The Axonpen System is indicated for the illumination and visualization of intracranial tissue and fluids and the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum.
Device Story
Axonpen System comprises a steerable, single-use neuroendoscope (Axonpen) and a reusable monitor (Axonmonitor). Device inputs include visual data from an integrated distal camera and LED light source. Surgeon operates mechanical controls on the handle to steer the distal tip (horizontal plane), extend the distal tube, and manage irrigation/aspiration. Aspiration utilizes regulated hospital vacuum; irrigation uses third-party saline. Axonmonitor (Axonbox + tablet) processes camera signals for real-time display and recording. Used in neurosurgery to visualize and evacuate intracranial tissue/fluids. Output provides surgeons with direct visualization of the surgical field, facilitating controlled tissue/fluid removal. Benefits include improved visualization and efficient evacuation of intracranial targets without requiring repositioning of the entire endoscope.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993), electrical safety (IEC 60601), and optical performance verification (noise reduction, contrast enhancement, resolution, distortion, dynamic range). Bench testing confirmed successful aspiration of fluid and clots without damaging adjacent tissue.
Technological Characteristics
Materials: 304 Stainless Steel, TPU, Polyurethane. Steerable neuroendoscope with integrated LED and camera. Connectivity: Wired connection to Axonmonitor (tablet-based). Power: Battery-operated monitor, hospital vacuum for aspiration. Software: Firmware for camera/light control and image processing. Sterilization: Single-use sterile endoscope; reusable monitor.
Indications for Use
Indicated for illumination, visualization, and controlled aspiration of intracranial tissue and fluids during surgery of the Ventricular System or Cerebrum.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
{0}------------------------------------------------
November 27, 2020
ClearMind Biomedical % Craig Coombs President Coombs Medical Device Consulting, Inc 1100 Pacific Marina. Suite 806 Alameda, California 94501
Re: K201308
Trade/Device Name: Axonpen, Axonmonitor, Axonbox, Tablet Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: October 27, 2020 Received: October 28, 2020
### Dear Craig Coombs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Adam D. Pierce, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201308
Device Name Axonpen System
#### Indications for Use (Describe)
The Axonpen System is indicated for the illumination of intracranial tissue and fluids and the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## A. Device Information:
| Category | Comments |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | ClearMind Biomedical<br>5F, No. 167, Fuxing N. Rd., Songshan Dist., Taipei City<br>105, Taiwan, R.O.C.<br>+886-2-22697417 |
| Correspondent Contact<br>Information: | Craig Coombs<br>Coombs Medical Device Consulting<br>1100 Pacific Marina, Suite 806<br>Alameda, CA 94501<br>Tel: 510-995-8499 |
| Device Common Name: | Endoscope, neurological |
| Device Classification & Name: | 21 CFR 882.1480, Neurological endoscope |
| Device Classification &<br>Product Code: | Class II,<br>GWG |
| Device Proprietary Name: | Axonpen™ System |
## Predicate Device Information:
| Predicate Device: | Artemis™ Neuro Evacuation Device |
|-------------------------------------------|----------------------------------|
| Predicate Device Manufacturer: | Penumbra Inc. |
| Predicate Device Premarket Notification # | K171332 |
| Predicate Device Common Name: | Endoscope, neurological |
| Predicate Device Classification & Name: | 21 CFR 882.1480, Neurological |
| | endoscope |
| Predicate Device Classification & | Class II, |
| Product Code: | GWG |
#### Predicate Device Information:
| Predicate Device: | Artemis™ Eye System |
|-------------------------------------------|-------------------------------|
| Predicate Device Manufacturer: | Penumbra Inc. |
| Predicate Device Premarket Notification # | K190719 |
| Predicate Device Common Name: | Endoscope, neurological |
| Predicate Device Classification & Name: | 21 CFR 882.1480, Neurological |
| | endoscope |
| Predicate Device Classification & | Class II, |
| Product Code: | GWG |
## B. Date Summary Prepared
24 Nov 2020
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for ClearMind Biomedical. The logo consists of a blue abstract shape on the left, resembling a stylized letter "C" or a series of connected arrows. To the right of the shape is the company name "CLEARMIND" in gray, with the word "BIOMEDICAL" in smaller, lighter gray letters underneath. The overall design is clean and modern, suggesting a focus on clarity and innovation in the biomedical field.
