CARESCAPE ONE and CARESCAPE DOCK FO, CARESCAPE ECG, CARESCAPE TEMP, CARESCAPE PRES
K200494 · GE Healthcare · MHX · Mar 29, 2020 · Cardiovascular
Device Facts
Record ID
K200494
Device Name
CARESCAPE ONE and CARESCAPE DOCK FO, CARESCAPE ECG, CARESCAPE TEMP, CARESCAPE PRES
Applicant
GE Healthcare
Product Code
MHX · Cardiovascular
Decision Date
Mar 29, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility. The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters. The CARESCAPE ONE can be used as a standalone monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring. The CARESCAPE ONE can be connected as an accessory to a compatible CARESCAPE monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition. Visual and audible alarms, user controls, and user interface on the CARESCAPE ONE are not active in this mode. The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.
Device Story
Multi-parameter physiological patient monitor; used in professional healthcare facilities and intra-hospital transport. Operates as standalone monitor or acquisition module for compatible CARESCAPE host monitors. Inputs: ECG, invasive/non-invasive pressure, SpO2, temperature, impedance respiration, CO2 via plug-and-play Active Cable Modules (ACM). Transforms inputs into waveforms, vital sign values, and alarms. Standalone mode: provides display, alarms, and user interface. Accessory mode: transmits data to host monitor; local UI/alarms disabled. Operated by licensed healthcare practitioners or trained personnel. Benefits: flexible bedside/transport monitoring; enables continuous patient data acquisition across care settings.
Clinical Evidence
No clinical studies required for substantial equivalence. Bench testing confirmed design specifications, software, hardware, and electromagnetic compatibility. Clinical report provided for GE TruSignal SpO2 sensors in neonates to support existing performance data; no changes to pulse oximetry algorithm or hardware compared to predicate.
Technological Characteristics
Multi-parameter monitor with 7-inch TFT LCD. Uses Active Cable Modules (ACM) for parameter acquisition via custom USB-style connectors. Internal battery and AC power. Linux OS. Complies with IEC 60601-1, IEC 60601-2-27, IEC 80601-2-30:2018, IEC 80601-2-49:2018, ISO 80601-2-55:2018, ISO 80601-2-56:2017+A1:2018, ISO 80601-2-61:2017, and IEC 62133-2:2017. IP44 water ingress rating.
Indications for Use
Indicated for monitoring hemodynamic (ECG, ST segment, arrhythmia, invasive/non-invasive BP, SpO2, pulse rate, temperature) and respiratory (impedance respiration, CO2) parameters in adult, pediatric, and neonatal patients. Not for use in controlled MR environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
March 29, 2020
GE Healthcare Joel Kent Senior Regulatory Affairs Manager 8200 West Tower Avenue Milwaukee, Wisconsin 53223
Re: K200494
Trade/Device Name: CARESCAPE ONE Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, BZQ, CCK, DPS, DPZ, DQA, DQK, DRT, DSI, DSI, DSK, DXN, FLL, MLD Dated: February 25, 2020 Received: February 28, 2020
Dear Joel Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200494
Device Name CARESCAPE ONE
#### Indications for Use (Describe)
The CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.
The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.
The CARESCAPE ONE can be used as a standalone monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arthythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.
The CARESCAPE ONE can be connected as an accessory to a compatible CARESCAPE monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition. Visual and audible alarms, user controls, and user interface on the CARESCAPE ONE are not active in this mode.
The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time.
The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.
Contraindications for using CARESCAPE ONE: The CARESCAPE ONE is not intended for use within a controlled MR environment.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a circular border. The border has decorative flourishes resembling water droplets or stylized leaves at the top, bottom, and sides of the circle.
GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223
## 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
### Owner/Contact/Date (807.92(a)(1):
| Date: | March 10, 2020 |
|---------------------------|------------------------------------------------------------------------------------------------------------------|
| Owner/Submitter: | GE Medical Systems Information Technologies, Inc. 8200 West Tower Avenue Milwaukee, Wisconsin 53223 |
| Primary Contact Person: | Joel Kent Senior Regulatory Affairs Manager GE Healthcare Phone: 617-851-0943 E-mail: joel.kent@ge.com |
| Secondary Contact Person: | Monica Morrison Regulatory Affairs Director GE Healthcare Phone: 608-515-3077 E-mail: monica.morrison@ge.com |
Device names (807.92(a)(2)):
Trade Name: CARESCAPE ONE Common/Usual Name: Multiparameter patient monitor (monitor, physiological, patient (with arrhythmia detection or alarms)
{4}------------------------------------------------
Classification Names: 21 CFR 870.1025 Arrhythmia detector and alarm (including ST-
segment measurement and alarm)
21 CFR 868.2375 monitor, breathing frequency
21 CFR 868.1400 Carbon dioxide gas analyzer
21 CFR 870.2340 Electrocardiograph
21 CFR 870.2710 oximeter, ear
21 CFR 870.2700 oximeter
21 CFR 870.1425 Programmable diagnostic computer
21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm)
21 CFR 870.1025 detector and alarm, arrhythmia
21 CFR 870.1100 alarm, blood-pressure
21 CFR 870.1110 computers, blood-pressure
21 CFR 870.1130 system, measurement, blood-pressure, noninvasive
21 CFR 880.2910 Clinical electronic thermometer
21 CFR 870.1025 monitor, St Segment with alarm
Product Code: MHX
| Subsequent Product Codes: |
|---------------------------|
| BZQ |
| CCK |
| DPS |
| DPZ |
| DQA |
| DQK |
| DRT |
| DSI |
| DSJ |
| DSK |
| DXN |
| FLL |
| MLD |
{5}------------------------------------------------
| Predicate Device(s)<br>(807.92(a)(3)): | The primary predicate for this submission is K190008,<br>CARESCAPE ONE<br><br>Additional predicates/reference devices:<br>K071073, Patient Data Module (PDM)<br>K191249, CARESCAPE B450<br>K191149, CARESCAPE B650<br>K191323, CARESCAPE B850 |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Description<br>(807.92(a)(4)): | CARESCAPE ONE, with CARESCAPE Software version 3<br>belongs to the CARESCAPE V3 patient monitor family. The<br>concept of the CARESCAPE ONE is to provide a flexible<br>bedside monitor that can also be used during intra-hospital<br>transport. The flexibility of the CARESCAPE ONE allows the<br>user to configure the monitor's vital sign acquisition for only th<br>parameters they require. This is achieved using plug and play<br>Active Cable Modules (ACM) that connect via medical grade<br>USB ports on the CARESCAPE ONE monitor. Note that the<br>USB ports are not compatible with commercial USB items on<br>the market due to a custom connector design. Each ACM is<br>dedicated to measuring a particular vital sign, currently we have<br>ECG/Respiration, Invasive Blood Pressure, Temperature, SpO2,<br>or CO2. The only exception is the Non-Invasive Blood Pressure<br>(NIBP) measurement which does not require a separate ACM<br>since the capability to measure NIBP is built-in to the<br>CARESCAPE ONE monitor itself. CARESCAPE ONE<br>provides the users the acquired display values, waveforms,<br>alarms and status messages in compact footprint monitor that<br>runs on an internal battery as well as AC power when connected<br>to the docking station.<br><br>When connected to a compatible host monitor, CARESCAPE<br>ONE operates as an acquisition device. In this mode, the<br>CARESCAPE ONE screen and user interface is effectively<br>disabled and it transmits data received from the Active Cable<br>Modules to the host monitor, which is responsible for managing<br>clinical configuration settings, and displaying values,<br>waveforms, alarms, and status messages. |
CARESCAPE ONE is compatible with the following 510(k) cleared host monitors:
- CARESCAPE B450 (K191249) •
- CARESCAPE B650 (K191149)
- CARESCAPE B850 (K191323)
{6}------------------------------------------------
Intended Use: (807.92(a)(5)):
## Indications
The CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.
The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.
The CARESCAPE ONE can be used as a standalone monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, noninvasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.
The CARESCAPE ONE can be connected as an accessory to a compatible CARESCAPE monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition. Visual and audible alarms, user controls, and user interface on the CARESCAPE ONE are not active in this mode.
