Optomed Smartscope M5 digital camera with EY4 optics module is intended to capture digital images and video of the fundus of the human eye. Optomed Smartscope M5 with optics module ES2 is intended to capture images and video of the surface of the human eye and surrounding areas.
Device Story
Handheld digital ophthalmic camera system; consists of M5 handset, attachable optics modules (EY4 for fundus, ES2 for anterior segment), and charging/transfer cradle. EY4 module uses white/infrared LEDs and fixation target LEDs to capture retinal images. ES2 module uses white/cobalt blue LEDs for anterior imaging. Device utilizes 5MP CMOS sensor; stores images/video on flash memory card; transfers data to PC via USB. Operated by clinicians in clinical settings. Provides digital visualization of ocular structures to assist in ophthalmic examination and documentation. Benefits include portable, non-mydriatic-capable imaging for eye health assessment.
Clinical Evidence
Bench testing only. Evaluated optical radiation safety per ISO 15004-2:2007, including photometric luminance, UV/IR radiation, and blue-light weighted radiance. Device classified as Class 2 instrument per ISO 15004-2:2007. Performance verified through side-by-side comparison with predicate and evaluation by development team and practicing physicians.
Technological Characteristics
Handheld ophthalmic camera; 5MP CMOS sensor; 2.4" TFT LCD display. EY4 module: white/NIR LEDs, fixation target LEDs. ES2 module: white/cobalt blue LEDs, internal baffle/diffuser. Connectivity: USB 1.1. Power: 4.8V Ni-MH battery. Standards: IEC 60601-1 (3rd ed), IEC 60601-1-2, ISO 15004-2:2007, EN 62471:2008. Dioptre compensation: -20D to +20D.
Indications for Use
Indicated for capturing digital images and video of the human eye fundus (EY4 module) or the surface of the eye and surrounding areas (ES2 module). Intended for prescription use; can be used with or without mydriatic agents.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Predicate Devices
Optomed Smartscope M5 with optics modules EY3 with ES1 (K110986)
Submission Summary (Full Text)
{0}------------------------------------------------
13.12.2013 1/3
DEC 18 2013
K132186
#### 510(k) Su nmary
| Date: | December 13, 2013 | |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) owner's name: | Optomed Oy | |
| Address: | Hallituskatu 13-17 D, 90100 Oulu, Finland | |
| Contact person: | Jyri Leskelä, Quality Manager | |
| Company phone: | +358 20 741 3380 | |
| Fax number: | +358 42 588 1187 | |
| Device name: | Trade name: Optomed Smartscope M5 digital camera with<br>Optomed Smartscope EY4 optics module and Optomed<br>Smartscope ES2 optics module<br>Common/usual name: Ophthalmic camera<br>Classification name: camera, ophthalmic; ac-powered (21 CFR<br>886.1120, Product code: HKI) | |
| Predicate device: | Optomed Smartscope M5 with optics modules EY3 with ES1<br>(510(k) number: K110986, Product code: HKI) | |
| Indications for Use: | Optomed Smartscope M5 digital camera with EY4 optics module is<br>intended to capture digital images and video of the fundus of the<br>human eye. Optomed Smartscope M5 with optics module ES2 is<br>intended to capture images and video of the surface of the human<br>eye and surrounding areas. | |
| Point of comparison | Optomed Smartscope M5 with EY4 and ES2 | Optomed Smartscope M5 with EY3 and ES1 |
