Solitaire Platinum Revascularization Device, 6x40 mm
K160641 · Micro Therapeutics, Inc. d/b/a ev3 Neurovascular · NRY · Mar 8, 2017 · Cardiovascular
Device Facts
Record ID
K160641
Device Name
Solitaire Platinum Revascularization Device, 6x40 mm
Applicant
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
Product Code
NRY · Cardiovascular
Decision Date
Mar 8, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K160641 · Mar 8, 2017
Solitaire Platinum Revascularization Device, 6x40 mm
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular
STRATIS Registry (Systematic Evaluation of Patients TReated with Neurothrombectomy Devices for AcuTe Ischemic Stroke)
The sponsor leveraged clinical performance and safety data from a subgroup of 296 patients treated with the 4x40 Solitaire device within the STRATIS registry to support the substantial equivalence of the new 6x40 device dimension.
STRATIS Registry; Prospective, multi-center, non-randomized, observational registry; Follow-up/Duration: 90 days
Patients diagnosed with acute ischemic stroke due to large intracranial vessel occlusion treated with Medtronic neurothrombectomy devices; Sample Size: 984 total enrolled; 296 in the 4x40 device subgroup; Number of Sites: 55
Not applicable for this study
Symptomatic ICH, Mortality at 90 days, Successful revascularization (TICI 2b-3), Functional Independence (mRS 0-2) at 90 days
Indications for Use
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
Device Story
Device is a neurovascular thrombus retriever; used to restore blood flow in large intracranial vessels (ICA, M1/M2 segments, basilar, vertebral arteries). Input: physician-guided delivery via microcatheter to occlusion site. Operation: distal nitinol stent-like component expands to engage/remove thrombus; platinum-iridium markers provide radiopacity for fluoroscopic guidance. Output: mechanical removal of clot. Used in clinical/hospital settings by physicians. Benefits: revascularization in acute ischemic stroke patients. Device is a portfolio expansion of predicate; identical materials/design/manufacturing.
Clinical Evidence
No new clinical study performed. Leveraged confirmatory data from the prospective, multi-center, non-randomized STRATIS Registry (n=296 patients treated with 4x40 Solitaire). Primary endpoints: safety (symptomatic ICH 2.8%, 90-day mortality 16.2%) and performance (successful revascularization TICI 2b-3 90.4%, functional independence mRS 0-2 60.3% at 90 days). Device-related SAEs were 0.3%.
Technological Characteristics
Nitinol stent and push-wire; 90% Platinum/10% Iridium radiopaque markers; PTFE push-wire shrink tubing; PTFE/Grilamid introducer sheath. Sterilization: Ethylene Oxide. Connectivity: None (mechanical). Dimensions: 6-40-10 mm. No software/algorithm.
Indications for Use
Indicated for patients experiencing ischemic stroke within 8 hours of symptom onset due to large intracranial vessel occlusion, who are ineligible for or failed IV t-PA therapy.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three heads, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the eagle. The eagle is depicted in black and white, and the text is in a simple, sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 8, 2017
Micro Therapeutics, Inc. d/b/a ev3 Neurovascular Jennifer Correa Senior Regulatory Affairs Specialist 9775 Toledo Way Irvine, California 92618
Re: K160641
Trade/Device Name: Solitaire™ Platinum Revascularization Device (6x40 mm) Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NRY Dated: February 16, 2017 Received: February 21, 2017
Dear Ms. Correa:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
William J.