#### C. Description of Device
The Axonpen System consists of a neuroendoscope (Axonpen) and a monitor (Axonmonitor) to view the image from the distal end of the Axonpen.
The Axonpen is a steerable, single-use, neuroendoscope. It has an integrated camera and LED light source to allow the surgeon to view the surgical field at the distal end of the Axonpen.
The Axonpen has a working channel that can accept third-party neuroendoscopic tools with an OD of ≤2.5mm and a length of at least 41cm. That same channel can be used for irrigation and aspiration of the surgical target site. The Axonpen is connected to a hospital source vacuum to provide the aspiration feature and is connected to a thirdparty saline infusion bag to provide irrigation solution. The distal 2.57cm of the Axonpen can be steered to the left or right of the axis in the horizontal plane. Additionally, the end of the distal tube of the Axonpen can be extended an additional 3.2mm to facilitate aspiration of fluid and tissue.
The usable length of the Axonpen can be adjusted by setting the Stopper mechanism, which can limit the depth penetration of the device according to the physician's preference.
The steering, irrigation, aspiration, and distal tube extension are all hand-controlled with mechanical features in the handle of the Axonpen.
The Axonpen is connected by a cable to the Axonmonitor. The Axonmonitor is a combination of the Axonbox, which contains the firmware for the Axonpen camera and light, and a third-party tablet. The software on the tablet allows viewing of the camera image along with recording capabilities. The Axonmonitor is battery operated and provides all the electrical power needed by the Axonpen. The Axonmonitor is supplied with a battery charger that is plugged into the mains and can be charging when the Axonpen System is being used.
#### D. Indications for Use
The Axonpen™ System is indicated for the illumination and visualization of intracranial tissue and fluids and the controlled aspiration of tissue and/or fluid during surgery of the Ventricular System or Cerebrum.
{5}------------------------------------------------
K201308: Axonpen™ System
510(k) Summary
### E. Tabular Comparison of Application and Predicate Devices in Regard to Substantial Equivalence
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Axonpen™ System is<br>indicated for the<br>illumination and<br>visualization of intracranial<br>tissue and fluids and the<br>controlled aspiration of<br>tissue and/or fluid during<br>surgery of the Ventricular<br>System or Cerebrum. | The Artemis™ Neuro<br>Evacuation Device is used<br>for the controlled<br>aspiration of tissue and/or<br>fluid during surgery of the<br>Ventricular System or<br>Cerebrum in conjunction<br>with a Penumbra<br>Aspiration Pump. | The Artemis™ Eye System<br>is indicated to provide<br>visualization and<br>illumination of intracranial<br>tissue and fluids during<br>diagnostic and<br>therapeutic procedures. | The application device<br>Indications for Use is a<br>combination of the two<br>predicates. This<br>combination does not<br>raise new questions of<br>safety or efficacy that<br>were not raised in the<br>predicates |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for ClearMind Biomedical. The logo consists of a blue abstract shape on the left, followed by the word "CLEARMIND" in gray, and the word "BIOMEDICAL" in a smaller font size below it. The abstract shape appears to be two stylized arrows pointing in opposite directions.
