Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.
Device Story
The Optomed Smartscope M5 with FA optics module is a handheld ophthalmic camera used in medical environments to capture fluorescein angiograms. The device consists of the M5 camera body and an interchangeable FA optics module attached via bayonet connector. The FA module utilizes blue and infrared light sources and internal fixation targets to guide light through the pupil to the fundus. The camera sensor captures images/video, which are stored on a memory card or transferred to a PC via USB or WLAN. The device is operated by trained healthcare professionals. Captured images are used for documentation and clinical consultation, assisting physicians in visualizing retinal vasculature to support diagnostic decision-making. The device is powered by a rechargeable Ni-MH battery and includes an eye cup for patient contact.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), optical safety (IEC 62471, ANSI Z80.36-2016), biocompatibility (ISO 10993-5/10), optical performance (ISO 10940), and usability (IEC 62366-1).
Technological Characteristics
Handheld ophthalmic camera with interchangeable optics modules. FA module uses blue (Osram Oslon LB H9GP) and NIR (OSRAM Oslon SFH-4715) LEDs. 5MP color CMOS sensor. 2.4" TFT-LCD display. Diopter compensation -20D to +20D. 40-degree FOV. Connectivity: USB 1.1, WLAN. Power: 4.8V 1000mAh Ni-MH battery. Eye cup material: Momentive LIM 6030AB. Software: IEC 62304 compliant. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10940, ISO 15004-1/2.
Indications for Use
Indicated for capturing fluorescein angiograms of the human eye in patients requiring such imaging. Prescription use only. Must be operated by properly trained personnel.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Optomed Smartscope M5 digital camera / Optomed Smartscope EY4 optics module (K132186)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Optomed Oyj Jyri Leskelä Director Q&RA Yrttipellontie 1 Oulu, 90230 Finland
Re: K201325
Trade/Device Name: Optomed Smartscope M5 with Optomed Smartscope FA Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: May 7, 2020 Received: May 18, 2020
Dear Jyri Leskelä:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Elvin Ng
Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201325
Device Name Optomed Smartscope M5 Optomed Smartscope FA
Indications for Use (Describe)
Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| 510(k) submitter: | Optomed Oyj | |
|--------------------------------|---------------------------------------------------|------------------------------------------------------------------------------|
| Address: | Yrttipellontie 1, FI-90230 Oulu, Finland | |
| Company phone: | +358 20 741 3380 | |
| Contact person: | Mr. Jyri Leskelä, Quality and Regulatory Director | |
| Date: | October 15, 2020 | |
| Subject device: | Trade name: | Optomed Smartscope FA |
| | Common/usual name: | Ophthalmic camera |
| | Classification name: | camera, ophthalmic, ac-powered (21 CFR 886.1120) |
| | Regulatory Class: | II |
| | Product Code: | HKI |
| Predicate device:<br>(Primary) | Trade name: | Optomed Smartscope M5 digital camera<br>Optomed Smartscope EY4 optics module |
| | 510(k) number: | K132186 |
| | Regulatory Class: | II |
| | Product code: | HKI |
| Predicate device: | Trade name: | Kowa Genesis Df |
| | 510(k) number: | K091098 |
| | Regulatory Class: | II |
| | Product code: | HKI |
Image /page/3/Picture/4 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a circle with a smaller circle inside of it.
{4}------------------------------------------------
### Description of the Device
Optomed Smartscope M5 camera with Smartscope FA optics module is designed for use in a medical environment. Captured images and video are used for documentation and consultation. Optomed Smartscope M5 camera has a memory card where captured images and recorded videos are saved. Optomed Smartscope M5 camera is used with interchangeable optics modules previously cleared Smartscope EY4 and Smartscope ES2 and subject device Smartscope FA. Optics modules are attached to the camera with bayonet connectors.
Optomed Smartscope FA optics module is designed for fundus fluorescein angiography imaging of the human eye. Optomed Smartscope FA optics module includes two light sources: Blue light and infrared light, optical component that reflects light to the optical path and lenses that guide light through the pupil to the back of the eye and back to the camera sensor. Optomed Smartscope FA optics module has nine internal fixation targets for the patient to fixate on during imaging. Optomed Smartscope FA optics module receives power from Optomed Smartscope M5 digital camera.
Optomed Smartscope M5 camera has a WLAN module inside and when WLAN is used, Optomed Smartscope M5 camera transfers captured images and recorded videos to the PC automatically immediately after imaging. Images and videos can also be transferred to PC from the memory card when the camera is placed on Charging Station and the USB cable is connected between Charging Station and the PC. The image data transfer method to PC is similar as with any other USB mass storage device.
Optomed Smartscope M5 camera has a rechargeable Ni-MH battery that is charged when the camera is placed on Charging Station, which is connected to the mains by power supply cable. Charging Station can also be used as an external battery charger for the spare battery included in sales case. When Optomed Smartscope M5 camera is not used, it may be stored on Charging Station. Storing the device on Optomed Smartscope M5 camera Charging Station is not harmful for the battery.
