K200573 · Medical Instinct Deutschland GmbH · DZE · May 14, 2021 · Dental
Device Facts
Record ID
K200573
Device Name
BoneTrust Mini Implant System
Applicant
Medical Instinct Deutschland GmbH
Product Code
DZE · Dental
Decision Date
May 14, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BoneTrust® Mini and Mini+ Implants are intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of permanent endosseous dental implant(s). Use of BoneTrust® Mini and Mini+ Implants is not to exceed one hundred and eighty (180) days. BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments are intended for use with BoneTrust® Mini / Mini+ Dental Implants to provisional prosthetic reconstructions during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments is not to exceed one hundred and eighty (180) days. BoneTrust® Mini Ball Attachments are intended to be used with BoneTrust® Mini / Mini+ Dental Implants to support and/or retain provisional removable dental prostheses during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Ball Attachments is not to exceed one hundred and eighty (180) days.
Device Story
BoneTrust® Mini Implant System comprises two-piece, root-form dental implants and associated abutments (Esthetic, Crown Base, Ball Attachments) for temporary prosthetic support. Implants are self-tapping, manufactured from Grade 4 titanium (machined or sand-blasted/acid-etched). Abutments are titanium alloy (Ti-6Al-4V). System uses external torx connection for implant-abutment interface. Intended for use in partially or fully edentulous mandibles and maxilla during healing phase of permanent implants; maximum duration 180 days. Implants supplied sterile (beta-radiation); abutments supplied non-sterile for end-user sterilization. Clinicians place implants surgically; prosthetic components are attached to support provisional restorations. System aids in restoring chewing function during healing period. Benefits include temporary stabilization of provisional prostheses, facilitating patient comfort and function while permanent implants integrate.
Clinical Evidence
No clinical studies were performed. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility assessment, bacterial endotoxin testing (USP <85>), sterilization validation (ISO 11137), shelf-life validation, and surface analysis (SEM/EDX).
Technological Characteristics
Implants: Grade 4 titanium (ISO 5832-2/ASTM F67). Abutments: Ti-6Al-4V alloy (ISO 5832-3/ASTM F136). Connection: External torx. Surface: Machined or sand-blasted/acid-etched/passivated. Sterilization: Beta-radiation (implants); end-user (abutments). Dimensions: Diameters 2.3-2.5mm, lengths 10.0-13.0mm.
Indications for Use
Indicated for partially or fully edentulous patients requiring temporary support for provisional prosthetic devices (fixed or removable) during the healing phase of permanent endosseous dental implants. Use limited to 180 days.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Sterngold 2.2mm Angled Micro ERA Dental Implant System (K092434)
BoneTrust® Implant Systems, Medical Instinct Deutschland GmbH (K182313)
MIS LOCKIT Abutments System, OT-Equators and Ball Attachments, MIS Implants Technologies Ltd. (K182228)
Temporary Snap Abutment, Nobel Biocare AB (K161435)
MS SA Implant System, OSSTEM Implant Co. Ltd. (K122171)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Medical Instinct Deutschland GmbH % André Weingerl RA Consultant WRC Consulting Am Hohstetter 1a Steißlingen, Baden Württemberg 78256 GERMANY
Re: K200573
Trade/Device Name: BoneTrust® Mini Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 8, 2021 Received: April 12, 2021
Dear André Weingerl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200573
Device Name BoneTrust® Mini Implant System
### Indications for Use (Describe)
The BoneTrust® Mini and Mini+ Implants are intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of permanent endosseous dental implant(s). Use of BoneTrust® Mini and Mini+ Implants is not to exceed one hundred and eighty (180) days.
BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments are intended for use with BoneTrust® Mini / Mini+ Dental Implants to provisional prosthetic reconstructions during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments is not to exceed one hundred and eighty (180) days.
BoneTrust® Mini Ball Attachments are intended to be used with BoneTrust® Mini / Mini+ Dental Implants to support and/or retain provisional removable dental prostheses during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Ball Attachments is not to exceed one hundred and eighty (180) days.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div style="display:flex; align-items:center;"> <span> <span style="font-size:1.5em;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div style="display:flex; align-items:center;"> <span> <span style="font-size:1.5em;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K200573
## VOLUME 006
## 510(k) Summary
Date of Submission 2021-14-05
APPLICANT: Medical Instinct Deutschland GmbH Graseweg 24 37120 Bovenden Germany Tel.: +49 5593 95196 Fax.: +49 5593 95195
> E-Mail: info@medical-instinct.de Internet: www.medical-instinct.de
CONTACT PERSON: Jan Worlitz Managing Director Tel.: +49 5593 95196 E-Mail: worlitz@medical-instinct.de
{4}------------------------------------------------
#### 1. Device Name
| Trade Name: | BoneTrust® Mini Implant System |
|----------------------|----------------------------------------------------------------|
| Classification Name: | Endosseous dental implant - Endosseous dental implant abutment |
#### 2. Classification Product Code / Subsequent Code
BoneTrust® Mini Implant system can be classified according to following Device Name and Product Code:
| Product Code: | DZE |
|-------------------------|-----------------|
| Device Class: | 2 |
| Classification Panel: | Dental |
| Regulation number: | 21 CFR 872.3640 |
| Secondary Product Code: | NHA |
## 3. Predicate device(s)
#### Primary Predicate Device .
| EM Provisional; Hiossen, Inc. | | #K191751 |
|-------------------------------------------------------------------|--|----------|
| • Reference Devices | | |
| Sterngold 2.2mm Angled Micro ERA Dental Implant System, Sterngold | | #K092434 |
| BoneTrust® Implant Systems, Medical Instinct Deutschland GmbH | | #K182313 |
| MIS LOCKIT Abutments System, OT-Equators and Ball Attachments, | | #K182228 |
| MIS Implants Technologies Ltd. | | |
| Temporary Snap Abutment, Nobel Biocare AB | | #K161435 |
| MS SA Implant System, OSSTEM Implant Co. Ltd. | | #K122171 |
{5}------------------------------------------------
#### Device Description 4.
The BoneTrust® Mini Implant System is a two-piece dental implant system including various sizes of threaded root-form dental implants and abutments to provide temporary support of prosthetic restorations in edentulous or partially edentulous patients during the healing phase of permanent endosseous dental implant(s). The maximum duration of intraoral use of all members of the BoneTrust® Mini Implant System is 180 days
#### 4.1. Implants Description
The BoneTrust® Mini Implants are root form implants manufactured from commercially pure titanium (Grade 4) per ISO 5832-2 and are supplied either with a machined surface or a sand-blasted, acid-etched surface treatment. All BoneTrust® Mini Implant types feature an external torx connection for fitment of BoneTrust® Mini cement-type abutments or ball type attachments
The following types of BoneTrust® Mini Implants are available:
## BoneTrust® Mini implant
Image /page/5/Figure/8 description: The image shows a close-up of a dental implant. The implant is made of a metallic material and has a screw-like shape. The implant is designed to be inserted into the jawbone to support a dental prosthesis, such as a crown, bridge, or denture.
The BoneTrust® Mini Implant is a self-tapping straight body implant manufactured from commercially pure titanium with a machined surface. The implant features an external torx Implant-abutment connection.
## BoneTrust® Mini+ Implant
Image /page/5/Figure/11 description: The image shows a close-up of a dental implant. The implant is made of metal and has a screw-like shape. The top of the implant is smooth and rounded, while the bottom is threaded. The implant is designed to be inserted into the jawbone to support a dental prosthesis, such as a crown, bridge, or denture.
The BoneTrust® Mini+ Implant is a self-tapping tapered body implant manufactured from commercially pure titanium with a microstructured surface by means of sandblasting / acid-etching surface-treatment followed by passivation. The implant features an external torx Implant-abutment connection.
| Description | BoneTrust® Mini implant | BoneTrust® Mini+ Implant |
|-----------------------------------|---------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Design / Shape | Screw type implant, straight Body, self-tapping | Screw type implant, tapered Body, self-tapping |
| Sizes - Diameter<br>x length (mm) | 2.3 x 11.5<br>2.3 x 13.0 | 2.5 x 10.0<br>2.5 x 13.0 |
| Cuff height (mm) | 2.25 | 3.50 |
| Material | Unalloyed titanium grade 4<br>(ISO 5832-2:1999 / ASTM F67-06:2006) | Unalloyed titanium grade 4<br>(ISO 5832-2:1999 / ASTM F67-06:2006) |
| Surface | Machined surface | Microstructured blasted etched surface passivated |
| Prosthetic Platform | Ball Attachment (Denture O-Ball),<br>Abutments for temporary cement-type restorations | Ball Attachment (Denture O-Ball),<br>Abutments for temporary cemented restorations |
| Abutment Connection | External Torx Connection | External Torx Connection |
| Packaging /<br>Sterilization | Sterile packed in blister; Sterilized by irradiation | Sterile packed in blister; Sterilized by irradiation |
| Duration of use | Max 180 days | Max 180 days |
{6}------------------------------------------------
## 4.2. Abutments
BoneTrust® Mini Abutments are pre-manufactured single-piece Abutments directly connected to the BoneTrust® Mini Dental Implants and intended for use as an aid in the temporary prosthetic rehabilitation during the healing phase of permanent dental implants. BoneTrust® mini Abutments include Ball-type connections intended for support of removable overdentures and bridges as well as Cement-type connections for fixed temporary prosthetic restorations.
### 4.2.1. BoneTrust® Mini Esthetic Abutment
Image /page/6/Picture/5 description: The image shows a gray, cone-shaped object. The object appears to be made of a smooth material, possibly plastic or metal. The cone shape is wider at the bottom and tapers towards the top. There are some horizontal lines near the bottom of the cone, creating a layered effect.
The BoneTrust® Mini Esthetic Abutments are pre-manufactured Abutments manufactured from Ti-6Al-4V Alloy as per ISO 5832-3 / ASTM F136 standard. The Abutments are available in straight shape. They are characterized by an anatomical garland-shaped course which replaces the need for customization to a greatest possible extent. They are intended for single tooth restoration as well as bridge restorations. The abutment might be shortened up to a minimum abutment post height of 4 mm. The abutments are not intended for angular correction, correction of diameter or taper.
| Description | Mini Esthetic Abutment | | |
|----------------------------|------------------------------------------------------------------|----------------------|------------------------------------------------------------------------|
| Angulation (°) | 0 | | |
| Available Sizes | Platform Diameter | Gingival Height (GH) | |
| | 2.7 mm | 1.0 mm | |
| Total Abutment Height [mm] | 5.5 | | |
| Design | External torx connection | | |
| Material | Titanium alloy Ti-6AI-4V (ISO 5832-3 / ASTM F136) | | |
| Surface | Machined | | |
| Packaging / Sterilization | packed in an individual PE bag - Must be sterilized by the user. | | |
| Duration of use | Max 180 days | | |
| End User Modification | Wall thickness/diameter | | No modifications allowed |
| | Post height | | Shortening to a minimum Abut-<br>ment Post Height of 4 mm al-<br>lowed |
| | Angulation and divergence | | No modifications allowed |
| | Gingival height | | No modifications allowed |
| | Connection platform | | No modifications allowed |
| | Method of modification | | Hand Tools only |
{7}------------------------------------------------
## 4.2.2. BoneTrust® Mini Crown Base Abutment
Image /page/7/Figure/3 description: The image shows a close-up of a metal dental implant. The implant is cylindrical and has a series of threads running along its length. The threads are evenly spaced and appear to be designed to grip the bone. The implant has a smooth, polished surface and is likely made of titanium or a titanium alloy.
BoneTrust® Mini Crown Base Abutments are premanufactured Abutments manufactured from Ti-6Al-4V Alloy as per ISO 5832-3 / ASTM F136 standard. The Abutments are available in straight shape and intended to be used for dentures and screw retained bridge constructions. The abutment might be shortened up to a minimum abutment post height of 4 mm. The abutments are not intended for angular correction, correction of diameter or taper.
| Description | Mini Crown Base Abutment | | | |
|----------------------------|------------------------------------------------------------------|----------------------|------------------------------------------------------------------------|--|
| Angulation (°) | 0 | | | |
| | Platform Diameter | Gingival Height (GH) | | |
| Available Sizes | 2.7 mm | 1.0 mm | | |
| Total Abutment Height [mm] | 7.5 | | | |
| Connection Type | External torx connection | | | |
| Material | Titanium alloy Ti-6Al-4V (ISO 5832-3 / ASTM F136) | | | |
| Surface | machined | | | |
| Packaging / Sterilization | packed in an individual PE bag - Must be sterilized by the user. | | | |
| Duration of use | Max 180 days | | | |
| | Wall thickness/diameter | | No modifications allowed | |
| | Post height | | Shortening to a minimum Abut-<br>ment Post Height of 4 mm al-<br>lowed | |
| End User Modification | Angulation and divergence | | No modifications allowed | |
| | Gingival height | | No modifications allowed | |
| | Connection platform | | No modifications allowed | |
| | Method of modification | | Hand Tools only | |
{8}------------------------------------------------
#### 4.2.3. BoneTrust® Mini Ball-Attachments
Image /page/8/Picture/3 description: The image shows a dental implant and its abutment. The implant is a screw-shaped device that is surgically placed into the jawbone to support a dental prosthesis, such as a crown, bridge, or denture. The abutment is a connector that attaches to the implant and provides a base for the prosthesis. The implant and abutment are made of titanium, a biocompatible material that is well-tolerated by the body. The implant has a rough surface to promote bone growth and integration, while the abutment has a smooth surface to prevent plaque accumulation.
BoneTrust® Mini Ball-Attachments are indicated for the use of temporary nonsplinted, removable dentures in the treatment of partially or totally edentulous patients to restore chewing function. They consist of the Ball type Attachment and the O-Ring matrix. The Ball type Attachment is manufactured from Ti-6Al-4V Alloy as per ISO 5832-3 / ASTM F136 standard with a machined surface. This attachment is mounted on the implant by a screw-locking mechanism and serves as the male part of the removable BoneTrust® Mini Ball-Attachment system. The retaining female counterpart, the O-rinq matrix is manufactured from Ti-6Al-4V Alloy as per ISO 5832-3 / ASTM F136 with a machined surface. The O-Ring matrix is incorporated into an existing prosthesis via polymerization, engages the outside of the ball shape and allows retention of the prosthesis to the denture.
All components of the BoneTrust® Mini Ball-Attachment Abutments are not intended for any kind of modification by the user.
| Description | Mini Ball-Attachments |
|---------------------------|------------------------------------------------------------------|
| Angulation (°) | 0 |
| Platform diameter (mm) | 2.7 |
| Body diameter (mm) | 2.3 |
| Connection Type | Internal locking screw fixation |
| Material | Titan Grade 5 (ISO 5832-3 / ASTM F136) |
| Surface | machined |
| Packaging / Sterilization | packed in an individual PE bag - Must be sterilized by the user. |
| Duration of use | Max 180 days |
#### 4.3. BoneTrust® Mini Retaining Screw
Image /page/8/Picture/8 description: The image shows a close-up of a small, metallic screw. The screw has a cylindrical head and a threaded shaft. The metal appears to be a light gray color. The screw is likely used for fastening or joining materials together.
The BoneTrust® Mini Retaining Screw is manufactured from Ti-6Al-4V Alloy as per ISO 5832-3 / ASTM F136 standard with a machined surface. It is used to firmly connect BoneTrust® Mini Abutments with BoneTrust® Mini Implants via the implant's and abutment's torx-connection.
The BoneTrust® Mini Retaining Screw is delivered sterile and supplied with each BoneTrust® Mini Implant.
The BoneTrust® Mini Retaining Screw is not intended for any kind of modification by the user.
| Description | Mini Retaining Screw |
|---------------------------|------------------------------------------------------|
| Angulation (°) | 0 |
| Total Length (mm) | 4.4 |
| Diameter (mm) | 1.8 |
| Connection Type | M1,4x0,25 |
| Material | Titanium alloy Ti-6AI-4V (ISO 5832-3 / ASTM F136) |
| Surface | machined |
| Packaging / Sterilization | Sterile packed in blister; Sterilized by irradiation |
| Duration of use | Max 180 days |
{9}------------------------------------------------
#### 5. Indications for Use
The BoneTrust® Mini and Mini+ Implants are intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of permanent endosseous dental implant(s). Use of BoneTrust® Mini and Mini+ Implants is not to exceed one hundred and eighty (180) days.
BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments are intended for use with BoneTrust® Mini / Mini+ Dental Implants to provisional prosthetic reconstructions during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Mini Esthetic abutments and BoneTrust® Mini Crown Base Abutments is not to exceed one hundred and eighty (180) days.
BoneTrust® Mini Ball Attachments are intended to be used with BoneTrust® Mini / Mini+ Dental Implants to support and/or retain provisional removable dental prostheses during the healing phase of permanent endosseous dental implants. Use of BoneTrust® Ball Attachments is not to exceed one hundred and eighty (180) days.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for "medical instinct". The words are in gray, except for the plus sign in the middle, which is orange. The logo also includes a registered trademark symbol in the upper right corner.
#### 6. Technological Characteristics and Substantial Equivalence
## 6.1. Comparison of indications for use and Technological Characteristics
The indications for use as well as technologics of the proposed devices are compared to the primary predicate and reference devices in the following tables:
| Company | Proposed Device:<br>Medical Instinct Deutsch-<br>land GmbH | Primary Predicate De-<br>vice:<br>Hiossen Inc. | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutsch-<br>land GmbH | Reference Device:<br>OSSTEM Implant Co.<br>Ltd. | Result |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Name | BoneTrust® Mini / Mini+<br>Implant System | EM Provisional | Sterngold 2.2mm Angled<br>Micro ERA Dental Implant<br>System | BoneTrust® Implant Sys-<br>tems | MS SA Implant System,<br>Type: Narrow ridge | -- |
| 510(K)<br>number | -- | K191751 | K092434 | K182313 | K122171 | -- |
| Intended<br>use | The BoneTrust® Mini and<br>Mini+ Implants are intended<br>to be loaded immediately in<br>partially or fully edentulous<br>mandibles and maxilla to<br>serve as temporary support<br>for provisional prosthetic<br>device during the healing<br>phase of permanent endos-<br>seous dental implant(s). Use<br>of BoneTrust® Mini and<br>Mini+ Implants is not to<br>exceed one hundred and<br>eighty (180) days. | The EM Provisional is<br>intended to be loaded<br>immediately in partially or<br>fully edentulous mandibles<br>and maxilla to serve as<br>temporary support for<br>provisional prosthetic<br>device during the healing<br>phase of permanent en-<br>dosseous dental im-<br>plant(s). | The Sternqold 2.2mm Angled<br>ERA dental implants are intend-<br>ed for long term as well as<br>temporary surgical implantation<br>in the bone of the patient's upper<br>or lower arch to provide immedi-<br>ate load or delayed load of<br>prosthetic systems, such as<br>artificial teeth, in order to restore<br>the patient's chewing function.<br>Immediate loading of the ERA<br>Implant should only occur when<br>the position of the implants<br>provides adequate bone quantity<br>and quality to allow proper<br>immediate mechanical stabiliza-<br>tion of the self-tapping screw<br>into the bone and where occlu-<br>sal and lateral forces can be<br>limited with appropriate occlusal<br>design and a soft diet. | Bone Trust® Dental Im-<br>plants are medical devices<br>intended to be surgically<br>placed in the bone of the<br>maxillary and/or mandibular<br>arches to provide support for<br>prosthetic restorations<br>(crowns, bridges or overden-<br>ture) in edentulous or partial-<br>ly edentulous patients to<br>restore a patients' chewing<br>function.<br>Bone Trust® Dental Im-<br>plants are in-tended for<br>immediate or delayed load-<br>ing after 12 weeks. | The MIS SA Implant (Nar-<br>row Ridge) is intended to<br>use in the treatment of<br>missing mandibular central<br>and lateral incisors to<br>support prosthetic device;<br>such as artificial teeth, in<br>order to restore chewing<br>function in partially eden-<br>tulous patients. The MIS<br>SA Implant (Narrow<br>Ridge) is intended for<br>single use only. It is in-<br>tended for delayed loading | Similar to primary<br>predicate device.<br>The primary predi-<br>cate device contains<br>limitations of implant<br>duration in the label-<br>ing.<br>The additional limita-<br>tion of implant dura-<br>tion is similar to<br>reference devices<br>and does not change<br>the intended use of<br>the device |
## 6.1.1. BoneTrust® Mini / Mini+ Implants
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for "medical instinct". The words are in lowercase letters and are gray, except for the plus sign between "ins" and "tinct", which is orange. A small "®" symbol is located to the right of the word "instinct".
| Company | Proposed Device:<br>Medical Instinct Deutsch-<br>land GmbH | Primary Predicate De-<br>vice:<br>Hiossen Inc. | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutsch-<br>land GmbH | Reference Device:<br>OSSTEM Implant Co.<br>Ltd. | Result |
|--------------------------------------|------------------------------------------------------------|------------------------------------------------|--------------------------------------------|-------------------------------------------------------------|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implant<br>material | Titanium Grade 4 ASTM<br>F67 | Titanium Grade 4 ASTM<br>F67 | Commercially pure Titanium | Titanium Grade 4 ASTM<br>F67 | Titanium Alloy Ti-6Al-4V<br>(ASTM F 136) | Identical to reference<br>device K182313 |
| Implant<br>Type | Screw-type | Screw-type | Screw-type | Screw-type | Screw-type | Identical |
| Design | straight body / tapered Body | straight body | straight body | straight body / tapered body | straight body | Identical to reference<br>device K182313 |
| Screw type | Self-tapping | Self-tapping | Self-tapping | Self-tapping | Self-tapping | Identical |
| Distal End | With thread | With thread | With thread | With thread | With thread | Identical |
| Single or<br>two piece | Two piece design | One piece design | Two piece design | Two piece design | One piece design | Identical to reference<br>devices K092434<br>and K182313 |
| Implant to<br>abutment<br>connection | Eternal Torx Implant / Abut-<br>ment fixture | Not applicable (One piece<br>design) | External hex Implant / Abutment<br>fixture | Cylindrical Internal Hexagon<br>or and conical torx | Not applicable (One piece<br>design) | The implant-<br>Abutment connection<br>geometry does not<br>raise any concerns in<br>regard to safety and<br>effectiveness of the<br>implant abutment<br>interface |
{12}------------------------------------------------
Image /page/12/Picture/1 description: The image shows the logo for "medical instinct". The words are in lowercase letters and gray. The plus sign in the middle of the word "instinct" is orange. There is a registered trademark symbol in the upper right corner of the logo.
| Company | Proposed Device:<br>Medical Instinct Deutsch-<br>land GmbH | Primary Predicate De-<br>vice:<br>Hiossen Inc. | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutsch-<br>land GmbH | Reference Device:<br>OSSTEM Implant Co.<br>Ltd. | Result |
|-------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Implant<br>Dimensions<br>(Diameter X<br>Length) | Ø2.3 X 11.5<br>Ø2.3 X 13.0<br>Ø2.5 X 10.0<br>Ø2.5 X 13.0 | Ø1.8 mm X 10.0 mm<br>Ø1.8 mm X 13.0 mm<br>Ø1.8 mm X 15.0 mm<br>Ø2.5 mm X 10.0 mm<br>Ø2.5 mm X 13.0 mm<br>Ø2.5 mm X 15.0 mm | Ø2.2 mm X 10.0 mm<br>Ø2.2 mm X 13.0 mm<br>Ø2.2 mm X 15.0 mm | Ø3.4 X 8<br>Ø3.4 X 10<br>Ø3.4 X 11,5<br>Ø3.4 X 13<br>Ø3.4 X 14,5<br>Ø4 X 6.5<br>Ø4 X 8<br>Ø4 X 10<br>Ø4 X 11.5<br>Ø4 X 13<br>Ø4 X 14.5<br>Ø5 X 6.5<br>Ø5 X 8<br>Ø5 X 10<br>Ø5 X 11.5<br>Ø5 X 13 | Ø2.5 mm X 8.5 mm<br>Ø2.5 mm X 10.0 mm<br>Ø2.5 mm X 11.5 mm<br>Ø2.5 mm X 13.0 mm<br>Ø2.9 mm X 8.5 mm<br>Ø2.9 mm X 10.0 mm<br>Ø2.9 mm X 11.5 mm<br>Ø2.9 mm X 13.0 mm | Sizes are within<br>range of primary<br>predicate device |
| Surface<br>treatment | Machined /<br>Microstructured blasted<br>etched surface passivated | Machined | Microstrucured, acid-etched<br>Surface | Microstructured blasted<br>etched surface, passivated | Sand blasting and acid<br>etching | Identical to K191751<br>and K182313 |
| Sterilization | Beta Radiation | Gamma Radiation | Gamma Radiation | Beta Radiation | Radiation sterile | Identical to K182313 |
{13}------------------------------------------------
| Company | Candidate:<br>Medical Instinct Deutschland<br>GmbH | Reference Device:<br>Nobel Biocare AB | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutschland<br>GmbH | Result |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | BoneTrust® Mini Esthetic Abut-<br>ments, Crown Base Abutments | Temporary Snap Abutment | Sterngold 2.2mm Angled Micro<br>ERA Dental Implant System | BoneTrust® Esthetic Abutments | -- |
| Code | NHA | NHA | DZE | NHA | -- |
| 510(K) number | -- | K161435 | K092434 | K182313 | -- |
