The EndoGI Biliary Stent System is intended for delivery of stent(s) to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Device Story
EndoGI Biliary Stent System consists of delivery catheter preloaded with two straight, radiopaque plastic stents featuring 65° duodenal bends. Used by physicians in clinical settings to treat biliary strictures or stones. Device utilizes combination of rapid-exchange and over-the-wire techniques to deploy two stents in single procedure, reducing need for multiple delivery system insertions. Safety lock mechanisms integrated into handle. Stents provide drainage, splinting, or patency. Output is physical placement of stents within biliary tract. Benefits include procedural efficiency by enabling dual-stent placement via single passage over guidewire.
Clinical Evidence
No human clinical data. Evidence consists of bench testing (kink resistance, flow rate, bond strength, dimensional verification, compression/tensile testing) and GLP-like porcine animal study (n=6). Animal study validated successful deployment of two stents and retrieval in single procedure; average procedure time 20 minutes; no local tissue reactions observed.
Technological Characteristics
10 Fr delivery system; two 110 mm radiopaque plastic stents; compatible with 0.035" guidewires. Materials include various polymers and stainless steel. Sterilization via Ethylene Oxide (EtO). Single-use. Mechanical features include safety lock handle and dual-stent preloading capability.
Indications for Use
Indicated for patients requiring biliary tract drainage, bile duct splinting during healing, or maintenance of bile duct patency in cases of stricture or stone presence.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
AdvaniX™ Biliary Stent with NaviFleX™ RX Delivery System (K101314)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
EndoGI Medical, Ltd. % Bosmat Friedman Regulatory Consultant ProMedoss, Inc 3521 Hatwynn Rd. Charlotte, North Carolina 28269
Re: K193600
Trade/Device Name: EndoGI Biliary Stent System Regulation Number: 21 CFR 876.5010 Regulation Name: Biliary Catheter And Accessories Regulatory Class: Class II Product Code: FGE Dated: May 27, 2020 Received: May 28, 2020
## Dear Bosmat Friedman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Daniel G. Walter, Jr. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193600
Device Name EndoGI Biliary Stent System
Indications for Use (Describe)
The EndoGI Biliary Stent System is intended for delivery of stent(s) to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 1 SUBMITTER
Applicant's Name:
EndoGI Medical Omri Naveh, CEO +972-52-4625474 +972-4-6098600 omri@EndoGI-Medical.com
### Contact Person:
Bosmat Friedman Regulatory Affairs Consultant 3521 Hatwynn Rd. Charlotte, NC 28269 Phone: 647-975-3974 bosmat.f@promedoss.com
## Date Prepared:
December 20, 2019
#### 2 DEVICE
Trade Name: EndoGI Biliary Stent System
| Classification Code: | Name: Catheter, Biliary, Diagnostic |
|----------------------|----------------------------------------|
| | Product Code: FGE |
| | Regulation No: 876.5010 |
| | Class: 2 |
| | Review Panel: Gastroenterology/Urology |
#### 3 PREDICATE DEVICE
Primary predicate device:
- . AdvaniX™ Biliary Stent with NaviFleX™ RX Delivery System, manufactured by Boston Scientific, cleared under K101314; Product Code: FGE.
Reference device:
- Marathon Stent with Fusion Technology, manufactured by Cook Ireland, cleared under K060624; Product code: FGE.
#### 4 DEVICE DESCRIPTION
The EndoGI System incorporates two preloaded stents. Both stents are identical in design with a straight body and one end incorporating a 65° bend (duodenal bend). The stents are comprised of a biocompatible radiopaque plastic which allows visualization of the stents post deployment. Both stents are pre-mounted on the delivery system and are delivered via near-the-wire and over-the-wire methods.
{4}------------------------------------------------
The EndoGI System is available 10 Fr and includes two 110 mm stents. The system is compatible with a guidewire of up to 0.035". The safety lock mechanisms are incorporated within the delivery device handle.
#### 5 INDICATIONS FOR USE
The EndoGI Biliary Stent System is intended for delivery of stent(s) to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
#### SUBSTANTIAL EQUIVALENCE 6
The EndoGI Biliary Stent System is substantially equivalent to the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System. Both devices have similar indications for use; the main difference is that the EndoGI System allows the insertion of two stents on one delivery device while the predicate requires two separate delivery device insertions in order to achieve the same results.
