ENFit enteral syringe

K161979 · Jiangyin Caina Technology Co., Ltd. · PNR · Mar 22, 2017 · Gastroenterology, Urology

Device Facts

Record IDK161979
Device NameENFit enteral syringe
ApplicantJiangyin Caina Technology Co., Ltd.
Product CodePNR · Gastroenterology, Urology
Decision DateMar 22, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.5980
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

The proposed device is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups.

Device Story

Disposable enteral feeding syringe; sizes 1ml-60ml; incorporates female ENFit connector for interface with enteral access devices. Used in clinical or home settings by clinicians or laypersons to measure and dispense fluids into patient gastrointestinal systems. Manual operation; provides fluid delivery for patients unable to swallow. Sterilized via ethylene oxide; single-use. Benefits include standardized connection to prevent misconnections and accurate fluid administration.

Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993), connector performance (ISO 80369-3/20), syringe performance (ISO 7886-1), seal strength (ASTM F88), and package integrity (ASTM F1140).

Technological Characteristics

Disposable enteral syringe; sizes 1ml-60ml; female ENFit connector per ISO 80369-3:2016. Sterilized via ethylene oxide (SAL 10^-6). Materials meet ISO 10993 biocompatibility standards. Manual operation; no software or electronic components.

Indications for Use

Indicated for patients of all age groups physically unable to eat and swallow, requiring delivery of fluids into the gastrointestinal system. Intended for use by clinicians or laypersons under clinician supervision in clinical or home settings.