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|--------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technology | The Axonpen neuro-<br>endoscope provides single<br>channel for irrigation,<br>aspiration and tool<br>placement for<br>neurosurgical field. The<br>distal tip is steerable.<br>Relies upon regulated<br>Hospital vacuum for<br>aspiration.<br>The distal end of the<br>scope has a camera + LED<br>lighting.<br>The reusable Axonmonitor<br>displays and records live<br>imaging captured by the<br>Axonpen camera | Provides aspiration of<br>neurosurgical targets via a<br>tube that is placed in a<br>neuroendoscope's<br>working channel. The<br>tube includes a macerator<br>to facilitate the<br>movement of tissue and<br>fluids through tube.<br>Provides proprietary<br>vacuum pump for the<br>aspiration | The Artemis Eye is a<br>neuro-endoscope<br>consisting of a camera,<br>two channels for irrigation<br>and aspiration/drainage,<br>and a working channel for<br>a surgical tool.<br>Artemis Eye Tablet - a<br>reusable component that<br>connects to the Artemis<br>Eye and displays live<br>imaging captured by the<br>Artemis Eye camera. | Technology is identical.<br>The application device is a<br>combination of the two<br>Artemis predicates. The<br>application device does<br>not have an internal<br>macerator, rather it relies<br>on using a larger diameter<br>aspiration tube to<br>facilitate evacuation of<br>tissue and fluids. |
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
| Manufacturer<br>Provided<br>System<br>Components | Neuro endoscope<br>(Axonpen™) that provides<br>working channel for<br>endoscopic tools, along<br>with irrigation and<br>aspiration.<br>Handpiece has integrated<br>camera & lighting.<br>Reusable Axonmonitor<br>tablet style display for<br>camera image and<br>recording.<br>Battery Charger. | Hand piece that provides<br>powered clot maceration<br>within its aspiration tube.<br>Aspiration pump &<br>pump tubing to hand<br>piece | A neuro endoscope that<br>has an Artemis Eye<br>camera at its distal end.<br>Channels for suction and<br>irrigation if required<br>Working channel to allow<br>passage of the Artemis<br>Neuro Evacuation device.<br>Tablet style display for<br>camera image<br>Battery Charger | Both the predicate<br>devices (in combination)<br>and the application device<br>are neuroendoscopes.<br>They both have distal<br>cameras and lighting that<br>are displayed on a tablet<br>sized monitor. They both<br>have at least one working<br>channel for<br>neuroendoscopic tools,<br>irrigation and<br>aspiration/drainage.<br>The predicate device has a<br>macerator in its aspiration<br>device along with its own<br>vacuum source. |
| Hospital<br>Supplied<br>Components | Vacuum source<br>Vacuum regulator<br>Irrigation fluid<br>Waste bucket | Endoscope<br>Vacuum source<br>Vacuum regulator | Camera<br>Lighting<br>Irrigation fluid<br>drain | Achieves same clinical<br>effect |
{7}------------------------------------------------
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image contains the logo for Clearmind Biomedical. The logo consists of a blue abstract symbol on the left, followed by the word "CLEARMIND" in gray, and the word "BIOMEDICAL" in a smaller font size below it. The abstract symbol appears to be two stylized arrows pointing to the left.
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aspiration<br>Source | Regulated hospital<br>vacuum | Penumbra Aspiration<br>Pump | Not described | The predicate and<br>application devices have<br>different sources of<br>regulated vacuum. The<br>source of the vacuum<br>does not raise new<br>questions of safety or<br>efficacy. |
| Aspiration<br>Control | Fingertip controlled via<br>pressing button on<br>Axonpen handle. | Fingertip controlled via<br>vacuum regulator hole on<br>Wand handle. | Not Described | Vacuum activation<br>controlled by user |
| Aspiration<br>regulation | Manually regulated at<br>hospital vacuum regulator | Manually regulated at<br>Artemis System Pump | Not Described | Vacuum regulated by user |
| Mechanism<br>of Clot<br>Aspiration | Vacuum aspiration aided<br>by large aspiration tube<br>inner diameter (0.248" ID)<br>and extender tube | Vacuum Aspiration aided<br>by a battery-powered<br>macerator in Evacuator<br>and ability to advance<br>20mm beyond distal end<br>of Artemis Eye neuro<br>endoscope | None | Both devices use vacuum<br>to aspirate clots and<br>tissue. The predicate uses<br>a macerator to facilitate<br>passage of debris from<br>the surgical field. The<br>application device uses a<br>larger diameter tube (i.e.<br>0.248" ID vs 0.048" —<br>0.100″ ID) to facilitate<br>debris passage. |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for Clearmind Biomedical. The logo consists of a blue abstract shape on the left, resembling a stylized letter C or a pair of overlapping arrows. To the right of the shape is the company name, "CLEARMIND," in a bold, gray sans-serif font. Below "CLEARMIND" is the word "BIOMEDICAL" in a smaller, lighter blue font.
