Disposable Enteral Feeding Syringe with Enfit Connector
Applicant
Shinva Ande Healthcare Apparatus Co., Ltd.
Product Code
PNR · Gastroenterology, Urology
Decision Date
Sep 12, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Disposable Enteral Feeding Syringe with Enfit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral feeding syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
Device Story
Disposable enteral feeding syringe; sizes 1ml-60ml; manual operation. Components: barrel with ENFit connector, piston (plunger and plunger stopper), lubricant, optional tip cap. Piston made of polyisoprene synthetic rubber (DEHP/latex-free). Device uses interference fit between piston and barrel for fluid sealing; manual pushing/pulling of plunger enables fluid suction and infusion. Used in clinical or home settings by clinicians or laypersons (under supervision) to deliver fluids into the gastrointestinal system. Benefits patients unable to eat or swallow by providing a standardized, secure fluid transfer mechanism via ISO 80369-3 compliant connectors.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench performance testing, including biocompatibility (cytotoxicity, irritation, sensitization, acute systemic toxicity), sterilization validation (ISO 11135), shelf-life/accelerated aging, and mechanical testing per ISO 80369-3 and ISO 7886-1.
Indicated for patients of all age groups who are physically unable to eat and swallow, requiring delivery of fluids into the gastrointestinal system.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 12, 2020
Shinva Ande Healthcare Apparatus Co., Ltd. % Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. 8-9th Floor, R&D Building, No. 26 Quinglan Street, Panyu District Guangzhou, 510006 CHINA
Re: K193657
> Trade/Device Name: Disposable Enteral Feeding Syringe with Enfit Connector Classification Number: 21 CFR 876.5980 Classification Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PNR Dated: August 5, 2020 Received: August 10, 2020
Dear Mike Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K193657
### Device Name
Disposable Enteral Feeding Syringe with Enfit Connector
Indications for Use (Describe)
The Disposable Enteral Feeding Syringe with Enfit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral feeding syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a stylized letter 'A' in black against a white background. The 'A' is designed with a curved top, resembling a swooping shape, and a solid, bold base. A small circle with a number '1' inside is positioned in the upper right corner of the 'A', possibly indicating a trademark or design element.
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
#### SUBMITTER l.
SHINVA ANDE Healthcare Apparatus Co., Ltd.
No.77 Development Zone North Road, Zibo, Shandong, 255086, China Phone: +86-533-3917821
Fax: +86-533- 3918218
Primary Contact Person: Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technical Service Co., Ltd. Tel: (+86)-20-6231 6262 Fax: (+86) -20-8633 0253
| Secondary Contact Person: Xiaolei Tian |
|-------------------------------------------|
| R&D Director |
| SHINVA ANDE Healthcare Apparatus Co., Ltd |
| Tel: (+86)-533-3917821 |
| Fax: (+86)- 533- 3918218 |
Sep 12, 2020 Date prepared
### II. DEVICE
| Name of Device: | Disposable Enteral Feeding Syringe with Enfit<br>Connector |
|-----------------|-------------------------------------------------------------|
| Common Name: | Disposable Enteral Feeding Syringe with Enfit<br>Connector |
| Model/size: | 1ml, 2 ml,2.5 ml, 3ml, 5ml, 10ml, 20ml, 30ml, 50ml,<br>60ml |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a stylized, black-and-white logo. The logo features a curved, abstract shape resembling a stylized letter 'A' or a swooping bird. In the upper right corner of the shape, there is a small, filled circle with a white number '1' inside, suggesting a trademark or branding element.
| Classification Number: | 21 CFR 876.5980 |
|------------------------|---------------------------------------|
| Classification Name: | Gastrointestinal tube and accessories |
| Device Panel | Gastroenterology/Urology |
| Regulatory Class: | II |
| Product Code: | PNR |
## III. PREDICATE DEVICE
Predicate device K161979, ENFit Enteral Syringe This predicate has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
The proposed device is a disposable enteral feeding syringe provided in a variety of size from 1ml-60ml (1ml, 2ml, 2.5ml, 3ml, 5ml, 10ml, 30ml, 50ml and 60ml). The device is composed by barrel with ENFit connector, piston (plunger and plunger stopper), lubricant and tip cap (with or without). The piston is made of polyisoprene synthetic rubber, without DEHP or natural latex. This device incorporates a female ENFit connector for connection to an enteral access device with a male ENFit connector.
They are two types of syringe:
Side connector ENFit syringe and central connector ENFit syringe. The size of the central connector ENFit syringes range from 1ml to 5ml, and they are the low dose tip syringes; and side connector ENFit syringes range from 10ml to 60ml, and they are the standard syringes.
