K170216 · Tactile Systems Technology, Inc. (Dba Tactile Medical) · JOW · May 26, 2017 · Cardiovascular
Device Facts
Record ID
K170216
Device Name
Flexitouch System
Applicant
Tactile Systems Technology, Inc. (Dba Tactile Medical)
Product Code
JOW · Cardiovascular
Decision Date
May 26, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5800
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Flexitouch System and garments for legs, arms, trunk, and chest are intended for use by medical professionals and patients who are under medical supervision, for the treatment of many conditions such as: - Lymphedema - Primary lymphedema - Post mastectomy edema - Edema following trauma and sports injuries - Post immobilization edema - Venous insufficiency - Reducing wound healing time - Treatment and assistance in healing stasis dermatitis, venous stasis ulcers, or arterial and diabetic leg ulcers The Flexitouch System and garments for the head and neck are intended for use by medical professionals and patients who are under medical supervision for the treatment of head and neck lymphedema.
Device Story
Flexitouch System PD32-G3 is a pneumatic sequential compression device for lymphedema and edema management. System comprises a pneumatic controller and multi-chambered garments (legs, arms, trunk, chest, head, neck). Controller inflates up to 32 garment chambers via tubing harness to deliver sequential calibrated gradient pressure. Used in clinical or home settings under medical supervision. Device utilizes pressure-sensing technology to control chamber inflation, replacing time-based control. Healthcare providers prescribe therapy; patients operate device to perform treatment. Sequential compression assists lymphatic drainage and wound healing. Benefits include reduced edema and improved healing outcomes for chronic venous/arterial conditions.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical performance testing, including electrical safety (IEC 60601-1, IEC 60601-1-2), software verification/validation, biocompatibility (ISO 10993), shipping (ASTM D4169), and pressure verification testing.
Technological Characteristics
Pneumatic sequential compression controller with up to 32 chambers. Enclosure and manifold constructed of plastic. Operates on 100–240 VAC 50/60 Hz. Features pressure-sensing control and full-color LCD interface. Microprocessor-based digital/analog electronics. Biocompatible materials per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, ASTM D4169.
Indications for Use
Indicated for medical professionals and patients under medical supervision for treatment of lymphedema (primary, post-mastectomy, head/neck), edema (post-trauma, sports injury, post-immobilization), venous insufficiency, and wound healing (stasis dermatitis, venous stasis ulcers, arterial/diabetic leg ulcers).
Regulatory Classification
Identification
A compressible limb sleeve is a device that is used to prevent pooling of blood in a limb by inflating periodically a sleeve around the limb.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
May 26, 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Tactile Systems Technology, Inc. (dba Tactile Medical) Thomas Dold Vice President. Quality and Regulatory Affairs 1331 Tyler Street NE. Suite 200 Minneapolis, Minnesota 55413
Re: K170216
Trade/Device Name: Flexitouch® System, Model PD32-G3 Regulation Number: 21 CFR 870.5800 Regulation Name: Compressible Limb Sleeve Regulatory Class: Class II Product Code: JOW, PPS Dated: April 21, 2017 Received: April 25, 2017
Dear Thomas Dold:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Fernando Aguel
Fernando Aguel
-s
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170216
Device Name
Flexitouch System, Model PD32-G3
#### Indications for Use (Describe)
The Flexitouch System and garments for legs, arms, trunk, and chest are intended for use by medical professionals and patients who are under medical supervision, for the treatment of many conditions such as:
- Lymphedema
- Primary lymphedema
- Post mastectomy edema
- · Edema following trauma and sports injuries
- · Post immobilization edema
- · Venous insufficiency
- · Reducing wound healing time
· Treatment and assistance in healing stasis dermatitis, venous stasis ulcers, or arterial and diabetic leg ulcers
The Flexitouch System and garments for the head and neck are intended for use by medical professionals and patients who are under medical supervision for the treatment of head and neck lymphedema.
Type of Use (Select one or both, as applicable)
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------|
| <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Tactile Medical. The logo features a stylized letter "T" in blue and gray, with the top of the "T" forming a roof-like shape. To the right of the "T" is the word "Tactile" in blue, with the letters in a bold, sans-serif font. Below "Tactile" is the word "MEDICAL" in gray, with the letters in a smaller, sans-serif font. A blue line underlines the word "MEDICAL".
