K201015 · Acutus Medical, Inc. · DQK · Dec 1, 2020 · Cardiovascular
Device Facts
Record ID
K201015
Device Name
AcQMap High Resolution Imaging and Mapping System
Applicant
Acutus Medical, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Dec 1, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart. OR When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.
Device Story
AcQMap System is a computer-based diagnostic recording system for use in EP labs; operates outside sterile field. Inputs: ultrasound data from AcQMap catheters, ECG/EGM signals, and impedance-based localization data from patient electrodes. System transforms inputs into 3D cardiac chamber reconstructions, catheter position tracking, and electrical activity maps (charge density or voltage-based). Output displayed on workstation for physician use. Enables identification of arrhythmia sources and guidance for conventional EP catheters. Benefits: improved visualization of complex arrhythmias compared to conventional mapping alone. Used by physicians to inform clinical decision-making during cardiac procedures.
Clinical Evidence
No new clinical testing performed. Clinical evidence for the original AcQMap System (K170948) incorporated by reference. The DDRAMATIC-SVT study (prospective, non-randomized, open-label, 84 patients) demonstrated safety and effectiveness for the system's intended use.
Technological Characteristics
Console-based system with workstation and auxiliary interface box. Sensing: ultrasound (10 MHz), impedance-based localization (15-50 kHz), and ECG/EGM recording (0.05-500 Hz). Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-25, IEC 60601-2-37. Class I, Type CF defibrillator-protected. IP20 ingress protection. Power: 100-230 VAC.
Indications for Use
Indicated for patients prescribed electrophysiology procedures for the visualization of cardiac anatomy and mapping of complex arrhythmias.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
Predicate Devices
AcQMap High Resolution Imaging and Mapping System, Model 900100 (K193013)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 1, 2020
Acutus Medical, Inc. Serena Sanginthirath Senior Regulatory Affairs Specialist 2210 Faraday Ave, Suite 100 Carlsbad. California 92008
Re: K201015
Trade/Device Name: AcQMap High Resolution Imaging and Mapping System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, IYO, ITX Dated: October 26, 2020 Received: October 27, 2020
Dear Serena Sanginthirath:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and probibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### ACQMAP® HIGH RESOLUTION IMAGING AND MAPPING SYSTEM, MODEL 900100
#### 510(K) Premarket Notification
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K201015
#### Device Name
AcQMap® High Resolution Imaging and Mapping System. Model 900100
#### Indications for Use (Describe)
The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
OR
When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to.
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### Date Prepared: April 20, 2020
#### General Information
### Applicant:
Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Phone: 1-442-232-6080 Fax: 1-442-232-6081
#### Contact Person:
Serena Sanginthirath Senior Regulatory Affairs Specialist Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008, U.S.A Phone: 1-442-232-6178 FAX: 1-442-232-6081 Email: Serena.Sanginthirath@acutus.com
#### Device Information
Trade Name: AcQMap" High Resolution Imaging and Mapping System, Model 900100
### Generic/Common Name:
Programable diagnostic computer and Ultrasonic pulsed echo imaging system
### Classification:
Class II / 21 CFR § 870.1425 and Class II / 21 CFR § 892.1560
# Product Code(s):
DQK, IYO, ITX
{4}------------------------------------------------
## PREDICATE DEVICE
AcQMap High Resolution Imaging and Mapping System, Model 900100 (K193013)
## DEVICE DESCRIPTION
The AcQMap High Resolution Imaging and Mapping System, Model 900100 operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation and the AcQMap Auxiliary Interface Box.
The AcQMap High Resolution Imaging and Mapping System, Model 900100 ("AcQMap System Model, 900100") is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.
The AcQMap System hardware consists of three functional subsystems:
- . Ultrasound imaging,
- . ECG and EGM recording; and
- . Impedance based electrode Localization.
The AcQMap System, Model 900100 is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter (cleared under K170819). The AcQMap System provides:
- 3-D cardiac chamber reconstruction - Contact and non-contact (ultrasound),
- . Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters,
- Cardiac electrical activity as waveform traces,
- . Contact LAT and voltage amplitude maps
- . Remapping of the chamber at any time during the procedure; and
- . Dynamic, three-dimensional, charge density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation.
The AcQMap System, Model 900100 is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.
Additionally, the AcQMap System allows physicians to perform traditional contact mapping activities, including establishing a coordinate system, localizing conventional electrophysiology catheters relative to one another within the coordinate system, recording contact electrograms, and initiating a procedure without the AcQMap Catheter present. Based on the information captured in the contact electrograms, the physician may decide to treat an arrythmia without deploying the AcQMap Catheter.
The AcQMap System is used with the AcQMap Patient Electrode Kit.
