K200786 · Venus Concept USA, Inc. · GEX · Jun 9, 2020 · General, Plastic Surgery
Device Facts
Record ID
K200786
Device Name
Venus Epileve
Applicant
Venus Concept USA, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jun 9, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venus Epileve is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for: - Hair removal; - Permanent hair reduction (defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen); and - Treatment of pseudofolliculitis barbae.
Device Story
Tabletop diode laser system for hair removal and pseudofolliculitis barbae treatment; utilizes 800nm infrared laser energy. Operates via selective photothermolysis; melanin in hair follicles absorbs laser energy, causing thermal denaturation. System includes console with power supply, laser driver, controller, and LCD touch screen; detachable applicator with sapphire light guide and integrated contact cooling (water/TEC). Operated by clinicians in dermatology/surgery settings. Clinician selects parameters via GUI; triggers energy delivery via applicator button or footswitch. Integrated safety features include emergency stop, user authentication, and software/hardware output limits. Benefits include effective hair reduction with enhanced patient comfort via active skin cooling.
Clinical Evidence
Bench testing only. No clinical data presented. Performance testing included software validation, device verification, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), and laser safety (IEC 60825-1). Biocompatibility of patient-contacting materials confirmed per ISO 10993.
Technological Characteristics
Diode laser (800±15 nm); fiber-coupled output. Sapphire light guide tip with aluminum anodized shell. Integrated water cooling and TEC for contact skin cooling. Tabletop console; 15-inch LCD touch display. Safety features: emergency stop, user login/password. Biocompatible materials (ISO 10993). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60825-1, IEC 60601-2-22, IEC 62304.
Indications for Use
Indicated for hair removal, permanent hair reduction, and treatment of pseudofolliculitis barbae in all Fitzpatrick skin types, including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
June 9, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Venus Concept USA Inc. % Elissa Burg Regulatory Consultant BioVision Ltd Had Nes 183 Had Nes, 1295000 Israel
Re: K200786
Trade/Device Name: Venus Epileve Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: March 19, 2020 Received: March 26, 2020
Dear Elissa Burg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200786
Device Name
Venus Epileve
Indications for Use (Describe)
The Venus Epileve is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for:
- · Hair removal:
- Permanent hair reduction (defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen); and
- · Treatment of pseudofolliculitis barbae.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
## VENUS EPILEVE
#### 510(k) number K200786
| Applicant Name: | Venus Concept USA Inc.<br>1880 N Commerce Pkwy,<br>Ste 2, Weston, FL33326, USA<br>Tel: +972-549599215 |
|-----------------------|--------------------------------------------------------------------------------------------------------|
| Contact Person: | Dr. Yoni Iger<br>VP QA/RA/CA<br>Venus Concept USA Inc. |
| Date Prepared: | June 3, 2020 |
| Trade Name: | Venus Epileve |
| Classification Name: | 21 CFR 878.4810 Laser surgical instrument for use in general and plastic<br>surgery and in dermatology |
| Product Code: | GEX |
| Classification: | Class II Medical Device |
| Classification Panel: | General & Plastic Surgery |
| Predicate Device: | Venus Concept's Velocity Diode Laser System (K162765) |
#### Intended Use/Indication for Use:
The Venus Epileve is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for:
- Hair removal; ●
- Permanent hair reduction (defined as the long-term stable reduction in the number ● of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen); and
- Treatment of pseudofolliculitis barbae
{4}------------------------------------------------
### Device Description:
The Venus Epileve device consists of a tabletop console (main unit) and one detachable diode laser applicator. The system console is the main enclosure that holds and contains all external and internal system components: a power supply unit, laser diode driver, controller unit, and touch screen with LCD display. The console incorporates a water cooling system with a pump in order to constantly keep the tip of the applicator, light guide, chilled.
The detachable diode laser applicator is connected to the system via a cable passed through the applicator connectors located in the designated compartment on the system's rear side. When not in use, the applicator can be secured within the applicator cradle on the top panel of the console. The applicator delivers the laser energy to a patient's skin through the applicator's tip made of a sapphire light guide framed in aluminum anodized shells. The applicator is supplied with only one spot size light guide.
