BD PhaSeal Optima Closed System Drug Transfer Device
K201099 · Becton, Dickinson and Company · ONB · Aug 12, 2020 · General Hospital
Device Facts
Record ID
K201099
Device Name
BD PhaSeal Optima Closed System Drug Transfer Device
Applicant
Becton, Dickinson and Company
Product Code
ONB · General Hospital
Decision Date
Aug 12, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.
Device Story
BD PhaSeal Optima Injector (N40-O) is a sterile, single-use closed system drug transfer device (CSTD) component. It facilitates the reconstitution and transfer of hazardous drugs (e.g., antineoplastics) in healthcare settings. The device utilizes a single-lumen cannula and a collet-style fitting with self-sealing elastomeric double membranes to create an airtight, leakproof connection. When the cannula is inserted, it perforates the membranes; upon retraction, the membranes reseal, preventing drug vapor escape and environmental contaminant ingress. The N40-O model incorporates a new ratchet mechanism in the needle hub/grip for secure connection and a needle hub cover to restrict access. It connects to external devices via Luer Lock. The device is operated by healthcare professionals. It benefits patients and staff by minimizing exposure to hazardous drug aerosols and spills and maintaining drug sterility for up to 168 hours.
Clinical Evidence
Bench testing only. Performance testing included attachment/detachment force, fragmentation, and system flow rate. All tests met predetermined acceptance criteria.
Indicated for use as a closed system drug transfer device (CSTD) for the reconstitution and transfer of antineoplastic or other hazardous drugs in healthcare settings to minimize exposure to drug vapors, aerosols, and spills, and to prevent microbial ingress.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
BD PhaSeal™ Optima Closed System Transfer Device (K181221)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
August 12, 2020
Becton, Dickinson and Company Amy Honev Sr. Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K201099
Trade/Device Name: BD PhaSeal Optima Closed System Drug Transfer Device-Injector (N40-0) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: July 9, 2020 Received: July 13, 2020
Dear Amy Honey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201099
Device Name
BD PhaSeal™ Optima Closed System Drug Transfer Device - Injector (N40-O)
Indications for Use (Describe)
The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------|
| <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the BD logo. The logo consists of an orange circle with a stylized sunburst design on the left and the letters "BD" in blue on the right. The sunburst design has a central circle with rays emanating outwards, and the letters "BD" are in a bold, sans-serif font.
## K201099
510(k) Summary (21 CFR §807.92)
BD PhaSeal™ Optima Closed System Drug Transfer Device – Injector (N40-O)
| Submitter | Submitter Name: | Becton, Dickinson and Company |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Information | Submitter Address: | 1 Becton Drive<br>Franklin Lakes, NJ 07417 |
| | Contact Person: | Amy Honey<br>Sr. Staff Regulatory Affairs Specialist |
| | Email Address: | amy.honey@bd.com |
| | Phone Number: | (801) 304-3908 |
| | Date of Preparation: | August 11, 2020 |
| Subject Device | Trade Name: | BD PhaSeal™ Optima Closed System Drug Transfer Device<br>- Injector (N40-O) |
| | Common Name: | Closed System Drug Transfer Device (CSTD) |
| | Regulation Number: | 21 CFR §880.5440 |
| | Regulation Name: | Intravascular Administration Set |
| | Regulatory Class: | Class II |
| | Product Code: | ONB |
| | Classification Panel: | General Hospital |
| Predicate | Trade Name: | BD PhaSeal™ Optima Closed System Transfer Device |
| Device | 510(k) Reference: | K181221 |
| | Common Name: | Closed System Drug Transfer Device (CSTD) |
| | Regulation Number: | 21 CFR §880.5440 |
| | Regulation Name: | Intravascular Administration Set |
| | Regulatory Class: | Class II |
| | Product Code: | ONB |
| | Classification Panel: | General Hospital |
