The IMDS NHancer™ Guide Wire Support Catheter with Hydrophilic Coating is a guidewire exchange and infusion device designed for use in the vascular system. The IMDS NHancer™ guidewire support catheter is intended to support a guidewire during access of vasculature and allows for exchange of guidewires and provides a conduit for the delivery of diagnostic contrast agents.
Device Story
NHancer is a single-use, over-the-wire (OTW) guidewire support catheter used during percutaneous transluminal coronary intervention (PTCI) procedures. The device consists of a single-lumen shaft, a distal tip with a radiopaque marker band, a proximal female luer hub, and a torquer for locking onto a 0.014-inch guidewire. The distal shaft features a hydrophilic coating to reduce friction. Operated manually by a physician in a clinical setting, the catheter is advanced over a guidewire through a guiding catheter or sheath into the target vessel. It provides mechanical support for guidewire navigation, enables guidewire exchange, and serves as a conduit for contrast media injection. The device benefits patients by facilitating access to occluded vessels during interventional procedures.
Clinical Evidence
Bench testing only. Performance testing included crossability, guidewire friction, torqueability, flow rate, marker visibility, guidewire exchangeability, and contrast administration. Additional bench tests verified mechanical integrity (pull/burst strength, kink resistance, luer testing) and biocompatibility (ISO 10993). All tests met acceptance criteria.
Indicated for use in the vascular system to support a guidewire during vascular access, facilitate guidewire exchange, and provide a conduit for diagnostic contrast agent delivery.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Spectranetics QuickCross® Support Catheters (K033678)
Submission Summary (Full Text)
{0}------------------------------------------------
TraditIonal 510(k) Submission for NHancer
Image /page/0/Picture/1 description: The image shows the text "K121077 Page 1 of 5 IMDS". The text appears to be handwritten. The letters are in black ink on a white background. The word "IMDS" is in a different font than the rest of the text.
## 005_510(k) Summary
DEC 2 1 2012
#### A. Manufacturer - and Contact Information
| Manufacturer: | IMDS Operations B.V.<br>Ceintuurbaan Noord 150<br>9301 NZ Roden<br>The Netherlands<br>Telephone: +31 50 8200 230<br>Fax: +31 50 8200 231 |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------|
|---------------|------------------------------------------------------------------------------------------------------------------------------------------|
Contact Person: Mr. E. Schulting Managing Director Telephone: +31 50 8200 230 e-mail: eschulting@imds.nl
Date of preparation: 04 April 2012
#### B. Reason for Premarket Notification
This Premarket Notification is being submitted for the NHancer™ guidewire support catheter which is a new device being manufactured by IMDS Operations B.V.
#### C. Device Information
| Trade Name: | NHancerTM |
|----------------------|-----------------------------------------------------------|
| Common Name: | 0.014" guidewire support catheter |
| Classification Name: | Percutaneous Catheter (21 CFR 870.1250, product Code DQY) |
#### D. Device Description
The purpose of the NHancer™ guidewire support catheter is to facilitate the placement of a guide wire through an occluded vessel during Percutane Transluminal Coronary Intervention procedures. The NHancer™ is a single use device, consisting of a shaft, a distal tip and a female luer (hub) on the proximal end. On the shaft are two depth markings applied to indicate the length of the device that is in the body. On the distal tip of the device a radiopaque marker band is applied. The distal part of the shaft is fitted with a hydrophilic coating. On the shaft a torquer is placed. With this component the catheter can be locked on to the guide wire that is in the device. The torquer is initially placed on a strain relief which is located on the proximal end of the shaft, near the female luer (hub).
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the text "K121077 Page 2 of 5 IMDS". The text appears to be handwritten. The letters and numbers are black against a white background. There is a logo next to the word "of".
#### E. Intended Use
The IMDS NHancer™ Guide Wire Support Catheter with Hydrophilic Coating is a guidewire exchange and infusion device designed for use in the vascular system. The IMDS NHancer™ guidewire support catheter is intended to support a guidewire during access of vasculature and allows for exchange of guidewires and provides a conduit for the delivery of diagnostic contrast agents.
#### F. Principle of Operation / Technology
The NHancer™ guidewire support catheter is operated manually or by manual process.
