K123208 · Canon Inc. -Medical Equipment Group · HKI · Mar 19, 2013 · Ophthalmic
Device Facts
Record ID
K123208
Device Name
DIGITAL RETINAL CAMERA
Applicant
Canon Inc. -Medical Equipment Group
Product Code
HKI · Ophthalmic
Decision Date
Mar 19, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Indications for Use
The CR-2 Plus AF is intended to be used for taking digital images of the retina of the human eye without a mydriatic. The CR-2 Plus AF has the following photography modes: color, red free, cobalt digital and fundus autofluorescence (FAF).
Device Story
Digital Retinal Camera CR-2 Plus AF captures digital images of human retina without mydriatic. Device integrates Canon EOS Digital Camera; features autofocus, automatic shooting, and automatic switching between anterior segment and fundus imaging. Replaces manual diopter compensation knob with focus ring adjustment. Used in clinical settings for applications like telemedicine and electronic filing. Healthcare providers view images immediately to assist in diagnostic decision-making and documentation. Benefits include increased imaging efficiency and non-mydriatic operation.
Clinical Evidence
Bench testing only. Performance testing, software validation, electrical safety, and electromagnetic compatibility testing were conducted to demonstrate substantial equivalence. No clinical data provided.
Technological Characteristics
Ophthalmic camera; non-mydriatic; digital imaging via integrated Canon EOS camera. Features autofocus, automatic shooting, and automatic switching between anterior segment and fundus modes. Focus ring replaces manual diopter compensation knob. Complies with US Performance Standard for ophthalmic equipment. Software-controlled image acquisition.
Indications for Use
Indicated for capturing digital retinal images in humans without mydriasis. Supports color, red-free, cobalt digital, and fundus autofluorescence (FAF) imaging modes.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "Canon" in a bold, sans-serif font. The letters are black, and the background is white. The word is centered in the image and takes up most of the space.
## 5. 510(k) Summary
## K123208
## MAR 1 9 2013
| Submitter: | Canon, Inc. - Medical Equipment Group<br>30-2 Shimomaruko, 3-chome<br>Ohta-ku, Tokyo 146-8501 Japan |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Naoyasu Asaka<br>Manager<br>TEL: 81-3-3758-2111<br>FAX: 81-3-5482-3960<br>asaka.naoyasu@canon.co.jp |
| Contact/Application<br>Correspondent | Izumi Maruo<br>MIC International<br>4-1-17 Hongo, Bunkyo-ku,<br>Tokyo, 113-0033, Japan<br>TEL: 81-3-3818-8577<br>FAX: 81-3-3818-8573<br>maruo@mici.co.jp |
| Date Prepared: | October 10, 2012 |
| Trade Name: | Digital Retinal Camera CR-2 Plus AF |
| Common Name: | Ophthalmic camera |
| Classification Name: | Ophthalmic camera. (21 CFR 886.1120, Product code HKI) |
| Regulation Class | II |
| Predicate Device: | K111612 Canon Digital Retinal Camera CR-2 Plus |
| Device Description: | The Digital Retinal Camera CR-2 Plus AF is used for taking digital<br>images of a human retina without a mydriatic. Canon EOS Digital<br>Camera is mounted to the CR-2 Plus AF. Images can be viewed<br>immediately, and procedures of imaging are more efficient with many<br>different applications such as telemedicine and electronic filing. The<br>CR-2 Plus AF is equipped with autofocus/automatic<br>shooting/automatic switching function from anterior segment image to<br>fundus image. |
| Statement of<br>Intended Use: | The CR-2 Plus AF is intended to be used for taking digital images of<br>the retina of the human eye without a mydriatic. The CR-2 Plus AF has<br>the following photography modes: color, red free, cobalt digital and<br>fundus autofluorescence (FAF). |
. .
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "Canon" in a bold, sans-serif font. The letters are black, and the background is white. The word is slightly curved, with the "C" being the highest point and the "n" being the lowest point. The letters are evenly spaced and easy to read.
Modification from predicate device
The CR-2 Plus AF is modified from the CR-2 Plus by adding following functions;
- Autofocus of the fundus image .
- . Automatic shooting of the fundus image
- Automatic switching from anterior segment image to fundus image .
In addition, the "diopter compensation knob" is removed since diopter compensation is performed using the focus ring instead of the diopter compensation knob in the CR-2 Plus AF.
The CR-2 Plus AF has the same intended use and fundamental technological characteristics as the CR-2 Plus. However, the CR-2 Plus AF has some different technological characteristics compared to the CR-2 Plus as described above. In order to evaluate safety and effectiveness of the CR-2 Plus AF, non-clinical tests were performed. In conclusion, result of the testing demonstrated that the CR-2 Plus AF does not raise any new safety and effectiveness concerns compared to the CR-2 Plus.
Non-clinical tests were conducted to evaluate safety and effectiveness of the CR-2 Plus AF as follows. Performance testing, Software Validation, Electrical safety, and Electromagnetic Compatibility testing have been performed. The unit complies with the US Performance Standard for ophthalmic equipment. The CR-2 Plus AF met all requirements of the standards.
Canon. Inc. - Medical Equipment Group concluded that the CR-2 Plus AF is substantially equivalent to the predicate device listed above. This conclusion is based on the identical intended use and fundamental technological characteristics, and the similarities in the functional design. Although the CR-2 Plus AF has some different technological characteristics from the predicate, the non-clinical testing results indicated that the CR-2 Plus AF met all requirements of recognized or voluntary standard. Based on the test results. the CR-2 Plus AF does not raise any new safety and effectiveness concerns compared to the CR-2 Plus.
Statement of Substantial Equivalence
Summary of Non-Clinical/Test Data:
Conclusion:
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three tail feathers, representing the department's commitment to health, services, and people. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle.
March 19, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Canon Inc. Medical Equipment Group % Izumi Maruo MIC International Corp. 4-1-17 Hongo Bunkyo-ku, Tokyo 113-0033
Re: K123208
Trade/Device Name: Digital Retina Camera CR-2 Plus AF Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI Dated: March 5, 2013 Received: March 7, 2013
Dear Izumi Maruo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Izumi Maruo
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia YAlexander -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): K123208
Device Name: Digital Retinal Camera CR-2 Plus AF
Indications For Use:
The Digital Retinal Camera CR-2 Plus AF is intended to be used for taking digital images of the retina of the human eye without a mydriatic. The CR-2 Plus AF has the following photography modes: color, red free, cobalt digital and fundus autofluorescence (FAF).
Prescription Use X (part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use_ (21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ka N. To
2013.03.14 08:54:22 -04'00'
(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:________________________________________________________________________________________________________________________________________________________________
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.