The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time.
The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility.
Contraindications for using CARESCAPE ONE: The CARESCAPE ONE is not intended for use within a controlled MR environment.
{7}------------------------------------------------
The CARESCAPE ONE that is the subject of this submission is Technology (807.92(a)(6)): based on the primary predicate CARESCAPE ONE (K190008).
> The main differences with respect to the predicate discussed throughout the submission are summarized below:
- Add the capability for CARESCAPE ONE to be utilized as an acquisition device when connected with host monitors (CARECSAPE Monitors). CARESCAPE ONE can be used as an acquisition module for compatible host devices, in addition to use as a stand-alone monitor. There are no changes to the device itself specifically to enable compatibility with the host monitors, as it was already built into the device design. The labeling and instructions have been updated to reflect the use with compatible host devices. Compatibility with CARESCAPE ONE was already cleared in the additional predicate devices CARESCAPE B650 K191149, CARESCAPE B450 K191249 and CARESCAPE B850 K191323. The acquisition module functionality in CARESCAPE ONE, as discussed in K190008, is substantially equivalent to the Patient Data Module (K071073). - Minor software updates for maintenance/bug fixes only. No changes to function, operation, parameters monitored, or algorithms.
- Compliance demonstrated with newer versions of FDA recognized standards. No actual changes to the device were required to meet the updated standards, other than one update to the labeling for the water ingress protection level (IP44). - Minor updates to the compatible accessories list.
The CARESCAPE ONE remains substantially equivalent to the predicates and the device itself (software and hardware) remains nearly identical to the version cleared under K190008. There are no changes to the monitored parameters or algorithms.
A summary of the main changes compared to the predicate are listed below in the comparison table.
{8}------------------------------------------------
| Product Comparison versus Predicate Main features | | | |
|---------------------------------------------------|--|--|--|
| | | | |
| Specification | PREDICATE CARESCAPE ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The CARESCAPE ONE is a multi-parameter physiological patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.<br><br>The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.<br><br>The CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.<br><br>The CARESCAPE ONE is indicated for use on adult, pediatric, and neonatal patients and on one patient at a time.<br><br>The CARESCAPE ONE is indicated for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in the proper use of the equipment in a professional healthcare facility. | The CARESCAPE ONE is both a multi-parameter physiological patient monitor and an accessory to a multi-parameter patient monitor intended for use in multiple areas and intra-hospital transport within a professional healthcare facility.<br><br>The CARESCAPE ONE is indicated for the monitoring of hemodynamic (including ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, and temperature), and respiratory (impedance respiration and CO2 airway gas) physiological parameters.<br><br>The CARESCAPE ONE can be used as a standalone monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition and monitoring.<br><br>The CARESCAPE ONE can be connected as an accessory to a compatible CARESCAPE monitor. In this mode of operation, the CARESCAPE ONE provides ECG, ST segment, arrhythmia detection, invasive pressure, non-invasive blood pressure, SpO2, pulse rate, temperature, impedance respiration, and CO2 airway gas parameter acquisition. Visual and audible alarms, user controls, and user interface on the CARESCAPE ONE are not active in this mode. | Equivalent.