K190719 · Penumbra, Inc. · GWG · Aug 30, 2019 · Neurology
Device Facts
Record ID
K190719
Device Name
Artemis Eye System
Applicant
Penumbra, Inc.
Product Code
GWG · Neurology
Decision Date
Aug 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1480
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.
Device Story
Artemis Eye System provides visualization and illumination of intracranial tissue/fluids during neurosurgical procedures. System comprises single-use neuro-endoscope (Artemis Eye) and reusable tablet (Artemis Eye Tablet). Endoscope features integrated camera, irrigation/drainage pathways, and working channel for surgical tools. Tablet displays live video feed from endoscope. Used by neurosurgeons in clinical settings (e.g., OR) for diagnostic and therapeutic procedures. Physician operates device to navigate intracranial space; visual output on tablet guides surgical maneuvers and decision-making. System benefits patients by enabling minimally invasive visualization of intracranial structures.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included dimensional/visual inspection, design feature testing, simulated use, and destructive testing. Biocompatibility (ISO 10993-1), sterilization (EN ISO 11135:2014), electrical safety/EMC (IEC/EN 60601-1 series), and software verification/validation were successfully completed.
Technological Characteristics
Neuro-endoscope probe made of stainless steel. Illumination via surface mount LED. System includes reusable tablet display (≥ 1920x1280 resolution). Connectivity: wired connection between endoscope and tablet. Sterilization: EO. Software: Major level of concern. Complies with IEC/EN 60601-1, 60601-1-2, 60601-1-6, and 62366.
Indications for Use
Indicated for visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures in patients requiring neuroendoscopic intervention.
Regulatory Classification
Identification
A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.
Predicate Devices
Medtronic High Resolution Channel Neuroendoscope (K002572)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 30, 2019
Penumbra, Inc. Aditi Kolla Regulatory Affairs Specialist One Penumbra Place Alameda, California 94502
Re: K190719
Trade/Device Name: Artemis Eye System Regulation Number: 21 CFR 882.1480 Regulation Name: Neurological Endoscope Regulatory Class: Class II Product Code: GWG Dated: July 30, 2019 Received: August 1, 2019
Dear Aditi Kolla:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Matthew Krueger Assistant Director DHT5A: Division of Neurosurgical. Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K190719
Device Name Artemis™ Eye System
Indications for Use (Describe)
The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "Penumbra" in a bold, red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern, with a focus on the company name and a stylized initial.
### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990,
Penumbra, Inc. is providing the summary of Substantial Equivalence for the Artemis™ Eye System.
### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Aditi Kolla Regulatory Affairs Specialist Phone: (510) 995-2010 FAX: (510) 217-6414 Email: akolla@penumbrainc.com
#### Date of Preparation of 510(k) Summary 1.3
August 29, 2019
### Device Trade or Proprietary Name 1.4
Artemis™ Eye System
### Primary Device Classification 1.5
Regulatory Class: II Classification Panel: Neurology Common Name: Neurological Endoscope Classification Name: Endoscope, neurological Regulation Number: 21 CFR 882.1480 Product Code: GWG
### 1.6 Predicate & Reference Devices
| 510(k)<br>Number | Clearance Date | Device | Name of<br>Manufacturer |
|------------------|----------------------|--------------------------------------------------------|-------------------------|
| Predicate Device | | | |
| K002572 | November 16,<br>2000 | Medtronic High<br>Resolution Channel<br>Neuroendoscope | Medtronic PS Medical |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Penumbra, a healthcare company. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
| 510(k)<br>Number | Clearance Date | Device | Name of<br>Manufacturer |
|------------------|-----------------------|-------------------------------------|-------------------------|
| K162475 | September 16,<br>2016 | Trice Mi-eye 2, Mi-eye 2<br>Monitor | Trice Medical Inc. |
### 1.7 Predicate Comparison
| Table 1: Predicate Device Comparison | | | |
|--------------------------------------|--|--|--|
|--------------------------------------|--|--|--|
| Attribute | Predicate Device | Subject Device |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Medtronic High Resolution<br>Channel Neuroendoscope | Penumbra Artemis Eye System |
| 510(k) No. | K002572 | To be determined |
| Classification | Class II, GWG (882.1480) | Class II, GWG (882.1480) |
| Indications For Use | The High Resolution Channel<br>Neuro-endoscope is indicated for<br>diagnostic and intraoperative<br>procedures where the physician<br>desires direct vision of intracranial<br>tissue where cerebrospinal fluid<br>(CSF) may be contacted. | The Artemis Eye System is<br>indicated to provide visualization<br>and illumination of intracranial<br>tissue and fluids during diagnostic<br>and therapeutic procedures. |
| Intended Use | Intended for visualization of<br>intracranial tissue during diagnostic<br>and intraoperative procedures. | SAME |
| Portability | Hand-Held | SAME |
| Weight | Not published in the 510(k)<br>Summary | ≤ 300 g |
| Video Output | LCD monitor | SAME |
| Illumination Light<br>Source | 1 mm CAMLite Light Cable | Surface Mount LED |
| Neuroendoscope Probe<br>Material | Stainless Steel | SAME |
| Working Length | 13 cm or 21.6 cm | 15 cm |
| Total Length | 17.2 cm or 25.8 cm | 25 cm |
| Working Tube/Probe<br>OD | 4.2 mm | 6 mm |
| Working Channel<br>Diameter | 2.15 mm | 3.1 mm |
| Camera Resolution | High (30,000 optical fibres) or<br>Standard (10,000 optical fibres) | 200 pixels x 200 pixels |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the word "Penumbra" in a red font, followed by a red circle with a white "P" inside. The word "Penumbra" is written in a sans-serif font. The "P" in the circle is also in a sans-serif font. The logo is likely for a company or organization called Penumbra.
