K191149 · GE Healthcare Finland Oy · MHX · Jan 21, 2020 · Cardiovascular
Device Facts
Record ID
K191149
Device Name
CARESCAPE B650
Applicant
GE Healthcare Finland Oy
Product Code
MHX · Cardiovascular
Decision Date
Jan 21, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CARESCAPE B650 is a multi-parameter patient monitor intended for use in multiple areas and intra hospital transport within a professional healthcare facility. The CARESCAPE B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B650 is indicated for monitoring of: · hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation), · respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and · neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission). The CARESCAPE B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network. The CARESCAPE B650 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B650 also shows alarms from other ECG sources. The CARESCAPE B650 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices. The CARESCAPE B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Device Story
Multi-parameter patient monitor for professional healthcare facilities; monitors hemodynamic, respiratory, and neurophysiological signals via connected acquisition modules (PDM, CARESCAPE ONE, E-Modules). Transforms physiological inputs into real-time waveforms, numeric trends, and alarm notifications; utilizes EK-Pro V14 algorithm for arrhythmia detection. Operated by clinicians; provides visual/audible alerts and data for clinical decision-making. Supports stand-alone use or network integration for remote viewing and data management. Benefits include continuous patient status oversight and alarm management to reduce fatigue. Updated hardware (ARM Cortex-A9) and software platform (CSP V3) compared to predicate; includes USB support for remote alarm devices (CARESCAPE RAD).
Clinical Evidence
No clinical data required; substantial equivalence supported by bench testing, including software verification/validation, hardware performance, electromagnetic compatibility, electrical safety, and usability testing.
Technological Characteristics
15-inch TFT LCD display; ARM Cortex-A9 processor; rechargeable Li-ion battery. Supports wired (10/100baseT) and wireless (802.11abgn) networking with WPA2-Enterprise and IEEE 802.11r fast roaming. Modular design with E-module slots and ePort interface. Software-based arrhythmia detection (EK-Pro V14). Complies with IEC 60601-1-8 for alarm defaults.
Indications for Use
Indicated for monitoring adult, pediatric, and neonatal patients in professional healthcare facilities, including intra-hospital transport. Monitors hemodynamic, respiratory, and neurophysiological status. Contraindicated for use in controlled MR environments.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
January 28, 2020
GE Healthcare Finland Oy Joel Kent Senior Regulatory Affairs Manager Kuortaneenkatu 2 00510 Helsinki Finland
Re: K191149
Trade/Device Name: CARESCAPE B650 Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX, BZK, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS, DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL, GWJ, GWQ, KOI, KRB, MLD, NHO, NHP, NHQ, OLT, OLW, OMC, ORT Dated: December 18, 2019 Received: December 20, 2019
Dear Joel Kent:
This letter corrects our substantially equivalent letter of January 21, 2020.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Jennifer W. Shih -S
Jennifer Shih Assistant Director (Acting) Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K191149
Device Name CARESCAPE B650
#### Indications for Use (Describe)
The CARESCAPE B650 is a multi-parameter patient monitor intended for use in multiple areas and intra hospital transport within a professional healthcare facility.
The CARESCAPE B650 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE B650 is indicated for monitoring of:
· hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation),
· respiratory (impedance respiration, airway gases (CO2, O2, N2O, and anesthetic agents), spirometry, gas exchange), and
· neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).
The CARESCAPE B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE B650 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B650 also shows alarms from other ECG sources.
The CARESCAPE B650 also provides other alarms, trends, snapshots and calculations, and can be connected to displays, printers and recording devices.
The CARESCAPE B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
Contraindications for using the monitor
The CARESCAPE B650 is not intended for use in a controlled MR environment.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo, which consists of the letters 'GE' intertwined within a circular border. The letters are stylized with curved lines, giving them a classic and elegant appearance. The logo is presented in black and white, with the letters and border in black against a white background.
