K193662 · Eezcare Medical Corp · JOW · Feb 11, 2021 · Cardiovascular
Device Facts
Record ID
K193662
Device Name
Ezlymph, Ezlymph M
Applicant
Eezcare Medical Corp
Product Code
JOW · Cardiovascular
Decision Date
Feb 11, 2021
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
EzLymph and EzLymph M are intended to treat lymphedema.
Device Story
EzLymph and EzLymph M are portable, mains-powered intermittent pneumatic compression (IPC) pump systems used in home settings to treat lymphedema. The system consists of a pump, connecting air hoses, and compression garments (arm, full leg, half leg). The pump delivers compressed air to inflate and deflate bladders within the garments, mimicking natural muscle contractions to improve circulation and move excess lymphatic and venous fluid. The device operates via a mechanical pressure-control knob and on/off switch; it contains no microprocessor or control panel. Patients or caregivers operate the device to apply uniform or sequential compression cycles (180 seconds). By stimulating muscles and reducing edema, the device aims to relieve lymphedema discomfort. The system is non-networked and lacks connectivity features.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), usability engineering (IEC 62366-1), risk management (ISO 14971), home healthcare standards (IEC 60601-1-11), and pump/garment pressure cyclic functionality testing.
Technological Characteristics
Mains-powered pneumatic pump; ABS plastic case; biocompatible nylon/TPU garments. Operates via mechanical control (no microprocessor). Provides uniform or sequential compression (30-90 mmHg). Dimensions: 280L x 140W x 70H mm. Weight: 1.7 kg. Standalone device; no connectivity/network capability. IP21 ingress protection. Sterilization not specified; garments are single-patient use.
Indications for Use
Indicated for the treatment of lymphedema in patients requiring prescription-based intermittent pneumatic compression therapy.
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a stacked format.
February 11, 2021
Eezcare Medical Corp % John Gillespy President FDA 510k Consultants, LLC 1100 Del Lago Cir, Ste 104 Palm Beach Gardens, Florida 33410
Re: K193662
Trade/Device Name: Ezlymph, Ezlymph M Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW Dated: January 7, 2021 Received: January 11, 2021
Dear John Gillespy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Fernando Aguel Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193662
Device Name EzLymph and EzLymph M
| Indications for Use (Describe) |
|--------------------------------|
|--------------------------------|
EzLymph and EzLymph M are intended to treat lymphedema.
Type of Use (Select one or both, as applicable):
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5 – 510(k) Summary
| | 1. 510(k) Submitter: | Eezcare Medical Corp<br>No. 3-1, Minquan St., New Taipei City 236<br>Taipei, Tu-Cheng District, Taiwan 23679<br>Phone: 757-224-0177<br>Email: michelle.mitchell@eezcare.com.tw | | |
|--|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|--|
| | 2. Company Contact: | Michelle C. Mitchell | | |
| | 3. Date of Submission: | December 29, 2019 | | |
| | 4. 510(k) Preparer: | John F. Gillespy, MBA<br>FDA 510k Consulting, LLC<br>Palm Beach Gardens, FL 33410<br>Phone: 386-243-4332<br>Email: john@fda510kconsultants.com | | |
| | 5. Device Classification: | Trade name:<br>Common name:<br>Device:<br>Class:<br>Product Code: | EzLymph, EzLymph M<br>Lymphedema Pump System<br>Sleeve, Limb, Compressible<br>II<br>JOW | |
| | 6. Predicate: | Applicant:<br>Device:<br>510(k) Number: | ArjoHuntleigh AB (Sweden)<br>Hydroven 3 Pump, Hydroven FPR<br>Pump, Hydroven Garments<br>K910188 | |
| | Reference: | Applicant:<br>Device:<br>510(k) Number: | Tactile Medical<br>Flexitouch System<br>K170216 | |
- 7. Device Description... EzLymph and EzLymph M is a pump system intended to apply intermittent pneumatic compression (IPC) to arms and legs to relieve the discomfort of lymphedema (see image of device in Figure 1 below).
The device helps move the flow of lymphatic and venous fluid throughout the extremities by stimulating muscles within the affected extremity. This action mimics natural muscle contractions, improves circulation, and moves excess fluid back to the circulatory system so edema can be eliminated from the body.