<br><br>The predicate device is standalone Patient Monitor, which includes patient signal acquisition and processing, patient data display, alarm signal generation, and user input functions.<br><br>The new device includes the standalone functionality and behavior, plus includes the ability to delegate alarm signal generation, patient data display, and user input functions to a connected "host" CARESCAPE Patient Monitor. The Patient Data Module (K071073) is a reference device which is similar to the CARESCAPE ONE used as an acquisition module since it connects to a CARESCAPE Monitor host and can also be used in patient transport and then reconnected to a host.<br><br>The changes have been made to clarify that the CARESCAPE ONE can be used both as a standalone monitor and a compatible module/accessory for a host CARESCAPE Monitor.<br><br>Verification of interoperability/ compatibility was completed. Compatibility of CARESCAPE ONE with the host monitors was also cleared in the following submissions: CARESCAPE B650 K191149, CARESCAPE B450 K191249 and CARESCAPE B850 K191323.<br><br>This change is substantially equivalent to the predicate device. |
| Contraindications for using CARESCAPE ONE: | The CARESCAPE ONE is not intended for use within a controlled MR environment. | The CARESCAPE ONE is | |
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
| | | The CARESCAPE ONE is<br>indicated for use under the<br>direct supervision of a<br>licensed healthcare<br>practitioner, or by personnel<br>trained in the proper use of the<br>equipment in a professional<br>healthcare facility.<br><br>Contraindications for using<br>CARESCAPE ONE:<br>The CARESCAPE ONE is<br>not intended for use within a<br>controlled MR environment. | |
| Patient type | Adult, pediatric & neonatal | Adult, pediatric & neonatal | Identical |
| Use<br>environments | Within a professional<br>healthcare facility (Not<br>intended for MRI) | Within a professional<br>healthcare facility (Not<br>intended for MRI) | Identical |
| Intrahospital<br>transport within<br>a professional<br>healthcare<br>facility. | Yes | Yes | Identical |
| Acquisition<br>Mode<br>(accessory to a<br>compatible host<br>monitor) | N/A | Connects to a compatible<br>"host" CARESCAPE Patient<br>Monitor.<br><br>The CARESCAPE ONE<br>provides parameter<br>acquisition for the host patient<br>Monitor. Visual and audible<br>alarms, user controls, and user<br>interface functions are not<br>active on the CARESCAPE<br>ONE and instead those<br>functions are provided by the<br>host CARESCAPE Patient<br>Monitor. | Equivalent.<br><br>The predicate device is standalone<br>Patient Monitor, which includes<br>patient signal acquisition and<br>processing, patient data display, alarm<br>signal generation, and user input<br>functions.<br><br>The new device includes the<br>standalone functionality and behavior,<br>plus includes the ability to act as an<br>acquisition device to a host monitor.<br>This is equivalent to the reference<br>device Patient Data Module (K071073)<br>which acts as an acquisition module to<br>a host device and was a predicate to<br>K190008. This acquisition mode<br>allows it to delegate alarm signal<br>generation, patient data display, and<br>user input functions to a connected<br>"host" CARESCAPE Patient Monitor.<br>The CARESCAPE ONE can be used<br>both as a standalone and a compatible<br>module for a host CARESCAPE<br>Monitor. Verification of<br>interoperability/compatibility was<br>completed. Compatibility of<br>CARESCAPE ONE with the host<br>monitors was also cleared in the<br>following submissions: CARESCAPE<br>B650 K191149, CARESCAPE B450<br>K191249 and CARESCAPE B850<br>K191323. This change is substantially<br>equivalent to the predicate device. |
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
