K162765 · Venus Concept , Ltd. · GEX · Jan 11, 2017 · General, Plastic Surgery
Device Facts
Record ID
K162765
Device Name
Venus Velocity Diode Laser System
Applicant
Venus Concept , Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Jan 11, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venus Velocity is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for: - Hair removal; - Permanent hair reduction (defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen); and - Treatment of pseudofolliculitis barbae.
Device Story
Venus Velocity is a diode laser system for hair removal and treatment of pseudofolliculitis barbae. Device consists of a console and a detachable diode laser applicator with interchangeable sapphire light guides. System uses a water-cooling pump to regulate handpiece/tip temperature, minimizing patient discomfort and burn risk. Operated by clinicians in dermatology or plastic surgery settings. Clinician selects energy parameters via software interface; laser energy is delivered to the skin to target hair follicles. Safety features include emergency stop, user authentication, and software-controlled energy output limits. Benefits include permanent hair reduction and treatment of skin conditions associated with hair growth. Device performance is verified through electrical safety, electromagnetic compatibility, and biocompatibility testing.
Clinical Evidence
Bench testing only. Includes electromagnetic compatibility and electrical safety testing per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. Biocompatibility of patient-contacting materials confirmed per ISO 10993-1 and ISO 10993-5. Software verification and validation performed. System verification confirmed energy outputs meet specifications.
Technological Characteristics
Diode laser system; console with water-cooling pump; detachable applicator; interchangeable sapphire light guides (small, medium, large). Energy source: diode laser. Safety features: emergency stop, user login/password, software-controlled energy limits. Biocompatible materials per ISO 10993. Electrical/EMC compliance per IEC 60601 series.
Indications for Use
Indicated for hair removal, permanent hair reduction, and treatment of pseudofolliculitis barbae in all Fitzpatrick skin types, including tanned skin.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Beijing Anchorfree Technology's Diode Laser Hair Removal System (K141973)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with three curved lines above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 11, 2017
Venus Concept Ltd. % Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market St., 29th Floor Philadelphia, Pennsylvania 19103
Re: K162765
Trade/Device Name: Venus Velocity Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use In General And Plastic Surgery and In Dermatology Regulatory Class: Class II Product Code: GEX Dated: November 29, 2016 Received: November 29, 2016
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K162765
Device Name
Venus Velocity
Indications for Use (Describe)
The Venus Velocity is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for:
- Hair removal:
- Permanent hair reduction (defined as the long-term stable reduction in the number of hairs regrowing when ● measured at 6, 9, and 12 months after the completion of a treatment regimen); and
- Treatment of pseudofolliculitis barbae.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
## Venus Concept Ltd.'s Venus Velocity
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Venus Concept Ltd. 4556 N. Hiatus Road Sunrise, FL 33351 Phone: 416.907.0115 Facsimile: 954.572.5680
Contact Person: Tal Bresler-Stramer, Ph.D., RAC, Vice President, QA/RA
Date Prepared: November 29, 2016
## Name of Device and Classification/Product Code
Trade/Proprietary Name: Venus Velocity
Common or Usual Name: Diode laser system
Classification Name: 21 C.F.R. § 878.4810, Laser surgical instrument for use in general and plastic surgery and in dermatology
Product Code: GEX
# Predicate and Reference Devices
Predicate Devices: Palomar Medical Technologies, Inc.'s Palomar Vectus Laser (K120622); Lumenis Ltd. LightSheer® Desire (K151947)
Reference Devices: Alma Lasers Modified Diode Laser Module with SHR Treatment Mode for use with the Family of Soprano XL Multi-Application Platforms (K112031): Sandstone Medical's Cheveux Diode Laser System (K100893); Milesman's Milesman Premium (K073300); ILOODA Company's Vikini (K151232); Beijing Anchorfree Technology's Diode Laser Hair Removal System (K141973).
# Intended Use / Indications for Use
The Venus Velocity is intended for all Fitzpatrick skin types, including tanned skin, for use in dermatology, general and plastic surgery applications for:
- Hair removal;
- o Permanent hair reduction (defined as the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen); and
- . Treatment of pseudofolliculitis barbae.
# Technological Characteristics
The Venus Velocity consists of a console, a detachable diode laser applicator, and three interchangeable sapphire light quides/tips (small, medium, and large sizes). A water system and
{4}------------------------------------------------
pump in the console distribute water to cool the handpiece and tips during device use, decreasing the likelihood of burns, discomfort, or pain during treatment. The system is also provided with a water filling kit.
The device incorporates several safety features, including an emergency laser stop button, user login and password protection, and other software and hardware settings to mitigate the risk of improper energy release and ensure that system outputs are within specifications. The device also comes with goggles and patient eye protectors.
## Performance Data
Performance testing of the subject device supports its safety and effectiveness profile for the proposed indications. Electromagnetic compatibility and electrical safety testing was conducted per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-22, and IEC 60825-1. All results were passing. In addition, the patient-contacting materials are biocompatible per ISO 10993-1 and ISO 10993-5. The Venus Velocity also underwent software verification and validation, with results demonstrating that the software is appropriate for release. System verification testing further confirmed that the system performs as intended, and that the energy outputs of the device meet specifications.
# Substantial Equivalence
The Venus Velocity has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices, the Palomar Vectus and Lumenis LightSheer Desire laser systems. The minor technological differences between the Venus Velocity and its predicates raise no new issues of safety or effectiveness. The key energy parameters are very similar for the devices, where the available wavelength, spot sizes, pulse durations, fluence outputs, and peak power for the Venus Velocity are within the ranges previously cleared for the predicates. In addition, the higher frequency available with the Venus Velocity does not raise different questions of safety or effectiveness because other laser-based devices for similar indications have also been cleared with the same 10 Hz maximum frequency. Performance data demonstrate that the Venus Velocity performs as intended, and further support that the device performs as intended. Thus, the Venus Velocity is substantially equivalent to its predicates.
# Conclusions
The tests outlined above demonstrate that the device performs as intended. The Venus Velocity has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. Minor differences between the subject and predicate devices do not present any new or different types of safety or effectiveness questions. In sum, the Venus Velocity is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.