K181221 · Becton, Dickinson and Company · ONB · Nov 30, 2018 · General Hospital
Device Facts
Record ID
K181221
Device Name
BD PhaSeal Optima Closed System Transfer Device
Applicant
Becton, Dickinson and Company
Product Code
ONB · General Hospital
Decision Date
Nov 30, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.
Device Story
Sterile, single-use closed system drug transfer device (CSTD) for reconstitution/transfer of antineoplastic/hazardous drugs; system components include Protector, Injector, Connector, and Infusion Adapter. Operation involves push on-pull off connection via collet-style fitting; liquid transfer occurs through double-membrane interface using self-sealing elastomeric membranes. Single lumen cannula perforates membranes during transfer; retraction seals system to prevent environmental contaminant ingress and drug vapor/aerosol escape. Used in healthcare settings by clinicians; minimizes individual/environmental exposure to hazardous drugs; prevents microbial ingress for up to 168 hours. Performance substantiated up to 10 penetrations.
Clinical Evidence
Bench testing only. Testing included microbial ingress, airtight/leakproof connections (TiCl4/Fluorescein), assembly/removal force, expansion chamber burst, packaging integrity (ASTM D4169, F88, F2096), hazardous vapor containment, EO residuals (ISO 10993-7), EO sterilization (ISO 11135), and pyrogenicity. Biocompatibility evaluation per ISO 10993-1 included cytotoxicity (ISO 10993-5), sensitization (ASTM F2148, ISO 10993-10), intracutaneous reactivity (ISO 10993-10, USP <88>), acute systemic toxicity (ISO 10993-11), pyrogenicity (USP <151>, ISO 10993-11), and hemolysis (ISO 10993-4). All tests passed.
Indicated for the reconstitution and transfer of antineoplastic or other hazardous drugs in the healthcare setting, providing an airtight and leakproof closed system to minimize exposure to drug vapors, aerosols, and spills, and to prevent microbial ingress for up to 168 hours.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
Predicate Devices
BD PhaSeal™ Closed System Transfer Device (K123213)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
November 30, 2018
Becton, Dickinson and Company Amy Honey Staff Regulatory Affairs Specialist 1 Becton Drive Franklin Lakes, New Jersey 07417
Re: K181221
Trade/Device Name: BD PhaSeal Ontima Closed System Transfer Device Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: October 22, 2018 Received: October 24, 2018
Dear Amy Honey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -S
Tina Kiang, Ph.D. for Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181221
Device Name
BD PhaSeal Optima Closed System Transfer Device
#### Indications for Use (Describe)
The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the BD logo. The logo consists of two parts: a circular orange symbol on the left and the letters "BD" in blue on the right. The circular symbol has a white design inside, resembling a stylized sun or flower. The letters "BD" are in a bold, rounded font.
## K181221
# 510(k) Summary (21 CFR §807.92)
# BD PhaSeal™ Optima Closed System Transfer Device
| Submitter<br>Information | Submitter Name: | Becton, Dickinson and Company |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| | Submitter Address: | 1 Becton Drive<br>Franklin Lakes, NJ 07417 |
| | Contact Person: | Amy Honey |
| | Email Address: | Staff Regulatory Affairs Specialist<br>amy.honey@bd.com |
| | Phone Number: | Phone: (801) 304-3908 |
| | Fax Number: | Fax: (801) 304-3963 |
| | Date of Preparation: | November 30, 2018 |
| Subject Device | Trade Name: | BD PhaSeal™ Optima Closed System Transfer Device |
| | Common Name: | Closed Antineoplastic and Hazardous Drug<br>Reconstitution and Transfer System |
| | Regulation Number: | 21 CFR §880.5440 |
| | Regulation Name: | Intravascular Administration Set |
| | Regulatory Class: | II |
| | Product Code: | ONB |
| | Classification Panel: | General Hospital |
| Predicate Device | Trade Name: | BD PhaSeal™ Closed System Transfer Device |
| | 510(k) Reference: | K123213 |
| | Common Name: | Closed Antineoplastic and Hazardous Drug<br>Reconstitution and Transfer System |
| | Regulation Number: | 21 CFR §880.5440 |
| | Regulation Name: | Intravascular Administration Set |
| | Regulatory Class: | II |
| | Product Code: | ONB |
| | Classification Panel: | General Hospital |