{5}------------------------------------------------
## INDICATIONS FOR USE
The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either charge density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
OR
When used with conventional electrophysiology catheters, the AcQMap System provides information about the electrical activity of the heart and about catheter location during the procedure.
## COMPARISON TO PREDICATE DEVICES
The subject device has the same intended use and fundamental scientific technology as the predicate device. Therefore, the purpose of the submission is replacing the cleared Patient Electrode Kit with the AcQMap Patient Electrode Kit. Table 1 provides a comparison of the predicate AcQMap System classification against the subject device. Table 2 provides a comparison of the indications for use with the predicate device. Table 3 provides a comparison of the technological characteristics against the predicate device relating to the device accessory.
| Table 1. Comparison of Classification with the Predicate Device | | |
|-----------------------------------------------------------------|----------------------------------------------------------------------------------|---------------------------------------------------------------------|
| Characteristics | Predicate Device | Subject Device |
| | AcQMap® High Resolution Imaging and<br>Mapping System, Model 900100<br>(K193013) | AcQMap® High Resolution Imaging and<br>Mapping System, Model 900100 |
| 510(k) Number | K193013 | TBD |
| Classification/ | Class II | Identical |
| Regulation<br>Number/ | 21 CFR § 870.1425<br>21 CFR § 892.1560 | |
| Regulation Name/ | Programable diagnostic computer<br>Ultrasonic pulsed echo imaging system | |
| Product Code | DQK, IYO, ITX | |
{6}------------------------------------------------
| Table 2. Comparison of Indications for Use with the Predicate Device | | |
|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristics | Predicate Device | Subject Device |
| | AcQMap® High Resolution Imaging and Mapping System, Model 900100 (K193013) | AcQMap® High Resolution Imaging and Mapping System, Model 900100 |
| 510(k) Number | K193013 | TBD |
| Indications for<br>Use | The AcQMap System is intended for use in patients for whom<br>electrophysiology procedures have been prescribed.<br><br>When used with the AcQMap Catheters, the AcQMap System is<br>intended to be used to reconstruct the selected chamber from<br>ultrasound data for purposes of visualizing the chamber<br>anatomy and displaying electrical impulses as either charge<br>density-based or voltage-based maps of complex arrhythmias<br>that may be difficult to identify using conventional mapping<br>systems alone.<br><br>AND<br><br>When used with the specified Patient Electrodes, the AcQMap<br>System is intended to display the position of AcQMap Catheters<br>and conventional electrophysiology (EP) catheters in the heart.<br><br>OR<br><br>When used with conventional electrophysiology catheters, the<br>AcQMap System provides information about the electrical<br>activity of the heart and about catheter location during the<br>procedure. | Identical |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device | | |
| Characteristics | Predicate Device | Subject Device |
| | AcQMap High Resolution Imaging and Mapping System, Model 900100 (K193013) | AcQMap High Resolution Imaging and Mapping System, Model 900100 |
| Patient Anatomy | Intracardiac Structures | Identical |
| Testing to Support<br>Substantial<br>Equivalence | Software V/V Electromagnetic and Electrical Safety Verification Testing, Accuracy Testing, and Animal Testing | Identical |
| System Safety<br>Standards | IEC 60601-1:2005 /A1:2012 IEC 60601-1-2:2014 IEC 60601-1-6:2010/A1:2013 IEC 60601-2-25:2015 IEC 60601-2-37:2015 | Identical |
| Physical Characteristics | | |
| System<br>Components | Console Workstation Workstation Cable Auxiliary Interface Box ECG Input Cable Ampere Ablation Catheter Adapter Cable Ampere RF Generator Adapter Cable ECG Output Cable Ablation Reference Cable Ablation Electrogram Cable ECG w/Snaps Cable ECG POST Cable 2mm Pin Jumper Set Patient Electrode Kit (800365-003) | All system components are identical, apart from the new Patient Electrode Kit (800605). Biocompatibility testing and performance testing on the new Patient Electrode Kit demonstrates that the AcQMap System and patient electrodes perform as intended. There are no different questions of safety or effectiveness. |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) | | |
| Characteristics | Predicate Device | Subject Device |
| | AcQMap® High Resolution Imaging and Mapping System, Model 900100 (K193013) | AcQMap® High Resolution Imaging and Mapping System, Model 900100 |
| Visual/Mapping<br>Characteristics | 3-D cardiac chamber reconstructions - Contact and non-contact (ultrasound); Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters; Cardiac electrical activity as waveform traces; Contact LAT and voltage amplitude maps; Remapping of the chamber at any time during the procedure; Dynamic, three-dimensional, Charge Density maps overlaid on the cardiac chamber reconstruction to show chamber-wide electrical activation. | Identical |
| Visualization<br>Device/Catheter | AcQMap Catheter (electrodes & transducers) or Conventional electrophysiology catheters | Identical |
| Physical Characteristics - Console/Amplifier Comparison | | |
