Canon non-mydriatic retinal camera CR series (CR-2 plus AF and CR-2 AF) is intended to be used for taking digital images of the retina of the human eye without mydriatic. The CR-2 plus AF has the following photography modes: color, red-free, cobalt digital and fundus autofluorescence (FAF). The CR-2 AF has the following photography modes: color, red-free, cobalt digital and additional infrared photography for anterior segment.
Device Story
Canon non-mydriatic retinal camera CR series (CR-2 AF and CR-2 Plus AF) captures digital retinal images; 45-degree field of view; autofocus; automatic shooting; automatic switching from anterior segment to retinal imaging. CR-2 Plus AF includes color, red-free, cobalt, and fundus autofluorescence (FAF) modes; uses Xenon tube light source. CR-2 AF includes color, red-free, cobalt, and infrared anterior segment photography; uses white LED light source. Operated by clinicians in clinical settings. Software (RICS or CU) manages image browsing, processing, storage, and DICOM/JPEG export. Output viewed by healthcare providers to assist in ophthalmic assessment. Benefits include non-mydriatic imaging, reducing patient discomfort and procedure time.
Clinical Evidence
Bench testing only. Includes electrical safety (ANSI/AAMI ES 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and ophthalmic instrument standards (ISO 15004-1, ISO 10940, ANSI Z80.36). Biocompatibility evaluated per ISO 10993. Software validated per FDA guidance.
Technological Characteristics
Ophthalmic camera; 45-degree field of view; 35mm working distance. Light sources: Infrared LED (observation), White LED (CR-2 AF flash), Xenon tube (CR-2 Plus AF flash). Connectivity: DICOM/JPEG export. Software: RICS or CU. Standards: ANSI/AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, ISO 15004-1, ISO 10940, ANSI Z80.36, ISO 10993.
Indications for Use
Indicated for taking digital images of the human retina without mydriatic. Intended for prescription use.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 23, 2020
Canon Inc. Akira Hirai General Manager 9-1, Imaikami-cho Nakahara-ku, Kawasaki, Kanagawa 211-8501 Japan
Re: K201122
Trade/Device Name: Canon non-mydriatic retinal camera CR series (Model CR-2 AF and CR-2 Plus AF) Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: June 18, 2020 Received: June 22, 2020
Dear Akira Hirai:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Elvin Ng Acting Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number
K201122
Device Name
Canon non-mydriatic retinal camera CR series (Model CR-2 AF and CR-2 Plus AF)
# Indications for Use (Describe)
Canon non-mydriatic retinal camera CR series (CR-2 plus AF and CR-2 AF) is intended to be used for taking digital images of the retina of the human eye without mydriatic. The CR-2 plus AF has the following photography modes: color, red-free, cobalt digital and fundus autofluorescence (FAF). The CR-2 AF has the following photography modes: color, red-free, cobalt digital and additional infrared photography for anterior segment.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M5.72727 10.8182L13 3.63636L12.3636 3L5.18182 10.1818L3.63636 8.63636L3 9.27273L5.72727 12L5.72727 10.8182Z" fill="black"></path> </svg> </span> |
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
#### 510(k) Summary
## Submitter
Canon Inc. 9-1, Imaikami-cho, Nakahara-ku, Kawasaki, Kanagawa 211-8501, Japan
#### Contact person
Mr. Akira Hirai Canon Inc. TEL: 81-3-3758-2111 FAX: 81-44-739-6695 Email: hirai.akira@mail.canon
#### Date prepared
July 21, 2020
Submission type Special 510(k)
## Name of Device
| Trade/Device Name: | Canon non-mydriatic retinal camera CR series (Model CR-2 AF and CR-2 Plus AF) |
|--------------------|-------------------------------------------------------------------------------|
| Regulation Number: | 21 CFR 886.1120 |
| Regulation Name: | Ophthalmic Camera |
| Regulatory Class: | Class II |
| Product Code: | HKI |
## Predicate Device
| K123208 | |
|--------------------|------------------------------------|
| Trade/Device Name: | Digital Retinal Camera CR-2 Plus A |
| Regulation Number: | 21 CFR 886.1120 |
| Regulation Name: | Ophthalmic Camera |
| Regulatory Class: | Class II |
| Product Code: | HKI |
#### Device Description
The Canon non-mydriatic retinal camera CR series is composed of two retinal cameras and its control software and is used for taking digital images of a human retina without mydriatic.
Two retinal cameras: The CR-2 Plus AF and CR-2 AF
- Both models take digital photographic retinal images of patient's eyes across an angle of view of 45 degrees, ● and have autofocus and automatic shooting of retinal image, and automatic switching from anterior segment image to retinal image.
- The CR-2 Plus AF camera is identical to the camera cleared under K123208. The photography mode includes color, red-free, cobalt and fundus autofluorescence (FAF).
- The CR-2 AF camera is modified version of the predicate CR-2 Plus AF (K123208). Most of the specifications are identical to the predicate; however, the photography mode does not include fundus autofluorescence (FAF) photography.
- . Both the CR-2 AF and the CR-2 Plus AF uses infrared light to observe the anterior segment, but CR-2 AF can save such an image.
{4}------------------------------------------------
Two Software: Retinal image control software (RICS) and Non-myd RC Capture Utility software (CU) programs
- . RICS software program supports browsing, processing, and storage function of the images. It also supports the output of the images to the DICOM storage server, export in DICOM or JPEG format and output to a printer.
- CU software program is a simplified version of RICS, without most of the graphic user interface and the database. It allows user to capture and transfer the retinal images from the retinal camera to the PC.
Both RICS and CU software programs are compatible with both CR-2 AF and CR-2 Plus AF cameras.