GE Healthcare Finland Oy Kuortaneenkatu 2 00510 Helsinki
Finland
T: +358 10 39411
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
## Owner/Contact/Date (807.92(a)(1):
| Date: | January 21, 2020 |
|------------------------------|-------------------------------------------------------------------------------------------------------|
| Owner/Submitter: | GE Healthcare Finland Oy. |
| | Kuortaneenkatu 2 |
| | 00510 Helsinki |
| | FINLAND |
| | Phone: +358 10 39411 |
| Primary Contact Person: | Joel Kent |
| | Senior Regulatory Affairs Manager |
| | GE Healthcare |
| | Phone: 617-851-0943 |
| | E-mail: joel.kent@ge.com |
| Secondary Contact Person: | Anssi Ruokonen |
| | Regulatory Affairs Leader |
| | GE Healthcare Finland Oy |
| | Kuortaneenkatu 2 |
| | 00510 Helsinki |
| | Finland |
| | Phone: + 358 10 394 3686 |
| | E-mail: Anssi.ruokonen@ge.com |
| Device names (807.92(a)(2)): | |
| Trade Name: | CARESCAPE B650 |
| Common/Usual Name: | Multiparameter patient monitor (monitor, physiological, patient (with arrhythmia detection or alarms) |
{5}------------------------------------------------
Classification Names: 21 CFR 870.1025 Arrhythmia detector and alarm (including STsegment measurement and alarm) 21 CFR 868.1850 spirometer, monitoring (w/wo alarm) 21 CFR 868.1730 computer, oxygen-uptake 21 CFR 868.2375 monitor, breathing frequency 21 CFR 868.2600 monitor, airway pressure (includes gauge and/or alarm) 21 CFR 868.1500 analyzer, gas, enflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, desflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, sevoflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1500 analyzer, gas, isoflurane, gaseous-phase (anesthetic conc.) 21 CFR 868.1700 analyzer, gas, nitrous-oxide, gaseous phase (anesthetic conc.) 21 CFR 868.1620 analyzer, gas, halothane, gaseous-phase (anesthetic conc.) 21 CFR 868.1400 analyzer, gas, carbon-dioxide, gaseous-phase 21 CFR 868.1720 analyzer, gas, oxygen, gaseous-phase 21 CFR 870.2340 electrocardiograph 21 CFR 870.2710 oximeter, ear 21 CFR 870.2700 oximeter 21 CFR 870.1425 computer, diagnostic, programmable 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm) 21 CFR 870.1025 detector and alarm, arrhythmia 21 CFR 870.1100 alarm, blood-pressure 21 CFR 870.1110 computer, blood-pressure 21 CFR 870.1435 computer, diagnostic, pre-programmed, single-function 21 CFR 870.1130 system, measurement, blood-pressure, noninvasive 21 CFR 870.2910 thermometer, electronic, clinical 21 CFR 882.1900 stimulator, auditory, evoked response 21 CFR 882.1400 full-montage standard electroencephalograph 21 CFR 868.2775 stimulator, nerve, peripheral, electric 21 CFR 870.1915 probe, thermodilution 21 CFR 870.1025 monitor, st segment with alarm 21 CFR 882.1400 non-normalizing quantitative electroencephalograph software 21 CFR 882.1400 index-generating electroencephalograph software 21 CFR 882.1400 reduced- montage standard electroencephalograph
{6}------------------------------------------------
21 CFR 882.1400 burst suppression detection software for electroencephalograph
| Product Code: | MHX |
|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subsequent Product Codes: | BZK, BZL, BZQ, CAP, CBQ, CBR, CBS, CCK, CCL, DPS,<br>DPZ, DQA, DQK, DRT, DSI, DSJ, DSK, DXG, DXN, FLL,<br>GWJ, GWQ, KOI, KRB, MLD, NHO, NHP, NHQ, OLT, OLW,<br>OMC, ORT |
| Predicate Device(s)<br>(807.92(a)(3): | K131223 CARESCAPE Monitor B650 |