Heetderks -S
Digitally signed by William J. Heetderks -
DN: c=US, o=U.S. Government, ou=HHS,
ou=FDA, ou=People,
0.9.2342.19200300.100.1.1=0010149848
cn=William J. Heetderks -S
Date: 2017.03.08 16:48:39-05'00'
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160641
#### Device Name
Solitaire™ Platinum Revascularization Device (6x40 mm)
#### Indications for Use (Describe)
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> × Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | K160641 |
|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | Micro Therapeutics, Inc. d/b/a ev3 Neurovascular<br>9775 Toledo Way<br>Irvine, CA 92618<br>Establishment Registration No. 2029214 |
| Contact Person: | Jennifer Correa<br>Senior Specialist, Regulatory Affairs<br>Telephone: (949) 297-9563<br>E-mail: Jennifer.L.Correa@medtronic.com |
| Date Summary<br>Prepared: | March 1, 2017 |
| Trade Name of<br>Device: | SolitaireTM Platinum Revascularization Device (6x40 mm) |
| Common Name of<br>Device: | Catheter, Thrombus Retriever |
| Classification of<br>Device: | 21 CFR 870.1250 - Class II |
| Product Code: | NRY |
| Predicate Device: | SolitaireTM Platinum Revascularization Device<br>510(k)#: K153071 |
| Performance Data: | The following non-clinical bench testing and confirmatory<br>clinical data supports the subject 6-40-10 SolitaireTM Platinum<br>Revascularization Device:<br>• Total System Length<br>• Marker-to-Marker Length<br>• Delivery Force<br>• Resheathing Force<br>• Multiple Resheathing Durability<br>• Body Finger Marker Coil Tensile<br>• Radial Force<br>• Radiopacity<br>• Ease of Use (Fluorosafe Marker Location)<br>• Particulate<br>• Leveraged confirmatory clinical data using the cleared<br>4x40 Solitaire devices |
| Conclusion: | The materials of construction, design, and manufacturing/<br>packaging processes for the subject 6-40-10 SolitaireTM Platinum<br>Revascularization Device are identical to the predicate SolitaireTM<br>Platinum Revascularization Device (K153071) therefore;<br>biocompatibility, sterilization, packaging, and shelf life<br>validations have been leveraged or adopted. Non-clinical animal<br>testing was performed to demonstrate safety, effectiveness, and |
{4}------------------------------------------------
usability of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical bench and animal testing confirmed that the subject 6-40-10 Solitaire™ Platinum Revascularization Device does not introduce new risks or increase existing risks. Confirmatory clinical data using the cleared 4x40 Solitaire devices (predicate 4-40-10 Solitaire™ Platinum and reference 4-40 Solitaire™ 2) was leveraged to support substantial equivalence of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. The subject Solitaire™ Platinum Revascularization Device should be determined to be substantially equivalent to the predicate Solitaire™ Platinum Revascularization Device (K153071) based on non-clinical bench and animal testing, confirmatory clinical data, similarities in design, principles of operation, and indications for use.
#### Device Description:
The subject 6-40-10 Solitaire™ Platinum Revascularization Device is designed to restore blood flow in patients experiencing ischemic stroke due to large intracranial vessel occlusion. The Solitaire™ Platinum Revascularization Device is designed for use in the neurovasculature such as the Internal Carotid Artery (ICA), M1 and M2 segments of the middle cerebral artery, basilar, and the vertebral arteries. The distal nitinol portion of the Solitaire™ Platinum Revascularization Device facilitates clot retrieval. The Solitaire™ Platinum Revascularization Device has Platinum Iridium radiopaque markers on the working cell length, proximal and distal ends. The subject Solitaire™ Platinum Revascularization Device is a portfolio expansion to the predicate Solitaire™ Platinum Revascularization Device (K153071).
### Indications for Use:
The Solitaire™ Platinum Revascularization Device is intended to restore blood flow by removing thrombus from a large intracranial vessel in patients experiencing ischemic stroke within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.
### Device Comparison:
The table below provides a comparison of the technological characteristics associated with the portfolio expansion of the subject 6-40-10 Solitaire™ Platinum Revascularization Device and the predicate Solitaire™ Platinum Revascularization Device (K153071).
