K170948 · Acutus Medical, Inc. · DQK · Oct 16, 2017 · Cardiovascular
Device Facts
Record ID
K170948
Device Name
AcQMap High Resolution Imaging and Mapping System
Applicant
Acutus Medical, Inc.
Product Code
DQK · Cardiovascular
Decision Date
Oct 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1425
Device Class
Class 2
Indications for Use
The AcQMap® High Resolution Imaging System is intended for use in patients for whom electrophysiology procedures have been prescribed. When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone. AND When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
Device Story
Diagnostic recording system for EP labs; uses ultrasound data and electrical signals to reconstruct 3D cardiac chamber anatomy. Input: ultrasound data from AcQMap Catheter; electrical signals from catheters. Processing: system reconstructs chamber surface and generates 3D dipole density or voltage-based maps of electrical activation. Output: 3D anatomical maps, catheter positions, and waveform traces displayed on workstation. Used by physicians to identify arrhythmia sources. Benefits: enables visualization of complex arrhythmias difficult to identify with conventional systems; allows remapping during procedures.
Clinical Evidence
Prospective, non-randomized, open-label study (DDRAMATIC-SVT) conducted at eight sites outside the U.S. with 84 patients. Evaluated safety and performance of the system in creating right and/or left atrial dipole density activation maps for supraventricular tachycardia. Results demonstrated the system is safe and effective for its intended use.
Technological Characteristics
System components: Console, Workstation, Patient Interface Unit, Ablation Interface Unit. Sensing: Ultrasound transducers and electrical electrodes. Connectivity: Standalone diagnostic recording system. Power: 110–240 V, 50/60 Hz. Software: Computer-based diagnostic recording system for 3D reconstruction and mapping.
Indications for Use
Indicated for patients undergoing electrophysiology procedures. Used for reconstruction of cardiac chambers from ultrasound data, visualization of anatomy, and display of electrical impulses (dipole density or voltage-based maps) for complex arrhythmias. Also indicated for displaying the position of AcQMap and conventional EP catheters in the heart.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 16, 2017
Acutus Medical, Inc. Brenda Clay Regulatory Affairs Manager 2210 Faraday Ave., Suite 100 Carlsbad, California 92008
Re: K170948
Trade/Device Name: AcQMap High Resolution Imaging and Mapping System Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DOK, IYO, ITX Dated: September 15, 2017 Received: September 18, 2017
Dear Brenda Clay:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
M.A. Wilhelm
for Bram D. Zuckerman, M.D.
Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170948
#### Device Name
AcQMap® High Resolution Imaging and Mapping System
Indications for Use (Describe)
The AcQMap® High Resolution Imaging System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Notification K170948
# GENERAL INFORMATION [807.92(a)(1)]
#### Applicant:
Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1-442-232-6080 Fax: 1-442-232-6081
#### Contact Person:
Brenda Clay Regulatory Affairs Manager Acutus Medical, Inc. 2210 Faraday Ave., Suite 100 Carlsbad, CA 92008 USA Phone: 1- 442-232-6161 FAX: 1- 442-232-6081 Email: Brenda.Clay@acutus.com
# Date Prepared: March 29, 2017
## DEVICE INFORMATION [807.92(a)(2)]
#### Trade Name:
AcQMap® High Resolution Imaging and Mapping System
### Generic/Common Name:
Programable diagnostic computer and Ultrasonic pulsed echo imaging system
### Classification:
Class II/21 CFR § 870.1425 and Class II/21 CFR § 892.1560
### Product Code:
DOK, IYO, ITX
{4}------------------------------------------------
# PREDICATE DEVICE(S) [807.92(a)(3)]
- EnSite Velocity™ Cardiac Mapping System (K130594); and o
- iLabTM Ultrasound Imaging System (K130243) ●
# DEVICE DESCRIPTION [807.92(a)(4)]
The AcQMap System operates outside of the sterile field and consists of the AcQMap Console, the AcQMap Workstation, the AcQMap Patient Interface Unit, and the AcOMap Ablation Interface Unit.
