The NeVa™ PV Thrombectomy Device is indicated for: - The non-surgical removal of emboli and thrombi from peripheral blood vessels. - Temporary use in peripheral vessel occlusion. - Use with aspiration and with the injection or infusion of contrast media and other fluids.
Device Story
NeVa™ PV Thrombectomy Device is a sterile, single-use catheter system for peripheral vascular revascularization. Device consists of a self-expanding nitinol basket attached to a core pusher wire. During procedure, device is delivered via microcatheter to target vessel, deployed across clot, and retrieved to remove emboli/thrombi. Radiopaque markers (platinum) at distal/proximal ends and along basket aid fluoroscopic visualization. Used by interventionalists in clinical settings. Proprietary 'drop zone' net pattern in nitinol basket optimizes clot capture. Device enables non-surgical restoration of blood flow, potentially reducing patient morbidity associated with peripheral vessel occlusion.
Clinical Evidence
Non-clinical testing included bench and GLP animal studies. Bench testing: physical/dimensional inspection, tensile/torque strength, delivery/retrieval forces, re-sheathing durability, navigation/steerability, flexibility/kink resistance, radiopacity, corrosion resistance, and clot retrieval testing. Biocompatibility per ISO 10993-1, sterilization per ISO 11137-1, and packaging per ISO 11607-1/2. Physician usability testing performed. No clinical human data provided.
Technological Characteristics
Self-expanding closed-cell nitinol basket; platinum radiopaque markers; core pusher wire. Compatible with 0.021" and 0.027" microcatheters. Sterile, single-use. Biocompatibility per ISO 10993-1; sterilization per ISO 11137-1; packaging per ISO 11607-1/2. No electronic components or software.
Indications for Use
Indicated for non-surgical removal of emboli and thrombi from peripheral blood vessels and temporary use in peripheral vessel occlusion in patients requiring revascularization.
Regulatory Classification
Identification
An embolectomy catheter is a balloon-tipped catheter that is used to remove thromboemboli, i.e., blood clots which have migrated in blood vessels from one site in the vascular tree to another.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of the Department of Health & Human Services seal. To the right of the symbol is a blue square containing the acronym "FDA" in white letters, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
January 15, 2021
Vesalio Sigi Caron Regulatory and Clinical Consultant Biologics and Medical Device Consulting Group PO Box 7605 Mammoth Lakes, California 93546
Re: K201085
Trade/Device Name: NeVa™ PV Thrombectomy Device Regulation Number: 21 CFR 870.5150 Regulation Name: Embolectomy Catheter Regulatory Class: Class II Product Code: QEW, KRA Dated: December 15, 2020 Received: December 17, 2020
## Dear Sigi Caron:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K201085
Device Name NeVa PV Thrombectomy Device
#### Indications for Use (Describe)
The NeVa™ PV Thrombectomy Device is indicated for:
- · The non-surgical removal of emboli and thrombi from peripheral blood vessels.
- · Temporary use in peripheral vessel occlusion.
- · Use with aspiration and with the injection or infusion of contrast media and other fluids.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 1. 510(K) SUMMARY
This 510(k) summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(K) NUMBER | K201085 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DATE PREPARED | January 13, 2021 |
| APPLICANT | Vesalio<br>Gustavo R. Prado, PhD<br>gprado@vesalio.com<br>3200 West End Ave, Ste 500<br>Nashville, TN 37203 |
| OFFICIAL<br>CORRESPONDENT | Sigi Caron, MBA RAC<br>Biologics and Medical Device Consulting Group<br>4260 Paul Sweet Rd.<br>Santa Cruz, CA 95065<br>Tel: (310) 614-3120<br>Fax: (888) 295-1535<br>sigi@BioMDG.com |
| TRADE NAME | NeVa™ PV Thrombectomy Device |
| COMMON NAME | Embolectomy Catheter |
| DEVICE CLASSIFICATION | Product Codes:<br>QEW - Catheter, Embolectomy (21 CFR §870.5150)<br>KRA - Catheter, Continuous Flush (21 CFR) §870.1210)<br>Class: II |
| PREDICATE DEVICE | ReVive™ PV Thrombectomy Device (K132281) |
| REFERENCE DEVICE | Trevo Provue Thrombus Retriever (K122478)<br>Solitaire Platinum Revascularization Device (K160641) |
# DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:
The NeVa™ PV Thrombectomy Device is a sterile, single-use thrombectomy device designed to for use in the peripheral vasculature to restore blood flow via non-surgical removal of emboli and thrombi from peripheral vessels in the vasculature. The NeVa™ PV device consists of a self-expanding nitinol basket attached to a core pusher wire that can be delivered through a micro catheter, deployed across a clot, and then removed with the clot to enable revascularization of the occluded vessel. To aid in fluoroscopic visualization, radiopaque markers are positioned at the distal end, proximal end, and along the nitinol basket of the NeVa™ PV device. A flexible
{4}------------------------------------------------
atraumatic soft tip attached to the distal end of the nitinol basket is radiopaque for enhanced fluoroscopic visualization.
The NeVaTM PV device is provided sterile and is intended for single patient use. The device is not intended to be re-sterilized or re-used. No accessories are provided with the NeVaTM PV device. The device is intended to be used in conjunction with the appropriately sized microcatheter as indicated in the proposed Instructions for Use.