| Indications for use | Optomed Smartscope M5 digital camera with EY4 optics module is intended to capture digital images and video of the fundus of the human eye. Optomed Smartscope M5 with optics module ES2 is intended to capture images and video of the surface of the human eye and surrounding areas. | Optomed Smartscope M5 camera with optics modules EY3 and ES1 is digital ophthalmoscope intended to capture digital images and video of the fundus of the human eye and surrounding area. |
| Usage | Prescription use | Prescription use |
| Use condition | Intended to use without mydriatic but can be used also with mydriatic | Intended to use without mydriatic but can be used also with mydriatic |
| Observation light source | Visible and infrared LED. EY4 LEDs:<br>White: OSRAM Oslon LUW-H9GP<br>NIR: OSRAM Oslon SFH-4715<br>Target LEDs: OSRAM LR QH9F<br>ES2 LEDs: White : Osram Advanced Power Topled LW G6SP-EAFA-JKQL-1<br>Blue : Osram Advanced Power Topled LB G6SP-V2BB-35-1 | Visible and infrared LED. EY3 LEDs:<br>White: LWW5SM-KY-QK.<br>NIR: SFH4232 (bin DA)<br>ES1 LEDS:: White : Osram Advanced Power Topled LW G6SP-EAFA-JKQL-<br>Blue : Osram Advanced Power Topled LB G6SP-V2BB-35-1 |
| Observation and display system | 2.4" active matrix color TFT LCD | 2.4" active matrix color TFT LCD |
| Photographing light source | Visible and infrared LED | Visible and infrared LED |
| Camera specification | Color CMOS camera maximum resolution 5Mp. ES2 uses maximum resolution. EY4 uses 1,77 Mp. | Color CMOS camera maximum resolution 5Mp. ES1 uses maximum resolution. EY3 uses 2,76 Mp. |
| Dioptre compensation (patient) | at least -20 D to +20 D | at least -20 D to +20 D |
| Picture angle | 40 degrees | Over 40 degrees |
| Storage media | Flash memory card | Flash memory card |
| Image data format | JPEG, MPEG-1, MPEG-4 | JPEG, MPEG-1, MPEG-4 |
| Weight | Camera: 400g, EY4: 300g, ES2: 92g | Camera: 400g, EY3: 180g, ES1: 80g |
| Power consumption | Re-chargeable Ni-MH Battery 4.8V;<br>Charging unit 44 VA | Re-chargeable Ni-MH Battery 4.8V;<br>Charging unit 44 VA |
| Output terminals and | USB (1.1) terminal (B-connector) | USB (1.1) terminal (B-connector) |
## Device description, Intended use & Effectiveness:
Optomed Smartscope M5 digital camera with EY4 optics module is intended to capture digital images and video of the fundus of the human eye.
The device set for retinal imaging consists of:
- Camera handset M5 .
- Attachable ophthalmic lens EY4 ●
- Eye cup for EY4 .
- Cradle for charging and image transfer
- Slit Lamp Adapter
Optomed Smartscope M5 with optics module ES2 is intended to capture images and the surface of the human eye and surrounding areas.
. :
i l
The device set for anterior ophthalmic imaging consists of
- Camera handset M5
- Attachable anterior ophthalmic lens ES2
- Cradle for charging and image transfer
Optomed O
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the word "OPTOMED" in a bold, sans-serif font. The "O" in "OPTOMED" is replaced with an eye symbol, where the pupil is a solid black circle and the iris is a white circle surrounded by a black outline. The letters are all capitalized and evenly spaced.
EY4 has an LED light source with visible white and infrared light. Also red light target LED's are used to eye position fixation during imaging. ES2 has an LED light source with white light and cobalt blue light. Image data is stored on the Flash memory card using 5 megapixel CMOS sensor and transferred to the PC by using USB connection. Device has rechargeable battery. Table 1 below includes a summary of the technical information used in the substantial equivalence comparison.
Optomed Oy,
Hallituskatu 1 90100 Oulu Finland 14
: ్ర
XP/VISTA/7
XP/VISTA/7.1 ..