| Device Description<br>(807.92(a)(4)): | CARESCAPE B650 V3 is a new version of a portable multi-<br>parameter patient monitoring system. The CARESCAPE B650<br>V3 includes the monitor with built-in CPU, power unit, a 15 inch<br>touch display, the CARESCAPE Software version 3 and the<br>battery. CARESCAPE B650 V3 is equipped with so called ePort<br>interface that supports use of PDM or CARESCAPE ONE<br>patient data acquisition modules for patient monitoring.<br>CARESCAPE B650 V3 is also equipped with two module slots<br>where patient data acquisition modules (E-Modules), can be<br>connected to perform patient monitoring. The CARESCAPE<br>B650 V3 includes features and subsystems that are optional or<br>configurable. |
| Intended Use: (807.92(a)(5): | Indications (from labeling)<br>The CARESCAPE B650 is a multi-parameter patient monitor<br>intended for use in multiple areas and intra hospital transport<br>within a professional healthcare facility.<br>The CARESCAPE B650 is intended for use on adult, pediatric,<br>and neonatal patients and on one patient at a time.<br>The CARESCAPE B650 is indicated for monitoring of:<br>• hemodynamic (including ECG, ST segment, arrhythmia<br>detection, ECG diagnostic analysis and measurement, invasive<br>pressure, non-invasive blood pressure, pulse oximetry, cardiac<br>output (thermodilution and pulse contour), temperature, mixed<br>venous oxygen saturation, and central venous oxygen<br>saturation),<br>• respiratory (impedance respiration, airway gases (CO2, O2,<br>N2O, and anesthetic agents), spirometry, gas exchange), and<br>• neurophysiological status (including electroencephalography,<br>Entropy, Bispectral Index (BIS), and neuromuscular<br>transmission). |
{7}------------------------------------------------
The CARESCAPE B650 can be a stand-alone monitor or interfaced to other devices. It can also be connected to other monitors for remote viewing and to data management software devices via a network.
The CARESCAPE B650 is able to detect and generate alarms for ECG arrhythmias: atrial fibrillation, accelerated ventricular rhythm, asystole, bigeminy, bradycardia, ventricular couplet, irregular, missing beat, multifocal premature ventricular contractions (PVCs), pause, R on T, supra ventricular tachycardia, tachycardia, trigeminy, ventricular bradycardia, ventricular fibrillation/ventricular tachycardia, ventricular tachycardia, and VT>2. The CARESCAPE B650 also shows alarms from other ECG sources.
The CARESCAPE B650 also provides other alarms, trends, snapshots and events, and calculations, and can be connected to displays, printers and recording devices.
The CARESCAPE B650 is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in proper use of the equipment in a professional healthcare facility.
### Contraindications for using the monitor
The CARESCAPE B650 is not intended for use in a controlled MR environment.
Technology (807.92(a)(6)): CARESCAPE B650 with CARESCAPE Software version 3 incorporates updated hardware and a new software platform.
> The fundamental function and operation of the proposed CARESCAPE B650 V3 monitor are unchanged compared to CARESCAPE Monitor B650 with ESP V2 software (K131223). There are no new types of monitored parameters introduced compared to the predicate B650 monitor.
A summary of the main changes compared to the predicate are listed below in the comparison table.