System components include (a) a mains-powered, portable pneumatic pump with a pressure-control knob and on/off switch (the device contains no control panel or microprocessor circuit); (b) connecting air hoses; and (c) compression garments attached
{4}------------------------------------------------
to patient limbs. In addition, garment inserts allow increase in circumference of arm and leg garments.
The pump intermittently releases compressed air through the tubing into inflating and deflating bladders within cuffs that are wrapped separately around the patient's upper and lower limbs (full arm, full leg, and half leg).
EzLymph and EzLymph M is intended for home use under the care of personnel authorized and trained to treat lymphedema. The device is intended for patients at risk for the medical condition. The system is portable.
- 8. Mechanism of Action... Intermittent pneumatic compression of air inflates and deflates cuffs attached to affected extremities.
- 9. Indications For Use... Intended to treat lymphedema.
The device is intended for prescription use only.
Image /page/4/Picture/6 description: The image shows an EzLymph pump and garments. The pump is white with a blue panel and a timer dial. The garments are gray with orange trim and have tubes attached to them. The garments appear to be for legs.
# 10. Comparison To Predicate & Reference Devices
The subject and predicate devices share the same:
- Intended use – To apply intermittent pneumatic compression to arms and legs to relieve the discomfort of lymphedema.
- Indications for use – To treat lymphedema (the predicate also claims other indications, as noted in Table 5).
- System design - Portable pump, air hoses, compression garments.
- Number of models – Two.
- Garment styles - Arm, leg.
- Compression methods – Uniform, sequential.
- Compression cycle lengths - 180 seconds.
- Air chambers - Single chamber, multiple chamber.
- Multiple garments – Both devices allow treatment with 2 garments.
- Mode of operation - Continuous.
- Energy sources - Mains power.
- Materials – ABS plastic (pump case) and biocompatible fabric (garments).
- Reusability – Reusable pump and air tubing; garments are single-patient use.
{5}------------------------------------------------
Minor differences between the subject and predicate devices include:
- Number of compression chambers – While both offer single- and multiple-chamber models, for the latter the subject provides 4 chambers compared to the predicate's 3 chambers.
- Compression cycle sequence – Both offer 180-second cycles except that the predicate increases cycle length to 120 seconds for cycles that include calf or thigh; since application is continuous over long time periods (days and weeks, not hours or minutes), this difference is immaterial.
The reference device demonstrates substantial equivalence for one physical characteristic, pressure range, for EzLymph.
| Characteristics | Subject Device | Predicate Device¹ | SE Comparison |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Device Name | EzLymph, EzLymph M | Hydroven 3 Pump,<br>Hydroven FPR Pump,<br>Hydroven Garments | NA |
| Manufacturer | Ezcare Medical Corp<br>(Taiwan) | ArjoHuntleigh AB<br>(Sweden) | NA |
| Estab Registration # | 3003801933 | 3007420694 | NA |
| 510k Number | Applied For | K910188 | NA |
| Device Photo (Pump<br>& Air Hoses) | Image: EzLymph pump and air hoses | Image: Hydroven 3 pump and air hoses | SE--both portable pumps |
| Device Photo<br>(Calf/Thigh<br>Garment) | Image: EzLymph calf/thigh garment | Image: Hydroven calf/thigh garment | SE--same basic design |
| Class | II | II | SE |
| Regulation # | 870.5800 | 870.5800 | SE |
| Product Code | JOW | JOW | SE |
| Common Description | Lymphedema Pump | Lymphedema Pump | SE |
| Medical Specialty | Cardiovascular | Cardiovascular | SE |
| Submission Type | 510(k) | 510(k) | SE |
| Intended Use | To apply intermittent<br>pneumatic compression to<br>arms and legs to relieve the<br>discomfort of lymphedema. | To apply intermittent<br>pneumatic compression to<br>arms and legs to treat<br>conditions listed in IFU. | SE--subject intended use is<br>subset of predicate's |