{2}------------------------------------------------
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Administration<br>Acres of can and<br>4<br>Acres Access characterial | Stanner<br><br> | 1<br>Achelond Arthul<br>****<br>1 | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------------------------|--|
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|-----------------------------------|--|
| Exposure parameters | "Exempt Group" (no risk) LED product<br>according to EN 62471:2008<br>Group 2 instrument according to ISO<br>15004-2:2007 | Class 1 LED according to IEC 60825-<br>1:2001<br>Group 1 instrument according to ISO<br>15004-2:2007 |
|---------------------|---------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Standards | IEC 60601-1:2005 (3rd edition)<br>IEC 60601-1-2:2007 (3rd edition)<br>EN 62471:2008<br>ISO 15004-2:2007 | IEC 60601-1:1988+A1+A2<br>IEC 60601-1-2:2001+A1 IEC 60601-1-<br>4:2000<br>EN (IEC) 60825-1:2001<br>+A1:2002+A2:2001<br>ISO 15004-2:2007 |
Technical Information was gathered by side-by-side comparison and labeling of the devices. Optomed Smartscope M5 with optics modules EY4 and ES2 and its predicate was also evaluated by Optomed development team and practicing physicians throughout the product development.
Following laboratory bench tests demonstrate Optical radiation safety of the Optomed Smartscope M5 with EY4 and ES2:
- 1. Optical radiation hazards as defined by standard ISO 15004-2, Ophthalmic instruments -- Fundamental requirements and test methods -- Part 2: Light hazard protection, First edition 2007-02-15.
- 2. Maximum photometric luminance
- the optical radiation hazards as defined by ISO 15004 for: 3. (i) ultraviolet and infrared radiation, and (ii) visible light and near IR radiation (blue-light weighted radiance and aphakic weighted radiance).
Optomed Smartscope M5 with optics modules EY4 and ES2 is classified as Class 2 instrument according to the international standard ISO 15004-2:2007.
Optomed Smartscope EY3 eye optics is a predecessor of EY4 eye optics The modifications in M5 camera with EY4 optics compared to M5 camera with EY3 optics are:
- Resolution is improved to meet the requirements of ISO10940 (Fundus camera standard) .
- Target LED's for eye position fixation during imaging
- 9 pcs of LEDs at different viewing angles o
- o Allows stabilizing the eye of the patient for a specific direction.
Optomed Smartscope ES1 eye optics is a predecessor of ES2 eye optics. The modifications in M5 camera with ES2 optics compared to M5 camera with ES1 optics are:
- Illumination concept is improved
Finland
- New diffuser structure to improve image quality o
- Separate baffle is included in the illumination path to block excess reflections of o stray light from the sides of the module
. i .
- to enhance image contrast : O
to allow shorter camera integration times to reduce n - O
Based on the comparison and supportive information, the subject device is subs equivalent to the predicate device.
Od.Oy
katu 13-17,D
Oulu
www.optomed.com
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo features a stylized symbol with three abstract human figures connected in a circular shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is printed in a simple, sans-serif font to the right of the symbol.
Food and Drug Administration 10903 New Hampshire Avenu Document Control Center - WO66 Silver Spring, MD 20993-0002
December 18, 2013
Optomed Oy % Mr. Jyri Leskelä Quality Manager Hallituskatu 13-17 FI 90100, Oulu Finland
Re: K132186
Trade/Device Name: Optomed Smartscope M5 digital camera with Optomed Smartscope EY4 optics module and Optomed Smartscope ES2 optics module Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: November 4, 2013 Received: November 6, 2013
Dear Mr. Leskelä:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{4}------------------------------------------------
Page 2 - Mr. Jyri Leskelä
- « « « » » » » » « » « » « » »
Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-204] or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Kesia Y.Alexander -S
for Malvina B. Eydelman, M.D.
Director
Division of Ophthalmic, and Ear, Nose, Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
#### 510(k) Number (if known): K132186
Device Name: Optomed Smartscope M5 digital camera, Optomed Smartscope EY4 optics module, Optomed Smartscope ES2 optics module
#### Indications for Use:
.
Optomed Smartscope M5 digital camera with EY4 optics module is intended to capture digital images and video of the fundus of the human eye.
Optomed Smartscope M5 with optics module ES2 is intended to capture images and video of the surface of the human eye and surrounding areas.
Prescription Use AND/OR (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Andrew Yang -S
2013.12.15
21:44:49 -05'00'
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.