K200269 · Orbusneich Medical Trading, Inc. · LOX · Mar 4, 2020 · Cardiovascular
Device Facts
Record ID
K200269
Device Name
Sapphire II PRO
Applicant
Orbusneich Medical Trading, Inc.
Product Code
LOX · Cardiovascular
Decision Date
Mar 4, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.0-1.25mm configurations) is indicated for: • balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (≥70% stenosis) for the purpose of improving myocardial perfusion The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.5-4.0mm configurations) is indicated for: • balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion • balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction The Sapphire® II PRO Balloon Dilatation Catheter is also indicated for: • percutaneous transluminal angioplasty in the peripheral vasculature, including renal, femoral, popliteal, infra-popliteal, tibial, and peroneal arteries.
Device Story
Over-the-wire balloon dilatation catheter; used for percutaneous transluminal coronary angioplasty (PTCA) and peripheral vascular procedures. Device features proximal Y-type hub, distal balloon, and internal lumen for 0.014-inch guidewire. External lumen allows inflation with dilute contrast media. Radiopaque platinum marker band aids positioning. Hydrophilic coating on distal section. Operated by physicians in clinical settings. Device advanced to stenosis; balloon inflated to dilate vessel, improving myocardial perfusion or treating occlusion. Benefits include restoration of blood flow in stenotic coronary or peripheral arteries.
Clinical Evidence
Bench testing only. No clinical data. Performance testing included dimensional verification, balloon rated burst pressure, fatigue, compliance, bond strength, tip pull strength, radiopacity, coating integrity, shaft burst, inflation/deflation time, flexibility, kinking, torque strength, and particulate evaluation. Biocompatibility testing included cytotoxicity, intracutaneous reactivity, sensitization, acute systemic toxicity, pyrogenicity, hemocompatibility (hemolysis, complement activation, PTT, platelet/leukocyte counts), and genotoxicity. All results met acceptance criteria.
Indicated for patients with coronary artery or bypass graft stenosis (≥70% for 1.0-1.25mm; symptomatic ischemia for 1.5-4.0mm), acute myocardial infarction, or peripheral vascular disease (renal, femoral, popliteal, infra-popliteal, tibial, peroneal arteries).
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 04, 2020
OrbusNeich Medical Trading Inc. John Pazienza Senior Director, Engineering 5363 NW 35th Avenue Fort Lauderdale, Florida 33309
# Re: K200269
Trade/Device Name: Sapphire II PRO Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: Percutaneous Transluminal Coronary Angioplasty (Ptca) Catheter Regulatory Class: Class II Product Code: LOX, LIT Dated: January 31, 2020 Received: February 4, 2020
Dear Mr. John Pazienza:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Gregory O'Connell Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200269
Device Name Sapphire II PRO Balloon Dilatation Catheter
Indications for Use (Describe)
The Sapphire® II PRO balloon dilatation catheter (1.0-1.25mm configurations) is indicated for:
· balloon pre-dilatation of a stenotic portion of a coronary artery or bypass graft stenosis (270% stenosis) for the purpose of improving myocardial perfusion.
The Sapphire® II PRO balloon dilatation catheter (1.5-4.0mm configurations) is indicated for:
· balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion.
· balloon dilatation of a coronary artery occlusion for the treatment of acute myocardial infarction.
The Sapphire® II PRO balloon dilatation catheter is also indicated for:
· percutaneous transluminal angioplasty in the peripheral vasculature, including renal, fomoral, popliteal, tibial, and peroneal arteries.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This 510(k) Summary is submitted in accordance with 21 CFR 807.92(c).