| Table 1: Predicate Device Comparison | | |
|--------------------------------------|-----------------------------------------------------|---------------------------------------------------|
| Attribute | Predicate Device | Subject Device |
| Trade Name | Medtronic High Resolution<br>Channel Neuroendoscope | Penumbra Artemis Eye System |
| Video Output/ Tablet<br>Dimensions | Not Applicable | 31.0 x 20.0 x 3.8 cm<br>(12.2 x 7.9 x 1.5 in.) |
| Display Screen<br>Resolution | Not Applicable | ≥ 1920 pixels x 1280 pixels |
| Video Output/ Tablet<br>Battery Life | Not Applicable | 3 hours (may vary based upon<br>usage) |
| Accessories | Peelaway Introducer Sheath | Stopcock |
| Neuro-Endoscope Use | Single-Use | SAME |
| Sterilization | EO | SAME |
| Shelf-Life | 36-Months | Artemis Eye: 12-Month data<br>currently available |
### Device Description 1.8
The Artemis Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures. The System consists of two components:
- Artemis Eye a single-use neuro-endoscope consisting of a camera, two pathways for . irrigation and/or drain, and a working channel for a surgical tool.
- Artemis Eye Tablet a reusable component that connects to the Artemis Eye and . displays live imaging captured by the Artemis Eye camera.
### 1.9 Indications for Use
The Artemis™ Eye System is indicated to provide visualization and illumination of intracranial tissue and fluids during diagnostic and therapeutic procedures.
## 1.10 Summary of Non-Clinical Data
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding substantial equivalence of the devices follows.
Included in this section is a summary description of the testing, which substantiates the performance of the subject Artemis Eye System.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
## 1.10.1 Sterilization
The subject Artemis Eye has proved to be sterile in accordance with EN ISO 11135:2014.
## 1.10.2 Biocompatibility Testing
Biocompatibility test studies were selected in accordance Guidance for Industry and FDA Staff: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016). All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. The following tests were successfully conducted on the patient-contacting Artemis Eye component of the subject device:
| Test | Method | Results |
|-------------------------------------------|----------------------------------------------------|------------------|
| In Vitro Cytotoxicity | ISO Elution Test (MEM Extract) | Non-Toxic |
| Sensitization | Magnusson-Kligman Method | Non-Sensitizing |
| Irritation (Intracutaneous<br>Reactivity) | ISO Intracutaneous (Intradermal)<br>Injection Test | Non-Irritant |
| Systemic Toxicity (Acute) | | |
| Acute Systemic Toxicity | ISO Acute Systemic Injection Test | Non-Toxic |
| Material Mediated<br>Pyrogen | USP Material-Mediated Rabbit<br>Pyrogen Test | Non-pyrogenic |
| Hemo-compatibility | | |
| In-Vitro Hemolysis | ASTM Method (Indirect & Direct<br>Contact) | Non-Hemolytic |
| Coagulation | PT and PTT Test | Non-Thrombogenic |
# 1.10.3 Bench-Top Performance (Design Verification)
Performance testing was conducted to evaluate the physical and mechanical properties of the Artemis Eye System and to demonstrate substantial equivalence to the predicate device. The following tests were performed and all tests passed successfully:
- Dimensional / visual inspection ●
- . Design Feature Testing
- . Simulated Use
- Destructive Testing .
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "Penumbra" in red font. To the right of the word is a red circle with a white "P" inside. The "P" has a horizontal line through it. The logo is simple and modern.
## 1.10.4 Shelf-life
The physical and mechanical performance of the Artemis Eye was evaluated for 12 months through accelerated aging studies to demonstrate device stability for the duration of the labeled shelf-life. Test results confirm the subject device will maintain device performance for the entirety of the proposed shelf-life.
## 1.10.5 Software Verification & Validation
Software verification and validation testing and documentation for the Artemis Eye System was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (issued May 11, 2005). The software for this device was considered as a "major" level of concern.
## 1.10.6 Electrical Safety/EMC Testing
Electrical safety and EMC testing were conducted on the Artemis Eye System. The system complies with the requirements of IEC/EN 60601-1, IEC/EN 60601-1-2, IEC 60601-1-6, and IEC/EN 62366.
## 1.11 Summary of Substantial Equivalence
The subject Artemis Eye System is substantially equivalent to the predicate device. The subject device has an identical intended use as the predicate device. The subject and the predicate devices differ slightly in regards to minor technological variations, while maintaining the same fundamental scientific technology. However, these differences do not raise different questions of safety and effectiveness.
The device testing described in the 510(k) Summary demonstrate the subject devices are substantially equivalent to the predicate device in regards to operating principle, fundamental technology and device performance. It also demonstrates the Artemis Eye System should perform as intended in the specified use conditions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.