SPRINTER LEGEND 1.25MM RX AND SPRINTER LEGEND 1.25MM OTW BALLOON
Applicant
Medtronic, Ireland
Product Code
LOX · Cardiovascular
Decision Date
Oct 22, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sprinter Legend RX and OTW 1.25mm Balloon Dilatation Catheter is indicated as a pre-dilatation catheter for enlarging coronary luminal diameters during PCI procedures.
Device Story
The Sprinter Legend 1.25mm RX and OTW are balloon dilatation catheters used in percutaneous coronary intervention (PCI). The device consists of a semicompliant balloon mounted on a delivery catheter (Rapid Exchange or Over-the-Wire). The physician inserts the catheter over a guidewire to the target coronary lesion under fluoroscopic guidance using radiopaque markers. The balloon is inflated via a luer port using pressurized fluid to a defined diameter to dilate the stenotic vessel. The device serves as a pre-dilatation tool to enlarge the coronary lumen, facilitating subsequent procedures such as stent placement. It is used in a clinical setting by interventional cardiologists. The device benefits patients by successfully enlarging coronary luminal diameters, improving blood flow (TIMI 3 flow), and preparing the lesion for further intervention.
Clinical Evidence
Prospective, multi-center, single-arm study (N=51 subjects, 54 lesions). Primary endpoint: procedural success (delivery, dilatation, no perforation/dissection/arrhythmia, TIMI 3 flow). Results: 100% procedural success, 100% device success, 100% lesion success. MACE rate 2.0%. Mean pre-procedure MLD 0.623mm improved to 2.173mm post-procedure. Data supports safety and efficacy for pre-dilatation.
Technological Characteristics
Semicompliant balloon on RX or OTW delivery catheter. RX: single lumen stainless hypotube. OTW: bifurcated luer for inflation and guidewire. Radiopaque markers for fluoroscopic visualization. Biocompatibility per ISO 10993-1:2009. Sterilization via Ethylene Oxide (EtO).
Indications for Use
Indicated for patients undergoing percutaneous coronary intervention (PCI) procedures requiring pre-dilatation to enlarge coronary luminal diameters, including those with chronic total occlusions (CTOs).
Regulatory Classification
Identification
Standard PTCA Catheter: A PTCA catheter is a device that operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. A PTCA balloon catheter has a single or double lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of radiographic marker to facilitate fluoroscopic visualization of the balloon during use. A PTCA catheter is intended for balloon dilatation of a hemodynamically significant coronary artery or bypass graft stenosis in patients evidencing coronary ischemia for the purpose of improving myocardial perfusion. A PTCA catheter may also be intended for the treatment of acute myocardial infarction; treatment of in-stent restenosis (ISR) and/or post-deployment stent expansion. Cutting/scoring PTCA Catheter: A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.
Special Controls
*Classification.* Class II (special controls). The special control for this device is “Class II Special Controls Guidance Document for Certain Percutaneous Transluminal Coronary Angioplasty (PTCA) Catheters.” See § 870.1(e) for the availability of this guidance document.(b)
*Cutting/scoring PTCA Catheter* —(1)*Identification.* A cutting/scoring PTCA catheter is a balloon-tipped catheter with cutting/scoring elements attached, which is used in those circumstances where a high pressure balloon resistant lesion is encountered. A cutting/scoring PTCA catheter is intended for the treatment of hemodynamically significant coronary artery stenosis for the purpose of improving myocardial perfusion. A cutting/scoring PTCA catheter may also be indicated for use in complex type C lesions or for the treatment of in-stent restenosis.(2)
*Classification.* Class III (premarket approval). As of May 28, 1976, an approval under section 515 of the act is required before this device may be commercially distributed. See § 870.3.
{0}------------------------------------------------
·
# 510(k) Summary
| Date Prepared: | October 21, 2010 | OCT 2 2 2010 |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Applicant: | Medtronic Ireland<br>Parkmore Business Park West<br>Galway<br>Ireland | |
| Primary Correspondent: | John-Paul Hughes<br>Regulatory Affairs Specialist<br>Medtronic Ireland<br>Parkmore Business Park West<br>Galway<br>Ireland<br>Phone: (011-353-91) 708579<br>Fax: (011-353-91) 708672<br>Email: john-paul.hughes@medtronic.com | |
| Alternate Correspondent: | Gerardine Finn<br>Medtronic Inc.<br>Medtronic Vascular<br>3576 Unocal Place<br>Santa Rosa<br>California 95403<br>USA<br>Phone: (707) 566 1548<br>Fax: (707) 566 1549<br>Email: gerardine.finn@medtronic.com | |
| Proprietary Name: | - Sprinter Legend 1.25mm Rapid Exchange Balloon<br>Dilatation Catheter<br>- Sprinter Legend 1.25mm Over-the-Wire Balloon Dilatation<br>Catheter | |
.