# Indications for Use
Canon non-mydriatic retinal camera CR series (CR-2 plus AF and CR-2 AF) is intended to be used for taking digital images of the retina of the human eye without mydriatic. The CR-2 plus AF has the following photography modes: color, red-free, cobalt digital and fundus autofluorescence (FAF). The CR-2 AF has the following photography modes: color, red-free, cobalt digital and additional infrared photography for anterior segment.
The Indication for Use for predicate device, Digital Retinal Camera CR-2 Plus AF is the follows:
"The Digital Retinal Camera CR-2 Plus AF is intended to be used for taking digital images of the retina of the human eye without a mydriatic. The CR-2 Plus AF has the following photography modes: color, red free, cobalt digital and fundus autofluorescence (FAF)."
It appears the intended for use of Canon non-mydriatic retinal camera CR series and predicate device is identical.
# Discussion of Substantial Equivalence
The CR-2 AF is modified from the predicate CR-2 Plus AF (K123208) and most of the specifications are identical to the predicate CR-2 Plus AF, the major differences are in the following aspects:
- The photography mode: The CR-2 AF has the following photography modes: color, red-free, cobalt . and infrared photography for anterior segment. The CR-2 AF doesn't have the fundus autofluorescence photography (FAF) mode while the CR-2 Plus AF (K123208) has.
- . The light source: The CR-2 Plus AF (K123208) needs to equip a large power supply box, capacitor and Xenon tube for the FAF photography because the FAF photography needs high light intensity for photographing images but the CR-2 AF doesn't have FAF photography feature and therefore the CR-2 AF adopts a White LED as a light source, whose output is lower than that of Xenon tube.
It is verified that there is no new risk associated with the light source change. The light hazard protection evaluation of CR-2 AF was evaluated to meet the eye safety limits of the Group 1 instrument based on ANSI Z80.36:2016 - Ophthalmics - Light Hazard Protection for Ophthalmic Instruments.
- The graphic user interface: Software CU is derived from the RICS by modifying the graphic user interface and removing the database from the RICS. CU can save the retinal images to the designated folder to transfer the images to the 3rd party software.
{5}------------------------------------------------
Therefore, the above differences between the proposed device device do not affect the basic design principle, usage, effectiveness and safety of use of the subject device.
| | Differences between both retinal camera models are shown in Table below. | | |
|--|--------------------------------------------------------------------------|--|--|
| | | | |
| | | Proposed | Predicate |
|-------------------------------|------------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | | CR-2 AF | CR-2 Plus AF |
| 510(k) Submitter<br>[Number] | | Canon Inc.<br>K201122 | Canon Inc.<br>[K123208] |
| Product Code | | HKI | HKI |
| Indications for Use | | The CR-2 AF is intended to be used for taking<br>digital images of the retina of the human eye<br>without mydriatic. | The CR-2 Plus AF is intended to be used for<br>taking digital images of the retina of the<br>human eye without mydriatic.<br>The CR-2 Plus AF has the following<br>photography modes: color, red free, cobalt<br>digital and fundus autofluorescence (FAF). |
| Device design | | | |
| Light<br>source | Observation<br>Flash | Infrared LED<br>White LED | Infrared LED<br>Xenon tube |
| Angular field of view | | 45/43 ° (digital magnification)<br>(35°when S.P switch is ON) | 45/43 ° (digital magnification)<br>(35°when S.P switch is ON) |
| Actual image size | | cp13.7 mm (on sensor array) | cp13.7 mm (on sensor array) |
| | Min. diameter of pupil<br>required | cp4mm<br>(cp3.3mm when S.P switch is ON) | cp4mm<br>(cp3.3mm when S.P switch is ON) |
| | Working distance (WD) | 35mm | 35mm |
| Focusing and<br>Photographing | | Focusing by aligning the split lines<br>automatically and photographing<br>automatically | Focusing by aligning the split lines<br>automatically and photographing<br>automatically |
| Switching function | | Automatically Switching function from<br>anterior segment to fundus after anterior<br>segments is aligned. | Automatically Switching function from<br>anterior segment to fundus after anterior<br>segments is aligned. |
| Photography mode | | color, red free, cobalt digital and infrared for<br>anterior segment | color, red free, cobalt digital and fundus<br>autofluorescence (FAF) |
| Anterior segment | | Observation and capture | Observation |
| Eye fixation lamp | | Internal (during observation of eye front<br>image and retinal image),<br>External | Internal (during observation of eye front<br>image and retinal image),<br>External |
| External dimensions | | W305×D500×H473mm | W305×D500×H513mm |
| Weight | | Approx.15kg | Approx.19.9kg |
| Control software | | Selectable from the followings:<br>Retinal imaging control software;<br>Non-myd RC Capture Utility | Retinal imaging control software |
## Performance Data
Non-clinical tests including Electrical safety, Electromagnetic Compatibility, Performance testing, and Software Verification and Validation were conducted to evaluate safety and effectiveness of the CR-2 AF. The CR-2 AF complies with the Recognized Consensus Standard, specifically ANSI/AAMI ES 60601-1. IEC 60601-1-2. IEC 60601-1-6. ISO 15004-1, ISO 10940 and ANSI Z80.36. A qualitative assessment of the function of the anterior segment infrared image for CR-2 AF camera has been performed. The contact parts with patient in the CR-2 AF are the same materials as those used in the predicate device CR-2 Plus AF (K123208) and have been evaluated according to ISO10993 series that is concluded to have no new biocompatibility concern. The software was validated according to the FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (May 2005).
# Conclusion
Based on the performance data, identical intended use and functional technological characteristics and the similarities in functional design, the proposed CR-2 AF is substantially equivalent to the CR-2 Plus AF (K123208) and does not raise any new questions regarding safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.