| Reason for<br>Submission | The reason for this submission is to introduce a new product offering-the subject BD<br>PhaSeal™ Optima Injector (N40-O)-to the BD PhaSeal™ Optima family of devices. | |
| Device<br>Description | BD PhaSeal™ Optima Closed System Drug Transfer Devices (CSTD) are sterile,<br>single use closed system drug transfer devices intended for the reconstitution and<br>transfer of antineoplastic or other hazardous drugs in the healthcare setting. The BD | |
| | PhaSeal™ Optima system is comprised of four devices—Protector, Injector,<br>Connector, and Infusion Adapter. | |
| | The closed transfer of liquid drugs takes place through a double membrane utilizing<br>self-sealing elastomeric membranes that are tightly fitted together through the collet-<br>style fitting on each of the BD PhaSeal™ Optima system devices. During use, the<br>single lumen cannula of the Injector perforates the double membranes for the transfer<br>of liquids. When the cannula is retracted the membranes seal off the transfer of<br>environmental contaminants into the system and/or escape of drug or vapor<br>concentrations outside the system, thereby minimizing the individual and<br>environmental exposure to drug vapor, aerosols, leaks and spills. The BD PhaSeal™<br>Optima system prevents microbial ingress for up to 168 hours. Performance of the<br>self-sealing membrane has been substantiated up to 10 penetrations. | |
| | Device labeling includes the following statement: "The ability to prevent microbial<br>ingress for up to 7 days should not be interpreted as modifying, extending, or<br>superseding a manufacturer's labeling recommendations for the storage and<br>expiration dating of the drug vial. Refer to drug manufacturer's recommendations and<br>USP compounding guidelines for shelf life and sterility information." | |
| Indications for<br>Use | The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug<br>transfer device (CSTD) that mechanically prohibits the transfer of environmental<br>contaminants into the system and the escape of drug vapor concentrations outside<br>the system, thereby minimizing individual and environmental exposure to drug vapor,<br>aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial<br>ingress for up to 168 hours. | |
| Technological<br>Characteristics | Technological characteristics of the subject device are substantially equivalent to the<br>predicate device. The subject BD PhaSeal™ Optima Injector N40-O achieves its<br>intended use based on the same technology and principles of operation as the<br>predicate device. | |
| | The changes to the device include modifications to the needle hub and grip<br>components to incorporate a ratchet mechanism that locks when rotated during<br>clockwise rotation, and a new needle hub cover component has been added to<br>restrict access to the needle hub. These differences were assessed in the<br>performance tests that were carried out and there were no new questions of safety<br>and effectiveness. | |
| | In addition, there were changes to the colorants and silicone formulation used in the<br>subject device. The colorant used in the subject device is used in different<br>components of the predicate device; therefore, there were no new risks introduced.<br>The silicone lube material is unchanged; therefore, there are no new biocompatibility<br>risks introduced to the device. | |
| | A comparison of the subject and predicate device technological characteristics is<br>provided in the table below. | |
{4}------------------------------------------------
| Attribute | SUBJECT<br>BD PhaSeal™™ Optima Closed System<br>Transfer Device – Injector (N40-O) | PREDICATE (K181221)<br>BD PhaSeal™™ Optima Closed System<br>Transfer Device (Injector [N35-O] device) |
|----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The BD PhaSeal™™ Optima system is an<br>airtight and leakproof closed system drug<br>transfer device (CSTD) that mechanically<br>prohibits the transfer of environmental<br>contaminants into the system and the<br>escape of drug vapor concentrations outside<br>the system, thereby minimizing individual | The BD PhaSeal™™ Optima system is an<br>airtight and leakproof closed system drug<br>transfer device (CSTD) that mechanically<br>prohibits the transfer of environmental<br>contaminants into the system and the escape<br>of drug vapor concentrations outside the<br>system, thereby minimizing individual and |