The device also comes with a footswitch, goggles, patient eye protectors, and water filling kit.
## Technological Characteristics:
The Venus Epileve delivers laser energy to the patient's skin via the diode laser applicator connected to the console. The console utilizes diode laser module as sources of optical energy and the optical output is fiber-coupled through the applicator to the treatment area.
Pulsed laser energy emitted by the system is absorbed by melanin in the hair follicles, which have greater optical absorption at the selected laser wavelength than the surrounding tissue. The selective absorption leads to localized heating and thermal denaturation of the target with minimal effect on the surrounding tissues.
The short- and long-term impact of the selective photo thermolysis on hair subparts caused by diode laser energy emitted into the skin leads to hair removal and permanent hair reduction. The laser applicator has an integrated contact skin cooling system to enhance safety and comfort during the treatment.
The device incorporates several safety features, including an emergency laser stop button, user login and password protection, and other software and hardware settings to mitigate the risk of improper energy release and ensure that system outputs are within specifications.
## Performance Data:
Venus Concept conducted several performance tests to demonstrate that the Venus Epileve system complies with performance standards and that it functions as intended.
- Performance Bench Testing: Several performance tests were performed, including software . validation and device verification tests in order to evaluate the Venus Epileve system's outputs per specifications, and as compared to the predicate device's specifications. The
{5}------------------------------------------------
results demonstrated that the differences in the technological characteristics of the subject and predicate device do not raise new types of safety or effectiveness concerns.
- Electrical Safety and Electromagnetic Compatibility: In addition, the system was tested per ● the applicable electrical safety and electromagnetic compatibility standards listed below, and all results were passing.
- . IEC 60601-1:2012 Ed. 3.1. Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Ed. 4. General requirements for basic safety and essential . performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-6: 2013 Ed.3.1, General requirements for basic safety and essential . performance - Collateral standard: Usability
- IEC 60825-1:2014 Ed.3, Safety of laser products Part 1: Equipment classification and . requirements
- IEC 60601-2-22:2007 Ed. 3 Particular requirements for basic safety and essential . performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
- . IEC 62304:2015 Medical device software - Software life cycle processes
- Software Testing: The software was also subjected to verification and validation testing, and . results demonstrated that the system performed as intended and that the energy outputs of the device meet specifications.
These performance tests demonstrated that the device meets the system requirements and do not raise new types of safety or effectiveness concerns. In addition, the patient-contacting materials are biocompatible per ISO 10993.
### Substantial Equivalence:
The following table compares the Venus Epileve system to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.
{6}------------------------------------------------
# Venus Concept Ltd.'s Venus Epileve Substantial Equivalence Chart
# Table 1. Technical Specifications
| Parameter | Venus Epileve<br>Venus Concept Ltd.<br>(Subject Device) | Venus Velocity<br>Venus Concept Ltd.<br>(Predicate Device) | |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(k) Number | Pending | K162765 | |
| Product<br>Class, Code | Class II, GEX | Class II, GEX | |
| Regulation<br>Number | 21 C.F.R. § 878.4810 | 21 C.F.R. § 878.4810 | |
| Regulation | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (878.4810) | Laser surgical instrument for use in general and plastic surgery and in<br>dermatology (878.4810) | |
| Manufacturer | Venus Concept Ltd | Venus Concept Ltd | |