{5}------------------------------------------------
| | Predicate Device<br>Solitaire™ Platinum<br>Revascularization<br>Device (K153071) | Subject Device<br>6-40-10 Solitaire™<br>Platinum<br>Revascularization<br>Device | Rationale for Difference<br>(if applicable) |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The Solitaire™<br>Platinum<br>Revascularization<br>Device is intended to<br>restore blood flow by<br>removing thrombus<br>from a large<br>intracranial vessel in<br>patients experiencing<br>ischemic stroke<br>within 8 hours of<br>symptom onset.<br>Patients who are<br>ineligible for<br>intravenous tissue<br>plasminogen<br>activator (IV t-PA) or<br>who fail IV t-PA<br>therapy are<br>candidates for<br>treatment. | Same | N/A |
| Principles of<br>Operation | The device is used in<br>the neurovasculature<br>to restore blood flow<br>by removing<br>thrombus. | Same | N/A |
| Method of<br>Supply | Stored within<br>dispenser coil, Tyvek<br>pouch, and shipping<br>carton. | Same | N/A |
| Sterilization<br>Method | Ethylene Oxide | Same | N/A |
| | Predicate Device<br>Solitaire™ Platinum<br>Revascularization<br>Device (K153071) | Subject Device<br>6-40-10 Solitaire™<br>Platinum<br>Revascularization<br>Device | Rationale for Difference<br>(if applicable) |
| Device Size(s) | 4-20-10<br>4-40-10<br>6-20-10 | 6-40-10 | The subject 6-40-10<br>Solitaire™ Platinum<br>Revascularization Device<br>shares similar technological<br>characteristics to the<br>predicate 6-20-10 (6 mm<br>stent diameter) and the 4-40-<br>10 (40 mm stent length) of<br>the predicate Solitaire™<br>Platinum Revascularization<br>Devices. Bench testing has<br>demonstrated that the subject<br>6-40-10 Solitaire™ Platinum<br>Revascularization Device is<br>within the existing ranges of<br>the 4x40 Solitaire devices for<br>device length, radial outward<br>force and delivery /<br>resheathing forces. |
| | | | |
| Stent | Nitinol. | Same | N/A |
| Push-wire | Nitinol. | Same | N/A |
| Distal Marker<br>Coil | 90% Platinum/<br>10% Iridium | Same | N/A |
| Body Finger<br>Marker Coil | 90% Platinum/<br>10% Iridium | Same | N/A |
| Push-wire<br>Shrink Tubing | PTFE | Same | N/A |
| Introducer<br>Sheath | PTFE/Grilamid | Same | N/A |
{6}------------------------------------------------
## Sterilization and Shelf Life:
The subject 6-40-10 Solitaire™ Platinum Revascularization Device is sterilized using the identical validated, Ethylene Oxide (EO) sterilization cycle as the predicate Solitaire™ Platinum Revascularization Device (K153071). The materials of construction, design, manufacturing process, packaging, and sterile load configurations are identical to the predicate Solitaire™ Platinum Revascularization Device (K153071). Therefore, no additional sterilization validation testing is required.
The subject 6-40-10 Solitaire™ Platinum Revascularization Device and subsequent packaging remain functional and maintain sterility for up to two (2) years. The materials
{7}------------------------------------------------
of construction, design, manufacturing process, packaging, and sterile load configurations are identical to the predicate Solitaire™ Platinum Revascularization Device (K153071). Therefore, no additional shelf life validation testing is required.
## Biocompatibility:
The subject 6-40-10 Solitaire™ Platinum Revascularization Device does not introduce any new materials into the finished device nor the manufacturing processes. The material of the subject 6-40-10 Solitaire™ Platinum Revascularization Device is identical to the material of the predicate Solitaire™ Platinum Revascularization Device (K153071). Biocompatibility data for the Solitaire device family was tested for the Solitaire™ FR Revascularization Device. The biocompatibility data for the Solitaire™ FR was adopted for the subject 6-40-10 Solitaire™ Platinum Revascularization Device. A summary of the biocompatibility testing completed for the Solitaire™ FR Revascularization Device includes:
| Test<br>Category | Test<br>Description | Method | Acceptance Criteria | Conclusion |
|-------------------------------|---------------------------------------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Cytotoxicity | L929 MTT<br>Cytotoxicity | ISO 10993-5 | Viability is $\geq$ 70%. | Acceptance criteria<br>met |
| Sensitization | Guinea Pig<br>Maximization<br>Sensitization | ISO 10993-10 | Test article does not<br>elicit a sensitization<br>response. | Acceptance criteria<br>met |
| Irritation | Intracutaneous<br>Irritation Test | ISO 10993-10 | Differences in the<br>mean test and control<br>scores of the extract<br>dermal observations<br>are $<$ 1.0. | Acceptance criteria<br>met |
| Acute<br>Systemic<br>Toxicity | Acute Systemic<br>Injection Test | ISO 10993-11 | No abnormal clinical<br>signs and weight loss<br>in excess of 10%.<br>Temperature rise<br>$\geq$ 0.5°C | Acceptance criteria<br>met |
| | Materials<br>Mediated<br>Rabbit Pyrogen | | | |
| Test<br>Category | Test<br>Description | Method | Acceptance Criteria | Conclusion |
