The TrueTear Intranasal Tear Neurostimulator provides a temporary increase in tear production during neurostimulation to improve dry eye symptoms in adult patients with severe dry eye symptoms.
Device Story
TrueTear is a prescription, non-implantable intranasal neurostimulator for dry eye. Device consists of a base unit, disposable tip, and charging case. User inserts tip into nose; base unit delivers low-level electrical stimulation to nasal sensory neurons to acutely increase tear production. Base unit includes an RFID transceiver to authenticate tips and track usage; a BLE module transmits one-way summary data to an optional mobile app. Device is intended for home use by patients. Output is electrical stimulation; clinical benefit is improved dry eye symptoms via increased tear production.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence was supported by bench testing, including electrical output characterization, biocompatibility (ISO 10993), electrical safety/EMC (IEC 60601 series), and software validation (IEC 62304).
Technological Characteristics
Base unit with BLE/RFID; disposable tip with stainless steel 316 (ASTM MIM-316L) contact surface. Electrical stimulation via low-level waveform. Powered by sealed battery with charging case. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and IEC 60601-2-10. Software follows IEC 62304 (moderate level of concern).
Indications for Use
Indicated for adult patients with severe dry eye symptoms to provide a temporary increase in tear production during neurostimulation.
Regulatory Classification
Identification
An intranasal electrostimulation device for dry eye symptoms is a prescription non-implantable, electrostimulation device intended to increase tear production for improvement in dry eye symptoms.
Special Controls
In combination with the general controls of the FD&C Act, the intranasal electrostimulation device for dry eye symptoms is subject to the following special controls:
- (1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
- (2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
- (3) Patient-contacting components of the device must be demonstrated to be biocompatible.
- (4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
- (5) Software verification, validation and hazard analysis must be performed.
- (6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
- a. Summaries of electrical stimulation parameters.
- b. Instructions on how to correctly use and maintain the device.
- c. