K191751 · Hiossen, Inc. · DZE · Dec 16, 2019 · Dental
Device Facts
Record ID
K191751
Device Name
EM Provisional
Applicant
Hiossen, Inc.
Product Code
DZE · Dental
Decision Date
Dec 16, 2019
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).
Device Story
EM Provisional is a one-piece, screw-form endosseous dental implant; manufactured from titanium alloy (Ti-6Al-4V). Device serves as temporary support for provisional prosthetic devices during healing phase of permanent implants. Surgically placed in upper or lower jaw arches by dental professionals. Machined surface finish. Supplied sterile via radiation. Device provides temporary stabilization; allows immediate loading in edentulous sites. Clinical benefit includes maintaining prosthetic function during osseointegration of permanent implants.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by non-clinical performance data, including biocompatibility, sterilization validation, surface characterization, fatigue testing, compressive strength, and pull-out testing, all of which rely on the established performance of the predicate device.
Indicated for patients with partially or fully edentulous mandibles and maxilla requiring temporary support for provisional prosthetic devices during the healing phase of permanent endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 16, 2019
Hiossen, Inc. Peter Lee QA/RA Manager 85 Ben Fairless Drive Fairless Hills, Pennsylvania 19030
Re: K191751
Trade/Device Name: EM Provisional Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: September 16, 2019 Received: September 17, 2019
Dear Peter Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for
Srinivas Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a stylized font, with the "H" in red and the rest of the letters in gray. There is a red swoosh above the "H". Below the word "Hiossen" is the word "Implant" in smaller, gray letters.
85 Ben Fairless Drive airless Hills, PA 19030 888-768-0001 www.hiossen.com
# Section 5 Indication for Use Statement - 1 PAGE
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
Device Name EM Provisional
Indications for Use (Describe)
The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).
Type of use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
[ Over-The-Counter Use (21 CFR 807 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Service (301) 443-6740 EF
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters. The word "Implant" is written in gray letters below the word "Hiossen."
5 Ben Fairless Drive Hills, PA 19030 www.hiossen.com
# Section 6 510(k) Summary - 4 PAGES
| (1) Submitter Information: | |
|----------------------------|--|
|----------------------------|--|
| Submitted by: | Hiossen, Inc.<br>85 Ben Fairless Drive<br>Fairless Hills, PA 19030 |
|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Contact Person:<br>Telephone Number:<br>Fax Number: | Peter Lee<br>267-759-7031<br>267-759-7031 |
| Date Prepared: | November 8, 2019 |
| (2) Device Name:<br>· Proprietary Name:<br>• Classification Name:<br>• CFR Number:<br>• Device Class:<br>• Product Code: | EM Provisional<br>Implant, Endosseous, Root-form<br>872.3640<br>Class II<br>DZE |
(3) Predicate Device:
| Device | 510(k) | Manufacturer |
|-------------------------|---------|------------------|
| MS System (Provisional) | K072871 | OSSTEM Co., Ltd. |
| Hiossen Implant System* | K140934 | Hiossen, Inc. |
*Reference device
# (4) Description of Device:
The EM Provisional is a dental implant made of titanium alloy (Ti-6Al-4V) and is supplied sterile. The surface is machined finished and is intended to be surgically placed in the bone of the upper or lower jaw arches.
The EM Provisional is similar to other commercially available products based on the intended use, the technology used, the claims, the material composition employed and performance characteristics. It is substantially equivalent in design, function and intended use to the MS System (Provisional), OSSTEM Implant Co., Ltd, K072871
| Device | Diameter(mm) | Thread Length(mm) |
|----------------|--------------|-------------------|
| EM Provisional | Ø1.8, Ø2.5 | 10.0, 13.0, 15.0 |
# (5) Indication For Use:
The EM Provisional is intended to be loaded immediately in partially or fully edentulous mandibles and maxilla to serve as temporary support for provisional prosthetic device during the healing phase of a permanent endosseous dental implant(s).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters. The "H" and the top part of the "i" are red, while the rest of the letters are gray. Below the word "Hiossen" is the word "Implant" in smaller, gray letters.
85 Ben Fairless Drive Fairless Hills, PA 19030 888-768-0001 www.hiossen.com
# (6) Substantial Equivalence:
The information and date provided in this submission established the EM Provisional is substantially equivalent to the primary predicate device, OSSTEM MS System (Provisional) in (510(k) number K072871).
| | Proposed Device<br>EM Provisional | Predicate Device<br>MS System (Provisional) | Reference Device<br>HIOSSEN IMPLANT<br>SYSTEM |
|---------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | EM Provisional | MS System (Provisional) | HIOSSEN IMPLANT<br>SYSTEM |
| Manufacturer | Hiossen, Inc. | Osstem Co., Ltd. | Hiossen, Inc. |
| 510(K) No. | New device | K072871 | K140934 |
| Design | Image: EM Provisional Design | Image: MS System (Provisional) Design | Image: HIOSSEN IMPLANT SYSTEM Design |
| Intended use | The EM Provisional is<br>intended to be loaded<br>immediately in partially or<br>fully edentulous mandibles<br>and maxilla to serve as<br>temporary support for<br>provisional prosthetic<br>device during the healing<br>phase of permanent<br>endosseous dental<br>implant(s). | The MS system<br>(Provisional) is intended to<br>be loaded immediately in<br>partially or fully edentulous<br>mandibles and maxilla to<br>serve as temporary<br>support for provisional<br>prosthetic device during<br>the healing phase of<br>permanent endosseous<br>dental implants. | The HIOSSEN Implant<br>System is indicated for<br>use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including;<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and final or<br>temporary abutment<br>support for fixed<br>bridgework. It is intended<br>for delayed loading. Ultra<br>wide Fixture System is<br>intended to be used in the<br>molar region. |
| Structure | • Screw form<br>• One piece fixture and<br>abutment | • Screw form<br>• One-piece of Fixture and<br>Abutment | • Internal Hex-connected<br>• Submerged Fixture<br>• Tapered body shape &<br>Straight body shape |
| Diameters | 1.8, 2.5 | 1.8, 2.5 | 3.5~6.8 |
| Lengths | 10.0, 13.0, 15.0 | 10.0, 13.0, 15.0 | 6.2~18.2 |
| Material | Titanium alloy Ti-6Al-4V<br>(ASTM F 136) | Titanium alloy Ti-6Al-4V<br>(ASTM F 136) | Pure Titanium Grade 4<br>(ASTM F67) |
| Surface | Machined finished | Machined finished | - SA(Sandblasted and<br>Acid etched) |
| Sterilization | Radiation Sterile | Radiation Sterile | Radiation Sterile |
| Packaging | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>tamper-evident outer<br>package. | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>outer box | Secured in plastic ampule,<br>housed in Tyvek-lidded<br>blister tray, placed in a<br>tamper-evident outer<br>package. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Hiossen Implant. The logo is composed of the letters "HI" in red, followed by "OSSEN" in gray. The word "IMPLANT" is written in smaller gray letters below the word "OSSEN". There is a registered trademark symbol next to the word "OSSEN".
| S.E. | The proposed device is substantially equivalent with respect to the intended use, structure, material, surface, sterilization and design with the legally marketed predicate device as well as the sterilization and packaging of the legally marketed reference device. The proposed device does not pose any new or increased risks as compared to the legally marketed predicate and reference devices. |
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
### Substantial Equivalence Conclusion
In accordance with the Federal Food Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, HIOSSEN, INC. concludes since the EM Provisional has the same design, intended use, structure, diameters, lengths, material, surface, sterilization as the predicate device, MS System (Provisional) (K072871) and same sterilization and packaging to the reference Hiossen Implant System (K140934) it is substantially equivalent. The propose device does not pose any new or increased risk as compared to both the legally marketed predicate and reference devices.
#### Non-Clinical Performance Data
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include data from the following tests:
#### Biocompatibility Testing
The EM Provisional is manufactured using the same manufacturing process and same well known and well established material as the predicate device, the MS System (Provisional), K072871, therefore we reason it was not necessary to re-test biocompatibility in order to support the biological safety of the EM Provisional.
#### Sterilization Validation
The EM Provisional (manufactured using the same manufacturing process, material and utilizes the same sterilize barrier system) is gamma irradiated under the same conditions and process as the predicate device, the MS System (Provisional), K072871, therefore we reason it was not necessary to re-test validation in order to support the sterilization validity of the EM Provisional.
#### Surface Treatment Characterization Testing
The EM Provisional surface is manufactured using the same manufacturing process, material and is its surface is machined finished just as the predicate device, the MS System (Provisional), K072871, therefore no additional character testing was necessary to support the equivalency of the EM Provisional.
#### Fatique Testing
The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test fatigue on the proposed device since both are the same device.
#### Compressive Strength Testing
The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test compressive strength on the proposed device since both are the same device.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Hiossen Implant. The word "Hiossen" is written in a combination of red and gray letters, with the "H" and the top part of the "i" in red, and the rest of the letters in gray. Below the word "Hiossen" is the word "IMPLANT" in gray, with a large amount of space between each letter.
85 Ben Fairless Drive Fairless Hills, PA 19030 888-768-0001 www.hiossen.com
Pull-out Testing The EM Provisional is manufactured using the same manufacturing process, material and same design, as the predicate MS System (Provisional), K072871. We reason based on the least burdensome principle that it was not necessary to re-test pull-out on the proposed device since both are the same device.
Clinical Performance Testing
No clinical performance report(s) is being submitted.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.