K200160 · Zoe Medical, Inc. · MWI · Feb 15, 2021 · Cardiovascular
Device Facts
Record ID
K200160
Device Name
740 SafeSAT
Applicant
Zoe Medical, Inc.
Product Code
MWI · Cardiovascular
Decision Date
Feb 15, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The 740 SafeSAT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring of adult and pediatric patients, excluding neonates. The Monitor features technology to facilitate the monitoring of non-invasive blood pressure, pulse rate and functional arterial oxygen saturation (SpO2) in a variety of clinical environments.
Device Story
740 SafeSAT is a portable, multiparameter patient monitor for spot-checking and continuous monitoring of adult/pediatric patients. Inputs include NIBP (oscillometric cuff) and SpO2 (red/infrared light absorption via finger sensors/probes). Device processes signals to output systolic/diastolic/mean arterial pressure, pulse rate, and SpO2 levels. Used in clinical environments by healthcare professionals; features color touch screen for configuration, alarm management, and trend display. Includes standby mode for workflow efficiency and serial/nurse call connectivity for EMR integration. Benefits include reliable vital sign tracking and alarm vigilance for variable acuity patients. Powered by AC or internal Li-Ion battery.
Clinical Evidence
Clinical studies conducted per ISO 80601-2-61 on healthy male/female adults (20-40 years) with varying skin tones. Evaluated SpO2 accuracy against blood gas analysis under motion and non-motion conditions. No adverse effects reported. Results demonstrate device meets performance claims and is substantially equivalent to the predicate.
Technological Characteristics
Portable multiparameter monitor; Li-Ion battery; color touch screen interface. SpO2 sensing: 2 LEDs (red/IR) with single photosensor (standard); single white LED with 2 photosensors (motion). Connectivity: Serial data output, nurse call relay, optional thermal strip chart recorder. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-8, IEC 80601-2-30, ISO 80601-2-61.
Indications for Use
Indicated for adult and pediatric patients, excluding neonates, for spot checking and continuous monitoring of non-invasive blood pressure, pulse rate, and functional arterial oxygen saturation (SpO2) by healthcare professionals.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 15, 2021
Zoe Medical, Inc. Jim Chickering Regulatory Affairs Manager 460 Boston Street Topsfield, Massachusetts 01983
Re: K200160
Trade/Device Name: 740 SafeSAT Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DQA, DXN Dated: February 8, 2021 Received: February 11, 2021
Dear Jim Chickering:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200160
Device Name
740 SafeSAT
#### Indications for Use (Describe)
The 740 SafeSAT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring of adult and pediatric patients, excluding neonates. The Monitor features technology to facilitate the monitoring of non-invasive blood pressure, pulse rate and functional arterial oxygen saturation (SpO2) in a variety of clinical environments.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------|--------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### SUBMITTER
Zoe Medical, Inc. 460 Boston Street Topsfield, MA 01983 USA
Phone: 978-887-1410 FAX: 978-887-1406
Establishment Registration: 3003294644
Contact Person: Jim Chickering, Regulatory Affairs Manager Date Prepared: January 3, 2020
#### DEVICE
Trade Name: 740 SafeSAT
Common Name: Patient Physiological Monitor
Classification Product Code: MWI - Monitor, Physiological, Patient (21 CFR 870.2300)
Secondary Product Codes: DQA - Oximeter (21 CFR 870.2700) DXN - NIBP Measurement System (21 CFR 870.1130)
Requlatory Class: II
Reason for Premarket Notification: New device
{4}------------------------------------------------
## PREDICATE DEVICE
The predicate device is the 740 SELECT patient physiological monitor (K130411).
| Predicate Scope | Predicate Device | 510(k) | Classification |
|-------------------|------------------|---------|----------------|
| Primary Predicate | 740 SELECT | K130411 | MWI |
The 740 SafeSAT subject device is an alternate construction of the predicate device. In this alternate construction, the Nellcor SpO2 OEM technology is replaced by the "SafeSAT" SpO2 OEM technology, which is manufactured by OSI Optoelectronics (Hawthorne, CA). Furthermore, the neonatal indication is removed from the subject device and the temperature monitoring capability, which is offered optionally in the predicate device, is not offered in the subject device.
No reference devices were used to help support a substantial equivalence determination.
#### DEVICE DESCRIPTION
The 740 SafeSAT physiological patient monitor is a portable and rugged non-invasive multiparameter device used for spot checking, continuous monitoring, and recording of blood pressure, pulse rate, and functional arterial oxygen saturation (SpO2) in a variety of clinical environments.
The 740 SafeSAT monitor employs SafeSAT SpO2 technology which includes the following components from OSI Optoelectronics:
| SafeSAT Component REF | Description |
|-----------------------|-------------------------------------|
| 5110 | SafeSAT SpO2 Adult Finger Sensor |
| 5120 | SafeSAT SpO2 Adult Disposable Probe |
| 5130 | SafeSAT SpO2 Adult Soft Tip Probe |
| 5210 | SafeSAT SpO2 Extension Cable |
| 5910 | SafeSAT SpO2 OEM Module |
The 740 SafeSAT monitor is compact and lightweight in design, allowing it to be easily carried or configured for use with a mobile stand, wall mount or on a table top.
A Lithium Ion (Li-lon) rechargeable battery is provided for automatic switching to back-up power if an unintended loss of AC power should occur as well as allowing the monitor to be used for clinical applications when portable monitoring is required.