During an interventional procedure, the physician will follow the standard procedure of placing a guide wire and introducer within a vessel. Then a guiding catheter or sheath would be advanced over the guide wire. Next, the NHancer™ guidewire support catheter would be inserted over the guide wire and through the hemostasis valve of the guiding catheter or sheath. The guide wire and NHancer™ guidewire support catheter would then be advanced to the target vessel. The NHancer™ guidewire support catheter can then be used for support and exchange of guide wires or for injection of contrast media.
#### G. Statement of Equivalence
The NHancer™ guidewire support catheter submitted in this 510(k) is substantially equivalent in intended use, design, technology, principles of operation, materials and performance to the Spectranetics QuickCross® Support Catheters, cleared under K033678.
{2}------------------------------------------------
K121077 Page 3 of 5
IMDS
### H. Design / Materials / Specifications
The NHancer™ guidewire support catheter in this submission uses a similar design, similar materials and specifications as the predicate device: Spectranetics QuickCross® Support Catheter. Differences in design, materials and/or specifications between the devices do not raise any issues of safety and effectiveness. Below is a table with a comparison of the design, materials and specifications between the NHancer™ guidewire support catheter and the predicate devices.
| # | Item | Predicate Device:<br>Spectranetics QuickCross<br>Support Catheter<br>Model Number: 518-032 | IMDS NHancer™ guidewire<br>support catheter<br>Model Number: 8101-B140P |
|----|--------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|
| 1 | Design / construction | OTW, single lumen | OTW, single lumen |
| 2 | Shaft material | Polyethylene | Pebax / Polyethylene |
| 3 | Strain Relief | present | present |
| 4 | Luer Lock or Hub | Female, 6% taper and screw<br>conform ISO 594-1 and -2. | Female, 6% taper and screw<br>conform ISO 594-1 and -2. |
| 4 | Number of radiopaque markers | 3 | 1 |
| 6 | Radiopaque marker material | Platinum | Platinum / Irridium |
| 7 | Effective Length or Usable<br>Length | 135 cm | 140 cm |
| 8 | Exit Marker location | none | 95 and 105 cm |
| 9 | Guide wire compatibility | 0.014 inch | 0.014 inch |
| 10 | Maximum Injection pressure | 300 psi | 300 psi |
| 11 | Outer Diameter Shaft | 0.76 mm | 0.75 mm |
| 12 | Minimum Guiding Catheter | 5 Fr | 5 Fr |
| 13 | Tip design / shape | Straight | Straight |
| 14 | Hydrophilic coating length | Distal 40 cm | Distal 25 cm |
| 15 | Hydrophilic coating material | not known | PU / PVP<br>hydrophilic coating |
{3}------------------------------------------------
K121077 Page 4 of 5
IMDS
#### ﻟــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ Performance
The NHancer™ guidewire support catheter in this submission and the Spectranetics QuickCross® Support Catheter, cleared under K033678, have similar performance characteristics. The following performance tests were conducted on these catheters. Comparison testing was performed on the IMDS NHancer™ guidewire support catheter and the Spectranetics QuickCross Support Catheter.
- 1) Crossability (both in-vitro as well as in-vivo)
- 2) Guide wire friction (in-vitro)
- 3) Guidewire Torqueability (both in-vitro as well as in-vivo)
- 4) Flow rate (in-vitro)
- Marker visibility (in-vivo) ട)
- 6) Guidewire exchangeability (in-vivo)
- 7) Contrast medium administration (in-vivo)
The performance of the NHancer™ guidewire support catheter is substantially equivalent to the performance of the predicate device (Spectranetics QuickCross® Support Catheter). The performance of the NHancer™ guidewire support catheter met all the acceptance criteria.
In addition, the following bench tests were performed on NHancer™ guidewire support catheter to assure proper performance. All test results met the acceptance criteria.
