K201519 · Microport Orthopedics, Inc. · JDI · Dec 10, 2020 · Orthopedic
Device Facts
Record ID
K201519
Device Name
PROFEMUR® GLADIATOR® Cemented Classic Stem
Applicant
Microport Orthopedics, Inc.
Product Code
JDI · Orthopedic
Decision Date
Dec 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PROFEMUR® GLADIATOR® Cemented Classic Stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients. 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2) inflammatory degenerative joint disease such as rheumatoid arthritis; 3) correction of functional deformity; and, 4) revision procedures where other treatments or devices have failed PROFEMUR® GLADIATOR® Cemented Classic Stems are single use implants intended for cemented arthroplasty.
Device Story
Cemented femoral hip stem; tapered-wedge design; reduced distal stem tip; glass-beaded body; proximal medial collar for rotational stability/subsidence protection; MicroPort 12/14 SLT taper trunnion for femoral head attachment; oval impaction feature for insertion. Used in total hip arthroplasty; implanted by orthopedic surgeons. Restores hip biomechanics, leg length, and head center offset. Benefits patient via pain relief and improved hip function.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical testing included proximal and distal fatigue strength, range of motion (ROM) analysis, and MRI safety analysis (ASTM F2052, F2119, F2182, F2503). Testing performed on worst-case constructs per ISO 7206-4 and ISO 7206-6.
Indicated for skeletally mature patients requiring total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity correction, or revision procedures.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
PROFEMUR® GLADIATOR® Thin HA Classic Stem w/ Collar (K200007)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
MicroPort Orthopedics Inc. Allen Mamaril Regulatory Affairs Specialist II 5677 Airline Road Arlington, Tennessee 38002
# Re: K201519
Trade/Device Name: PROFEMUR® GLADIATOR® Cemented Classic Stem Regulation Number: 21 CFR 888.3350 Regulation Name: Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JDI, KWL, KWY, LZO Dated: November 12, 2020 Received: November 13, 2020
# Dear Allen Mamaril:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
PROFEMUR® GLADIATOR® Cemented Classic Stems
Indications for Use (Describe)
PROFEMUR® GLADIATOR® Cemented Classic Stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
2) inflammatory degenerative joint disease such as rheumatoid arthritis;
3) correction of functional deformity; and,
4) revision procedures where other treatments or devices have failed
PROFEMUR® GLADIATOR® Cemented Classic Stems are single use implants intended for cemented arthroplasty.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue rounded square with a red star-like shape inside, followed by the text "MicroPort" in a bold, blue font. Below "MicroPort" is the word "Orthopedics" in a smaller, blue font, and below that is the tagline "Full Function, Faster" in a smaller, blue font.
# 510(k) Summary
| Submitted By: | MicroPort Orthopedics Inc.<br>5677 Airline Road<br>Arlington, TN 38002<br>USA<br>Telephone Number: (901) 290-5717 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | December 10, 2020 |
| Contact Person: | Allen Mamaril<br>Regulatory Affairs Specialist II<br>Telephone number: (901) 867-4551<br>Email: Allen.Mamaril@ortho.microport.com |
| Name of Device: | PROFEMUR® GLADIATOR® Cemented Classic Stem |
| Common Name: | Femoral Hip Stem |
| Device Classification Name | 21 CFR 888.3350 - Hip joint metal/polymer semi-constrained cemented<br>prosthesis; |
| | 21 CFR 888.3360 - Hip joint femoral (hemi-hip) metallic cemented or<br>uncemented prosthesis; |
| | 21 CFR 888.3390 - Hip joint femoral (hemi-hip) metal/polymer cemented<br>or uncemented prosthesis; |
| | 21 CFR 888.3353 - Hip joint metal/ceramic/polymer semi-constrained<br>cemented or nonporous uncemented prosthesis |
| Device Class: | Class II |
| Panel Code: | Orthopedics/87 |
| Product Codes: | JDI, KWL, KWY, LZO |
| Predicate Device: | PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE<br>10/14/2011) |
| Reference Device: | PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE<br>08/25/2009; K100866, SE 04/28/2010)<br>PROFEMUR® GLADIATOR® Thin HA Classic Stem w/ Collar<br>(K200007, SE 04/25/2020) |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue rounded square with a white circle inside, containing a red star-like shape. To the right of the square is the text "MicroPort" in a bold, blue font, with "Orthopedics" written below it in a smaller font. Below "Orthopedics" is the tagline "Full Function, Faster" in a smaller, lighter blue font.
### Device Description:
The PROFEMUR® GLADIATOR® Cemented Classic Stems present a tapered-wedge design and a reduced distal stem tip, identical to the legally marketed modular PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE 08/25/2009; K100866, SE 04/28/2010).
The subject stems are intended to be cemented and have a glass-beaded body, medial collar, proximal shoulder and neck region with a surface roughness identical to the legally marketed modular PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423. SE 08/25/2009; K100866. SE 04/28/2010).
The subject stems possess a proximal medial collar designed to assist rotational stability and offer protection against subsidence, identical to the cleared predicate PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE 08/25/2009; K100866, SE 04/28/2010). The trunnion is a MicroPort 12/14 SLT taper, with an as-machined geometry to form a taper lock with a mating femoral head implant. The trunnion is identical to the MicroPort femoral stem tapers of the predicate and reference devices.