Image /page/4/Picture/6 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a blue circle with a white dot in the center.
{5}------------------------------------------------
# Indications for Use
Optomed Smartscope M5 digital camera with Smartscope FA optics module is intended to capture fluorescein angiograms of the human eye.
Optomed Smartscope FA is classified as prescription device. Federal law restricts this device to sale by or on the order of a physician or licensed practitioner. The device may only be operated by persons who have been properly trained or who have the required knowledge and experience to do so. The device may only be used in accordance with its intended use.
Optomed Smartscope M5 camera is modular and can be used together with dedicated optics modules (Smartscope EY4 and Smartscope FA) to capture images from the fundus of the eye. Predicate device Kowa Genesis Df is used to fundus imaging as well. Changes implemented to subject device Smartscope FA do not affect safety and effectiveness of the product.
# Comparison of Technological Characteristics
Table 1 below includes a summary of the technical information used in the substantial equivalence discussion.
| Point of comparison | Optomed Smartscope M5<br>Optomed Smartscope FA | Optomed Smartscope M5<br>Optomed Smartscope EY4 | Kowa Genesis Df |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| 510(k) number | K201325 | K132186 | K091098 |
| Indications for use | Optomed Smartscope M5<br>digital camera with<br>Smartscope FA optics<br>module is intended to<br>capture fluorescein<br>angiograms of the human<br>eye. | Optomed Smartscope M5<br>digital camera with<br>Smartscope EY4 optics<br>module is intended to<br>capture digital images and<br>video of the fundus of the<br>human eye. | A hand-held mydriatic<br>retinal camera which<br>captures fundus image<br>(For color and FA retinal<br>photography) |
| Usage | Prescription use | Prescription use | Prescription use |
| Use condition | Intended to be used with<br>mydriatic drops. | Intended to be used<br>without mydriasis but can<br>be used also with<br>mydriatic drops. | Intended to be used with<br>mydriatic drops |
Image /page/5/Picture/9 description: The image shows the word "OPTOMED" in blue font. The two Os are stylized, with the second O having a circle around it and a dot in the center, resembling an eye. The font is bold and sans-serif, and the overall design is simple and modern. The background is a light gray color.
{6}------------------------------------------------
| Illumination source | Smartscope FA:<br>Blue: Osram Oslon LB<br>H9GP<br>NIR: OSRAM Oslon SFH-<br>4715<br>Target LEDs: OSRAM LR<br>QH9F | Smartscope EY4:<br>White: OSRAM Oslon<br>LUW-H9GP<br>NIR: OSRAM Oslon SFH-<br>4715<br>Target LEDs: OSRAM LR<br>QH9F | Observation light source:<br>white LED / blue LED<br>Photographing light source:<br>Xenon flash lamp 23WS |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Display system | 2.4", TFT-LCD, 240x320<br>pixels, 262 000 colors,<br>antireflective coating | 2.4", TFT-LCD, 240x320<br>pixels, 262 000 colors,<br>antireflective coating | Visual observation |
| Camera sensor<br>specification | Color CMOS camera<br>maximum resolution 5 Mp | Color CMOS camera<br>maximum resolution 5 Mp | Color CCD camera<br>2,000,000 pixels |
| Diopter compensation | From -20 D to +20 D | At least from -20 D to +20<br>D | -15D~+35D |
| Field of view | 40 degrees | 40 degrees | Horizontal: 30 degree<br>Vertical: 25 degree |
| Storage media | MicroSDHC memory card | MicroSDHC memory card | Flash memory card |
| Image data format | JPEG, MPEG-1, MPEG-4 | JPEG, MPEG-1, MPEG-4 | JPEG, uncompressed data |
| Weight | Smartscope M5: 400g<br>Smartscope FA: 300g | Smartscope M5: 400g<br>Smartscope EY4: 300g | Approx.1kg |
| Battery | Re-chargeable Ni-MH<br>battery,<br>HR4U700AAA, 4.8V, 1000<br>mAh | Re-chargeable Ni-MH<br>battery,<br>HR4U700AAA, 4.8V, 1000<br>mAh | Not battery-operated<br>device |
| Output terminals and data<br>collection | USB (1.1) terminal (B-<br>connector).<br>Compatible with<br>Windows® 7/8.1/10. | USB (1.1) terminal (B-<br>connector).<br>Compatible with<br>Windows® 7/8.1/10. | USB (1.1) terminal (B<br>connector)<br>Foot Switch connector<br>cable terminal |
| Exposure parameters | "Exempt Group" (no risk)<br>LED product according to<br>IEC 62471:2006 | "Exempt Group" (no risk)<br>LED product according to<br>IEC 62471:2006 | N/A |
Image /page/6/Picture/3 description: The image shows the word "OPTOMED" in blue font. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a circle with a smaller circle inside, and a white dot in the center.
