Intra Oral Sensor (model: IOX 1 / IOX 2) is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
Device Story
Intraoral sensor (IOX 1/IOX 2) acquires digital dental X-ray images; detects X-ray photons; converts to electrical image signals. System includes sensor, USB memory, holder, silicon cover. Used in dental clinics by dentists. Device senses X-ray exposure onset; automatically acquires/saves image data to PC via USB 2.0 interface. Output viewed on PC software for diagnostic assessment of teeth and oral tissues. Benefits include digital image capture replacing traditional film/screen systems, enabling storage and manipulation for improved diagnostic workflow.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per IEC 61223-3-4 and FDA guidance for solid-state X-ray imaging devices. Metrics: MTF >30% at 6 lp/mm; DQE >40% at 2.5 lp/mm. Electrical/mechanical safety verified per IEC 60601-1 and EMC per IEC 60601-1-2.
Indicated for dentists to collect dental x-ray photons and convert them into electronic impulses for storage, viewing, and manipulation for diagnostic purposes in patients requiring dental radiographic examinations.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 10, 2020
Dentium Co., Ltd (ICT Branch) % Dave Kim Medical Device Regulatory Affairs Mtech Group 7707 Fannin St. Ste 200, V111 HOUSTON TX 77054
Re: K201247
Trade/Device Name: Intra Oral Sensor (model: IOX 1/ IOX 2) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB, MUH Dated: May 4, 2020 Received: May 11, 2020
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201247
Device Name Intra Oral Sensor (model: IOX 1 / IOX 2)
#### Indications for Use (Describe)
Intra Oral Sensor (model: IOX 1 / IOX 2) is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K201247
The summary of 510(k) is being submitted in accordance with the requirements of 21 CFR Part 807.92.
| Date: | 6/8/2020 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT | Dentium Co.,Ltd (ICT Branch) |
| ADDRESS | 76, Changnyong-daero 256beon-gil, Yeongtong-gu<br>Suwon-si, Gyeonggi-do, 16229<br>Republic of Korea<br>++82-70-7098-6932 |
| Contact Person | Mr. Sang Woo Lee (swlee1@dentium.com) |
| Official<br>Correspondent | Dave Kim, MBA<br>Mtech Group<br>7707 Fannin St. Ste 200, Houston, TX 77054<br>TEL : 713-467-2607<br>Email : davekim@mtech-inc.net |
# Device Name
| Trade Name: | Intra Oral Sensor |
|--------------------|-------------------------|
| Model Name: | IOX1 / IOX2 |
| Regulation Name: | Stationary X-ray System |
| Regulatory Number: | 21 CFR 892.1680 |
| Regulatory Class: | II |
| Product code: | MQB, MUH |
| Panel: | Radiology |
# Predicate device
| Predicate device (K143000) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|
| Trade/Device Name: | RIO Sensor (RIS 500) |
| Regulation Number: | 21 CFR 892.1680 |
| Regulation Name: | Stationary x-ray system |
| Regulatory Class: | II |
| Product Code: | MQB |
| Reference Device (K090526) | |
| ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﮐﮯ ﺍﯾﮏ ﮨﮯ ﺍﻭﺭ ﺍﺱ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﮐﮯ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍﯾﮏ ﺍ | |
| Trade/Device Name: | EzSensor |
| Regulation Number: | 21 CFR 872.1800 |
| Regulation Name: | Extraoral source X-ray system |
| Regulatory Class: | II |
| Product Code: | MUH |
| | |
Dentium Co.,Ltd
{4}------------------------------------------------
# Descrintion
Intraoral Sensor (model: IOX 1 / IOX 2) by Dentium is a medical device that acquires digital images by detecting subject information through X-rays and converting them into electrical image signals to identify teeth and tissues in the mouth. The product consists of the Intraoral Sensor, USB Memory, Sensor Holder, Silicon Cover and Quick Guide.