| 510(k) Summary |
|----------------|
|----------------|
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|----------------------------------------|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Power Source<br>for Clot<br>Aspiration | Hospital sourced<br>regulated vacuum | Primary Cell Lithium Ion<br>Battery, 1.5 Volt DC Motor<br>contained within the<br>handle to power the<br>macerator<br>Proprietary vacuum<br>source | None | Same as previous |
| Steering | 10° Steerable in both<br>directions from axis along<br>horizontal plane. | Non steerable | Non steerable | Steerable application<br>device may allow for<br>clean-up of larger surgical<br>field without requiring<br>repositioning of the entire<br>endoscope. May pose<br>less risk than predicate<br>neuroendoscope. |
| Camera | Integrated, 80°<br>visualization at distal end. | None | Integrated, visualization at<br>distal end | Clinically identical |
| Lighting | Fiber optic Integrated LED<br>light source in Axonpen<br>handle. | None | Fiber optic Integrated LED<br>light source in handle | Identical |
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for CLEARMIND BIOMEDICAL. To the left of the text is a blue icon that looks like two arrows pointing in opposite directions. The text "CLEARMIND" is in gray, and the text "BIOMEDICAL" is in a smaller, light blue font below the word CLEARMIND.
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Display | Axonmonitor component<br>controls lighting and<br>visualization.<br>Direct electrical/ data<br>connection to Axonpen.<br>Video and still picture<br>recording.<br>Battery powered. | None | Artemis Eye Tablet - a<br>reusable component that<br>connects to the Artemis<br>Eye and displays live<br>imaging captured by the<br>Artemis Eye camera.<br>Battery powered. | Functionally identical |
| Handpiece<br>Dimensions | L: 33.0 cm<br>ID: 0.248"<br>OD: 0.256" (6.6mm) | L: 26.1 cm - 27.0 cm<br>ID: 0.048″ — 0.100″<br>OD: 00.058″ —0.109″ | Length: 25 cm<br>OD: 6 mm | The Axonpen and the<br>Artemis Eye<br>neuroendoscopes are<br>nearly the same outside<br>diameter |
| Wand Usable<br>Length | 5.0 to 10.0cm (12.0cm<br>with the adjustable<br>Stopper mechanism fully<br>compressed);<br>20.0cm with the Stopper<br>mechanism removed. | Increases usable length of<br>Artemis Eye by 2cm<br>because the Device can<br>extend beyond the distal<br>tip of the scope by 2cm. | 15cm | Application<br>neuroendoscope has a<br>shorter potential usable<br>length than the predicate.<br>No new questions of safety<br>or efficacy raised by this<br>3cm difference. |
| Patient<br>contacting<br>Wand<br>Materials | 304 Stainless Steel.<br>Polyurethane Fiber optic<br>line, TPU | 304 Stainless Stee | 304 Stainless Steel<br>Fiber optic line | Clinically identical; all<br>materials in all devices<br>have passed appropriate<br>biocompatibility testing. |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for ClearMind Biomedical. The logo consists of a stylized blue symbol on the left, resembling a stylized letter "C" or a curved arrow. To the right of the symbol is the text "CLEARMIND" in gray, with the word "BIOMEDICAL" in smaller, lighter gray letters underneath. The logo has a clean and modern design.
