 | 100-240<br>VAC +/-<br>10% | Single phase,<br>100-240V<br>50-60 Hz | Single phase,<br>100-240V<br>50-60 Hz |
| S.No | Parameters | SopranoICE<br>(K172193) | Lightsheer<br>Infinity<br>(K170179) | Primelase<br>Excellence<br>(K191321) | Primelase<br>Excellence<br>(Subject<br>device) |
| | | Hz, single<br>phase<br>220/230VAC,<br>6A, 50/60<br>Hz, single<br>phase | 15 A max. <br>50/60 Hz. | | |
{8}------------------------------------------------
High Technology Products SLU, Spain
### 5.7. Non-Clinical Testing
Non-clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed device complies with the following standards:
- 1. IEC 60601-1:2005 + AMD1:2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General 2. requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 62304:2006/AMD:2015 Medical device software software life cycle 3. processes
- 4. IEC 60601-2-22 Edition 3.1: Medical electrical equipment Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- 5. IEC60601-1-6 Edition 3.1: Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60825 1 Edition 2: Safety of laser products Part 1: Equipment 6. classification and requirements.
- 7. ISO 14971:2007 Medical devices Application of risk management to medical devices.
- 8. ISO 10993-5:2009 Biological Evaluation of Medical Device, Part 5-Tests for In Vitro cytotoxicity
- 9. ISO 10993-10:2010 Biological Evaluation of Medical Device, Part 10-Test for irritation and skin sensitization
#### 5.8. Clinical Testing
No clinical study is included in this submission.
{9}------------------------------------------------
High Technology Products SLU, Spain
#### 5.9. Conclusion
Based on the comparison and analysis above, the PRIMELASE Excellence device's intended use and technological characteristics do not raise new types of questions regarding safety and efficacy when compared to the predicates. Based on its technical characteristics, indications for use, and performance data, the PRIMELASE Excellence device is considered to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.