ADVANIX BILIARY STENT WITH NAVIFLEX RX DELIVERY SYSTEM
Applicant
Boston Scientific Corp
Product Code
FGE · Gastroenterology, Urology
Decision Date
Jul 23, 2010
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Device Story
Advanix Biliary Stent with NaviFlex RX Delivery System; plastic biliary stent and delivery catheter. Stent provides drainage of biliary duct/gallbladder; maintains patency; prevents closure. Stent shapes include center bend and duodenal bend; tapered leading end for papilla access; rounded trailing end. Delivery system includes guide catheter and outer push catheter with locking mechanism; radiopaque markers for visualization. Used in clinical setting by physicians; delivered to biliary tract. Stent constructed of polyethylene; guide catheter of polyvinylidene fluoride; push catheter of polyether block amide. Device facilitates bile flow; aids healing; manages strictures/stones.
Clinical Evidence
Bench testing only. Evaluated per EN ISO 10993-1:2009 for biocompatibility (cytotoxicity, sensitization, reactivity, systemic toxicity, genotoxicity, implantation). Performance testing included drainage lumen ID, stent dimensions, repositionability, deployment/trackability force, compatibility (guidewire, duodenoscope, caps), and mechanical integrity (tensile strength, buckling resistance, marker band strength). All tests passed.
Technological Characteristics
Stent: Polyethylene. Guide catheter: Polyvinylidene fluoride. Push catheter: Polyether block amide. Radiopaque markers included. Diameters: 7Fr, 8.5Fr, 10Fr. Lengths: 5cm-18cm. Delivery system working length: 202.5cm. Mechanical delivery via RX (rapid exchange) catheter system with locking mechanism. Biocompatibility per EN ISO 10993-1:2009.
Indications for Use
Indicated for patients requiring biliary drainage, bile duct splinting during healing, or maintenance of bile duct patency in the presence of strictures or stones.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
Microvasive® Biliary Stent and Delivery System (K965147)
Submission Summary (Full Text)
{0}------------------------------------------------
K 101314
Page 1 of 2
SECTION 5 510(k) SUMMARY
#### 510(k) SUMMARY
## 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752 Telephone: 508-683-4347 Fax: 508-683-5939
Contact: Elena Nieves Senior Regulatory Affairs Specialist Date Prepared: May 10, 2010
### 2. Proposed Device:
Trade Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Classification Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II
#### 3. Predicate Device:
Trade Name: Microvasive® Biliary Stent and Delivery System Manufacturer and Clearance Number: Boston Scientific Corporation, K965147 Classification Name: Biliary Catheters and Accessories Regulation Number: 876.5010 Product Code: FGE Classification: Class II
# 4. Proposed Device Description:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System consists of biliary plastic stents and delivery system catheters. They are sold separately or in a pre-loaded configuration, in which a stent comes attached to the catheter via a suture. The system is comprised of two (2) primary components: stent and delivery catheter with a locking mechanism. The Advanix Biliary Stent allows for drainage of the biliary duct and gall bladder by preventing closure and maintaining the patency of the biliary duct. These biliary stents are provided in center bend and duodenal bend shapes and have leading and trailing bars, a tapered leading end tip to facilitate access through the papilla, and a rounded tailing end to match the profile of the push catheter portion of the delivery system.
The Advanix Biliary Stent will be offered in 7Fr. 8.5Fr, and 10Fr diameters. The stent leneths for each diameter vary from 5cm-18cm stent lengths. The Advanix biliary stents are constructed of a Polyethylene material.
The NaviFlex RX Delivery System will be offered in 202.5cm working length. The delivery system consists of a guide catheter and outer push catheter. The guide catheter is constructed of Polyvinylidene fluoride material. The guide catheter also includes a radiopaque marker. The push catheter is composed of polyether block amide material and also has a Nylon repositioning suture.
Special 510(k) Premarket Notification, Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Proprietary and Confidential Information of Boston Scientific Corporation
JUL 2 3 2010
{1}------------------------------------------------
# 5. Intended Use:
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
# 6. Technological Characteristics:
The proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System has the same technological characteristics as the predicate Microvasive® Biliary Stent and Delivery System (K965147).