| Submitter: | OrbusNeich Medical Trading, Inc.<br>5363 NW 35th Avenue<br>Fort Lauderdale, FL 33309<br>Phone: 954.730.0711<br>Fax: 954.730.7601 |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John D. Pazienza |
| Date Prepared: | January 31, 2020 |
| Trade Name: | Sapphire II PRO Balloon Dilatation Catheter |
| Common Name: | Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheter |
| Classification Name: | Catheters, transluminal coronary angioplasty, percutaneous<br>21 CFR 870.5100(a) |
| Product Code: | LOX (primary)<br>LIT (secondary |
| Device Class: | Class II |
| Predicate Device: | OrbusNeich Sapphire II PRO (K180921; LOX, LIT; cleared 28-Jun-<br>2018) |
| Reference Devices: | Abbott Vascular Mini Trek OTW (K110617; LOX; cleared 02-Jun-<br>2011)<br>Boston Scientific Emerge OTW (K130391; LOX; cleared 10-Jul-2013)<br>Medtronic Sprinter Legend OTW (K103095; LOX; cleared 22-Oct-<br>2010) |
| Device Description: | The Sapphire II PRO Balloon Dilatation Catheter is now also available<br>as an over-the-wire balloon catheter with a working length of 150cm.<br>The semi-compliant balloons are available in diameters from 1.0-<br>1.25mm and lengths from 5-15mm with a rated burst pressure of 14<br>atmospheres. The catheter consists of proximal section with a Y-type<br>hub and distal section with a balloon near the distal tip. The straight<br>port of the hub is the guidewire entrance and the side port is used to<br>inflate and deflate the balloon. The external lumen provides for<br>inflation of the balloon with dilute contrast media solution.<br>The<br>internal lumen permits the use of a standard 0.014 inch guidewire to<br>facilitate advancement of the catheter to and through the stenosis to be<br>dilated. Hydrophilic lubricious coatings are applied to the distal<br>section. One radiopaque platinum marker band is located within the<br>balloon segment. The catheter is compatible with 5F or larger guiding<br>catheters and 4F or larger guiding sheaths. Two marked sections<br>(3mm in length) are located on the proximal shaft to indicate catheter<br>position relative to the tip of either guiding catheter or guiding sheath.<br>The design of this dilatation catheter does not incorporate a lumen for<br>distal dye injections or distal pressure measurements. |
| Intended Use: | The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.0-1.25mm<br>configurations) is indicated for:<br>• balloon pre-dilatation of a stenotic portion of a coronary artery or<br>bypass graft stenosis (≥70% stenosis) for the purpose of improving<br>myocardial perfusion<br>The Sapphire® II PRO Balloon Dilatation Catheter (Ø1.5-4.0mm<br>configurations) is indicated for:<br>• balloon dilatation of the stenotic portion of a coronary artery or<br>bypass graft stenosis in patients evidencing coronary ischemia for<br>the purpose of improving myocardial perfusion<br>• balloon dilatation of a coronary artery occlusion for the treatment<br>of acute myocardial infarction<br>The Sapphire® II PRO Balloon Dilatation Catheter is also indicated<br>for:<br>• percutaneous transluminal angioplasty in the peripheral<br>vasculature, including renal, femoral, popliteal, infra-popliteal,<br>tibial, and peroneal arteries. |
| Technological Characteristics: | At a high level, the subject and predicate devices are based on the same<br>technological elements:<br>• the same indications for use<br>• identical semi-compliant balloon<br>• nominal pressure of 6 atm<br>• rated burst pressure of 14 atm<br>• identical distal shaft<br>• hydrophilic coating<br>• 0.014" guidewire compatibility<br>• EO sterilization<br>The following technological differences exist between the subject and |
{4}------------------------------------------------
- predicate device: over-the-wire catheter shaft design ●
- specific materials selected for proximal shaft
- . exact dimensions of proximal shaft components and catheter
{5}------------------------------------------------
Performance Data:
Testing was leveraged from the predicate device (K180921) when applicable:
- Shelf-Life (Partial) ●
- Performance Testing (Partial) ●
- Dimensional Verification (Partial) O
- o Balloon Rated Burst Pressure (Partial)
- Balloon Fatigue O
- Balloon Compliance O
- Catheter Bond Strength (Partial) O
- Tip Pull Strength O
- Radiopacity o
- Coating Integrity o
Additional testing was performed to support the use of the Sapphire II PRO balloon dilatation catheter over-the-wire shaft design:
- Sterilization ●
- . Performance Testing
- o Visual Inspection
- o Dimensional Verification
- Balloon Preparation, Deployment, and Retraction O
- o Balloon Rated Burst Pressure
- o Shaft Burst
- Balloon Inflation and Deflation Time O
- o Catheter Bond Strength
- Flexibility and Kinking O
- Torque Strength O
- Particulate Evaluation O
- Biocompatibility
- Cytotoxicity O
- Intracutaneous Reactivity O
- Sensitization O
- Acute Systemic Toxicity o
- Pyrogenicity o
- O Hemocompatibility
- · Hemolysis
- Complement Activation
- Partial Thromboplastin Time .
- Platelet and Leukocyte Counts .
- o Genotoxicity
The Sapphire II PRO balloon dilatation catheter test results met all acceptance criteria and were similar to the predicate device.
Conclusion:
This information supports a determination of substantial equivalence between the Sapphire II PRO balloon dilatation catheter and the predicate device described above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.