The AcQMap High Resolution Imaging and Mapping System ("AcQMap System") is a diagnostic recording system. This computer-based system is intended for use in the Electrophysiology (EP) Lab, and it is capable of imaging, navigation and mapping of the atrial chambers of the heart.
The AcQMap System hardware consists of three functional subsystems:
- Ultrasound imaging, 0
- ECG and EGM recording; and o
- Impedance based electrode Localization. 0
The AcQMap System is used in conjunction with the associated AcQMap 3D Imaging and Mapping Catheter. The AcQMap System provides:
- o 3-D cardiac chamber reconstruction imaging,
- Three-dimensional position of the AcQMap Catheter and conventional electrophysiology catheters,
- Cardiac electrical activity as waveform traces, ●
- Remapping of the chamber at any time during the procedure; and o
- Three-dimensional, dipole density maps overlaid on the cardiac chamber o reconstruction to show chamber-wide electrical activation.
The AcQMap System is a diagnostic recording system consisting of ultrasound, electrical mapping components, a console, and a workstation. The AcQMap System is intended to create a surface reconstruction of the cardiac chamber as well as an electrical map of the substrate. The surface reconstruction and electrical map are then used by physicians to identify the source(s) of the arrhythmia.
# INDICATIONS FOR USE [807.92(a)(5)]
The AcQMap System is intended for use in patients for whom electrophysiology procedures have been prescribed.
When used with the AcQMap Catheters, the AcQMap System is intended to be used to reconstruct the selected chamber from ultrasound data for purposes of visualizing the chamber anatomy and displaying electrical impulses as either dipole density-based or voltage-based maps of complex arrhythmias that may be difficult to identify using conventional mapping systems alone.
# AND
When used with the specified Patient Electrodes, the AcQMap System is intended to display the position of AcQMap Catheters and conventional electrophysiology (EP) catheters in the heart.
{5}------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES [807.92(A)(6)]
Table 1 provides a comparison of the technological characteristics for the AcQMap System and the predicate devices.
{6}------------------------------------------------
510(k) Summary
| Characteristics | AcQMap® High Resolution<br>Imaging and Mapping System | EnSite™ Velocity™ Cardiac<br>Mapping System | iLab™ Ultrasound Imaging<br>System | Rationale for Substantial<br>Equivalence |
|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory | | | | |
| 510(k) Number | TBD | K130594 | K130243 | - |
| Classification/<br>Regulation<br>Number/Regulation<br>Name/Product<br>Code | Class II/21 CFR §<br>870.1425/Programable diagnostic<br>computer/DQK<br><br>Class II/21 CFR §<br>892.1560/Ultrasonic pulsed echo<br>imaging system/IYO, ITX | Class II/21 CFR §<br>870.1425/Programable<br>diagnostic computer/DQK | Class II/21 CFR §<br>892.1560/Ultrasound pulsed<br>echo imaging system/IYO,<br>ITX | The additional product codes<br>and regulation number for the<br>proposed device are required. |
| Indications for Use | The AcQMap System is intended<br>for use in patients for whom<br>electrophysiology procedures have<br>been prescribed.<br><br>When used with the AcQMap<br>Catheters, the AcQMap System is<br>intended to be used to reconstruct<br>the selected chamber from<br>ultrasound data for purposes of<br>visualizing the chamber anatomy<br>and displaying electrical impulses<br>as either dipole density-based or<br>voltage-based maps of complex<br>arrhythmias that may be difficult to<br>identify using conventional<br>mapping systems alone.<br><br>AND<br><br>When used with the specified<br>Patient Electrodes, the AcQMap<br>System is intended to display the<br>position of AcQMap Catheters and<br>conventional electrophysiology<br>(EP) catheters in the heart. | The EnSite™ Velocity™<br>Cardiac Mapping System is a<br>suggested diagnostic tool in<br>patients for whom<br>electrophysiology studies are<br>indicated.<br><br>*When used with the EnSite<br>Array™ Catheter, the EnSite™<br>Velocity™ Cardiac Mapping<br>System is intended to be used<br>in the right atrium of patients<br>with complex arrhythmias that<br>may be difficult to identify<br>using conventional mapping<br>systems alone.