The proposed device has the same intended use and is placed using the same methodology as the predicate device. Both the proposed and predicate device function in the same manner allowing for biliary drainage through the lumen.
The shape of the proposed Advanix Biliary Stent is the same as the predicate Microvasive Biliary Stent and Delivery System and also includes an additional center bend stent shape.
The materials of the proposed Advanix Biliary Stent have been modified to the polymer materials mentioned above in part 4 of this 510(k) Summary.
## 7. Performance Data:
In-vitro testing has been performed and all components, subassemblies, and/or full devices met the required specifications for the completed tests.
The proposed Advanix™ Biliary Stent was evaluated in accordance with EN ISO 10993-1:2009. The following tests were performed on the stent: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Systemic Toxicity -Acute Systemic Toxicity, Subacute Toxicity - Intravenous and Intraperitoneal, Genotoxicity - Ames Assay and Mouse Lymphoma, Implantation, and USP Physicochemical.
The NaviFlex™ RX Delivery System was evaluated in accordance with EN ISO 10993-1:2009. The following tests were performed on the delivery system: Cytotoxicity, Sensitization, Intracutaneous Reactivity, and USP Physicochemical.
The following tests were conducted on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System: Drainage Lumen ID, Stent Length, Stent OD, Stent Shape, Repositionability, Deployment Force, Trackability Force, Guidewire Compatibility, Duodenoscope Compatibility, Barb Flap Cover Compatibility, Biopsy Cap Compatibility, Delivery System Working Length, Delivery System OD, Guide Catheter Pushability, RX Port Buckling Resistance, Locking Mechanism Retention Ability, Locking Mechanism to Dual Lumen Catheter Tensile Strength, Dual Lumen Catheter to Single Lumen Catheter Tensile Strength, Pullwire Assembly to Guide Catheter Tensile Strength, Pullwire Tensile Strength, Guide Catheter RO Marker Band Strength, and Push Catheter RO Marker Band Strength.
#### 8. Conclusion:
All biocompatibility tests conducted on the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System passed. Therefore, the Advanix Biliary Stent with NaviFlex RX Delivery System is considered biocompatible for its intended use.
All device bench test results were acceptable. The data demonstrate that the Advanix™ Biliary Stent with NaviFlex™ RX Delivery System sufficiently meets the design specifications and is suitable for the intended use.
Boston Scientific Corporation has demonstrated that the proposed Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is substantially equivalent to Boston Scientific Corporation's currently marketed Microvasive® Biliary Stent and Delivery System (K965147).
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUM. S." written around the edge. Inside the circle is a stylized image of an eagle with its wings spread. The image is black and white and appears to be a scan of a document.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
Ms. Elena Nieves Sr. Regulatory Affairs Specialist Boston Scientific Corporation 100 Boston Scientific Way MARLBOROUGH MA 01752
JUL 2 3 2010
Re: K101314
Trade/Device Name: Advanix™ Biliary Stent with NaviFlex™ RX Delivery System Regulation Number: 21 CFR& 876.5010 Regulation Name: Biliary catheter and accessories Regulatory Class: II Product Code: FGE Dated: May 10, 2010 Received: May 11, 2010
Dear Ms. Nieves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related
ः ।
{3}------------------------------------------------
Page 2 -
adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the aurerse overns (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please rryou desire specific da 100 too too too too too fices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please the Concer 101 Donece and the beginne by reference to premarket notification" (21CFR Part note the regalation regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Tou may other of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Janine M. Morris
Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known):
Device Name:
Indications for Use:
K101314
AdvanixTM Biliary Stent with NaviFlex™ RX Delivery System
Advanix™ Biliary Stent with NaviFlex™ RX Delivery System is intended for delivery of the stent to the biliary tract for drainage of the bile ducts, for splinting of a bile duct during healing, or for providing bile duct patency in a stricture or past a stone.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hebert Demir
(Division Sign-Off) (Division Sign-Ont)
Division of Reproductive, Abdominal, and Radiological Device 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.