| Hemo-<br>compatibility | Hemolysis | ISO 10993-4 | No significant differences between the test article extract and negative control article results. The test article is considered non-hemolytic | Acceptance criteria met |
| | Partial Thromboplastin Time | | Clotting times are similar to the negative control and the reference material (HDPE) indicating the device materials are not an activator of the intrinsic coagulation pathway. | |
| | Platelet and Leukocyte Count | | Test article does not adversely affect the platelet and leukocyte components of the blood compared to the reference material. | |
| | Complement Activation C3a and SC5b-9 Assay | | Levels of C3a and SC5b-9 are comparable and less than the positive control. | |
| | Thrombosis | | Thrombo-resistance properties are acceptable in clinical use. | |
| Genotoxicity | Bacterial Mutagenicity Test | ISO 10993-3 | Test article is considered non-mutagenic | Acceptance criteria met |
| | In-vitro Mouse Lymphoma Assay | | Test article is considered non-mutagenic | |
| | In-vivo Mouse Micronucleus Assay | | Test article is considered non-mutagenic | |
{8}------------------------------------------------
{9}------------------------------------------------
# Performance Data – Bench:
A summary of the non-clinical bench testing performed for the subject 6-40-10 Solitaire™ Platinum Revascularization Device is summarized in the table below.
| Test Description | Method | Acceptance Criteria | Conclusion |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-------------------------|
| Total System Length | Total system length was measured from the distal tip of the distal marker to the proximal tip of the delivery system. | The stent length should be adequate for its intended use for delivery via standard microcatheters. | Acceptance criteria met |
| Marker-to-Marker Length | Marker-to-marker length is representative of the total stent length. Total stent length was measured 100% in process. | The stent length should be adequate for its intended use for delivery via standard microcatheters. | Acceptance criteria met |
| Delivery Force | Delivery force was measured through a representative tortuous anatomical model. | The stent must be below delivery force specification. | Acceptance criteria met |
| Resheathing Force | Resheathing force was measured through a representative tortuous anatomical model. | The stent must be below re-sheathing force specification. | Acceptance criteria met |
| Multiple Resheathing Durability | Multiple resheathing durability was evaluated on the ability to withstand delivery and withdrawal forces in a representative tortuous model beyond the recommended number of passes and re-sheathings allowed per the Instructions for Use. | The stent must be able to be deployed and resheathed up to four times. | Acceptance criteria met |
| Body Finger Marker Coil Tensile | Body finger marker coil tensile was performed to verify the strength of the laser weld of the Platinum/Iridium marker coil to the Nitinol body finger. | The body finger marker coil tensile should be greater than or equal to existing tensile strength specification. | Acceptance criteria met |
| Test Description | Method | Acceptance Criteria | Conclusion |
| Radial Force | Radial force was measured 100% in process. | The stent must be within existing radial force specification. | Acceptance criteria met |
| Radiopacity | Radiopacity was verified by both a material and dimensional analyses. | The radiopaque body markers must be visible using standard catheter laboratory equipment. | Acceptance criteria met |
| Ease of Use<br>(Fluorosafe Marker Location) | Ease of use (fluorosafe marker location) was measured from the distal tip of the distal marker to the distal end of the fluorosafe marker. | Less required fluoroscopy is better than more required fluoroscopy. | Acceptance criteria met |
| Particulate | Particulate was evaluated for generation under simulated use in a representative tortuous anatomical model. | The stent must not generate clinically unacceptable particulate. | Acceptance criteria met |
{10}------------------------------------------------
## Performance Data - Animal:
Non-clinical animal testing was conducted to verify the performance of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical animal testing was conducted to verify the subject 6-40-10 Solitaire™ Platinum Revascularization Device is safe and effective in an in-vivo Yorkshire cross swine model. Four (4) of the following eight (8) proposed target sites were evaluated:
- Right Vertebral ●
- Left Vertebral ●
- Right Internal Thoracic ●
- Left Internal Thoracic ●
- Right Costocervical ●
- Left Costocervical ●
- Renal ●
- Ascending Pharyngeal
Please note the renal and ascending pharyngeal are considered alternative back-up sites depending on the size of the swine, distal target zone diameter, length, consistency of vessel diameter, and the ability to harvest the vessel post procedure.
