K143000 · Ray Co., Ltd. · MQB · Jan 23, 2015 · Radiology
Device Facts
Record ID
K143000
Device Name
RIS500
Applicant
Ray Co., Ltd.
Product Code
MQB · Radiology
Decision Date
Jan 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
This system is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
Device Story
RIOSensor (RIS500) is a digital intraoral X-ray imaging system; consists of a CMOS solid-state sensor and PC-based software. Device detects onset of X-ray exposure; automatically acquires and saves image data to a PC. Used in dental clinics by dentists to replace traditional film/screen systems. Output is a digital image displayed on a PC for diagnostic manipulation. Benefits include real-time digital image capture, storage, and improved diagnostic workflow compared to film.
Clinical Evidence
Clinical evaluation performed by two licensed practitioners who observed and verified the device's performance in a clinical setting. Bench testing conducted per FDA guidance for solid-state X-ray imaging devices, including MTF (>30% at 6 lp/mm) and DQE (>40% at 2.5 lp/mm) per IEC 61223-3-4. All test results were satisfactory.
Indicated for dental patients requiring intraoral radiographic imaging for diagnostic purposes by dentists.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The caduceus is rendered in a dark color, while the text is in a lighter color. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2015
Ray Co., Ltd. % Mr. Andrew H. Paeng Consultant 4747 Hoen Avenue SANTA ROSA CA 95405
Re: K143000
Trade/Device Name: RIO Sensor (RIS 500) Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: October 28, 2014 Received: October 30, 2014
Dear Mr. Paeng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A. Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143000
## Device Name
Trade Name: RIOSensor Proprietary Name: RIS500 Common Name: Intraoral Imaging Unit
Indications for Use (Describe)
This system is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(K) Summary
{4}------------------------------------------------
## 510(k) Summary
The summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
Date:
| APPLICANT | RAY Co.,Ltd |
|----------------|--------------------------------------------------------------------------------------------------------------------------------|
| ADDRESS | #332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 443-823, Korea |
| Manufacturer | RAY Co.,Ltd<br>332-7, Samsung 1-ro, Hwaseong-si, Gyeonggi-do, 443-823, Korea<br>TEL : +82-31-605-1000<br>FAX : +82-2-6280-5534 |
| Contact Person | Kyungha Seo<br>e-mail: Kyungha.seo@raymedical.co.kr |
#### Device Name
Trade Name: RIOSensor Proprietary Name: RIS500 Common Name: Intraoral Imaging Unit
#### Classification
Classification name: solid state x-ray imager (flat panel/digital imager) Regulatory Number: 21 CFR 892.1680 Class: II Product code: MQB Panel: Radiology
#### Description
RIOSensor(Model RIS500) is intended to acquire real-time, clinical digital intraoral X-ray images using a solid-state imaging sensor. This system consists of the CMOS sensor and software for image display. This system senses the onset of the X-ray exposure and automatically acquires and save the image data to a PC (software).
{5}------------------------------------------------
#### Indication for use
This system is intended to collect dental x-ray photons and convert them into electronic impulses that may be stored, views and manipulated for diagnostic use by dentists.
"CAUTION: Federal (US) law restricts the sale of this device to, or on the order of, licensed professionals.
### Predicate device
- 1) Predicate device-1 Manufacturer: HUMANRAY Co., Ltd. Device: EzSensor T 510(k) Number: K121132
- 2) Predicate device-2 Manufacturer: E-WOO TECHNOLOGY Device: EzSensor 510(k) Number: K090526
#### Statement of Substantial Equivalence
| Parameter | Proposed Device | Predicated Device | Predicated Device |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufac<br>turer | RAY Co., Ltd | HUMANRAY Co., Ltd | E-WOO Technology Co., Ltd. |
| Device<br>Name | RIS500 | EzSensor P | EzSensor |
| 510(K)<br>Number | K143000<br>Traditional 510k | K121132<br>Special 510k | K090526<br>Traditional 510k |
| Feature | Image: Proposed Device Feature | Image: Predicated Device Feature | Image: Predicated Device Feature |
| Indications<br>for use | This system is intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, views and<br>manipulated for diagnostic use by<br>dentists. | EzSensor P, Intra-oral Imaging<br>System, isintended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, viewed, and<br>manipulated for diagnostic use by<br>dentists. | Indicated for intended to collect<br>dental x-ray photons and convert<br>them into electronic impulses that<br>may be stored, viewed, and<br>manipulated for diagnostic use by<br>dentist. |