<br><br>OR<br><br>*When used with the EnSite™<br>NavX™ Surface Electrode Kit,<br>the EnSite™ Velocity™<br>Cardiac Mapping System is<br>intended to display the<br>position of conventional<br>electrophysiology (EP)<br>catheters in the heart. | The iLab™ Ultrasound<br>Imaging System is intended<br>for ultrasound examinations<br>of intravascular pathology.<br>Intravascular ultrasound is<br>indicated in patients who are<br>candidates for transluminal<br>interventional procedures<br>such as angioplasty and<br>atherectomy. | All devices are intended for<br>electrophysiology procedures.<br>Both the AcQMap System and<br>EnSite System are indicated for<br>complex arrhythmias when<br>used with their respective<br>catheters. Both the AcQMap<br>System and iLab System<br>include the ultrasound<br>performance capabilities in the<br>indications for use statement.<br>The semantic differences in the<br>indications for use do not raise<br>different questions of safety or<br>effectiveness than the<br>predicates as demonstrated by<br>the AcQMap performance<br>testing. |
| | | | | |
| | able 1: "Comparison of Techniological Characteristics and | Tharacteristics with the Fredicate Devices (count.) tion | | ationale for Substant |
| tient Anator | acardiac Struct | cardiac Struct | ravascular Anator | ased on the performan Sting Continues and Children and China China There monimalian and Children and China Mesting salistics and Children and Children Checkives since so |
| festing to Suppor ubstantial Iquivalence | Safety Testing Testing Creating National Testing Comments National Testing Comments Nichinal Testing manus<br>oftware V/V lectromagnetic and Electrica<br>●<br>●<br>0 | Bench testing; a Animal testing | ctional/performan<br>eroperability testin<br>ctrical Safety tes<br>esting<br>and<br>0 | omplete performance testing onders parformance as alternale and and and the compressioners of the compressionelly of the contribution of the chives the chives the chivenest w |
| System Safet Standards | ●<br>● | ר הייתר משותר משותר משושים של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של המושב של<br>601 .1<br>●<br>● | Inknowr | Same |
| Characteristics | AcQMap® High Resolution<br>Imaging and Mapping System | EnSite™ Velocity™ Cardiac<br>Mapping System | iLab™ Ultrasound Imaging<br>System | Rationale for Substantial<br>Equivalence |
| Physical Characteristics | | | | |
| System<br>Components | Console, Workstation, Workstation<br>Cable Patient Interface Unit, Patient<br>Interface Unit Cable, Ablation<br>Interface Unit, Ablation Interface<br>Unit Cable, Auxiliary Catheter<br>Cable, Surface ECG Input Cable<br>and Patient Electrode Kit. | Amplifier and Display<br>Workstation (DWS)<br>Amplifier includes Navlink,<br>ArrayLink and CathLink<br>Modules, SJM ECG Cable,<br>Record Connect and<br>GenConnect.<br> | Unknown | Same |
| Visual/Mapping<br>Characteristics | 3-D cardiac chamber<br>reconstructions imaging, Three-dimensional position of<br>the AcQMap Catheter and<br>conventional electrophysiology<br>catheters, Cardiac electrical activity as<br>waveform traces, Remapping of the chamber at<br>any time during the procedure;<br>and Three-dimensional, dipole<br>density maps overlaid on the<br>cardiac chamber reconstruction<br>to show chamber-wide<br>electrical activation. | Three-dimensional<br>position of the catheter Display cardiac electrical<br>activity as waveform<br>traces; Three-dimensional maps<br>of the cardiac chamber | Combination of imaging<br>and soft tissue<br>visualization. Visualizes the catheter<br>relative to the anatomical<br>structure | The combination of three-<br>dimensional imaging<br>functionality and ultrasound<br>functionality for display of<br>dipole density maps does not<br>raise different questions of<br>safety or effectiveness as<br>demonstrated through<br>performance testing. |
| Visualization<br>Device/Catheter | AcQMap Catheter (electrodes and<br>transducers) | EnSite Array Catheter<br>(electrode) | BSC Ultra ICE - Ultrasound<br>Dx Catheter (transducer) | |
| Characteristics | AcQMap® High Resolution Imaging<br>and Mapping System | EnSite™ Velocity™ Cardiac<br>Mapping System | iLab™ Ultrasound Imaging<br>System | Rationale for Substantial<br>Equivalence |