{11}------------------------------------------------
Non-clinical animal testing was evaluated at 0-day (acute) and 30-day (chronic) time points. The non-clinical animal testing was designed to show no-clinical inferiority between the control 6-30 Solitaire™ 2 Revascularization Device and the test 6-40-10 Solitaire™ Platinum Revascularization Device. Non-clinical animal testing evaluated the following biological responses of vascular tissue specific to safety:
- . Luminal Thrombus
- Endothelial Cell Coverage
- Inflammation
- Fibrin
- Medial Smooth Muscle Cell (SMC) Loss ●
- Disruption of the Internal Elastic Lamina (IEL) and External Elastic Lamina (EEL)
- . Adventitia
- Neointimal Hyperplasia
- Elastic Fiber Separation ●
- . Lumen Area Reduction
## Performance Testing - Clinical:
## Type of data
A clinical study was not performed for the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Confirmatory clinical data using the cleared 4x40 Solitaire devices (predicate 4-40-10 Solitaire™ Platinum and reference 4-40 Solitaire™ 2) was leveraged to support substantial equivalence of the subject 6-40-10 Solitaire™ Platinum Revascularization Device. Leveraging of the 4x40 Solitaire device confirmatory clinical data is appropriate as the subject 6-40-10 Solitaire™ Platinum Revascularization Device is within the existing ranges of the 4x40 Solitaire devices for device length, radial outward force and delivery / resheathing forces. The confirmatory 4x40 Solitaire clinical data was collected as part of the Systematic Evaluation of Patients TReated with Neurothrombectomy Devices for AcuTe IschemicStroke (STRATIS) Registry.
## STRATIS Registry Design
A prospective, multi-center, non-randomized, observational registry evaluating the use of Covidien market-released neurothrombectomy devices in patients diagnosed with an acute ischemic stroke due to large intracranial vessel occlusion. The study used an independent Clinical Events Committee for adverse event adjudication and an independent Core Lab for procedural image assessment.
### Sample Size
Under the STRATIS Registry, 984 patients were enrolled that were treated with a Medtronic neurothrombectomy device at the discretion of treating physicians and as part of hospital standard practice at 55 sites within the United States. Of the 984 patients treated, a total of 296 were treated using only the cleared 4x40 device. Major patient eligible criteria for consent and enrollment into the Registry were to be treated with the Solitaire device within 8 hours of symptom onset.
{12}------------------------------------------------
## Confirmatory Clinical Data
The confirmatory clinical data leveraged from the STRATIS Registry 4x40 group included baseline demographics, safety data and performance data and showed that the cleared 4x40 Solitaire device has a similar clinical performance and safety profile to other cleared Solitaire device dimensions.
| STRATIS Registry 4x40 Subgroup | |
|----------------------------------------------------|----------------------------------------|
| Confirmatory Clinical Data | |
| Baseline Demographics | |
| Age (years) | 66.4 ± 15.2 (296)<br>67.0 (58.0, 77.0) |
| NIHSS at baseline | 16.8 ± 5.5 (296)<br>16.0 (12.0, 21.0) |
| Male Gender | 51.4% (152/296) |
| Safety Data | |
| Symptomatic ICH | 2.8% (7/252) |
| Mortality at 90 days | 16.2% (48/296) |
| Performance Data | |
| Successful<br>revascularization<br>(TICI 2b-3) | 90.4% (227/251) |
| Functional<br>Independence (mRS 0-2)<br>at 90 days | 60.3% (167/277) |
## Device Related Serious Adverse Events (SAEs)
| STRATIS Registry 4x40 Subgroup<br>Device-Related SAEs | |
|-------------------------------------------------------|--------------------|
| Study Device-Related<br>SAEs | 0.3% (1/296) [1]** |
** Subarachnoid hemorrhage
### Conclusion
Non-clinical bench, animal testing and the confirmatory clinical data demonstrated that the subject 6-40-10 Solitaire™ Platinum Revascularization Device is substantially equivalent to the predicate Solitaire™ Platinum Revascularization Device (K153071) and does not raise new questions of the safety and effectiveness of the device. Additionally, there are no changes to the indications for use or the fundamental scientific technology of the Solitaire™ Platinum Revascularization Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.