| Device<br>Description | RIOSensor(Model RIS500) is<br>intended to acquire real-time,<br>clinical digital intraoral X-ray<br>images using a solid-state imaging<br>sensor. This system consists of the<br>CMOS sensor and software for<br>image display. This system senses<br>the onset of the X-ray exposure<br>and automatically acquires and<br>save the image data to a PC<br>(software). | EzSensor P is a solid state x-ray<br>imager designed for dental<br>radiographic applications, The<br>EzSensot P digital intraoral sensor<br>provides digital image capture to<br>replace radiographic film/screen<br>system in general dental diagnostic<br>procedures. The captured digital<br>image is transferred to Personal<br>Computer via USB interface port. | The EzSensor is a solid state x-<br>ray imager designed for dental<br>radiographic applications, The<br>EzSensor provides digital image<br>capture for conventional<br>film/screen radiographic dental<br>examinations. The device is used<br>to replace radiographic film/screen<br>system in general dental diagnostic<br>procedures. The captured digital<br>image is transferred to Personal<br>Computer via USB interface port |
| Sensor<br>Dimension | Size 1: 39x25 mm<br>Size 2: 42x30 mm | Size "1.5": 38.7x29.2 mm<br>Size "2.0": 42.8x31.5 mm | Size "1.0": 35.7x25.2 mm<br>Size "1.5": 38.7x29.2 mm |
| Sensor<br>Thickness | 5.6 mm | 4.95 mm | 4.95mm |
| Active<br>Area(m<br>m) | Size 1: 39x25<br>Size 2: 42x30 | Size "1.5": 24.01x33.04<br>Size "2.0": 26.04x36.05 | Size "1.0": 20.02x30.03<br>Size "1.5": 24.08x31.85 |
| USB<br>Module | Integrated USB 2.0 module | Integrated USB 2.0 module | Integrated USB 2.0 module |
| Pixel<br>size | 20x20 μm | 20x20 μm | 35x35 μm |
| Pixel<br>Matrix | Size 1: 1000x1500 pixel<br>Size 2: 1300x1700 pixel | Size "1.5": 1200x1650 pixel<br>Size "2.0": 1300x1800 pixel | Size "1.0": 572x858 pixel<br>Size "1.5": 686x944 pixel |
| Pixel<br>Pitch | 20x20 μm | 20x20 μm | 35x35 μm |
| Theoretical<br>Resolution | 25 lp/mm | 25 lp/mm | 14.3 lp/mm |
| MTF | More than 30% at 6 lp/mm | More than 30% at 6 lp/mm | More than 30% at 6 lp/mm |
| DQE | More than 40% at 2.5 lp/mm | More than 40% at 2.5 lp/mm | More than 40% at 2.5 lp/mm |
{6}------------------------------------------------
The intended use, constructions, construction materials, technical characteristics and safety characteristics between RIS500 and Its predicate device are same. Ray Co.,Ltd Page
{7}------------------------------------------------
Accordingly we can claim the substantially equivalence of RIS500 to the predicate device.
#### Safety and Effectiveness Information:
Electrical, mechanical and environmental safety testing according to standard of IEC 60601-1(2005+ CORR.1(2006)+CORR.2(2007) was performed. EMC testing was conducted in accordance with the standard IEC 60601-1-2(2007).
The software of RIS500 has been validated according to FDA "Guidance for the Content d Premarket Submissions for Software Contained in Medical Devices" and applicable requirements contained in the guidance document.
Bench testing was conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 18, Performance Testing - Bench"
Bench testing is used to assess whether or not the parameter measured for describing functionalities related to imaging properties of the dental X-ray device and patient dosage satisfies the designated tolerance.
Performance (Imaging performance) testing was conducted according to standard of IEC 61223-3-4.
All test results were satisfactory.
Non-clinical considerations were conducted in accordance with FDA Guidance for the submissions of 510(k)'s for Solid State X-ray Imaging Devices". All test results were satisfactory.
The tests include the MTF(Modulation Transfer Function) and DQE(Detective Quantum Efficiency) of detector. MTF of detector shows the resolution more than 30% at 6 lp/mm and The DQE of detector shows the resolution more than 40% at 2.5 lp/mm.
Base on the Non-Clinical Test report, Even though the pixel size and active area of predicate detectors are different, the diagnostic image quality of RIS500 detector is equal or better than that of predicate device and there is no significant difference in efficiency and safety.
Clinical considerations were conducted according to FDA Guidance "Format for Traditional and Abbreviated 510(k)s, section 20".
For clinical testing. two licensed practitioners/clinicians observed and verified that Intraoral Imaging Unit from RIOSensor(Model name: RIS500).
#### Conclusions
In accordance with the federal Food. Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification. RAY Co., LTD concludes that the RIS500 is safe and effective to perform its intended use as well as substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.