| Device<br>Description | The BD PhaSeal™ Optima Closed System Transfer Device (CSTD) is a sterile single-<br>use closed system drug transfer device intended for the reconstitution and transfer of<br>antineoplastic or other hazardous drugs in the healthcare setting. The system is<br>comprised of four devices-Protector, Injector, Connector, and Infusion Adapter.<br>The closed transfer of liquid drugs takes place through a double membrane utilizing<br>self-sealing elastomeric membranes that are tightly fitted together through the collet-<br>style fitting on each of the BD PhaSeal™ Optima system devices. During use, the<br>single lumen cannula of the Injector perforates the double membranes for the transfer<br>of liquids. When the cannula is retracted the membranes seal off the transfer of<br>environmental contaminants into the system and/or escape of drug or vapor<br>concentrations outside the system, thereby minimizing the individual and<br>environmental exposure to drug vapor, aerosols, leaks and spills. The BD PhaSeal™<br>Optima system prevents microbial ingress for up to 168 hours. Performance of the<br>self-sealing membrane has been substantiated up to 10 penetrations. | |
| Indications for<br>Use | The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug<br>transfer device (CSTD) that mechanically prohibits the transfer of environmental<br>contaminants into the system and the escape of drug vapor concentrations outside<br>the system, thereby minimizing individual and environmental exposure to drug vapor,<br>aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial<br>ingress for up to 168 hours. | |
| Technological<br>Characteristics | Technological characteristics of the subject BD PhaSeal™ Optima Closed System<br>Transfer Device (CSTD) are substantially equivalent to that of the predicate BD<br>PhaSeal™ Closed System Transfer Device (CSTD) with respect to basic design and<br>function. The subject BD PhaSeal™ Optima Closed System Transfer Device (CSTD)<br>achieves its intended use based on the same technology and principles of operation<br>as the predicate BD PhaSeal™ Closed System Transfer Device (CSTD).<br>The key differences between the subject and predicate device are as follows: | |
| | • The action to connect mating devices within the subject BD PhaSeal™ Optima<br>CSTD system is a push on-pull off motion rather than the push-turn-push<br>motion of the predicate BD PhaSeal™ CSTD device. | |
| | • The connection between the subject BD PhaSeal™ Optima CSTD system<br>devices (Injector to Protector, Connector, or Infusion Adapter) is a collet-style<br>fitting with elastomeric double membranes rather than the bayonet fitting with<br>elastomeric double membranes of the predicate BD PhaSeal™ CSTD device. | |
| | • The Injector needle mechanism of the subject BD PhaSeal™ Optima CSTD<br>device is a collet-style fitting rather than the Safety Sleeve ErgoMotion™ of<br>the predicate BD PhaSeal™ CSTD device. | |
{4}------------------------------------------------
A comparison of the subject and predicate device technological characteristics is provided in the table below.
| Subject and Predicate Device Comparison Table | | |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Attribute | SUBJECT BD PhaSeal™ Optima Closed System Transfer Device | PREDICATE BD PhaSeal™ Closed System Transfer Device (K123213) |
| Indications for Use | The BD PhaSeal™ Optima system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The BD PhaSeal™ Optima system also prevents microbial ingress for up to 168 hours. | The PhaSeal™ system is an airtight and leakproof closed system drug transfer device (CSTD) that mechanically prohibits the transfer of environmental contaminants into the system and the escape of drug vapor concentrations outside the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols and spills. The PhaSeal™ system also prevents microbial ingress. |
| Devices of the CSTD System | Protector, Injector, Connector, Infusion Adapter | Protector, Injector, Connector, Infusion Adapter |
| Subject and Predicate Device Comparison Table (cont.) | | |
| Attribute | SUBJECT<br>BD PhaSeal™ Optima Closed System<br>Transfer Device | PREDICATE<br>BD PhaSeal™ Closed System Transfer<br>Device (K123213) |