The subject stems possess an oval impaction feature on the proximal shoulder for proper femoral stem insertion into the femoral canal, identical to the cleared reference PROFEMUR® GLADIATOR® Thin HA Classic Stem w/ Collar (K200007, SE 04/25/2020).
#### Materials
The subject PROFEMUR® GLADIATOR® Cemented Classic Stems are manufactured from forged cobalt chromium alloy (ASTM F799).
#### Geometry and Dimensions
The PROFEMUR® GLADIATOR® Cemented Classic Stems were designed to accommodate a wide range of patient anatomy and to provide an optimized fit to adequately restore the biomechanics of the respective hip (including restoration of the natural leg length and head center offset). The subject PROFEMUR® GLADIATOR® Cemented Classic Stems are offered in sizes 4, 6, 8, 10, and 12 with the stem length ranging from 125mm to 170mm, identical to the cleared reference PROFEMUR® GLADIATOR® Cemented Stems (K111910, SE 10/14/2011). The subject classic stem possesses two different neck options: standard (with a CCD neck angle of 135°) and extended/Varus 8° (with a CCD angle of 127°).
#### Intended Use:
PROFEMUR® GLADIATOR® Cemented Classic Stems are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
#### Indications for Use:
The subject device is indicated for the following conditions:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue rounded square with a white circle inside, containing a red star-like shape. To the right of the icon, the text "MicroPort Orthopedics" is displayed in blue, with "MicroPort" on the top line and "Orthopedics" on the second line. Below the company name, the tagline "Full Function, Faster" is written in a smaller, lighter blue font.
- 1) non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2) inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3) correction of functional deformity; and,
- 4) revision procedures where other treatments or devices have failed.
PROFEMUR® GLADIATOR® Cemented Classic Stems are single use implants intended for cemented arthroplasty.
# Comparison of Technological Characteristics with the Predicate devices:
Device comparison described in this premarket notification demonstrates that the subject PROFEMUR® GLADIATOR® Cemented Classic Stems are substantially equivalent to the identified predicate, PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE 08/25/2009; K100866, SE 04/28/2010), cleared by the FDA for commercial distribution in the United States. The subject devices were shown to be substantially equivalent and have equivalent technological characteristics to their predicate devices in areas including design, intended use, material composition, operational principles and function.
# Mechanical Testing:
MicroPort has evaluated the subject PROFEMUR® GLADIATOR® Cemented Classic Stems and data demonstrates substantial equivalence to the identified predicate PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE 08/25/2009; K100866. SE 04/28/2010). Applicable design verification testing and engineering analyses for the subject implants were completed in accordance with the following standards:
- . ASTM F799-19: Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants (UNS R31537, R31538, R31539)
- . ISO 7206-4:2010 Implants for surgery - Partial and total hip joint prostheses - Part 4: Determination of endurance properties and performance of stemmed femoral components
- ISO 7206-6:2013 Implants for surgery Partial and total hip joint prostheses Part 6: ● Endurance properties testing and performance requirements of neck region of stemmed femoral components.
- . ASTM F2052-6, "Standard test method for measurement of magnetically induced displacement force on passive implants in the magnetic resonance environment"
- . ASTM F2119-7 "Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants"
- . ASTM F2182-11a "Standard Test Method for Measurement of Radio Frequency Induced Heating near Passive Implants During Magnetic Resonance Imaging"
- . ASTM F2503-13 "Standard Practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment"
- EN ISO 21535 (2009): Non-active surgical implants Joint replacement implants Specific ● requirements for hip-joint replacement implants
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for MicroPort Orthopedics. The logo consists of a blue square with rounded corners, inside of which is a red star-like shape. To the right of the square is the text "MicroPort" in a larger, bold, blue font, and below that is "Orthopedics" in a smaller, blue font. Below "Orthopedics" is the text "Full Function, Faster" in a smaller, blue font.
The mechanical testing and engineering analyses performed that are applicable to the subject PROFEMUR® GLADIATOR® Cemented Classic Stems include:
- 1. Proximal Fatigue Strength Testing and Analysis
- 2. Distal Fatigue Strength Testing and Analysis
- 3. Range of Motion (ROM) Testing and Analysis
- 4. Magnetic Resonance Imaging (MRI) Safety Analysis
The data demonstrates that the subject PROFEMUR® GLADIATOR® Cemented Classic Stems are substantially equivalent to the legally marketed predicate, PROFEMUR® GLADIATOR® Cemented Stem (K111910, SE 10/14/2011) with PROFEMUR® Cobalt Chrome Modular Necks (K091423, SE 08/25/2009; K100866. SE 04/28/2010). Having met all acceptance criteria for mechanical testing performed on worstcase constructs, the subject PROFEMUR® GLADIATOR® Cemented Classic Stems do not introduce new or modified risks for safety and effectiveness compared to the predicate or other reference MicroPort Hip Systems and acetabular components which utilize comparable test methods and acceptance criteria.
#### Conclusion:
Based on the design features, the use of established well-known materials, feature comparisons, indications for use, principle of operations, and results of the mechanical testing and engineering analyses, the subject PROFEMUR® GLADIATOR® Cemented Classic Stems have shown to be substantially equivalent to the legally marketed predicate devices cited in this premarket notification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.