- 1) Usable length
- 2) Length of coated part
- Tip inner diameter 3)
- 4) Tip length
- Outer diameter 5)
- Pull strength after 5 rotations 6)
- Burst strength after 5 rotations 7)
- 8) Flow rate testing
- System Burst testing 9)
- 10) 6% taper Luer testing
- 11) Screw connection Luer testing
- 12) Female Hub
- 13) Exit marker location
- 14) Radiopaque marker location
- 15) Visual appearance
- 16) Force at break: shaft
- 17) Force at break: shaft/hub connection
- 18) Force at break: tip/shaft connection
- 19) Outer diameter radiopaque marker
- 20) Dimensions of packaging box
- 21) Product integrity after simulated transportation test
- 22) Pouch bubble test
- 23) Liquid dye test
- 24) Pouch seal strength
- 25} Pouch seal width
- 26) Label adhesive retention
- 27) Legibility of product information
- 28) Pouch delamination assessment after opening
- 29) Kink resistance
- 30) Torque testing
- 31) Coating Integrity and Particulate Evaluation
{4}------------------------------------------------
K 121077 Page 5 of 5
IMDS
#### J. Additional Safety Information
Biocompatibility testing was performed in accordance with the Blue Book Memorandum, G95-1. Use of International Standard ISO 10993, and Biological Evaluation of Medical Devices part 1: Evaluation and Testing. The NHancer™ guidewire support catheter is classified as an externally communication device. circulating blood, limited contact (up to 24 hrs), which resulted in the following tests:
- 1) Thrombosis testing
- 2) Hemolysis testing
- 3) Cytotoxicity (Growth Inhibition test)
- 4) Irritation {Intracutaneous Reactivity}
- 5) Sensitation (Local Lymph Node Assy)
- ର୍ଚ୍ଚ Systemic Toxicity (Acute Systemic Toxicity)
The NHancer " guidewire support catheter successfully passed all above listed biocompatibility tests.
The NHancer®1 guidewire support catheter is sterilized by means of irradiation (gamma irradiation). The sterility has been validated in accordance with the applicable parts of the ISO 11137 standard (part 1: 2006, part 2: 2006) and ISO 11737 standard (part 1: 2006, part 2: 2009) to provide a Sterility Assurance Level (SAL) of 10to.
The NHancer™ guidewire support catheter is certified to be non-pyrogenic in the unopened and undamaged package. An Endotoxin Kinetic Turbidimetric test has been performed and will be performed for future production batches in accordance with current Ph. Eur. Chapter 2.6.14 Method C, kinetic turbidimetric test, as well as current USP, the harmonized part of chapter 15 and chapter 161. Initial production batches of the NHancer™ guidewire support catheter successfully passed the Endotoxin Turbidimetric test.
#### K. Substantial Equivalence
The NHancer™ guidewire support catheter submitted in this 510(k) is substantially equivalent in the general intended use, design, materials, specifications, principles of operation / technology, and performance to the Spectranetics QuickCross Support Catheter cleared under K033678. Differences between the devices do not raise any issue of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized symbol that resembles a human figure embracing a bird or other winged creature. The image is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
DEC 2 1 2012
IMDS Operations B.V. c/o Mr. E. Schulting Managing Director Ceintuurbann Noord 150 9301 NZ Roden The Netherlands
Re: K121077
Trade/Device Name: NHancer guidewire support catheter Regulation Number: 21 CFR 870.1250 . Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: November 22, 2012 Received: November 26, 2012
Dear Mr. Schulting:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (IPMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{6}------------------------------------------------
Page 2 - Mr. Schulting
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincercly yours,
# Matthew G. Hillebrenner
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
{7}------------------------------------------------
Traditional 510(k) Submission for NHancer
Image /page/7/Picture/1 description: The image shows a logo with the letters "IMDS" in bold, sans-serif font. Above the letters is a stylized graphic of an animal, possibly a badger or similar creature, rendered in black and white. The animal is depicted in a simplified, almost abstract manner, with its head and upper body visible.
## 004_Indications for Use Statement
510(k) Number (if known): K121077
Device Name:
NHancer (model number: 8101-B140P)
Indications For Use:
The IMDS NHancer™ Guide Wire Support Catheter with Hydrophilic Coating is a guidewire exchange and infusion device designed for use in the vascular system. The IMDS NHancer™ guidewire support catheter is intended to support a guidewire during access of vasculature and allows for exchange of guidewires and provides a conduit for the delivery of diagnostic contrast agents.
#### Prescription Use _X_
AND/OR
Over-The-Counter Use _N/A_
(Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) . .
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
| Page 1 |
|--------|
|--------|
|--|--|
| (Division Sign-Off) |
|------------------------------------|
| Division of Cardiovascular Devices |
| 510(k) Number | K121077 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.