### INDICATIONS FOR USE:
The NeVa™ PV Thrombectomy Device is indicated for:
- The non-surgical removal of emboli and thrombi from peripheral blood vessels. ●
- Temporary use in peripheral vessel occlusion.
- Use with aspiration and with the injection or infusion of contrast media and other fluids. ●
# SUBSTANTIALLY EQUIVALENT TO:
The NeVa™ PV Thrombectomy Device is substantially equivalent in intended use and technological features to the ReVive™ PV Thrombectomy Device (K132281).
### SUMMARY OF TECHNOLOGICAL CHARACTERISTIC SIMILARITIES / DIFFERENCES:
The NeVa™ PV Thrombectomy Device and the predicate ReVive™ PV Thrombectomy Device are both sterile, single-use thrombectomy devices intended to restore blood flow via non-surgical removal of emboli and thrombi from peripheral vessels. Both the NeVa™ PV and the ReVive™ PV use a self-expanding nitinol basket to capture emboli and thrombotic material in occluded vessels. Both devices have a soft distal tip, integrated radiopaque markers for fluoroscopic visualization during the procedure, and are constructed of equivalent materials. Both devices are introduced by being loaded in an insertion sheath that is then delivered via a compatible microcatheter to the target vessel location for deployment.
The only notable difference between two the devices is that the NeVa™ PV Thrombectomy Device uses a different cell design (net pattern) in the nitinol basket structure. The NeVa PV incorporates a proprietary drop zone pattern to optimize capture of emboli and thrombotic material. The diameter of the basket remains similar in size, and the method of clot capture (entrapment) remains the same. There are no other technological differences between the two devices.
The NeVa™ PV device will also be offered in a range of configurations and sizes in order to provide more options to the interventionalist and to accommodate clot removal in different anatomical locations..
{5}------------------------------------------------
# BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
The NeVa™ PV Thrombectomy Device is substantially equivalent to the ReVive™ PV with respect to indications for use (intended use) and technical characteristics. The following table compares the features of the proposed NeVa™ PV device to its predicate.
| Element | New Device:<br>NeVa™ PV Thrombectomy Device | Primary Predicate:<br>ReVive™ PV Thrombectomy Device |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K201085 | K132281 |
| Indications for Use | The non-surgical removal of<br>emboli and thrombi from<br>peripheral blood vessels. Temporary use in peripheral<br>vessel occlusion. Use with aspiration and with the<br>injection or infusion of contrast<br>media and other fluids. | The non-surgical removal of emboli<br>and thrombi from peripheral blood<br>vessels. Temporary use in peripheral<br>vessel/graft occlusion. Use with aspiration and with the<br>injection or infusion of contrast<br>media and other fluids. The non-surgical removal of<br>thrombi from synthetic grafts. |
| Design Features | | |
| Use of basket to capture<br>clot? | Yes | Yes |
| Vessel Diameter Range | 2.0 - 6.0 mm | 1.5 - 5.0 mm |
| Basket Type | Self-Expanding Closed Cell Basket | Self-Expanding Closed Cell Basket |
| Basket Material | Nitinol | Nitinol |
| Basket - Maximum<br>Available Diameters | 4.0, 4.5, and 6.0 mm | 4.5 mm |
| Basket - Working<br>Lengths | 22 – 46 mm | 22 - 28 mm |
| Basket Radiopaque<br>Markers | Yes | Yes |
| Soft Distal Tip | Yes | Yes |
| Distal Tip Length | 5 mm | 6 mm |
| Distal /Proximal<br>Radiopaque Markers | Yes | Yes |
| Radiopaque Marker<br>Material | Platinum | Platinum |
| Core Pusher Wire | Yes | Yes |
| Pusher Wire Material | Nitinol | Nitinol |
| Compatible<br>Microcatheter Sizes | 0.021", 0.027" | 0.027" |
| Element | New Device:<br>NeVa™ PV Thrombectomy Device | Primary Predicate:<br>ReVive™ PV Thrombectomy Device |
| Loaded into Introducer<br>Sheath? | Yes | Yes |
| Provided Sterile? | Yes | Yes |
| Single Use? | Yes | Yes |
{6}------------------------------------------------
### PERFORMANCE STANDARDS:
No performance standards have been established by the Agency to date that apply to this device.
# SUMMARY OF NONCLINICAL TESTING:
Comprehensive design verification and validation testing was performed on the NeVa™ PV Thrombectomy Device. Results from bench and animal testing demonstrated substantial equivalence to the predicate. Testing performed included a combination of the following:
- Physical and Dimensional Inspection ●
- Tensile Strength ●
- Af Temperature ●
- Radial Forces
- Torque Strength / Torsion ●
- Delivery / Retrieval Forces ●
- Multiple Re-sheathing Durability ●
- Navigation/Steerability ●
- Flexibility/Kink Resistance
- Tip Deflection
- Radiopacity ●
- Corrosion Resistance ●
- Clot Retrieval Testing
- Physician Usability Testing ●
- Biocompatibility Testing (ISO 10993-1)
- Sterilization Validation (ISO 11137-1) ●
- Packaging Validation (ISO 11607-1 and -2) ●
- GLP Animal Study
Overall, testing confirms that the NeVa™ PV Thrombectomy Device can be used according to its intended use and in an equivalent manner to the predicate device.
CONCLUSION:
{7}------------------------------------------------
Based on the above information, Vesalio believes that the substantial equivalence of the NeVaTM PV Thrombectomy Device to the ReVive™ PV Thrombectomy Device has been demonstrated.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.