K141708 · Syneron Medical, Ltd. · OHV · Oct 6, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141708
Device Name
ULTRASHAPE SYSTEM
Applicant
Syneron Medical, Ltd.
Product Code
OHV · General, Plastic Surgery
Decision Date
Oct 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference.
Device Story
UltraShape System uses focused ultrasound energy to disrupt subcutaneous adipose tissue (SAT) for non-invasive aesthetic body contouring. System comprises a control unit and two ultrasonic transducers; transducers convert electrical signals into mechanical (acoustical) energy via piezoelectric ceramic elements. Device targets SAT while sparing neighboring tissues like blood vessels, nerves, and muscle. Real-time optical and acoustic feedback is provided via a tracking and guidance system. Operated by clinicians in a clinic setting. Output is focused ultrasound energy; healthcare providers use the system to achieve abdominal circumference reduction. Benefits include non-invasive fat reduction with minimal discomfort and high patient satisfaction.
Clinical Evidence
Clinical evaluation included two prospective, single-arm studies (n=102) and one additional safety study (n=21). Subjects (mean age 44, majority female) received three treatments at two-week intervals with follow-up at 2, 4, 8, and 12 weeks. Primary endpoint was abdominal circumference reduction. Results showed average midline reduction of -2.37 cm at 12 weeks with minimal weight change (<1%). Safety profile was positive; adverse events were mild to moderate, resolving within hours or days. No serious device-related adverse events reported.
Technological Characteristics
System includes console with computer and two ultrasonic transducers. Transducers utilize piezoelectric effect with ceramic elements. Features real-time optical/acoustic tracking and guidance. Biocompatibility per ISO 10993. Electrical safety per IEC 60601-1; EMC per IEC 60601-1-2. Focused ultrasound energy delivery.
Indications for Use
Indicated for non-invasive reduction in abdominal circumference in patients seeking aesthetic body contouring.
Regulatory Classification
Identification
A Focused Ultrasound Stimulator System for Aesthetic Use is a device using focused ultrasound to produce localized, mechanical motion within tissues and cells for the purpose of producing either localized heating for tissue coagulation or for mechanical cellular membrane disruption intended for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of an eagle with three profiles of human faces incorporated into its design. The eagle's head and beak are formed by the first face, and the body and wings are formed by the other two faces.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 6, 2014
Syneron Medical Ltd. c/o Ms. Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103
Re: K141708
Trade/Device Name: Syneron UltraShape System Regulation Number: 21 CFR 878.4590 Regulation Name: Focused ultrasound stimulator system for aesthetic use Regulatory Class: Class II Product Code: OHV Dated: September 12, 2014 Received: September 12, 2014
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Binita S. Ashar -S 2014.10.06 08:36:38 -04'00' Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement on last page |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| Indications for Use |
|---------------------|
|---------------------|
| 510(k) Number (if known) | K141708 |
|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | UltraShape System |
| Indications for Use (Describe) | The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference. |
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)<br/>Subpart C) </div> | Over-The-Counter Use (21 CFR 801 |
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review
instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information.
Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing
this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number. "
FORM FDA 3881 (1/14)
PSC Publishing Services (301) 443-6740 EF
{3}------------------------------------------------
### 510(k) SUMMARY
### Syneron Medical Ltd.'s UltraShape System
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Syneron Medical Ltd. P.O.B. 550 Industrial Zone Tavor Building Yokneam Illit, 20692 Israel Ruthie Amir, MD, Global Vice President of Clinical and Regulatory Affairs Phone: 972-73-244-2200 Facsimile: 972-73-244-2202
Contact Person: Janice Hogan Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103 Phone: 267 675 4600 Facsimile: 267 675 4601
Date Prepared: October 2, 2014
#### Name of Device
Syneron UltraShape System
#### Common or Usual Name
Focused Ultrasound Stimulator System for Aesthetic Use
## Classification
Focused Ultrasound Stimulator System for Aesthetic Use
21 CFR 878.4590, Class II, product code OHV
#### Predicate Devices
Syneron Contour I V3.1 (K133238)
#### Intended Use / Indications for Use
The UltraShape System delivers focused ultrasound energy that can disrupt subcutaneous adipose tissue (SAT) to provide a non-invasive approach to achieve a desired aesthetic effect. It is intended for non-invasive reduction in abdominal circumference.
#### Device Description
The UltraShape System is comprised of multiple components, including the control unit and two ultrasonic transducers. The UltraShape System selectively targets subcutaneous
{4}------------------------------------------------
adipose tissue via focused ultrasound for the purpose of non-invasive aesthetic body contouring. The transducer is an electro-mechanical device that converts an electrical signal into mechanical (acoustical) energy. The operating parameters of the UltraShape System achieve selective disruption of adipose tissue without damaging neighboring tissues such as blood vessels, nerves, or muscle.