# Appendix A 510(k) Summary
| Date Prepared: | May 25, 2017 |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Thomas Dold<br>Vice President, Quality and Regulatory Affairs<br>Tactile Systems Technology, Inc. (DBA Tactile Medical)<br>1331 Tyler Street NE, Suite 200<br>Minneapolis, Mn 55413<br>612-355-5100 |
| Contact Person: | Thomas Dold |
| Name of Device: | Flexitouch® System PD32-G3 |
| Classification: | Sleeve, Limb, Compressible (21 CFR 870.5800)<br>Sleeve, Head, and Neck, Compressible (21 CFR 870.5800) |
| Product Code(s): | JOW, PPS |
| Predicate Device: | Tactile Systems Technology, Inc. Flexitouch® System (K153311) |
| Reference Device: | Entré PD08-U (K143185) |
## Description of Device:
The Flexitouch system consists of two main components: a garment set and a pneumatic sequential controller. The garments are wrapped around affected region so that the garment fits snugly. The garments have multiple chambers that are filled with air to provide pneumatic compression therapy. Up to four (4) tubing harness assemblies containing eight (8) discrete individual air passage tubes connect individual garment chambers with the controller. Air passes through the tubes, delivering treatment via the sequential inflation of up to 32 air chambers in the garments.
## Indications for Use:
The Flexitouch® system and garments for legs, arms, trunk and chest are intended for use by medical professionals and patients who are under medical supervision, for the treatment of many conditions such as:
- Lymphedema
- Primary lymphedema ●
- Post mastectomy edema
- Edema following trauma and sports issues ●
- Post immobilization edema
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Tactile Medical. The logo features a stylized letter "T" with a gray vertical bar on the left and a blue triangular roof on top. To the right of the "T" is the word "Tactile" in blue, with the letters slightly overlapping. Below "Tactile" is the word "MEDICAL" in gray, with a thin blue line underneath.
- Venous insufficiency
- Reducing wound healing time ●
- Treatment and assistance in healing stasis dermatitis, venous stasis ulcers, arterial ulcers ● and diabetic leg ulcers
The Flexitouch System and garments for the head and neck are intended for use by medical professionals and patients who are under medical supervision for the treatment of head and neck lymphedema.
## Comparison to Technology Characteristics:
In addition to having the same indications for use, the Flexitouch® PD32-G3 has the same technology characteristics to the predicate devices with respect to design features, and operating principal as demonstrated in the table below.
The key differences between this device and the predicate Flexitouch® System PD32-U, include the use of pressure sensing technology rather than time based pressure control and a full color LCD user interface.
| Device<br>Characteristic | REFERENCE<br>Tactile Systems<br>Technology, Inc.<br>Entré PD08-U<br>(K143185) | PREDICATE<br>Tactile Systems Technology,<br>Inc. Flexitouch System®<br>PD32-U (K153311) | Tactile Systems Technology,<br>Inc. Flexitouch® System<br>PD32-G3<br>(New Device K170216) | Difference |
|-------------------------------------------|-------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|----------------|
| Electrical<br>Requirements | 100 VAC-264 VAC ~<br>50/60 Hz 14.4W | 100-240 VAC, 1.0 A, 50/ 60 Hz | 100–240 VAC 50/60 Hz | Same as PD32-U |
| Enclosure<br>material | All plastic construction | All plastic construction | All plastic construction | Same |
| Manifold<br>Assembly | Plastic construction with<br>integrated air channel | Plastic construction with<br>integrated air channel | Plastic construction with<br>integrated air channel | Same |
| User Interface | Tactile Pushbuttons | Tactile Pushbuttons | Tactile Pushbuttons | Same |
| Software/<br>Hardware | Analog and digital<br>electronic with<br>microprocessor | Analog and digital electronic with<br>microprocessor | Analog and digital electronic with<br>microprocessor | Same |
| Chambers | 8 Chambers | Up to 32 depending on garment<br>size and types | Up to 32 depending on garment<br>size and types | Same as PD32-U |
| Output | Sequential gradient<br>pressure | Sequential calibrated gradient<br>pressure | Sequential calibrated gradient<br>pressure | Same as PD32-U |
| Garment<br>Chamber<br>Pressure<br>Control | Pressure Based Control | Time Based Control | Pressure Based Control | Same as PD08-U |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Tactile Medical. The logo features a stylized letter "T" that is partially gray and partially blue, with a blue triangle on top, resembling a house. To the right of the "T" is the word "Tactile" in blue, with the letters "MEDICAL" in gray underneath. There is a horizontal gray line under the word "MEDICAL".
## Summary of Tests
Comparative performance testing was used to verify the equivalence between the Flexitouch PD32-G3, which is the subject of this submission and the predicate devices. The test results verify that the device met all pre-specified performance criteria and is substantially equivalent to the predicate devices for its intended use. Clinical testing was not required to establish substantial equivalence with the predicate devices.
The following non-clinical performance testing was performed in support of the substantial equivalence determination.
- Electrical safety and performance testing were according to standard IEC 60601-1: 2005 and standard IEC 60601-1-2:2014.
- Software verification and validation testing were conducted and documentation was provided ● as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
- Biocompatibility Testing - Cytotoxicity, Sensitization, Irritation per ISO 10993.
- Transportation Shipping Testing per ASTM D4169
- Device Storage Condition and Operational Condition Testing
- Pressure Verification Testing ●
## Conclusion
The data included in this submission demonstrates that the Flexitouch® PD32-G3 is substantially equivalent to the legally marketed predicate device, Flexitouch® PD32-U (K153311).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.