| Characteristic | Application Device:<br>Clearmind<br>Axonpen™ System | Predicate Device:<br>Penumbra<br>Artemis™ Neuro<br>Evacuation Device<br>(K171332) | Predicate Device:<br>Penumbra<br>Artemis™ Eye System<br>(K190719) | Impact on<br>Substantial Equivalence |
|----------------------------------|---------------------------------------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Aspiration<br>Tubing<br>Material | Polymers | Polymer | Not applicable | Clinically identical; all<br>materials in all devices<br>have passed appropriate<br>biocompatibility testing. |
| Irrigation<br>Tubing<br>Material | PVC | None,<br>provided via endoscope | Not applicable | Clinically identical; all<br>materials in all devices<br>have passed appropriate<br>biocompatibility testing. |
| Single use?<br>Sterile? | Axonpen- single use,<br>sterile<br>Display- reusable,<br>nonsterile | Hand Piece-single use,<br>sterile<br>Vacuum pump- reusable,<br>nonsterile. | Artemis Eye- single use,<br>sterile<br>Artemis Eye tablet-<br>reusable, nonsterile | In both cases the<br>components that are<br>inserted into the patient's<br>brain are single-use and<br>provided sterile and the<br>image monitors are<br>reusable.<br>The predicate's vacuum<br>pump is also reusable. |
{12}------------------------------------------------
#### F. Summary of Supporting Data
Biocompatibility data demonstrates that the device is in compliance with ISO 10993. Testing was conducted with fully manufactured Axonpen devices. The following tests were conducted and passed to demonstrate substantial equivalence with the predicate devices:
| Test Name | Standard | Methodology |
|--------------------------------------|-------------------|-----------------------------------------------------------------|
| Cytotoxicity | ISO 10993-5:2009 | L929 Mem Elution |
| Sensitization | ISO 10993-10:2010 | Kligman Maximization |
| Irritation | ISO 10993-10:2010 | Intracutaneous Injection: NaCl, CSO<br>extracts |
| Systemic toxicity | ISO 10993-11:2017 | Intravenous or intraperitoneal Injection:<br>NaCl, CSO extracts |
| Material<br>Mediated<br>pyrogenicity | ISO 10993-5:2009 | Rabbit pyrogen |
| Hemolysis | ASTM F756-17 | Rabbit blood |
The Axonpen System was demonstrated to be in compliance with all pertinent FDA recognized consensus standards including the ISO 8600 series (Endoscopes- Medical endoscopes and endotherapy devices), the ISO 60601 series (Medical electrical equipment- Part 1: General requirements for safety and essential performance) including IEC 60601-2-18 for endoscopes.
Bench testing demonstrated that the Axonpen System could successfully aspirate fluid and clots without damaging adjacent tissue.
#### Additional Testing included:
Optical Performance including:
- Image processing architecture verification Noise Reduction Contrast Enhancement Color Management (Performance) Display Nonlinear Transformations Image Intensity Uniformity Image Resolution Depth of Field Quantification of Geometric Distortion Dynamic Range of Visualization
Photobiological Safety Validation Water Ingress IPX Rating Packaging Verification Shelf life testing
{13}------------------------------------------------
Image /page/13/Picture/1 description: The image shows the logo for Clearmind Biomedical. The logo consists of a blue abstract symbol on the left, followed by the word "CLEARMIND" in gray. Below "CLEARMIND" is the word "BIOMEDICAL" in a smaller, lighter gray font. The abstract symbol appears to be two stylized arrows pointing in opposite directions.
#### G. Conclusion
Based on intended use, performance, and supporting documentation, Clearmind concludes that the Axonpen™ System is substantially equivalent in intended use, Indications for Use, technology, design, materials, physician use, and energy source as a combination of the predicate Artemis Eye System (K190719) and the Artemis Neuro Evacuation Device (K171332).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.