・
1
Medtronic Ireland
Meditorie Tieland
Sprinter Legend 1.25mm RX & Sprinter Legend 1.25mm OTW Balloon Dilatation Catheters
510(k) K103095
{1}------------------------------------------------
| CONFIDENTIAL<br>May not be reproduced without written permission from Medtronic Ireland | |
|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Sprinter Legend 1.25mm RX and Sprinter Legend 1.25mm OTW |
| Device Classification Name: | Cardiovascular Devices<br>Class II, 21 CFR Part 814 |
| Product Code: | 74 LOX |
| Device Description: | The Medtronic Sprinter Legend 1.25mm RX and OTW are balloon<br>dilatation catheters designed to perform Percutaneous Transluminal<br>Coronary Angioplasty (PTCA).<br><br>Sprinter Legend 1.25mm RX<br>The device consists of a semicompliant balloon mounted on a<br>Rapid Exchange delivery catheter. The Sprinter Legend 1.25mm<br>RX device has an effective length of 142cm. It will be available in<br>a single balloon diameter of 1.25mm and in balloon lengths from<br>6mm to 20mm.<br><br>Sprinter Legend 1.25mm OTW<br>The device consists of a semicompliant balloon mounted on an<br>Over-the-Wire delivery catheter. The Sprinter Legend 1.25mm<br>OTW device has an effective length of 152cm. Similar to Sprinter<br>Legend 1.25mm RX, Sprinter Legend 1.25mm OTW will also be<br>available in a 1.25mm balloon diameter and with balloon lengths<br>from 6mm to 20mm. |
| Indications For Use: | The Sprinter Legend RX and OTW 1.25mm Balloon Dilatation<br>Catheters are indicated as pre-dilatation catheters for enlarging<br>coronary luminal diameters during PCI procedures. |
| Substantially Equivalent Devices: | The Sprinter Legend 1.25mm RX and OTW Balloon Dilatation<br>Catheters use similar technology, materials and method of<br>operation to the approved predicates Sprinter Legend RX device<br>(1.5-4.0mm) (P790017/S096) and Sprinter OTW Balloon<br>Dilatation Catheter (P790017/S081). The indications for use for<br>these devices are similar to those of Sprinter Legend RX (1.5-<br>4.0mm) and the Sprinter OTW devices. The properties of the<br>Sprinter Legend 1.25mm RX and Sprinter Legend 1.25mm OTW<br>devices to treat disease are identical to previously approved PTCA<br>balloon dilatation catheters outlined above. |
:
י
{2}------------------------------------------------
Summary of Technological The Medtronic Sprinter Legend 1.25mm RX and OTW devices Characteristics: consist of a balloon at the distal end of the catheter, which can be inflated and deflated by a coaxial lumen, to a defined diameter at a specific pressure. This means that the inner shaft resides inside the distal outer shaft, leaving an annular space between the two for pressurized fluid to flow. For the OTW product the proximal end of the catheter has a luer for attachment to an inflation device. This bifurcated luer has two legs, one to control inflation and deflation of the catheter and one to allow insertion of a guidewire. The proximal section of the RX catheter is a single lumen stainless hypotube with a single luer port for attachment to an inflation device to control inflation and deflation of the catheter. Both catheters provide a lumen which enables the use of a guidewire to position the catheter. A radiopaque balloon marker, which facilitates imaging under fluoroscopy, enables accurate placement. Shaft markers for brachial and femoral techniques are in place. Summary of Studies: The following device integrity testing was completed using Sprinter Legend 1.25mm RX and or OTW devices: Catheter Dimensiona! Measurements & Profile . . Catheter Tensile Strength Radiopacity . Coating Testing . Balloon Compliance . System Pressure Capability . System Fatigue . Balloon Inflation and Deflation, Performance and Balloon . Preparation Interaction with Accessories . Wire Lumen Crush . Kissing Balloon Technique . Flexibility & Kink Test . Dye Flow through a Guide Catheter . Packaging Integrity Testing . The Medtronic Sprinter Legend 1.25mm RX and OTW Balloon Dilatation Catheters met all specified design and performance requirements. Summary of Clinical Data: Overview of Clinical Study Results of The Sprinter® Legend 1.25mm Trial Clinical data was required for the Sprinter Legend 1.25mm balloon diameter size to ensure that the technological characteristics of the small diameter balloon did not affect safety or effectiveness.
