The PROFEMUR® Hip System Modular Necks are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
Device Story
PROFEMUR® Hip System Modular Necks are cobalt chrome alloy components used in total hip arthroplasty. This submission expands the compatibility of these modular necks to include 28mm, 32mm, and 36mm alumina ceramic femoral heads. The device functions as a modular interface between the femoral stem and the femoral head. It is intended for use by orthopedic surgeons in a clinical/hospital setting. The device is implanted during hip replacement surgery to restore joint function and reduce pain. Clinical benefit is derived from the modularity allowing for anatomical reconstruction. The device is evaluated via mechanical testing, including burst strength, axial distraction, and fatigue testing, to ensure the integrity of the taper junction between the metal neck and ceramic head.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including pre- and post-fatigue burst strength testing, axial static distraction force (ASTM F2009-00), and static/cyclic fatigue strength testing (ASTM 2345-03).
Technological Characteristics
Materials: Cobalt-28Chromium-6Molybdenum alloy (ASTM F1537) for modular necks; high purity alumina (ISO 6474) for ceramic heads. Design: Modular taper junction. Sterilization: Ethylene oxide (ISO 11135) or radiation (ISO 11137). Testing standards: ASTM F2345-03, ASTM F2009-00, ISO 7206-6.
Indications for Use
Indicated for skeletally mature patients requiring total hip arthroplasty for non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, painful hip dysplasia), inflammatory degenerative joint disease (rheumatoid arthritis), correction of functional deformity, or revision procedures where other treatments have failed. Can be used in cemented or uncemented femoral and acetabular arthroplasty.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Klooneb
APR 2 8 2010
A Wright Medical Group Compa
# 510(k) Summary of Safety and Effectiveness
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® Hip System Modular Necks.
#### (a)(1) Submitted By
The submitter's name, address, telephone number, a contact person, and the date the summary was prepared:
| Submitter's Name: | Wright Medical Technology, Inc.<br>5677 Airline Rd. Arlington, TN 38002<br>800-238-7188 (phone), 901-867-4190 (fax) |
|-------------------|---------------------------------------------------------------------------------------------------------------------|
| Date: | April 28, 2010 |
| Contact Person: | Gregory Neal<br>Regulatory Affairs Specialist II |
(a)(2) Device Name Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name if known:
| Proprietary Name:<br>Common Name: | PROFEMUR® Hip System Modular Necks<br>Modular Neck |
|---------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| Classification Name and Reference: | 21 CFR 888.3353 Hip joint metal/ ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis.<br>Class II |
| Subject Device Product Code and Panel Code: | Orthopedics/87/LZO |
#### (a)(3) Predicate Device
Identification of the legally marketed device to which the submitter claims equivalence:
| Predicate Proprietary Name: | PROFEMUR® Hip System Modular Necks |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Classification and Number: | 888.3330 KWA Hip joint metal/ metal semi-<br>constrained, with an uncemented acetabular<br>component prosthesis Class III 510(k) K091423 |
| | PROFEMUR® LX 5/8 Coated Hip Stem<br>888.3353 LZO Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis Class II 510(k) K081090 |
| | CERAMIC FEMORAL HEAD<br>888.3353 LZO Hip joint metal/ceramic/polymer<br>semi-constrained cemented or nonporous uncemented<br>prosthesis Class II 510(k) K893685 |
## (a)(4) Device Description
Description of the device that is the subject of the premarket notification submission:
The subject modular femoral necks, manufactured from cobalt chrome alloy, were cleared for use in K091423 with metal femoral heads, but lacked testing for direct assembly to ceramic femoral heads. This 510(k) notification consists of the mechanical testing that allows for the clearance for use of the subject necks with the 28mm, 32mm,
{1}------------------------------------------------
and 36mm alumina ceramic femoral heads previously cleared in K893685, and removal of this previous contraindication (K091423), except for with the 28mm Long ceramic head.
Mean and minimum pre-fatigue burst strength testing results exceeded minimum values specified by FDA in the Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems - 1/10/95. Mean post-fatigue burst strength and mean axial static distraction force per ASTM F2009 was greater than that of the identical currently marketed ccramic femoral heads assembled to a titanium neck cone. Harder identical taper junctions demonstrated less fretting than softer identical junctions. Lower hardness identical junctions demonstrated resistance to fatigue failure per ISO 7206-6. Total Fretting wear for proximal and distal junctions was less than that of published metal-on-metal articulating wear results.