Regulatory Classification

Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized symbol that resembles three human profiles facing to the right, with flowing lines beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 May 5, 2017 Jiangsu Caina Medical Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China Re: K161979 Trade/Device Name: ENFit Enteral Syringe Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal Tubes and Accessories Regulatory Class: II Product Code: PNR Dated: February 15, 2017 Received: February 21, 2017 Dear Diana Hong: This letter corrects our substantially equivalent letter of March 22, 2017. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely. # Benjamin R. Fisher -S Beniamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161979 Device Name ENFit Enteral Syringe Indications for Use (Describe) The proposed device is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995 ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## Tab #7 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: ________ - 1. Date of Preparation: 03/17/2017 - Sponsor Identification 2. Jiangsu Caina Medical Co., Ltd. No.23, Huanxi Road, Zhutang Town, Jiangyin, Jiangsu, 214425, China Establishment Registration Number: 3005670221 Contact Person: Jun Lu Position: General Manager Tel: +86-510-86205183 Fax: +86-510-86215183 Email: jun.lu@cainamedical.com - 3. Designated Submission Correspondent Ms. Diana Hong (Primary Contact Person) Ms. Jing Cheng (Alternative Contact Person) ## Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net {4}------------------------------------------------ #### 4. Identification of Proposed Device Trade Name: ENFit Enteral syringe Common Name: Enteral Feeding Syringe with ENFit Connector Size: 1ml, 3ml, 5ml, 6ml, 10ml, 12ml, 20ml, 30ml, 35ml, 50ml, 60ml #### Regulatory Information Classification Name: Gastrointestinal tube and accessories; Classification: II; Product Code: PNR Regulation Number: CFR 876.5980; Review Panel: Gastroenterology/Urology; #### Indications for Use: ENFit Enteral Syringe is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups. #### Device Description The proposed device is a disposable enteral feeding syringe provided in a variety of sizes from 1ml~60ml. This device incorporates a female ENFit connection to an enteral access device with male ENFit connector. The propsoed syringe is sterilized by Ethylene Oxide Gas to achieve a SAL of 10 and supplied sterility maintenance package which could maintain the sterility of the device during the shelf life of five years. #### 5. Identification of Predicate Device Predicate device 510(k) Number: K161039 Manufacturer: NeoMed, Inc. Product Name: NeoConnector Oral/Enternal Feeding Syringes with ENFit Connector (12ml to 100ml) NeoConnector Low Dose Tip Oral/Enternal Feeding Syringes with ENFit Connector (0.5ml to 6ml) #### Non-Clinical Test Conclusion 6. Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: {5}------------------------------------------------ - > AAMI/CN3 (PS):2014, Small-bore connectors for liquids and gases in healthcare application-Part 3: Connectors for enteral applications; - > ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare application-Part 3: Connectors for enteral applications; - > ISO 80369-20:2015, Small-bore connectors for liquids and gases in healthcare application-Part 20: Common test methods: - > ISO 10993-5:2009, Biological Evaluation of Medical Device, Part 5-Tests for Vitro Cytotoxicity; - > ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for irritation and skin sensitization; - > ISO 10993-7:2008, Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals; - > ASTM F 88/F88M-09, Standard test method for seal strength of flexible barrier materials; - > ASTM F1140/F1140-13, Standard test methods for internal pressurization failure resistance of unrestrained packages - > USP37-NF32 <85> Bacterial Endotoxins Limit. - > ISO 7886-1:1993 Sterile hypodermic syringe for syring use-Part 1: Syringe for manual use Biocompatibility testing has demonstrated the biological safety of the proposed devices which may indirectly contact the patients. Product performance after Accelerated Aging test and Package Integrity after Accelerated Aging test evaluated the properties of the enteral feeding syringes afteraccelerated aging in support of the labeling. Connector testing performed on the proposed device included the items listed below, in accordance with ISO 80369-3:2016, Small-bore connectors for liquids and gases in healthcare applications -Part 3: Connectors for enteral applications, using the test methods provided in ISO 80369-20, Small-bore connectors for liquids and gases in healthcare applications – Part 20: Common test methods. The testing demonstrates the proposed devices conform to the requirements of ISO 80369-3:2016. | Individual test Defined in ISO 80369-3 | Requirement Defined in<br>AAMI/CN3 (PS):2014 | Test Method Defined<br>in ISO 80369-20 | |------------------------------------------|----------------------------------------------|----------------------------------------| | Fluid Leakage | Clause 6.2 | Annex C | | Stress Cracking | Clause 6.3 | Annex E | | Resistance to separation from axial load | Clause 6.4 | Annex F | | Resistance to separation from unscrewing | Clause 6.5 | Annex G | | Resistance to overriding | Clause 6.6 | Annex H | | Disconnection by unscrewing | Clause 6.7 | Annex I | {6}------------------------------------------------ Syringe testing performed on the proposed device included the items listed below, in accordance with ISO 7886-1:1993, Sterile hypodermic syringe for syring use-Part 1: Syringe for manual use, using the test methods provided in ISO 7886-1:1993. The testing demonstrates the proposed devices conform to the requirements of ISO 7886-1:1993. | Individual test Defined in ISO 7886-1:1993 | Requirement Defined in ISO 7886-1:1993 | |-------------------------------------------------|----------------------------------------| | Cleanliness | Clause 5 | | Limits for acidity or alkalinity | Clause 6 | | Limits for extractable metals | Clause 7 | | Lubricant | Clause 8 | | Tolerance on graduated capacity | Clause 9 | | Graduated scale | Clause 10 | | Barrel | Clause 11 | | Piston/plunger assembly | Clause 12 | | Dead space | Clause 14.1 | | Freedom from air and liquid leakage past piston | Clause 14.2 | - Substantially Equivalent (SE) Comparison 7. | Table 1 Comparison of Technology Characteristics | | | |--------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Proposed Device | Predicate Device | | Product Code | PNR | K161039<br>PNR | | Regulation<br>Number | CFR 876.5980; | CFR 876.5980; | | Intended Use | The proposed device is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the gastrointestinal system of a patient who is physically unable to eat and swallow. The enteral syringes are intended to be used in clinical or home care setting by users ranging from laypersons (under the supervision of a clinician) to clinicians, in all age groups. | The device is indicated for use as a dispenser, a measuring device, and a fluid transfer device. It is used to deliver fluids into the body orally or enterally. It is intended to be used in clinical or home care setting by users ranging from clinicians to laypersons (under the supervision of a clinician) in all age groups. | | Configuration | Barrel with ENFit connector | Barrel with ENFit connector | | | Purple Plunger | Purple Plunger | | | Piston | Piston | | | Tip cap | Tip cap | | Size | Low dose tip ENFit syringe 1ml, 3ml | Low dose tip ENFit syringe 0.5ml-6ml | | | Standard ENFit syringe 5ml-60ml | Standard ENFit syringe 12ml-100ml | | Sterile | Yes | Yes | | Single Use | Yes | Yes | ### Table 1 Comparison of Technology Characteristics {7}------------------------------------------------ | Biocompatibility | Conforms to the requirement of ISO 10993 series Standards | Conforms to the requirement of ISO 10993 series Standards | |------------------|-----------------------------------------------------------|-----------------------------------------------------------| | | | | | | No Cytoxicity | No Cytoxicity | | | No Irritation to Skin | No Irritation to Skin | | | No significant evidence of sensitization | No significant evidence of sensitization | #### Product Comparison Summary The proposed device and predicate device are intended for patients who require enteral nutrition due to illness or injury which prevents normal chewing and swallowing. These products are enteral syringes that have the same smalll-bore connector, similar intended use and product specification The difference between the proposed device and predicate device is the size of the syringe; both the proposed device and the predicate device comply with the relevant international Standards. Therefore the differences on syringe size will not raise new problems on safety and effectiveness of the proposed device. #### 8. Substantially Equivalent (SE) Conclusion Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...