| Indication For Use | EzLymph & EzLymph M are<br>intended to treat<br>lymphedema. | Intermittent Pneumatic<br>Compression (IPC) is effective<br>in treatment of following<br>clinical conditions, when<br>combined with an<br>individualized monitoring<br>program: Edema (dependent<br>& traumatic), Lymphedema,<br>(primary and secondary),<br>Chronic venous insufficiency,<br>Post phlebotic syndrome,<br>and Acute & chronic wounds<br>including venous leg ulcers &<br>post-surgical wounds. IPC<br>may also be beneficial in<br>management of: Fixed flexion<br>deformity, Arthritic<br>conditions, Lower limb pain<br>due to trauma or surgery,<br>and Lipedema. Selection<br>should be based upon a<br>holistic assessment of<br>patients' individual care<br>needs. | SE--subject intended use<br>(and IFU) is subset of<br>predicate's |
| Target Population | Sufferers of lymphedema | Sufferers of lymphedema<br>or other clinical conditions<br>listed in IFU | SE--subject intended use is<br>subset of predicate's |
| Anatomical Site | Arm/Leg | Arm/Leg | SE |
| Where Used | Home | Home | SE |
| Rx/OTC/Both | Rx Only | Rx Only | SE |
| Physical Characteristics -- Overall | | | |
| System Description | Portable pump, connecting<br>air hoses, & compressible<br>limb sleeves | Portable pump,<br>connecting air hoses, &<br>compressible limb sleeves | SE |
| Working Principle | Pneumatic pump supplies<br>compressed air to inflate<br>compression garments<br>attached to patient limbs | Pneumatic pump supplies<br>compressed air to inflate<br>compression garments<br>attached to patient limbs | SE |
| | | | |
| Garment Styles | Full Arm, Half Leg, Full Leg | Full Arm, Half Leg, Full Leg | SE |
| Garment Sizes | EzLymph--Arm (S,M), Leg<br>(Half,S,M,L);<br>EzLymph M--Arm (S,M), Leg<br>(Half,S,M,L) | Hydroven 3-- Single-<br>Chamber Arm (4 lengths),<br>Leg (6 lengths); Triple-<br>Chamber Arm (2 lengths),<br>Leg (6 lengths);<br>Hydroven FPR (all triple-<br>chamber)--Arm (2<br>lengths), Leg (6 lengths). | SE--both offer multiple<br>arm and leg sizes |
| Garment Inserts | Allow increase in<br>circumference of arm and<br>leg garments. | Allow increase in<br>circumference of arm and<br>leg garments. | SE |
| Number of Air<br>Chambers (Bladders) | EzLymph--1 (arm, leg) and 4<br>connected bags (arm, leg);<br>EzLymph M--4 independent<br>bags (arm, leg) | Hydroven 3--1 (arm, leg) &<br>3 connected bags (arm,leg);<br>Hydroven FPR -- 3 indepen-<br>dent bags (arm, leg). | SE--minor difference in<br>number of chambers |
| Compression --<br>Uniform | EzLymph (1 & 4-chamber<br>garments) -- Inflate 90 sec,<br>then deflate 90 sec. Cycle<br>time 180 sec. | Hydroven 3 (1-chamber<br>garments) -- Inflate 90 sec,<br>then deflate 90 sec. Cycle<br>time 180 sec. | SE |
| Compression --<br>Sequential | EzLymph M (4-chamber)--<br>Inflate sequentially from<br>distal (120, 90, 60, & 30 sec<br>respectively), with 10%<br>pressure decrease with<br>each subsequent chamber<br>moving up garment. Then<br>deflate from proximal (60<br>sec). Cycle time 180 sec. | Hydroven 3 & FPR (3-<br>chamber) -- Inflate<br>sequentially from distal<br>(114, 76, & 38 sec). Then<br>deflate from proximal (66<br>sec). Cycle time 180 sec. | SE--minor difference in<br>number of chambers |
| Multiple Garments | 2 garments can be used by<br>alternating for every 90 sec.<br>Cycle time is 180 sec<br>(EzLymph only). | 2 garments can be used by<br>alternating for every 90<br>sec. Cycle time is 180 sec<br>(Hydroven 3 only). | SE |
| Mode of Operation | Continuous | Continuous | SE |
| Pressure Range | EzLymph: 30-75 mmHg +/-<br>10%<br>EzLymph M: 30-90 mmHg<br>+/- 10% | Predicate: 30-100 mmHg<br>+/- 5%<br>Reference: 30-75 mmHg | EzLymph: SE--minor<br>difference w/reference<br>EzLymph M: SE--minor<br>difference w/predicate |
| Physical Characteristics -- Pump | | | |
| Dimensions (mm) | 280L x 140W x 70H | 270L x 150W x 105H<br>(Hydroven 3) or 130H<br>(FPR) | SE--minor difference |
| Weight | 1.7 kg (3.75lb) | 2.4 kg (5.3 lb) | SE--minor difference |
| Energy Source<br>(Mains) | AC 100V-230V (60/50 Hz) | AC 120V (60 Hz) | SE--minor difference |
| Fuse Rating | 1A 250V/T1A 250V | F500 mAH 250V | SE--minor difference |
## Table 5 - Comparison Table
<sup>1</sup> Comparison data for the reference device, Flexitouch (K170216), is shown in "Pressure Range" characteristic on Page 5.0-5.