| Dimensions | 99 cm L x 58 cm W x 76 cm D | |
| Weight Maximum | 80 kg | |
| Power<br>Requirement | 100-127 VAC, 50/60 Hz, 220-230 VAC, 50 Hz | Identical |
| Input Current | 4.6 A | |
| Fuse protection | 250 V, 6.3A, two high breaking capacity fuses | |
| Table 3: Comparison of Technological Characteristics Against the Predicate Device (Continued) | | |
| Characteristics | Predicate Device | Subject Device |
| | AcQMap® High Resolution Imaging and Mapping System, Model 900100 ( K193013) | AcQMap® High Resolution Imaging and Mapping System, Model 900100 |
| System Specifications | | |
| Safety<br>Information | IEC 60601-1, Class I, Type Defibrillator Protected CF, continuous<br>operation, no sterilization, equipment not suitable for use in the<br>presence of a flammable anesthetic mixture with air, oxygen or<br>nitrous oxide | Identical |
| Ingress<br>Protection | The Console is rated IP20 | Identical |
| Functional and Performance Characteristics | | |
| Ultrasound<br>Output | Frequency: 10 MHz+/-400 kHz<br>Maximum Voltage: 50V p-p<br>Maximum Power: 1 W peak | |
| Ultrasound<br>Performance | Single operating mode<br>Thermal Index less than 1.0<br>Mechanical Index less than 1.0 | Identical |
| Localization<br>Output | Frequency: Variable 15 kHz to 50 kHz<br>Maximum current: 1.2mA RMS | |
| ECG & EGM<br>Input | Bandwidth: 0.05 Hz to 500 Hz<br>Resolution: +/-1uV<br>Timing Accuracy: +/-1.6 microsecond | |
{7}------------------------------------------------
{8}------------------------------------------------
# AcQMap" High Resolution Imaging and Mapping System, Model 900100
{9}------------------------------------------------
# AcQMap" High Resolution Imaging and Mapping System, Model 900100
{10}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE
The indications for use of the subject device are identical to those of the predicate device. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the modified AcQMap High Resolution Imaging System, Model 900100, is substantially equivalent to the predicate device.
## PERFORMANCE DATA
All necessary bench testing was conducted on the modified AcQMap System to support a determination of substantial equivalence to the predicate device.
# NONCLINICAL TESTING SUMMARY
Design verification activities for functional testing were performed with their respective acceptance criteria to ensure that electrode patch modifications do not affect the safety or effectiveness of the device. All testing performed met the established performance specifications. Bench and animal testing were performed to verify the device met the pre-determined acceptance criteria. The following tests were performed.
- Performance Test
- Gel adhesion ।
- । Electrode impedance
- Wire insulation wall thickness .
- । Lead wire pull strength
- । Lead wire flex life
- Biocompatibility Test
- Shelf-Life Test
- Usability Test
- Non-GLP Animal Test
The changes to the Patient Electrode Kit were evaluated through design verification to show that the proposed AcQMap Patient Electrode Kit is acceptable for use and meets requirements.
The necessary bench testing was performed on the modified AcQMap High Resolution Imaging and Mapping System, Model 900100 to ensure that it conforms to the design specifications and to support a determination of substantial equivalence to the predicate device.
The collective results of the testing demonstrate that the design of the modified AcQMap High Resolution lmaging and Mapping System, Model 900100 meets its established performance specifications necessary for performance during its intended use.
The collective results of the nonclinical testing, either repeated for the modified device or incorporated by reference to the original AcQMap System 510(k) K193013, demonstrate that the materials chosen, the manufacturing processes, and design of the modified AcQMap High Resolution Imaging and Mapping System, Model 900100 meet the established specifications necessary for consistent performance during
{11}------------------------------------------------
Acutus Medical, Inc.
its intended use. In addition, the collective bench testing demonstrates that the proposed device does not raise different questions of safety or effectiveness when compared to the predicate device.
## CLINICAL TESTING SUMMARY
No further clinical testing is required to support the modified AcQMap High Resolution Imaging and Mapping System, Model 900100. The necessary clinical testing was completed for the original AcQMap System (K170948) and is incorporated by reference. That study, entitled, "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" was a prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S. The results for 84 patients demonstrated that the AcQMap System is safe and effective for its intended use.
# CONCLUSIONS
Nonclinical performance testing was conducted to the Patient Electrode Kit for the modified AcQMap System, to evaluate the overall performance of the device accessory. The clinical validation of the original AcQMap System (K170948) is applicable to the modified device. The collective results demonstrate that the modified AcQMap System, Model 900100 is safe and effective for its intended use. The subject AcQMap High Resolution Imaging and Mapping System, Model 900100 is safe and effective for its intended use and is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.