The 740 SafeSAT monitor is equipped with a color touch screen which is the primary interface for the user to control and configure the monitor (for patient type, alarms, settings and various clinical workflows). A dedicated message area provides patient alarm and technical messages. Date and time as well as the Monitor ID are displayed. A battery gauge provides battery capacity when the monitor is operating on battery power.
{5}------------------------------------------------
The Monitor supports both point-of-care and variable acuity monitoring. Spot vital signs measurements from multiple patients can be saved individually. The monitor can be used for continuous patient monitoring with full alarm support when increased vigilance is necessary based on patient acuity.
A Standby Mode provides enhanced bedside patient workflow and alarm management by permitting sensor attachment, automatic alarm suspension and a fast transition to active monitoring.
The color-coded Trends Display allows the user to recall all trended physiological patient measurement records. Trends can be recalled for saved snapshots or timed intervals.
The Monitor provides a screen keyboard for entry of patient data (e.g., age, weight, height, gender, date of birth, patient name, and patient identification number).
A serial data output is provided for connectivity to hospital EMR systems. An internal isolated relay switch is provided for attachment to nurse call systems. The Monitor supports an optional, externally connected, thermal strip chart recorder which is directly powered from a serial port.
The MAXNIBP non-invasive blood pressure (NIBP) parameter automatically inflates an occluding cuff and, using the oscillometric measurement technique, determines systolic, diastolic and mean arterial pressure as well as pulse rate. Measurement results along with user prompts and error messages are displayed in the dedicated message area on the display. Automatic, Manual and STAT measurement modes are available to cover a variety of clinical uses.
The pulse oximeter parameter (SpO₂) determines arterial hemoglobin oxygen saturation by measuring the absorption of red and infrared light passing through the tissue. Changes in absorption caused by pulsations of blood in the vascular bed are used to determine arterial saturation and pulse rate. The monitor can display a SpO2 plethysmograph waveform in spot and/or continuous monitoring mode. A signal strength bar provides relative quality of the SpO2 value based on the measurement site where the sensor is placed and the perfusion level of the patient. An audio pulse tone "beep" can be generated each time the SpO2 module detects a pulse.
A pulse rate is displayed and appropriately color-coded to indicate the source parameter (SpO2 or NIBP).
Alarm conditions are displayed as text messages in a dedicated area on the bottom of the display. A recurring series of tones are provided to indicate when an alarm limit has been violated, a condition exists that affects monitoring (e.g., battery low), or a condition exists that affects patient safety. Visual cues (flashing background) are displayed within a parameter cell to indicate when an alarm is currently active for that parameter.
{6}------------------------------------------------
# COMPARISON OF INDICATIONS FOR USE
The 740 SafeSAT is indicated for use as a bedside, portable device for use by health care professionals, clinicians and medically qualified personnel for spot checking, continuous monitoring and recording of adult and pediatric patients, excluding neonates. The Monitor features technology to facilitate the monitoring of non-invasive blood pressure, pulse rate and functional arterial oxygen saturation (SpO2) in a variety of clinical environments.
The removal of the neonatal population and the deletion of temperature monitoring from the indications statement in the subject device does not change the intended use of the device relative to the predicate. Temperature monitoring is an optional configuration of the predicate device.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The predicate device and subject device are substantially the same physical device. The subject device uses a different SpO2 OEM technology within the device, with similar patient-applied sensors.
Both the predicate device and subject device employ the well-known method of calculating the patient's SpO2 parameter value, in which the ratio of red to infrared light absorption is measured. This is achieved with 2 LED's (red and infrared) that are illuminated into the patient's tissue in an alternating sequence and a single photosensor that measures the red and infrared light levels in the same alternating sequence.
When sensor motion is detected, the subject device employs a second method for measuring the red and infrared light. This is achieved with a single white LED that is illuminated into the patient's tissue and 2 photosensors that are narrowly sensitive to red and infrared frequencies, which measure the red and infrared light levels simultaneously.
{7}------------------------------------------------
# NON-CLINICAL TESTING
The following tests were performed in accordance with design control regulations to demonstrate substantial equivalence with the predicate device:
- Electrical safety per IEC 60601-1 ●
- Electromagnetic Compatibility per IEC 60601-1-2 ●
- Usability per IEC 60601-1-6 ●
- Alarm safety and essential performance per IEC 60601-1-8 ●
- . Non-invasive Blood Pressure safety and essential performance per IEC 80601-2-30
- SpO2 safety and essential performance per ISO 80601-2-61 ●
- Software verification per FDA Software Guidance
- Mechanical and environmental safety ●
The test results demonstrated that all requirements and performance specifications were met.
## CLINICAL TESTING
Clinical studies were performed with all sensors in accordance with the ISO 80601-2-61 standard on pulse oximetry to determine accuracy with respect to blood testing. Additionally, clinical studies were performed in accordance with the ISO 80601-2-61 standard with the Adult Finger Sensor (REF 5110) in the presence of motion artifact. The studies were conducted on healthy male and female adults in the range of 20 to 40 years of age and with a range of skin tone, as outlined in the standard.
No adverse effects or complications were reported during the studies.
The clinical data demonstrates that the device exceeds its performance claims for SpO2 measurement accuracy under motion and non-motion conditions, and supports the determination of substantial equivalence of the subject device.
Given that the SpO2 measurement accuracy claims of the subject device are the same as those of the predicate device, the clinical data supports the substantial equivalence claim to the predicate device.
## CONCLUSIONS
The clinical performance evaluation and non-clinical testing demonstrate that the 740 SafeSAT patient physiological monitor is substantially equivalent to the predicate device with respect to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.