K201475 · CAO Group, Inc. · PBX · Oct 28, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201475
Device Name
Ultimate Contour, Ultimate Contour Mini
Applicant
CAO Group, Inc.
Product Code
PBX · General, Plastic Surgery
Decision Date
Oct 28, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
1) Application of non-thermal radio frequency for relief of minor muscle aches and pains, relief of muscle spasm, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite. 2) Application of ultrasound for non-invasive reduction in abdominal circumference on adults with a Body Mass Index (BMI) of 25 or greater.
Device Story
Ultimate Contour/Mini is a dual-modality aesthetic device combining radio-frequency (RF) and ultrasound. RF handpiece delivers energy to heat tissue, inducing collagen contraction and increasing local circulation for pain/spasm relief. Ultrasound handpiece uses a piezo transducer to focus energy in subdermal fatty tissue for abdominal circumference reduction. Operated by clinicians in medical offices; settings adjusted via interactive color LCD touchscreen. Device includes onboard tutorial videos. System uses single power source and control circuitry; handpiece attachment determines energy delivery mode. Benefits include non-invasive body contouring and temporary pain relief without anesthesia. Output affects clinical decision-making by providing a non-surgical option for fat reduction and muscle therapy.
Clinical Evidence
Single-site, single-arm study (n=42) evaluated waist circumference reduction. Mean age 42.8, mean baseline BMI 29.36. Primary endpoint: reduction ≥ 1 inch. Results showed average reduction of 2.0 inches at 4-week follow-up (p=0.05). Correlation between baseline and follow-up was 0.9901. Adverse events included ringing/buzzing in ears (n=7, 2 discontinued), skin rash, and changes in bowel/urine consistency; most resolved within 1-2 days. No permanent audiological issues identified.
Technological Characteristics
Dual-modality system: RF and 37kHz ultrasound. Materials: stainless steel, nylon plastic. Cooling: heatsink/air convection. Power: 100-240VAC, 50-60Hz. Form factors: free-standing (45 lbs) or table-top (5 lbs). Connectivity: standalone with LCD touchscreen. Software: Major level of concern. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6.
Indications for Use
Indicated for adults (BMI ≥ 25) for non-invasive abdominal circumference reduction via ultrasound, and for relief of minor muscle aches/spasms, improved local circulation, and cellulite appearance reduction via non-thermal radio frequency.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2020
CAO Group, Inc. Robert K. Larsen Regulatory Affairs Manager 4628 West Skyhawk Drive West Jordan, Utah 84084
Re: K201475
Trade/Device Name: Ultimate Contour, Ultimate Contour Mini Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: PBX, OHV Dated: July 7, 2020 Received: July 13, 2020
Dear Robert K. Larsen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201475
Device Name
Ultimate Contour, Ultimate Contour Mini
Indications for Use (Describe)
1) Application of non-thermal radio frequency for relief of minor muscle aches and pains, relief of muscle spasm, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite. 2) Application of ultrasound for non-invasive reduction in abdominal circumference on adults with a Body Mass Index (BMI) of 25 or greater.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☑</span> |
|----------------------------------------------|-------------------------------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | □ |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summarv
#### Applicant Information:
| Company Name: | CAO Group, Inc. |
|------------------|-----------------------------------------------------------|
| Company Address: | 4628 West Skyhawk Drive<br>West Jordan, Utah 84084 U.S.A. |
| Company Phone: | 1-801-256-9282 |
| Company Fax: | 1-801-256-9287 |
| Contact Person: | Robert K. Larsen |
October 27, 2020
Preparation Date:
Device Name:
#### Trade Name: Ultimate Contour and Ultimate Contour Mini Common Name 1: Massager, vacuum, radio frequency induced heat PBX Product Code 1: 878.4400 Regulation 1: Product Classification 1: Class II Common Name 2: Focused ultrasound stimulator for aesthetic use Product Code 2: OHV Regulation 2: 878.4590 Product Classification 2: Class II
#### Legally Marketed Predicate Devices for Substantial Equivalence:
Ultimate Contour, manufactured by CAO Group, Inc. (K171052)
UltraShape, manufactured by Syneron Medical, Ltd. (K141708)
#### Description of Submitted Device:
The Ultimate Contour is a medical device that combines the technologies of radio-frequency energy and ultrasound in one convenient unit. The Ultimate Contour utilizes a single power source and control circuitry that directs the emission of energy based on which operating screen is selected and which applicator handpiece is attached to the unit's delivery cable. Based on the attached handpiece, the Ultimate Contour delivers to the handpiece of the appropriate energy intensity based on the settings selected.