The Disposable Enteral Feeding Syringe with Enfit Connector relies on the interference matching between piston and inner chamber wall of the barrel to ensure the sealing of liquid in the inner chamber of the barrel. The plunger and the plunger stopper constitute a sealing assembly. By pushing and pulling the plunger, the plunger stopper
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a stylized letter 'A' in black, with a curved shape above it that resembles a swoosh or an arc. There is a small circle with a number '1' inside it, positioned in the upper right corner of the swoosh. The overall design is simple and modern, suggesting a logo or symbol for a brand or organization.
can be driven to move freely inside the barrel, similar to the piston structure, so as to realize the suction and infusion of fluid.
The proposed syringe is sterile. Sterile device was sterilized by Ethylene Oxide Gas to achieve a SAL of 10 ° and supplied sterility maintenance package which could maintain the sterility of the device during the shelf of five years.
#### INDICATION FOR USE V.
The Disposable Enteral Feeding Syringe with Enfit Connector is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral feeding syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Disposable Enteral Feeding Syringe with Enfit Connector has the same technological characteristics and fundamental design as the predicate devices. The Disposable Enteral Feeding Syringe with Enfit Connector and the predicate devices are all designed to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow.
The Disposable Enteral Feeding Syringe with Enfit Connector has same indications for use, patient population with the predicates.
At a high level, the subject and predicate device are based on the following same technological elements:
- . They are all disposable for single use.
- . Their operation mode are all manual use.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a stylized letter 'A' in black, with a curved design. Above and to the right of the 'A', there is a small circle with a number '1' inside, likely indicating a registered trademark symbol. The logo appears simple and modern.
- They all have the same configuration that composed by barrel with ENFit connector, piston (plunger and plunger stopper), lubricant and tip cap.
- . Their connection type is the small-bore connector featuring ISO 80369-3:2016/AMD 1:2019, Small-Bore Connectors for Liquids and Gases in Healthcare Applications - Part 3: Connectors For Enteral Applications.
The subject device has different sizes compared with the predicate device in low dose and standard dose type. This difference does not raise concerns of safety and effectiveness.
Therefore, the Disposable Enteral Feeding Syringe with Enfit Connector are substantially equivalent to the cleared predicate device (K161979).
### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
Biocompatibility of the Disposable Enteral Feeding Syringe with Enfit Connector was evaluated in accordance with the FDA Bguidance "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing," and ISO 10993-1:2009 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. The following tests were performed, as recommended:
- . Cytotoxicity
- . Irritation
- Sensitization
- Acute Systemic Toxicity
The Disposable Enteral Feeding Syringe with Enfit Connector contacts indirectly with the human body for a duration of less than 24 hours.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo that appears to be a stylized letter 'A' or 'D' in a bold, sans-serif font. The letter is black and has a dynamic, forward-leaning appearance, suggesting movement or progress. There is a small circle with a number '1' inside it, positioned in the upper right corner of the logo, possibly indicating a trademark or a version number. The logo is set against a white background, and there is a thin black line at the bottom of the image, possibly indicating the edge of a document or a border.
## Sterilization and shelf life testing
The sterilization method has been validated according to ISO 11135:2014 Sterilization Of Health-Care Products - Ethylene Oxide - Requirements For The Development, Validation And Routine Control Of A Sterilization Process For Medical Devices, which has thereby determined the routine control and monitoring parameters. The shelf life of the Disposable Enteral Feeding Syringe with Enfit Connector is determined based on the accelerated aging test. And after the accelerated aging test, the performance test and package integrity test performed on the proposed device.
### Bench performance testing
Connector testing performed on the proposed device, in accordance with ISO 80369-3:2016/AMD 1:2019, Small-bore connectors for liquids and gases in healthcare applications- Part 3: Connectors for enteral applications, using the test methods provided in ISO 80369-20:2015, Small-bore connectors for liquids and gases in healthcare applications- Part 20: Common test methods. The testing demonstrates the proposed devices conform to the requirements of ISO 80369-3:2016/AMD 1:2019. Syringe testing performed on the proposed device in accordance with ISO 7886-1:2017, sterile hypodermic syringes for single use — Part 1: Syringes for manual use. The testing demonstrates the proposed device conform to the requirements of ISO 7886-1: 2017.
Aging performance and sterile packaging integrity tests were also performed.
### Clinical Testing
Based on the similarities of the device specifications, intended use, indications for use between the proposed device Disposable Enteral Feeding Syringe and its predicate devices, no clinical studies were needed to support this 510(k) Premarket Notification.
### VIII. CONCLUSION
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a stylized letter 'A' in black. The 'A' is designed with a curved top, resembling a swoosh or an arc. There is a small circle with a number '1' inside it, positioned at the upper right of the 'A' symbol. The logo has a clean and modern appearance.
The proposed device Disposable Enteral Feeding Syringe with Enfit Connector is substantially equivalent to its predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.