{7}------------------------------------------------
| | Group 1 instrument<br>according to ANSI Z80.36-<br>2016 | Group 2 instrument<br>according to ISO 15004-<br>2:2007 | |
|-----------|---------------------------------------------------------|---------------------------------------------------------|-----|
| Standards | IEC 60601-<br>1:2005+A1:2012 (edition<br>3.1) | IEC 60601-<br>1:2005+A1:2012 (edition<br>3.1) | N/A |
| | IEC 60601-1-2:2014<br>(edition 4.0) | IEC 60601-1-2:2014<br>(edition 4.0) | |
| | IEC 60601-1-<br>6:2010+A1:2013 (edition<br>3.1) | IEC 60601-1-<br>6:2010+A1:2013 (edition<br>3.1) | |
| | IEC 62471:2006 | IEC 62471:2006 | |
| | ISO 15004-1:2006 | ISO 15004-1:2006 | |
| | ANSI Z80.36-2016 | ISO 15004-2:2007 | |
| | ISO 10940:2009 | ISO 10940:2009 | |
| | ISO 10993-5:2009 | ISO 10993-5:2009 | |
| | ISO 10993-10:2009 | ISO 10993-10:2009 | |
| | IEC 62304:2006+A1:2015 | IEC 62304:2006+A1:2015 | |
| | IEC 62366-1:2015 | IEC 62366-1:2015 | |
| | | | |
The modifications in Optomed Smartscope M5 camera with Smartscope FA optics module compared to Optomed Smartscope M5 camera with Optomed Smartscope EY4 optics module are:
Smartscope FA:
- . Illumination source blue led
- . Short pass filter in illumination path in front of blue led
- Barrier filter in imaging path
- . Optical safety Group 1 tested according to ANSI Z80.36-2016
The differences in Optomed Smartscope M5 camera with Smartscope FA optics module compared to Kowa Genesis Df are:
Smartscope FA:
- . Increased FOV (field of view)
- Higher image resolution
OPTOMED
{8}------------------------------------------------
## Performance Data
The following performance data is provided in support of the substantial equivalence determination.
### Electrical safety and electromagnetic compatibility (EMC)
Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to all suitable clauses of IEC 60601-1:2005+A1:2012 (safety) and IEC 60601-1-2:2014 (EMC). Optomed Smartscope M5 camera with Smartscope FA optics module fulfills the requirements of the standards.
#### Optical safety
Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to the standard IEC 62471:2006 and is classified as "Exempt Group" (NO RISK) LED products.
Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to the standard ANSI Z80.36-2016 and is classified as Group 1 ophthalmic instruments.
### Software Verification and Validation
The level of concern of the software is moderate. Software verification and validation were conducted to ensure the fulfillment of the system requirements and functional specifications. Optomed Smartscope M5 camera with Smartscope FA optics module complies with the standard IEC 62304:2006+A1:2015.
### Environmental testing
Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to ISO 15004-1:2006, IEC 60601-1 and IEC 60068-2 standards to verify the mechanical stress and ambient conditions for use and storage as prescribed for the device fulfills the requirements of the standard. In addition, Optomed Smartscope M5 camera with Smartscope FA optics module was tested according to IEC 60068-2 for high temperature, low temperature, shock, vibration and bump to verify transportation conditions.
#### Biocompatibility
Optomed Smartscope FA optics module has an eye cup that contacts skin around eye during normal usage of the device. The material of Optomed Smartscope FA optics module eye cup is Momentive LIM 6030AB. In vitro cytotoxicity tests according to ISO 10993-5:2009 and Irritation and skin
Image /page/8/Picture/15 description: The image shows the logo for Optomed. The logo is blue and features the word "OPTOMED" in a bold, sans-serif font. The "O" in "OPTOMED" is replaced by a stylized eye symbol, which is also blue. The background of the image is a light gray color.
{9}------------------------------------------------
sensitization tests according to ISO 10993-10:2009 were performed to the material of the eye cup, and the biocompatibility of Optomed Smartscope FA optics module is at suitable level.
#### Optical performance
Optomed M5 camera with Optomed Smartscope FA optics module fulfills the requirements of the standard ISO 10940:2009 Ophthalmic instruments - Fundus cameras.
#### Usability (Human Factors)
Optomed M5 camera with Optomed Smartscope FA optics module was designed and evaluated by following the principles depicted in the usability engineering process. The usage of Optomed M5 camera with Optomed Smartscope FA optics module was evaluated to be suitable for its intended use and the devices complies with the standards IEC 60601-1-6:2010+A1:2013 and IEC 62366-1:2015.
### Conclusions
Based on the provided performance data and the comparison, Optomed M5 camera with Optomed Smartscope FA optics module, is as safe, as effective and performs similar to the predicate devices. Optomed M5 camera with Optomed Smartscope FA optics module is substantially equivalent to the predicate devices.
Image /page/9/Picture/8 description: The image shows the word "OPTOMED" in blue. The "O" in "OPTOMED" is replaced with an eye symbol. The eye symbol is a blue circle with a white dot in the center.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.