Delivered radio frequency energy is accomplished with the multi-nodal RF handpiece that contains RF energy to the treatment area and works to gently heat the tissue to induce collagen contraction and increase local circulation. The elevated temperature further works to provide temporary relief from pain and muscle spasms. The unit features an interactive color LCD touchscreen for adjusting device settings, as well as for viewing onboard tutorial videos.
Delivered ultrasound is accomplished with a piezo transducer located in the ultrasound handpiece that is bonded to the handpiece contact surface. The handpiece is applied directly to the patient skin at the abdominal area and the emitted ultrasound energy focuses within the subdermal fatty tissue, causing reduction in waist circumference.
{4}------------------------------------------------
#### K201475
The unit features an interactive color LCD touchscreen for adjusting device settings, as well as for viewing onboard tutorial videos.
The device is presented as two model (005-00035) is an integrated free-standing system. The second model (005-00036) is a portable, table-top style units feature exactly the same internal electronics and components, the exact same software, the exact same attachments and accessories, and the exact same environment of use. The models differ only in the shape and dimensions of the mechanical enclosure of the control unit.
## Indications for Use of the Submitted Device:
The submitted device is indicated for -
- 1) Application of non-thermal radio frequency for relief of minor muscle aches and pains, relief of muscle spasm, temporary improvement of local blood circulation, and temporary reduction in the appearance of cellulite.
- 2) Application of ultrasound for non-invasive reduction in abdominal circumference on adults with a Body Mass Index (BMI) of 25 or greater.
| Summary of Technological Characteristics and Substantial Equivalence: | | | |
|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------------------------|
| | CAO Group, Inc.<br>Ultimate Contour | CAO Group, Inc.<br>Ultimate Contour<br>K171052 | Syneron Medical, Ltd.<br>UltraShape<br>K141708 |
| Ultrasound<br>Frequency | 37kHz | Not Present | 200kHz |
| Ultrasound<br>Power | 1.00 W/cm² | Not Present | 17 W/cm² |
| Ultrasound<br>Focal Distance | 6mm | Not Present | 15mm |
| Electrical<br>Power Input<br>(System) | 100-240VAC, 50-60Hz, 207VA | 100-240VAC, 50-60Hz,<br>207VA | 110-120/200-240VAC, 50/60Hz |
| Cooling<br>Method<br>(Handpiece) | Heatsink/Air Convection | Heatsink/Air Convection | Not available |
| Cooling<br>Method (Unit) | Heatsink / Fan air cooled | Heatsink / Fan air cooled | Not available |
| Application<br>Time | 10-40 minutes | 10-40 minutes | 30-60 minutes |
| Dimensions | (Model 005-00035): 19" x 19" x<br>50"<br>(Model 005-00036): 13.75" x<br>14.5" x 14" | (Model 005-00035): 19" x<br>19" x 50"<br>(Model 005-00036): 13.75"<br>x 14.5" x 14" | 25.6" x 38.6" x 78.75" |
| Weight | (Model 005-00035): 45.0 lbs.<br>(Model 005-00036): 5.0 lbs | (Model 005-00035): 45.0 lbs.<br>(Model 005-00036): 5.0 lbs | 285.0 lbs. |
{5}------------------------------------------------
| | CAO Group, Inc.<br>Ultimate Contour | CAO Group, Inc.<br>Ultimate Contour<br>K171052 | Syneron Medical, Ltd.<br>UltraShape<br>K141708 |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| User Interface | Interactive color LCD<br>touchscreen | Interactive color LCD<br>touchscreen | Interactive color LCD<br>touchscreen |
| Patient-<br>Contacting<br>Materials | Stainless Steel<br>Nylon Plastic | Stainless Steel<br>Nylon Plastic | Stainless Steel<br>ABS Plastic |
| Sterilization of<br>Patient-<br>Contacting<br>Materials | None Specified | None Specified | None Specified |
| Environment<br>of Use | Controlled medical office or<br>practice | Controlled medical office or<br>practice<br>K171052 | Controlled medical office or<br>practice<br>K141708 |
| 510(k)<br>Number | Pending this application | K171052 | K141708 |
| Indications for<br>Use | 1) Application of non-thermal<br>radio frequency for relief of<br>minor muscle aches and<br>pains, relief of muscle<br>spasm, temporary<br>improvement of local blood<br>circulation, and temporary<br>reduction in the appearance<br>of cellulite.<br><br>2) Application of ultrasound for<br>non-invasive reduction in<br>abdominal circumference on<br>adults with a Body Mass<br>Index (BMI) of 25 or greater. | Application of non-thermal<br>radio frequency for relief of<br>minor muscle aches and<br>pains, relief of muscle<br>spasm, temporary<br>improvement of local blood<br>circulation, and temporary<br>reduction in the<br>appearance of cellulite. | The UltraShape System delivers<br>focused ultrasound energy that<br>can disrupt subcutaneous<br>adipose tissue (SAT) to provide<br>a non-invasive approach to<br>achieve a desired aesthetic<br>effect. It is intended for non-<br>invasive reduction in abdominal<br>circumference. |
### Rationale for Substantial Equivalence:
The submitted device shares the same indications for use as the predicates. The submitted device utilizes the same energy type and means of energy application as the predicate devices. The submitted device is used by the same type of operator as the predicate devices. The submitted device provides for similar user interface and control mechanisms as the predicate devices. Ultrasound component of subject device has technological differences to compare to the predicate. Pre-clinical performance testing was performed to evaluate substantial equivalence of ultrasound component to the predicate.