| Physical Characteristics - Console/Amplifier Comparison | | | | |
| Dimensions | 68 cm L x 48.3 cm W x 72.6 cm D | 49 cm H x 46 cm W x 51 cm D | Unknown | Same |
| Weight<br>Maximum | 50 kg | 31 kg | 213 pounds (System weight) | |
| Power<br>Requirement | 110 – 240 V, 50/60 Hz | 100, 110/120, 220/240 V<br>~50/60 Hz | Unknown | |
| Input Current | 5A | 450W maximum (4.5A) | Unknown | |
| Fuse protection | 250 V, 6.3 A, two high breaking<br>capacity fuses | 250 V, 5 A VAC, two fuses | Unknown | |
Table 1: Comparison of Technological Characteristics with the Predicate Devic
{7}------------------------------------------------
510(k) Summary
Table 1: Comparison of Technological Characteristics with the Predicate Devices (con
{8}------------------------------------------------
510(k) Summary
{9}------------------------------------------------
510(k) SUMMAR
-
{10}------------------------------------------------
## SUBSTANTIAL EQUIVALENCE
The indications for use for the predicate devices is substantially equivalent to the proposed indications for use for the AcQMap System. Any differences in the technological characteristics between the devices do not raise any different questions of safety or effectiveness. Thus, the AcQMap System is substantially equivalent to the predicate devices.
## PERFORMANCE DATA [807.92(b)]
All necessary bench and clinical testing was conducted on the AcQMap System to support a determination of substantial equivalence to the predicate devices.
# [807.92(b)(1)] Nonclinical Testing Summary
The necessary bench testing was performed on the AcOMap System to ensure that it conforms to the design specifications and to support a determination of substantial equivalence to the predicate device. The bench testing performed included the following:
- Transportation Testing ●
- AcQMap Verification Testing ●
- Software Verification and Validation
- System Accuracy Testing
- Electromagnetic Compatibility and Electrical Safety Testing
- . AcQMap Catheter Validation Testing-Animal Study
The AcQMap System device was tested to verify that the device met the established performance specifications.
The collective results of the testing demonstrate that the design of the AcQMap System meets its established performance specifications necessary for performance during its intended use.
The collective results of the nonclinical testing demonstrate that the materials chosen, the manufacturing processes, and design of the AcQMap System meet the established specifications necessary for consistent performance during its intended use. In addition, the collective bench testing demonstrates that the AcQMap System does not raise different questions of safety or effectiveness for collecting data that enables the creation of 3D anatomic maps that display chamber-wide electrical activation when compared to the predicate devices.
# [807.92(b)(2)] Clinical Testing Summary
Although clinical testing is not required to demonstrate substantial equivalence to the predicate device for its intended use, Acutus performed a clinical study to demonstrate the safety and performance of the AcQMap High Resolution Imaging and Mapping System in gathering data to create right and/or left atrial dipole density activation maps in subjects with supraventricular tachycardia. The study entitled, "Dipole Density Right (and left) Atrial Mapping and Assessment of Therapy In Complex Supraventricular Tachycardia, (DDRAMATIC-SVT)" is a prospective, non-randomized, open-label study conducted at eight clinical sites outside the U.S. The results for 84 patients demonstrates that the AcQMap System is safe and effective for its intended use.
{11}------------------------------------------------
# CONCLUSIONS [807.92(b)(3)]
Extensive nonclinical performance testing was conducted on the AcQMap System to evaluate the overall performance of the device. Acutus also performed clinical validation testing to evaluate the performance of the AcQMap System with the AcQMap 3D Imaging and Mapping Catheter. The collective results demonstrate that the AcQMap System is safe and effective for its intended use.
### SUMMARY
The AcQMap System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.