# (a)(5) Intended Use of the Device
Statement of the intended use of the subject of the premarket notification submission, including a general description of the diseases or conditions that the device will diagnose, treat, prevent, cure, or mitigate, including a description, where appropriate, of the patient population for which the device is intended:
#### Intended Use
The PROFEMUR® Hip System Modular Necks are intended for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients.
#### Indications for Use
- non-inflammatory degenerative joint disease such as ostcoarthritis, avascular necrosis, ankylosis, protrusio 1. acetabuli, and painful hip dysplasia:
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and.
- 4. revision procedures where other treatments or devices have failed
Modular necks can be used during either cemented or uncemented femoral and acetabular arthroplasty.
# (a)(6) Technological Characteristics of the Device
Summary of the technological characteristics of the new device in comparison to those of the predicate device:
The indications for use for the subject PROFEMUR® Hip System Cobalt Chrome Modular Necks are identical to the indications previously cleared for the Predicate Device (K091423). The identical intended use for ceramic heads with a metal femoral component was cleared for the modular PROFEMUR® LX 5/8 Hip Stem (K081090). The design and materials of the PROFEMUR® Hip System Cobalt Chrome Modular Necks are unchanged from the design presented in K091423. The design and materials of the Ceramic Femoral Heads are unchanged from the design presented in K893685.
### (b)(1) Nonclinical Testing
Brief discussion of the nonclinical tests submitted, referenced, or relied on in the premarket notification submission for a determination of substantial equivalence:
PROFEMUR® Hip System Modular Necks meet requirements per pre and post-fatigue burst strength testing specified by FDA in the Guidance Document for the Preparation of Premarket Notifications for Ceramic Ball Hip Systems; per static, fatigue, and post-fatigue compressive burst testing of 28mm alumina ceramic heads conducted in accordance with ASTM 2345-03 - Determination of Static and Cyclic Fatigue Strength of Ceramic Modular Femoral Heads; per axial static distraction of ccramic femoral heads conducted in accordance with ASTM F2009-00 -Standard Test Method for Determining the Axial Disassembly Force of Taper Connections of Modular Prostheses; per ISO 7206-6 Partial and total hip joint prostheses-Part 6: Determination of endurance properties of head and neck region of stemmed femoral components; per ISO: 11135 Sterilization of health care products- Requirements for development, validation and routine control of sterilization process for medical devices-Ethylene oxide; per ISO: 11137 Sterilization of health care products- Requirements for development, validation and routine control of sterilization process for medical devices-Radiation sterilization; per ASTM F1537 Wrought Cobalt-28Chromium-6Molybdenum Alloys for Surgical Implants, and per ISO 6474 Ceramic Materials Based on High Purity Alumina.
{2}------------------------------------------------
# (b)(2) Clinical Testing
Brief discussion of the clinical tests submitted, referenced, or relied on in the premarket notification for a determination of substantial equivalence:
Clinical data was not provided for the class II modular neck devices.
#### (b)(3) Conclusions
Conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in (a)(3):
The indications for use of the PROFEMUR® Hip System Modular Necks are identical to the previously cleared predicate devices. The design features and materials of the subject devices are unchanged. The findinged. The findinental scientific technology of the modified device has not changed relative to the predicate devices. The safely and effectiveness of the PROFEMUR® Hip System Modular Necks are adequately supported by the substanial equivalence information, materials information, and analysis data provided within this Premarket Ny ific
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" surrounding the top and sides of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Wright Medical Technology, Inc. % Mr. Gregory Neal Regulatory Affairs Specialist II 5677 Airline Road Arlington, Texas 38002
APR 2 8 2010
Re: K100866
Trade/Device Name: PROFEMUR® Hip System Modular Necks Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: II Product Code: LZO Dated: March 23, 2010 Received: March 29, 2010
Dear Mr. Neal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{4}------------------------------------------------
Page 2 - Mr. Gregory Neal
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Oarbare Buelm
Mark N. Melkers Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "WRIGHT." in bold, sans-serif font. Underneath the word is a decorative design that looks like two stylized wings or swooshes. The text and design are black against a white background, creating a high-contrast image.
# Indications
510(k) Number: K100866
Device Name: PROFEMUR® Hip System Modular Necks
Indications For Use
- non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, l. ankylosis, protrusio acetabuli, and painful hip dysplasia;
- inflammatory degenerative joint disease such as rheumatoid arthritis; 2.
- correction of functional deformity; and, 3.
- revision procedures where other treatments or devices have failed 4.
Modular necks can be used during either cemented or uncemented femoral and acetabular arthroplasty.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Onetu
(Division Sign-Off) for mxm
Division of Surgica!, Orthopedic, and Restorative Devices
510(k) Number K100866
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.