Medtronic Ireland Mouronio Kalano
Sprinter Legend 1.25mm RX & Sprinter Legend 1.25mm OTW Balloon Dilatation Catheters 510(k) K103095
{3}------------------------------------------------
### Study Purpose:
To examine acute outcomes when the Sprinter Legend 1.25 mm OTW balloon catheter is used as a pre-dilatation catheter for enlarging coronary luminal diameters during PCI procedures with subsequent use, in most cases, of another PTCA catheter to complete the dilatation.
## Design:
This was a prospective, multi-center, single arm open label study that evaluated acute outcomes when the Sprinter Legend 1.25 mm balloon catheter was used as a predilatation catheter for enlarging coronary luminal diameters during percutaneous coronary intervention (PCI) procedures.
Eligibility was based upon the assessment of the lesion's stenosis being ≥70% and included chronic total occlusions (CTOs). The primary endpoint in this study was the Procedural Success consisting of the following:
- · Delivery of the balloon catheter to the target lesion and dilatation of the lesion using the study device.
- · No evidence of arterial perforation, flow limiting dissection, reduction in TIMI flow from baseline TIMI or clinically significant arrhythmias following pre-dilatation step with the . study device.
- · TIMI 3 flow post-dilatation at the conclusion of the PCI procedure.
The secondary endpoints were Device success. Lesion success, Inhospital
MACE, Target Lesion Failure (TLF), Thrombosis, Dissection or perforation.
Balloon Rupture and Arrhythmias.
Device success consists of the following:
· Delivery of the balloon catheter to the target lesion and dilatation of the lesion using the study device
· No evidence of arterial perforation, flow limiting dissection, reduction in TIMI flow from baseline TIMI or clinically significant arrhythmias following predilatation step with the study device.
Lesion Success consists of the following:
• Delivery of the balloon catheter to the target lesion and dilatation of the lesion using any PCI method
A total of 51 patients were enrolled in this study at Two (2) study
{4}------------------------------------------------
sites in the United States. Clinical follow-up was conducted postprocedure until discharge.
## Demographics:
Baseline demographic parameters for the enrolled population show that the mean age was 67.3 years and 90.2% (46/51) of the subjects were male. Additionally, 96.1% (49/51) subjects had hyperlipidemia and 96.1% (49/51) had hypertension.
### Results:
The primary and secondary endpoint results are presented in Table l and 2 below. The combined data collected for the primary endpoint indicates that the procedural success was 100.0% (51/51) for this trial. The rate of device success was 100% (51/51) and that of lesion success was also 100% (54/54).
| | | | | | Table 1: Effectiveness Measures |
|--|--|--|--|--|---------------------------------|
|--|--|--|--|--|---------------------------------|
| Primary Endpoint | Sprinter Legend 1.25mm<br>(N=51 Subjects)<br>(N=54 Lesions) |
|-------------------|-------------------------------------------------------------|
| Procedure Success | 100.0% (51/51) |
| Device Success | 100.0% (51/51) |
| Lesion Success * | 100.0% (54/54) |
* Percentage is based on number of lesions with available data. Percentage is based on available angiographic core lab data and if missing, site reported data of post Sprinter Legend TIMI and Perforation are applied.
## Table 2: Principal Safety and Effectiveness Results
| Secondary Endpoint | Sprinter Legend<br>(N=51 Subjects)<br>(N=54 Lesions) |
|---------------------------------------------------------|------------------------------------------------------|
| Major Adverse Cardiac Event (MACE) | 2.0% (1/51) |
| Death | 0.0% (0/51) |
| Cardiac Death | 0.0% (0/51) |
| Target Vessel MI (TVMI, Medtronic Historic Definition) | 2.0% (1/51) |
| TVMI (ARC Definition) | 15.7% (8/51) |
| Emergent CABG | 0.0% (0/51) |
| Clinically Driven Target Lesion Revascularization (TLR) | 0.0% (0/51) |
| Dissection | 0.0% (0/51) |
| Flow Limiting Dissection | 0.0% (0/51) |
| Clinically Driven Target Vessel Revascularization (TVR) | 0.0% (0/51) |
| Perforation | 0.0% (0/51) |
| Clinical/Angiographic | 0.0% (0/51) |
| Thrombus | 0.0% (0/51) |
Medtronic Ireland
Sprinter Legend 1.25mm RX & Sprinter Legend 1.25mm OTW Balloon Dilatation Catheters 510(k) K103095
{5}------------------------------------------------
| Arrhythmias | 0.0% (0/51) |
|-------------------------------------------|----------------|
| Target Lesion Failure (TLF) | 2.0% (1/51) |
| Target Vessel Failure (TVF) | 2.0% (1/51) |
| Stent Thrombosis (ARC Definite/Probable)1 | 0.0% (0/44) |
| Definite | 0.0% (0/44) |
| Probable | 0.0% (0/44) |
| Effectiveness Measures | |
| Procedure Success | 100.0% (51/51) |
| Device Success | 100.0% (51/51) |
| Lesion Success2 | 100.0% (54/54) |
Percentage is based on number of subjects with available data. except:
1 For subjects receiving stents
2 Percentage is based on number of lesions with available data
#### Conclusion:
The Sprinter® Legend 1.25 mm catheter trial met its primary endpoint of procedural success as well as the endpoints of lesion and device success and demonstrated acceptably low rates of adverse events.