{8}------------------------------------------------
| Specification | CARESCAPE Monitor B650<br>with ESP V2 software<br>(K131223) | CARESCAPE B650 with<br>CSP software version V3 | Differences | |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient type | Adult, pediatric & neonatal | Adult, pediatric & neonatal | Identical | |
| Use environments | Within a professional<br>healthcare facility (Not<br>intended for MRI) | Within a professional<br>healthcare facility (Not<br>intended for MRI) | Identical | |
| Size (H x W x D)<br>& Weight | 360 x 370 x 220 mm (14.2 x<br>14.6 x 8.7 in) and weight 9.2<br>kg (20.3 lbs) with battery but<br>without modules | 360 x 370 x 220 mm (14.2 x<br>14.6 x 8.7 in) and weight 9.2<br>kg (20.3lbs) with battery but<br>without modules. | Identical | |
| Processor | Freescale PowerPC MPC8270 | Freescale ARM Cortex-A9 | Equivalent<br>The CARESCAPE<br>Bx50 V3 monitors have<br>an updated common<br>CPU platform. | |
| Module Housing | Two optional E-module slots<br>and optional recorder. One<br>slide mount for acquisition<br>module | Two optional E-module slots<br>and optional recorder. One<br>slide mount for acquisition<br>module | Identical | |
| Display/screen | 15" Active matrix color TFT<br>LCD | 15" Active matrix color TFT<br>LCD | Identical | |
| Waveforms and<br>parameter<br>windows | Standard view: Up to 8<br>individual waveforms and up<br>to 20 parameter windows, if<br>horizontal parameter area<br>turned on. | Standard view: Up to 8<br>individual waveforms and up<br>to 20 parameter windows, if<br>horizontal parameter area<br>turned on. | Identical | |
| Modules | E-PSM; E-PSMP, E-PRESTN,<br>E-PRETN, E-RESTN, M-<br>PRESTN, M-PRETN, M-<br>RESTN, E-miniC, E-CO, E-<br>CAIO, E-CAIOV, E-<br>CAIOVX, E-COVX, E-COV,<br>E-P, E-PP, E-PT, E-COP, E-<br>COPSv, E-EEG, E-Entropy,<br>E-NMT, E-BIS, E-NSAT, E-<br>NSATX, E-Masimo, E-sCO,<br>E-sCAIO, E-sCOV, and E-<br>sCAIOV, E-PiCCO, PDM | E-BIS, E-COP, E-COPSv, E-<br>PiCCO, E-EEGX, E-Entropy,<br>E-Masimo, E-miniC, E-NMT,<br>E-NSATX, E-PP, E-PT, E-<br>sCAIO, E-sCAIOV, E-<br>sCAIOVX, E-sCO, E-sCOV,<br>E-sCOVX, PDM,<br>CARESCAPE ONE | Equivalent<br>Removed support for<br>several legacy E/M<br>multiparameter<br>hemodynamic<br>acquisition modules as<br>many newer acquisition<br>modules are included,<br>covering the same<br>parameters.<br>Added support for<br>CARESCAPE ONE and<br>E-EEGX acquisition<br>modules.<br>Added support for E-<br>sCAIOVX and E-<br>sCOVX modules | |
| Specification | CARESCAPE Monitor B650<br>with ESP V2 software<br>(K131223) | CARESCAPE B650 with<br>CSP software version V3 | Differences | |
| Available<br>parameters | ECG, ST segment, arrhythmia<br>detection, ECG diagnostic<br>analysis and measurement,<br>invasive pressure, non-<br>invasive blood pressure, pulse<br>oximetry, cardiac output<br>(thermodilution and pulse<br>contour), temperature, mixed<br>venous oxygen saturation, and<br>central venous oxygen<br>saturation, impedance<br>respiration, airway gases<br>(CO2, O2, N2O, and<br>anesthetic agents), spirometry,<br>gas exchange,<br>electroencephalography,<br>Entropy, Bispectral Index<br>(BIS), neuromuscular<br>transmission. | ECG, ST segment, arrhythmia<br>detection, ECG diagnostic<br>analysis and measurement,<br>invasive pressure, non-<br>invasive blood pressure, pulse<br>oximetry, cardiac output<br>(thermodilution and pulse<br>contour), temperature, mixed<br>venous oxygen saturation, and<br>central venous oxygen<br>saturation, impedance<br>respiration, airway gases<br>(CO2, O2, N2O, and<br>anesthetic agents), spirometry,<br>gas exchange,<br>electroencephalography,<br>Entropy, Bispectral Index<br>(BIS), neuromuscular<br>transmission. | Identical | |
| EK-Pro<br>arrhythmia<br>detection<br>algorithm | EK-Pro V13 | EK-Pro V14 | Equivalent<br>CARESCAPE B650 V3<br>uses an EK-ProV14<br>arrhythmia analysis<br>algorithm compared to<br>the EK-Pro V13 used in<br>the predicate monitors. | |