| Intended Use<br>and Indications<br>for Use | The Venus Epileve is intended for all Fitzpatrick skin types, including<br>tanned skin, for use in dermatology, general and plastic surgery<br>applications for:<br>• Hair removal;<br>• Permanent hair reduction (defined as the long-term stable reduction<br>in the number of hairs regrowing when measured at 6, 9, and 12<br>months after the completion of a treatment regimen); and<br>• Treatment of pseudofolliculitis barbae. | The Venus Concept Velocity Diode laser is intended for all Fitzpatrick<br>skin types, including tanned skin, for use in dermatology, general and<br>plastic surgery applications for:<br>• Hair removal;<br>• Permanent hair reduction (defined as the long-term stable reduction<br>in the number of hairs regrowing when measured at 6, 9, and 12<br>months after the completion of a treatment regimen); and<br>• Treatment of pseudofolliculitis barbae. | |
| Laser System - Technical Specifications | | | |
| Main<br>Components | • A tabletop laser console (with graphical user interface)<br>• An applicator with a sapphire light guide | • Floor standing laser console (with graphical user interface).<br>• An applicator with three sizes of sapphire light guides (small,<br>medium, large) | |
| Energy Type | Diode laser | Diode laser | |
| Wavelength | 800±15 nm nominal infrared (785-815 nm) | 800±15 nm nominal infrared (785-815 nm) | |
| Spot Size | 2.0 cm² (16.5mm x 12 mm) | • Small: 1.69 cm² (13mm x13mm)<br>• Medium: 3.5 cm² (13mm x27mm)<br>• Large: 7.0 cm² (20mm x 35mm) | |
| Pulse Energy [J] | 80 J | • Small: 10-170 J<br>• Medium: 14-140 J<br>• Large: 28-140 J | |
| Energy Density,<br>Fluence [J/cm²] | 4-40 J/cm² | • Small: 6-100 J/cm²<br>• Medium: 6-40 J/cm²<br>• Large: 4-20 J/ cm² | |
| Pulse Duration<br>[msec] | 5-170 ms | 5 - 170 ms | |
| Frequency [Hz] | Slide Mode: ≤10 Hz , Pulse Mode:≤ 3 Hz | • Small: Slide Mode: ≤10 Hz , Pulse Mode:≤ 3 Hz<br>• Medium: Slide Mode: ≤10 Hz , Pulse Mode:≤ 3 Hz<br>• Large: Slide Mode: ≤ 2 Hz , Pulse Mode:≤ 4 Hz | |
| Max Peak<br>Power [Watt] | 1600 W | 2400 W | |
| Light guide | Sapphire | Sapphire | |
| Cooling system | Water cooling, closed system<br>Light Guide cooled by TEC | Water cooling, closed system<br>Light Guide cooled by TEC | |
| Light guide<br>cooling | (+5) °C - (+15)°C | (+5) °C - (+15)°C | |
| Beam Delivery | Diode array laser located in applicator/ cooled sapphire contact tip<br>(reusable) | Diode array laser located in applicators/ cooled sapphire contact tip<br>(reusable) | |
| Beam Control | Applicator trigger button or footswitch (optional) | Applicators trigger button | |
| Electrical<br>Requirements | • Voltage: 100-240 VAC<br>• Frequency: 50-60 Hz<br>• Power: ≤0.6 KVA | • Voltage: 100-240 VAC<br>• Frequency: 50-60 Hz<br>• Power: ≤1.2 KVA | |
| Overall Weight | ~25 kg (55 lbs) | ~35 kg (77.2 lbs) | |
| System console<br>Dimensions | 45 x 56 x 35 cm<br>(W x D x H, not packed) | 65 x 55 x 110 cm<br>(W x D x H, not packed) | |
| Display | LCD touch display 15 inch | LCD touch display 10.4 inch | |
{7}------------------------------------------------
{8}------------------------------------------------
As described in the comparison table above, the subject Venus Epileve and the predicate Venus Velocity (K162765) devices share the same intended use and same indications for use, and similar technological characteristics and principles of operation. The differences in the system console design, and available energy parameters between the Venus Epileve and its predicate (K162765) do not present any new questions of safety or effectiveness, as is further supported by the wide range of diode lasers already cleared for the same indications.
Furthermore, the Venus Epileve device underwent performance testing, including bench testing, software validation testing, electrical safety according to IEC 60601-1, electromagnetic compatibility testing according to IEC 60601-1-2, laser safety testing according to IEC 60825-1 and others. These performance tests in addition to a bench test demonstrated that the differences in the technological characteristics between the subject and predicate devices do not raise new types of safety or effectiveness concerns.
### Conclusions:
Therefore, based on the same intended use and indications for use, similar technological characteristics, and principles of operation, the Venus Epileve device is substantially equivalent to its predicate device Venus Velocity (K162765).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.