Also, as evidenced by the angiographic core lab, 100% (54/54) of lesions had a resulting TIMI 3 coronary blood flow post-procedure further supporting the efficacy profile of this pre-dilatation catheter. The Sprinter Legend 1.25mm balloon successfully assisted subsequent percutaneous coronary interventions to enlarge coronary luminal diameters from a MLD preprocedure of 0.623mm to 2.173mm post-procedure.
In conclusion, the safety and efficacy data presented in this report illustrate the Sprinter Legend 1.25mm balloon may be used to successfully enlarge coronary luminal diameters during PCI procedures as a pre-dilatation balloon catheter.
{6}------------------------------------------------
| Biocompatibility Information: | The testing which supports the biocompatibility of the Medtronic<br>Sprinter Legend 1.25mm RX and OTW Balloon Dilatation<br>Catheter product family is consistent with International Standard<br>ISO10993-1:2009 "Biological Evaluation of Medical devices- Part<br>1: Evaluation and Testing."<br><br>When classified according to this standard, the catheter and dilator<br>included in the Medtronic Sprinter Legend 1.25mm RX and OTW<br>Balloon Dilatation Catheter product family are categorized as<br>external communicating devices with limited exposure i.e. whose<br>contact with circulating blood is ≤ 24 hours.<br><br>The following Biocompatibility tests were performed:<br>Cytotoxicity Using ISO Elution Method ASTM Hemolysis Study Material Mediated Pyrogen ISO Maximization Sensitization Study ISO Intracutaneous Study ISO Systemic Toxicity Study C3a and SC5b-9 Complement Activation Assay Study Physicochemical Tests ASTM Partial Thromboplastin Time Study (PTT) In Vivo Thromboresistance Study in the Dog, Jugular Vein The biocompatibility evaluation completed verifies that the<br>Medtronic Sprinter Legend 1.25mm RX and OTW Balloon<br>Dilatation Catheters are biocompatible. |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilization Validation: | The Medtronic Sprinter Legend 1.25mm RX and OTW Balloon<br>Dilatation Catheters will be sterilized using a validated Ethylene<br>Oxide (EtO) sterilization process. |
| Conclusion: | Through the data and information presented, Medtronic Ireland<br>considers the Medtronic Sprinter Legend 1.25mm RX and OTW<br>Balloon Dilatation Catheters to be substantially equivalent to the<br>predicate devices. |
·
.
・
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{7}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medtronic Ireland c/o Ms. Gerardine Finn Vice President, Regulatory Affairs Medtronic, Inc. 3576 Unocal Place Santa Rosa, CA 95403
OCT 2 2 2010
Re: K103095
Trade/Device Name: Sprinter Legend 1.25mm Rapid Exchange Balloon Dilatation Catheter Sprinter Legend 1.25mm Over-the-Wire Balloon Dilatation Catheter Regulation Number: 21 CFR 870.5100 Regulation Name: PTCA Catheters Regulatory Class: Class II (two) Product Code: LOX Dated: Not Dated Received: October 8, 2010
Dear Ms. Finn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{8}------------------------------------------------
Page 2 - Ms. Gerardine Finn
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### CONFIDENTIAL May not be reproduced without written permission from Medtronic Ireland
.
## Indication for Use
| 510(k) Number (if known): | K103095 | OCT 2 2 2010 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Device Name: | Medtronic Sprinter Legend 1.25mm RX and Sprinter Legend<br>1.25mm OTW Balloon Dilatation Catheters | |
| Indications for Use: | The Sprinter Legend RX and OTW 1.25mm Balloon<br>Dilatation Catheter is indicated as a pre-dilatation catheter<br>for enlarging coronary luminal diameters during PCI<br>procedures. | |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
-
Medtronic Ireland Sprinter Legend 1.25mm RX & Sprinter Legend 1.25mm OTW Balloon Dilatation Catheters 510(k) K103095
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.