### Performance Data:
### ELECTRICAL SAFETY AND ELECTROMAGNETIC COMPATIBILITY
The Ultimate Contour is demonstrated to comply with the performance requirements of IEC 60601-1: 3ª Edition, IEC 60601-1-2: 3rd Edition with gap analysis to 4th Edition, and IEC 60601-1-6.
### PERFORMANCE BENCH TESTING
Bench testing per internal verification testing demonstrates that the Ultimate Contour meets the essential performance requirements established for the device, including measurements of the radio frequency power
{6}------------------------------------------------
generated, and essential aspects of hardware and software performance. The software of this device is stated as a Major level of concern since failure of the software could present significant risk to the patient. All essential functions are demonstrated to work according to design specifications.
#### BIOCOMPATIBILITY
Materials used in the patient-contacting portions of this device are equivalent to legally marketed reference devices listed above that incorporate patient-contacting articles used for contacting similar tissues.
#### CLINICAL PERFORMANCE TESTING
A clinical evaluation of the device on a representation of the intended population was performed in a single-site single-arm study. The study evaluated the safety and effectiveness of the device in waist circumference reduction. A total of 42 subjects completed the treatment regimen of the device with one week between each treatment, and 1-week follow ups. The safety analysis included all subjects who participated at any point during the course of the study. Effectiveness was gauged by determining if waist circumference reduction was achieved relative to a value identified as clinically meaningful, a value of at least 1 inch of waist circumference reduction of the treatment regimen. The clinically meaningful value was identified based on published studies utilizing other devices for the same purpose. Measurements were conducted prior to and after each treatment, and at the 1-week follow-up visits. Clinicians were blinded relative to measuring pre-treatment and post-treatment.
The study included subjects age 18 or older who met all inclusion criteria and none of the exclusion criteria. The majority of study subjects were Caucasian females. Mean age of the subjects was 42.8 years. Mean baseline waist circumference prior to the first treatment was measured at 40.95 inches, and an average initial BMI of 29.36. The primary effectiveness endpoint measured the difference between waist circumference prior to the first treatment with the device. The safety endpoint assessed all adverse events and serious adverse events occurring during the study.
The study results for the primary endpoint demonstrated an average circumference reduction of 2.0 inches to the 4-week follow up visit, which was statistically significant compared to the target threshold of 1.0 inches (p=0.05). The greatest circumference reduction for this period was measured at 3.75 inches, and the least circumference reduction was measured at 0.75 inches. T-test and Kolmogorov-Smirnov tests demonstrated that the data can be considered normally distributed. Correlation analysis comparing the baseline measurements and 4-week follow-up measurements returned a correlation score of 0.9901. With all missing data imputed with a least-favorable regimen outcome (0.75 inch reduction), the effect of device treatment on circumference reduction remained greater than the 1.0 inch reduction threshold. The treatment was administered without anesthetic. 7 subjects reported a sensation of ringing or buzzing in the ears, which was substantial enough to cause 2 of the subjects to discontinue participation. Follow-up audiology assessment with these subjects identified no substantial or worsening audible capabilities. Most adverse events resolved within 1-2 days after the treatment. These included skin rash, changes in bowel movement, and changes in urine consistency. The events were either self-resolved with the use of over-the-counter products.
#### Conclusion:
Comparison of subject and predicate devices identified technological similarities and differences. Performance testing was performed to evaluate substantial equivalence of dissimilar technological characteristics. Based on technological comparison to predicate devices and the results of pre-clinical performance testing, the Ultimate Contour device is substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.