| Indications | BoneTrust® Mini Esthetic abutments<br>and BoneTrust® Mini Crown Base<br>Abutments are intended for use with<br>BoneTrust® Mini / Mini+ Dental<br>Implants to provide support for<br>provisional prosthetic reconstruc-<br>tions during the healing phase of<br>permanent endosseous dental<br>implants. Use of BoneTrust® Mini<br>Esthetic abutments and BoneTrust®<br>Mini Crown Base Abutments is not<br>to exceed one hundred and eighty<br>(180) days. | The Temporary Snap Abutment is<br>intended to be used to fabricate<br>and support provisional restora-<br>tions that aid in creating an esthetic<br>emergence through the gingiva<br>during the healing period and prior<br>to final restoration. The Temporary<br>Snap Abutment can be used for<br>cement retained or screw-retained<br>provisional restorations. The abut-<br>ments can be used for single-unit<br>and multi-unit restorations. Use of<br>the temporary Snap Abutment is<br>not to exceed one hundred and<br>eighty (180) days. | The Sterngold 2.2mm Angled ERA<br>dental implants are intended for<br>long term as well as temporary<br>surgical implantation in the bone of<br>the patient's upper or lower arch to<br>provide immediate load or delayed<br>load of prosthetic systems, such as<br>artificial teeth, in order to restore<br>the patient's chewing function.<br>Immediate loading of the ERA<br>Implant should only occur when the<br>position of the implants provides<br>adequate bone quantity and quality<br>to allow proper immediate mechan-<br>ical stabilization of the self-tapping<br>screw into the bone and where<br>occlusal and lateral forces can be<br>limited with appropriate occlusal<br>design and a soft diet. | Bone Trust® Abutments and Pros-<br>thetic parts are intended for use<br>with Bone Trust Dental Implants in<br>the maxillary and/or mandibular<br>arches to provide support for<br>crowns, bridges or overdenture for<br>edentulous or partially edentulous<br>patients. | Substantially<br>Equivalent to<br>K161435.<br>Indications of<br>K092434 include<br>temporary and<br>permanent use and<br>thus cover all indi-<br>cations claimed for<br>the proposed de-<br>vice. The more<br>restrictive indica-<br>tions of the pro-<br>posed device do<br>not raise any con-<br>cerns in regard to<br>safety and effec-<br>tiveness<br>Minor differences in<br>language do not<br>impact the safety<br>and effectiveness<br>of the subject de-<br>vice |
| Company | Candidate:<br>Medical Instinct Deutschland<br>GmbH | Reference Device:<br>Nobel Biocare AB | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutschland<br>GmbH | Result |
| Duration of use | Temporary use | Temporary use | Temporary use / Permanent use | Permanent use | Identical to<br>K161435<br>Indications of<br>K092434 include<br>temporary and<br>permanent use and<br>thus cover all indi-<br>cations claimed for<br>the proposed de-<br>vice. The more<br>restrictive indica-<br>tions of the pro-<br>posed device do<br>not raise any con-<br>cerns in regard to<br>safety and effec-<br>tiveness |
| Material | Titanium alloy Ti6AI4V (ASTM F136) | Abutments and screws -<br>Titanium vanadium alloy<br>(ASTM F1472, ASTM F136) | Wrought Titanium<br>6AL-4V ELI | Titanium alloy Ti6AI4V<br>(ASTM F136) | Identical |
| Surface | Machined | Machined | Machined | Machined | Identical |
| Size | Platform Diameter: 2.7mm<br>Body Diameter: 3.9mm, 4.2mm<br>Gingival Height: 1.0mm | Platform Diameter: Narrow Plat-<br>form (NP), Regular Platform (RP),<br>Wide Platform (WP)<br>Body Diameter: 4.0, 4.5, 6.0mm<br>Collar height: 1.5mm, 3mm | Platform Diameter: 2.2mm (Implant<br>thread major diameter)<br>Cuff Height: 0,76 – 4mm | Platform Diameter: ø 3.4 - 4.0 -<br>5.0mm<br>Body Diameter: 4.5mm<br>Gingival Height: 0.5 - 0.7 - 2.5 -<br>4.5mm | Reference device<br>K092434 has been<br>used to address<br>any dimensions not<br>cleared in the<br>reference Device<br>K161435 submis-<br>sion. |
| Company | Candidate:<br>Medical Instinct Deutschland<br>GmbH | Reference Device:<br>Nobel Biocare AB | Reference Device:<br>Sterngold | Reference Device:<br>Medical Instinct Deutschland<br>GmbH | Result…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.