{6}------------------------------------------------
{7}------------------------------------------------
## Physical Characteristics--Compression Garment
{8}------------------------------------------------
| Ingress Protection | IP21 | IPX0 | SE--minor difference |
|---------------------------------------|----------------------------------|--------------------------|------------------------------------------------|
| Control Panel | Mechanical | Not provided | SE |
| Operating Temp | 5C to 40C | 10C to 40C | SE--minor difference |
| Storage/Transport | -20C to 70C | -20C to 50C | SE--minor difference |
| Relative Humidity | <93% RH non-condensing | 30%-75% (non-condensing) | SE--minor difference |
| Garment Material<br>(Patient Contact) | Nylon fabric with TPU<br>coating | Not provided | SE--subject passed<br>biocompatibility testing |
| Case Material | ABS plastic | ABS plastic | SE |
#### Components & Accessories
| Components | Pump, air hoses, garments | Pump, air hoses, garments | SE |
|-------------|---------------------------|---------------------------|----|
| Accessories | User manual | User manual | SE |
## Non-Clinical Testing
EzLymph and EzLymph M passed the following non-clinical tests, all of which were performed to current FDA-recognized standards:
- Electrical Safety... IEC 60601-1:2005, Mod
- EMC... IEC 60601-1-2:2014
- Usability Engineering... IEC 62366-1:2015
- Risk Management... ISO 14971:2007
- Home Healthcare... IEC 60601-1-11:2015
The device passed pump functionality testing and garments pressure cyclic testing.
## Cybersecurity
EzLymph and EzLymph M is in compliance with FDA's guidance, "Management of Cybersecurity in Medical Devices" (2014). The device is not capable of connecting (wirelessly or hard-wired) to another device, to the Internet or other network, or to portable media (e.g., USB or CD).
#### Substantial Equivalence
EzLymph and EzLymph M successfully followed the pathway to Substantial Equivalence in the FDA guidance document, "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications" (2014). The steps are summarized below:
- The predicate device is legally marketed and was found substantially equivalent through 510(k) premarket submission.
- The subject and predicate devices have the same intended use.
- Technological differences between the subject and predicate were evaluated; none of the differences raised different issues of safety and effectiveness.
- The following methods for evaluation of the effects of different characteristics on safety and effectiveness were deemed acceptable-testing for electrical safety,
{9}------------------------------------------------
EMC, usability engineering, home healthcare, and mechanical performance; and risk management assessment. Evaluation methods were conducted to FDArecognized standards where applicable.
- Data from these tests demonstrated equivalence and support the indications for use.
In summary, all necessary testing has been performed and the results support the conclusion that EzLymph and EzLymph M is substantially equivalent to the legally marketed predicate, Hydroven, based on both (a) comparison of intended use, materials, technology, and design and (b) testing to FDA-recognized standards, and the device thus does not raise any concerns of safety or effectiveness.
Based on the information contained within this submission, the applicant concludes that EzLymph and EzLymph M is substantially equivalent to the identified predicate device and warrants clearance for marketing activities.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.