# Indication for use
Intra Oral Sensor (model: IOX 1 / IOX 2) is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
| Parameter | Proposed Device | Predicated Device | Reference Device |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufac<br>turer | DENTIUM Co., Ltd | RAY Co., Ltd | E-WOO Technology Co., Ltd. |
| Device<br>Name | Intra Oral Sensor<br>IOX1 / IOX2 | RIS500 | EzSensor |
| 510(K)<br>Number | K201247 | K143000 | K090526 |
| Feature | Image: Proposed Device Feature | Image: Predicated Device Feature | Image: Reference Device Feature |
| Indications<br>for use | Intra Oral Sensor (model: IOX 1 /<br>IOX 2) is intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, views and<br>manipulated for diagnostic use by<br>dentists. | This system is intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, views and<br>manipulated for diagnostic use by<br>dentists. | Indicated for intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, viewed, and<br>manipulated for diagnostic use by<br>dentist. |
| Device<br>Description | Intra Oral Sensor (model: IOX 1 /<br>IOX 2) is a medical device that<br>acquires digital images by<br>detecting subject information<br>through X-rays and converting<br>them into electrical image signals<br>to identify teeth and<br>tissues in the mouth. The product<br>consists of the Intraoral Sensor,<br>USB Memory, Sensor Holder,<br>Silicon Cover and Quick Guide | RIOSensor(Model RIS500) is<br>intended to acquire real-time,<br>clinical digital intraoral X-ray<br>images using a solid-state<br>imaging sensor. This system<br>consists of the CMOS sensor and<br>software for image display. This<br>system senses the onset of the X-<br>ray exposure and automatically<br>acquires and save the image data<br>to a PC (software). | The EzSensor is a solid state x-<br>ray imager designed for dental<br>radiographic applications, The<br>EzSensor provides digital image<br>capture for conventional<br>film/screen radiographic dental<br>examinations. The device is used<br>to replace radiographic<br>film/screen system in general<br>dental diagnostic procedures. The<br>captured digital<br>image is transferred to Personal<br>Computer via USB interface port |
| Sensor<br>Dimension | IOX 1: 36.7 x 24 mm<br>IOX 2: 42.6 x 29.1 mm | Size 1: 39x25 mm<br>Size 2: 42x30 mm | Size "1.0": 35.7x25.2 mm Size<br>"1.5": 38.7x29.2 mm |
| Sensor<br>Thickness | 5.1 mm | 5.6 mm | 4.95mm |
| Active<br>Area(m<br>m) | IOX 1: 20 x 30 mm<br>IOX 2: 26 x 36 mm | Size 1: 39x25<br>Size 2: 42x30 | Size "1.0": 20.02x30.03<br>Size "1.5": 24.08x31.85 |
| USB<br>Module | Integrated USB 2.0 module | Integrated USB 2.0 module | Integrated USB 2.0 module |
| Pixel size | 20x20 μm | 20x20 μm | 35x35 μm |
| Pixel Matrix | IOX 1: 1000x1500 pixel IOX 2:<br>1280x1801 pixel | Size 1: 1000x1500 pixel Size 2:<br>1300x1700 pixel | Size "1.0": 572x858 pixel Size<br>"1.5": 686x944 pixel |
| Pixel Pitch | 20x20 μm | 20x20 μm | 35x35 μm |
| Theoretical<br>Resolution | 25 lp/mm | 25 lp/mm | 14.3 lp/mm |
| MTF | More than 30 % at 6 lp/mm | More than 30% at 6 lp/mm | More than 30% at 6 lp/mm |
| DQE | More than 40 % at 2.5 lp/mm | More than 40% at 2.5 lp/mm | More than 40% at 2.5 lp/mm |
### Statement of Substantial Equivalence
{5}------------------------------------------------
The intended use, constructions, construction materials, technical characteristics and safety characteristics between Intra Oral Sensor (model: IOX 1 / IOX 2) and its predicate device are same.
The differences include the digital X-ray imager sizes and image viewing software. Performance testing was conducted for the subject device to access whether or not the parameter required for functionalities related to imaging properties of the dental X-ray device meets the designated acceptance criteria. All test results were satisfactory.
{6}------------------------------------------------
## Safetv and Effectiveness Information:
Electrical, mechanical and environmental safety testing according to standard of IEC 60601-1(2005+ CORR.1(2006)+CORR.2(2007), AMD 1: 2012 was performed. EMC testing was conducted in accordance with the standard IEC 60601-1-2(2014).
The software of Intra Oral Sensor (model: IOX 1 / IOX 2) has been validated according to FDA "Guidance for the Content d Premarket Submissions for Software Contained in Medical Devices" and applicable requirements contained in the guidance document.
Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench"
Bench testing is used to assess whether or not the parameter measured required for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance.
Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4. All test results were satisfactory.
Non-clinical considerations were conducted in accordance with FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices".
All test results were satisfactory.
The tests include the MTF(Modulation Transfer Function) and DOE(Detective Quantum Efficiency) of detector. MTF of detector shows the resolution more than 30 % at 6 lp/mm and The DQE of detector shows the resolution more than 40 % at 2.5 lp/mm.
Base on the Non-Clinical Test report, Even though the pixel size and active area of predicate detectors are different, the diagnostic image quality of Intra Oral Sensor (model: IOX 1 / IOX 2) is substantially equivalent to that of predicate device and there is no significant difference in efficiency and safety.
Clinical Data: Not required for a finding of substantial equivalence.
## Conclusions
In accordance with the federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. Dentium Co., Ltd concludes that the Intra Oral Sensor (model: IOX 1 / IOX 2) present no new risk and perform its intended use substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.