The purpose of this submission is to increase acoustic intensity and to add an additional smaller ultrasound transducer.
## Technological Characteristics
The UltraShape System is comprised of the system console, including the computer, and two ultrasonic transducers. The transducers deliver the focused ultrasound energy beam to the targeted treatment area, and real-time optical and acoustic feedback (optional) on the treatment is provided via the tracking and quidance system. The transducers functionality is based on the piezoelectric effect implemented with the ceramic element.
## Performance Data
The following nonclinical performance testing was conducted to support the substantial equivalence of the UltraShape System to the predicate device, consistent with FDA's "Class II Special Controls Guidance Document: Focused Ultrasound Stimulator System for Aesthetic Use" (2011). In all instances, the UltraShape System functioned as intended.
- . Biocompatibility testing in accordance with ISO 10993 for skin irritation, sensitization, cytotoxicity testing supported the biocompatibility of the patient-contacting components of the device.
- Beam profile testing demonstrated that the acoustic energy is delivered and . concentrated in the desired target location, at a focal depth highly similar to that of the predicate device.
- Acoustic power testing demonstrated that the acoustic power of the transducers is highly predictable with low variability.
- . In vitro acoustic and thermal measurements demonstrated the safety of non-targeted tissues both proximal and distal to the targeted region.
- . Software verification and validation was performed, and demonstrated that the software performs as intended.
- . Electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2) and electromagnetic immunity testing was conducted and demonstrated the electrical safety of the device.
- In vivo testing in an animal model was performed which demonstrated the treatment effects of the UltraShape, and supported its safety and efficacy profile for the intended use.
In addition, clinical evaluation of the device in the intended population was performed in two, separate, prospective, single arm studies. The studies evaluated the safety and
{5}------------------------------------------------
effectiveness of the UltraShape in 102 enrolled subjects. The majority of the subjects were female with mean age of 44 years. After baseline screening, patients underwent three treatments each at two weeks apart, with follow-up visits at 2, 4, 8 and 12 weeks after the The abdominal circumference reduction, weight, satisfaction and safety last treatment. results were assessed for the study subjects throughout the study visits. The results demonstrated that at final follow-up, subjects achieved an average of -2.37 cm circumference reduction at the midline. Similar treatment results were observed at the measurement points 2 cm above and 2 cm below the midline. It should also be noted that very little weight change was observed in the study subjects overall, with an average weight change measuring less than 1% at 12 weeks (or longer term follow up) for all subjects. Midline circumference reduction results in the weight stable subjects (-2.18 cm) were very similar to the results observed in the overall population. Therefore, the subject device demonstrated substantially equivalent performance over time compared to the predicate in the overall study population and in the weight stable subgroup. Additional sensitivity analyses, including last observation carried forward (LOCF) and worst case analyses, continue to demonstrate the robust treatment effect of the UltraShape in abdominal circumference reduction. The safety profile of the UltraShape device was maintained through long term follow up: the only events in the studies were mild to moderate in severity and resolved within a few hours or days. In addition, the majority of subjects in the UltraShape studies reported overall satisfaction with treatment. An additional prospective clinical study assessed the device safety profile in 21 patients. Subjects underwent 3 treatment sessions each 2 weeks apart; follow-up visits occurred at 2, 4, 8 and 12 weeks after the last treatment visit. The study results further supported the safety profile of device treatment with only one minor event from which the patient completely recovered without the need for any intervention. Subjects reported minimal discomfort and overall satisfaction with the treatments.
Therefore, multiple clinical studies with the UltraShape device demonstrated successful circumference reduction and a positive safety profile, with no device related serious adverse events. Thus, the clinical studies support substantial equivalence to the predicate.
# Substantial Equivalence
The UltraShape has the same indications for use, similar technological characteristics and the same principles of operation as its predicate device. The technological differences between the UltraShape and the Contour I V3.1 mainly consist of differences in acoustic intensity and the addition of a smaller transducer. Nonclinical and clinical studies of the UltraShape have demonstrated the safety and effectiveness profile of the UltraShape in the intended population. Thus, the UltraShape is substantially equivalent to the predicate device.
# Conclusion
Syneron's UltraShape System is a Focused Ultrasound Stimulator System for Aesthetic Use Class II device that has been evaluated in nonclinical testing in accordance with FDA's Special Controls Guidance Document. Testing demonstrated that the device performs as intended. The UltraShape device is substantially equivalent to its predicate device (K133238).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.