| Primary Device<br>Materials | Membrane (all devices)<br>Polyisoprene<br>Cannula (Injector)<br>Stainless Steel<br>Injector Lube<br>Silicone<br>Needle Hub (Injector)<br>Polypropylene<br>Spike (Protector)<br>Polypropylene<br>Spike Sleeve (Protector)<br>Thermoplastic Elastomer (TPE)<br>Connector Body<br>Polypropylene<br>Infusion Spike (Infusion Adapter)<br>Polypropylene<br>Infusion Port (Infusion Adapter)<br>Thermoplastic Elastomer (TPE) | Membrane (all devices)<br>Thermoplastic Elastomer (TPE)<br>Cannula (Injector)<br>Stainless Steel<br>Injector Lube<br>N/A<br>Needle Hub (Injector)<br>Polypropylene<br>Spike or Cannula (Protector)<br>Polypropylene or Stainless Steel<br>Spike Sleeve (Protector)<br>N/A<br>Connector Body<br>Polypropylene<br>Infusion Spike (Infusion Adapter)<br>Polypropylene<br>Infusion Port (Infusion Adapter)<br>Thermoplastic Elastomer (TPE) |
| Mating Method<br>(Action of Device<br>Connections) | Push on-Pull off | Push-Turn-Push |
| Connection<br>between Devices<br>within the System | Collet-style fitting with elastomeric double<br>membranes | Bayonet fitting with elastomeric double<br>membranes |
| Transfer<br>Mechanism<br>(responsible for<br>airtight &<br>leak-proof<br>connections): | Elastomeric double membrane | Elastomeric double membrane |
| Injector Needle<br>Mechanism | Collet-style fitting | Safety Sleeve, ErgoMotion™ |
| Injector<br>Connection to<br>Syringe | Luer Lock | Luer Lock |
| Connector<br>Connection to<br>Patient IV Line | Luer Lock connection or Spike Port | Luer Lock connection or Spike Port |
{5}------------------------------------------------
{6}------------------------------------------------
## Performance Tests
Performance testing was conducted to support a determination of substantial equivalence as compared to BD's own predicate device. Results of these tests demonstrate that the subject device is substantially equivalent to the predicate device. Any testing performed to a recognized standard is noted in the table below.
| Performance Test | Results |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Microbial ingress testing per FDA guidance document,<br>Intravascular Administration Sets Premarket Notification<br>Submissions [510(k)] | PASS |
| Airtight connections utilizing TiCl4 vapor test | PASS |
| Leakproof connections testing utilizing Fluorescein test | PASS |
| Protector assembly force and removal force testing | PASS |
| Protector expansion chamber burst testing | PASS |
| Packaging integrity and shelf life testing per the following:<br>ASTM D4169-16 Standard Practice for Performance<br>Testing of Shipping Containers and Systems<br>ASTM F88/F88M-15 Standard Test Method for Seal<br>Strength of Flexible Barrier Materials<br>ASTM F2096-11 Standard Test Method for Detecting<br>Gross Leaks in Packaging by Internal Pressurization<br>(bubble test) | PASS |
| Hazardous vapor containment using alcohol residuals<br>level criterion | PASS |
| EO residuals testing per ISO 10993-7 | PASS |
| EO sterilization validation (SAL of 10-6) per ISO 11135 | PASS |
| Pyrogenicity - bacterial endotoxin test (LAL) | PASS |
In addition, a biocompatibility evaluation was conducted on the subject device per ISO 10993-1:2009, Biological Evaluation of Medical Devices—Part 1: Evaluation and Testing Within a Risk Management Process. Based on the evaluation, the following biological tests were conducted:
| Biocompatibility Test | Results |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| Cytotoxicity per ISO 10993-5 Sensitization per ASTM F2148 and ISO 10993-10 Intracutaneous Reactivity per ISO 10993-10 and USP 39-NF 34 <88> Acute Systemic Toxicity per ISO 10993-11 Pyrogenicity (material-mediated rabbit pyrogen) per USP <151> and ISO 10993-11 Hemolysis per ISO 10993-4 | All materials are biocompatible. |
Per the design control requirements specified in 21 CFR 820.30, the subject device met all predetermined acceptance criteria for the above-listed performance tests, demonstrating substantial equivalence to the predicate device.
{7}------------------------------------------------
### Summary of Substantial Equivalence
Both the subject BD PhaSeal™ Optima Closed System Transfer Device and predicate BD PhaSeal™ Closed System Transfer Device rely on similar technology to achieve the intended use. Both devices utilize elastomeric double membranes to create the sealed environment. The main difference between the subject device and predicate device is the connection method used to attach and detach system components; however, the fundamental scientific technology used to assure an airtight and leak-proof connection remains unchanged. The results of performance and biological testing of the subject BD PhaSeal™ Optima Closed System Transfer Device met all predetermined acceptance criteria and demonstrate that the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.