| Monitored<br>Parameters | Parameters monitored by<br>CARESCAPE ONE include:<br>- hemodynamic (ECG, ST<br>segment, arrhythmia detection,<br>invasive pressures, NIBP,<br>temperature, and pulse<br>oximetry)<br>- respiratory (impedance<br>respiration, CO2) | Parameters monitored by<br>CARESCAPE ONE include:<br>- hemodynamic (ECG, ST<br>segment, arrhythmia<br>detection, invasive pressures,<br>NIBP, temperature, and pulse<br>oximetry)<br>- respiratory (impedance<br>respiration, CO2) | Identical |
| Medical<br>Standards | IEC 60601-1:2005 + C1:2006<br>+ C2:2007 + A1:2012<br>IEC 60601-1-2:2007<br>IEC 60601-1-2:2014-02 / EN<br>60601-1-2:2015<br>IEC 60601-1-6:2010 +<br>A1:2013<br>IEC 60601-1-8:2006 +<br>A1:2012<br>IEC 60601-2-27:2011<br>IEC 80601-2-30:2013<br>IEC 60601-2-34:2011<br>IEC 60601-2-49:2011<br>ISO 80601-2-55:2011<br>ISO 80601-2-56:2009<br>ISO 80601-2-61:2011<br>IEC 62304:2006 + A1:2015<br>IEC 62366:2017 + A1:2014<br>ISO 10993-1:2009<br>IEC 62133:2012 | IEC 60601-1:2005 + C1:2006<br>+ C2:2007 + A1:2012<br>IEC 60601-1-2:2007<br>IEC 60601-1-2:2014-02 / EN<br>60601-1-2:2015<br>IEC 60601-1-6:2010 +<br>A1:2013<br>IEC 60601-1-8:2006 +<br>A1:2012<br>IEC 60601-2-27:2011<br>IEC 80601-2-30:2018<br>IEC 60601-2-34:2011<br>IEC 80601-2-49:2018<br>ISO 80601-2-55:2018<br>ISO 80601-2-56:2017 +<br>A1:2018<br>ISO 80601-2-61:2017<br>IEC 62304:2006 + A1:2015<br>IEC 62366:2017 + A1:2014<br>ISO 10993-1:2009<br>IEC 62133-2:2017 | Equivalent.<br>Both the predicate and the new device<br>comply with the same standards,<br>except the new device complies with<br>the more recent edition of the<br>particulars for the following standards:<br>IEC 80601-2-30:2018<br>IEC 80601-2-49:2018<br>ISO 80601-2-55:2018<br>ISO 80601-2-56:2017+A1 2018<br>ISO 80601-2-61:2017<br>The battery is identical to the<br>predicate, but we now comply with the<br>most recent battery standard IEC<br>62133-2:2017.<br>Verification testing to show<br>compliance with the updated standards<br>are included in the submission. In<br>order to meet the new standard<br>requirements we did not require any<br>software or hardware changes<br>compared to the predicate<br>CARESCAPE ONE (K190008). We<br>have only changed our manuals to list<br>the new standards and our IP41 water<br>ingress specification and label has<br>changed to a higher rating to comply<br>with IEC 60601-2-61:2017. The new<br>pulse oximetry standard requires IPx2<br>instead of IPX1, but we tested to IP44 |
| Specification | PREDICATE CARESCAPE<br>ONE (K190008) | CARESCAPE ONE | Discussion of Differences |
| Parameters<br>Acquisition<br>Method | The CARESCAPE ONE<br>utilizes the Active Cable<br>Modules (ACMs) or<br>PARAMETERS, in which the<br>parameter electronics are<br>encapsulated into the<br>respective patient cables,<br>rather than inside the main<br>frame of the monitor. Only<br>the NIBP parameter is<br>integrated into the<br>CARESCAPE ONE monitor.<br><br>Parameter/Active Cable<br>Modules:<br>CARESCAPE TEMP -<br>Temperature<br>CARESCAPE PRES –<br>Invasive Pressure<br>CARESCAPE ECG - ECG<br>CARESCAPE SPO2 -<br>Trusignal SPO2<br>CARESCAPE SPO2 Nellcor -<br>Nellcor SPO2<br>CARESCAPE SPO2 Masimo -<br>Masimo SPO2<br>CARESCAPE CO2 - LoFlo | The CARESCAPE ONE<br>utilizes the Active Cable<br>Modules (ACMs) or<br>PARAMETERS, in which the<br>parameter electronics are<br>encapsulated into the<br>respective patient cables,<br>rather than inside the main<br>frame of the monitor. Only<br>the NIBP parameter is<br>integrated into the<br>CARESCAPE ONE monitor.<br><br>Parameter/Active Cable<br>Modules:<br>CARESCAPE TEMP -<br>Temperature<br>CARESCAPE PRES –<br>Invasive Pressure<br>CARESCAPE ECG - ECG<br>CARESCAPE SPO2 -<br>Trusignal SPO2<br>CARESCAPE SPO2 Nellcor -<br>Nellcor SPO2<br>CARESCAPE SPO2 Masimo -<br>Masimo SPO2<br>CARESCAPE CO2 - LoFlo | Identical |
| EK-Pro<br>arrhythmia<br>detection<br>algorithm | EK-Pro V14 | EK-Pro V14 | Identical |
| Size (H x W x<br>D) | 155 mm x 270 mm x 65 mm<br>(6.1 in x 10.6 in x 2.6 in)<br><br>Note: Excludes dock | 155 mm x 270 mm x 65 mm<br>(6.1 in x 10.6 in x 2.6 in)<br><br>Note: Excludes dock | Identical…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.