| | SUBJECT | PREDICATE (K181221) |
| Attribute | BD PhaSeal™ Optima Closed System<br>Transfer Device - Injector (N40-O) | BD PhaSeal™ Optima Closed System<br>Transfer Device (Injector [N35-O] device) |
| | and environmental exposure to drug vapor,<br>aerosols and spills. The BD PhaSeal™<br>Optima system also prevents microbial<br>ingress for up to 168 hours | environmental exposure to drug vapor,<br>aerosols and spills. The BD PhaSeal™ Optima<br>system also prevents microbial ingress for up<br>to 168 hours |
| Devices of the<br>CSTD System | Protector, Injector, Connector, Infusion<br>Adapter | Protector, Injector, Connector, Infusion<br>Adapter |
| | Injector Membrane Lube<br>Silicone | Injector Membrane Lube<br>Silicone |
| | Membrane<br>Polyisoprene | Membrane<br>Polyisoprene |
| | Injector Grip 1<br>Polypropylene + white colorant | Injector Grip 1<br>Polypropylene + white colorant |
| | Injector Grip 2<br>Polypropylene + white colorant | Injector Grip 2<br>Polypropylene + white colorant |
| | Needle Hub<br>Polypropylene + white colorant | Needle Hub<br>Polypropylene + blue colorant |
| Device<br>Components/<br>Materials | <b>Needle Hub Cover</b><br>Polypropylene + white colorant | N/A - does not incorporate a Needle Hub<br>Cover |
| | <b>Cannula</b><br>Stainless Steel | <b>Cannula</b><br>Stainless Steel |
| | <b>Cannula Lube</b><br>Silicone | <b>Cannula Lube</b><br>Silicone |
| | <b>Cap</b><br>Polypropylene | <b>Cap</b><br>Polypropylene |
| | <b>Collet</b><br>Polyoxymethylene (POM) | <b>Collet</b><br>Polyoxymethylene (POM) |
| | <b>Spring</b><br>Stainless Steel | <b>Spring</b><br>Stainless Steel |
| Mating Method<br>(action of device<br>connections) | Push on-Pull off | Push on-Pull off |
| Connection<br>between Devices<br>within the<br>System | Collet style fitting with elastomeric double<br>membranes | Collet style fitting with elastomeric double<br>membranes |
| Transfer<br>Mechanism<br>(responsible for<br>airtight & leak-<br>proof<br>connections) | Elastomeric double membrane | Elastomeric double membrane |
| Injector Needle<br>Safety<br>Mechanism | Collet style fitting | Collet style fitting |
| Attribute | SUBJECT<br>BD PhaSealTM Optima Closed System<br>Transfer Device – Injector (N40-O) | PREDICATE (K181221)<br>BD PhaSealTM Optima Closed System<br>Transfer Device (Injector [N35-O] device) |
| Injector<br>Connection to<br>External Device<br>(e.g. syringe) | Luer Lock | Luer Lock |
{5}------------------------------------------------
{6}------------------------------------------------
A risk analysis was performed for the modifications to the subject device in accordance with ISO 14971, Medical devices – Applications of risk management to medical devices, and possible risks were identified. Based on risk identification, verification and validation activities were carried out to ensure the risk acceptability criteria have been met and the risks have been mitigated.
| Performance Tests | Per the design control requirements specified in 21 CFR 820.30, and the risks identified, the following performance tests were conducted, and all predetermined acceptance criteria were met. | |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| | Performance Test | Results |
| | Attachment/detachment force | PASS |
| | Fragmentation | PASS |
| | System flow rate | PASS |
| | Testing results demonstrate that the subject device is substantially equivalent to the predicate device. | |
| Summary of Substantial Equivalence | The subject device uses the same technology as the predicate device to meet its intended use of minimizing individual and environmental exposure to drug vapors, aerosols, leaks, and spills during fluid transfer when mated to other devices within the system. The indications for use, technological characteristics, and performance testing results demonstrate that the subject BD PhaSeal™ Optima Injector (N40-O) is substantially equivalent to the predicate device. | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.