Bothe devices incorporate a combination of rapid exchange and over-the-wire techniques to deliver the stent(s). Comparative bench testing demonstrated that the stent components of both systems are substantially equivalent. Additional bench and animal testing on the EndoGI delivery device demonstrated that the device performs as intended and that no new safety and effectiveness concerns have been introduced due to the ability of the system to deliver two stents.
| Feature | EndoGI System | Advanix™ (K101314) | Comparison to Primary<br>Predicate |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Reg. Number | 876.5010 | 876.5010 | Same |
| Product Code | FGE | FGE | Same |
| Indication for<br>Use | The EndoGI is intended<br>for delivery of stent(s)<br>to the biliary tract for<br>drainage of the bile<br>ducts, for splinting of a<br>bile duct during healing,<br>or for providing bile<br>duct patency in a<br>stricture or past a stone. | Advanix™ Biliary Stent with<br>NaviFlex™ RX Delivery<br>System is intended for<br>delivery of the stent to the<br>biliary tract for drainage of<br>the bile ducts, for splinting of<br>a bile duct during healing, or<br>for providing bile duct<br>patency in a stricture or past a<br>stone. | Similar-with the addition<br>of the ability to place two<br>stents with one passage<br>over the guide wire |
| Principle of<br>Operation | Combination of rapid<br>exchange and over-the-<br>wire techniques to<br>deliver and place two<br>stents | Combination of rapid<br>exchange and over-the-wire<br>techniques to deliver a single<br>stent | Similar; The performance<br>of the EndoGI System was<br>evaluated via bench and<br>animal testing, the results<br>of which support our<br>substantial equivalency<br>claim |
{5}------------------------------------------------
| Feature | EndoGI System | Advanix™<br>(K101314) | Comparison to Primary<br>Predicate |
|-----------------------|---------------------------------------------------------|-----------------------------------------|----------------------------------------------------------------------------------------------|
| Material | Various polymers and<br>stainless steel | Various polymers and<br>stainless steel | Similar; delivery system<br>and stent successfully<br>passed all biocompatibility<br>testing |
| Single use | Yes | Yes | Same |
| Sterility | EtO | EtO | Same |
| Mechanical Properties | | | |
| Tensile test | Obtained results supports substantial equivalency claim | | Similar |
| Compression test | Obtained results supports substantial equivalency claim | | Similar |
| Flow Rate | Obtained results supports substantial equivalency claim | | Similar |
#### 7 PERFORMANCE DATA
## Biocompatibility:
The following biocompatibility tests were performed on the stent and delivery system: Stent:
- Systemic Toxicity (Implant Method) and Implant Evaluation Test in Rabbits -
- -Systemic Toxicity (Dual Route Repeated Exposure Method) Test
- Material Mediated Pyrogenicity Test (ISO/USP) -
- Acute Systemic Toxicity Test -
- -Guinea Pig Maximization Test
- -Intracutaneous Reactivity Test
- -Mouse Lymphoma Assay
- Ames Bacterial Reverse Mutation Assay -
- MEM Elution Cytotoxicity Assay -
## Delivery System:
For the delivery device:
- MEM Elution Cytotoxicity Assay -
- -Material Mediated Pyrogenicity Test
- -Acute Systemic Toxicity Test
- -Guinea Pig Maximization Test
- Intracutaneous Reactivity Test -
## Non-Clinical Performance Testing:
The EndoGI Biliary Stent System device has undergone and successfully passed the following tests:
- Mechanical Evaluation -
- -Kink Resistance
- -Flow Rate
- -Bond Strength
- -Dimensional Verification
{6}------------------------------------------------
- EndoGI Biliary Stent Compression Force (Bile) -
- -EndoGI Biliary Stent Tensile Test (Bile)
### Animal Performance Testing:
A GLP-like animal study on porcine model was conducted to evaluate the performance of the system and to validate its ability to successfully deploy two stents. A total of 6 repeats including 2 deployments and 6 stent retrievals were performed. The duration of the complete procedure (from delivery system insertion to the beginning of monitoring following retrieval) was 17 minutes - 29 minutes with an average of 20 minutes per repeat. The second stent deployment following the first stent was performed within 1-2 minutes with an average of 1 minute for second stent deployment.
The goals of the study were met: namely, the system was able to successfully deliver 2 stents utilizing a one-step procedure, no local tissue reaction was visualized, and successful stent removal was achieved.
#### 8 CONCLUSION
The EndoGI System has similar indications for use as the main predicate, the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System. Although the indications for use statement of the Advanix addresses the placement of only one stent, the product labeling clearly demonstrates that multiple stent placement with the Advanix is actively promoted. Furthermore, multiple stent placement is common practice and therefore the ability of the EndoGI System to allow for the placement of two stents does not raise new safety or effectiveness concerns. The main technological difference between the EndoGI delivery system and the predicates is due to the systems' ability to incorporate two preloaded stents in one system. We believe that no new safety and effectiveness concerns are raised due to this technological difference. With respect to mode of operation, the EndoGI System utilizes a combination of rapid exchange and over-the-wire techniques to deliver the stents to the desired location, this is identical to the mode of operation of the identified predicates. The company has provided sufficient comparative testing between the EndoGI System and Advanix predicate as well as additional pre-clinical bench and animal data to demonstrate our substantial equivalency claim. Consequently, it is clear that the EndoGI System is as safe and effective as its primary predicate without raising any new safety and/or effectiveness concerns.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.