| Graphical user<br>interface | GE Healthcare Common User<br>Interface (CUI) Requirements<br>and Style Version 6 | GE Healthcare HDX | A new color scheme for<br>the display and slight<br>visual adjustments have<br>been incorporated, in<br>line with a GE<br>Healthcare wide user<br>interface design<br>guideline. The overall<br>user interface layout,<br>structure, operations,<br>and workflow remains<br>the same as the<br>predicate. | |
| Printing | Built-in or central and<br>networked laser printer<br>Printings for waveforms,<br>alarms waveforms, numeric<br>trends | Built-in or central and<br>networked laser printer<br>Printings for waveforms,<br>alarms waveforms, numeric<br>trends | Identical | |
| Specification | CARESCAPE Monitor B650<br>with ESP V2 software<br>(K131223) | CARESCAPE B650 with<br>CSP software version V3 | Differences | |
| Mounting options | Multiple GCX mounting<br>systems, Roll Stand, Quick<br>Mount | Multiple GCX mounting<br>systems, Roll Stand, Quick<br>Mount | Identical | |
| Alarms | Alarm management core<br>functionalities:<br>Classification and notification<br>of alarms<br>Adjustment of alarm settings<br>Alarm On/Off functionality<br>and audio silencing | Alarm management core<br>functionalities:<br>Classification and notification<br>of alarms<br>Adjustment of alarm settings<br>Possibility to set critical alarm<br>limits<br>Alarm On/Off functionality<br>and audio silencing | Equivalent<br>The predicate already<br>included the options to<br>configure some alarm<br>settings. In V3, more<br>flexibility was added for<br>the user to adjust alarm<br>priorities and alarm<br>criteria for additional<br>parameter alarms,<br>including additional<br>alarm delay options,<br>critical alarm options,<br>alarm acknowledgement<br>and latched alarm<br>indicators, and a pause<br>monitoring & central<br>function.<br>In general, options for<br>tailored/specific alarm<br>management schemes<br>have been added<br>supporting clinicians in<br>their goals of reducing<br>alarm fatigue while<br>maintaining safety.<br>Default setting are<br>according to IEC 60601-<br>1-8. | |
| Specification | CARESCAPE Monitor B650<br>with ESP V2 software<br>(K131223) | CARESCAPE B650 with<br>CSP software version V3 | Differences | |
| Remote Alarm<br>Device | Alarm Interface of ESP V2<br>SW was not utilized | Alarm Interface to the Remote<br>Alarm Device, CARESCAPE<br>RAD | Equivalent<br>The predicate device did<br>not support the<br>connector type of the<br>Remote Alarm Box with<br>Remote Light (RAB<br>RL). Therefore, the<br>Alarm Interface of ESP<br>V2 could not be utilized<br>The CARESCAPE RAD<br>is the new USB<br>accessory for use with<br>the CARESCAPE B650<br>V3.<br>As the proposed<br>CARESCAPE B650 V3<br>provides USB<br>connection, now the<br>Alarm Interface of the<br>CARESCAPE B650 V3<br>is utilized for remote<br>alarming.<br>The CARESCAPE RAD<br>is intended for relaying<br>primary alarm signals<br>from the host device to<br>an external distributed<br>alarm system, i.e. a<br>nurse call system or a<br>remote alarm light. The<br>CARESCAPE B650 V3<br>alarm functionality is<br>unaffected when using<br>the CARESCAPE RAD<br>as the new accessory<br>simply receives data<br>from the monitor (one-<br>way communication) to<br>indicate when an alarm<br>is active or not. | |
| | Intrahospital<br>transport within a<br>professional<br>healthcare facility. | Yes | Yes | Identical |
| | Battery operation | Rechargeable Lithium-Ion<br>batteries | Rechargeable Lithium-Ion<br>batteries | Identical |
| | Specification | CARESCAPE Monitor B650<br>with ESP V2 software<br>(K131223) | CARESCAPE B650 with<br>CSP software version V3 | Differences |
| | Networking<br>capability | CARESCAPE Network<br>LAN/VLAN<br>Optional WLAN | CARESCAPE Network<br>LAN/VLAN<br>Single wire network<br>configuration supported for<br>CARESCAPE Networks<br>Optional WLAN | Equivalent<br>The single wire network<br>configuration simplifies<br>the installation and<br>maintenance of<br>CARESCAPE B650 V3<br>patient monitors. The<br>single wire network<br>configuration has no<br>impact on clinical<br>monitoring. |
| | Network interface…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.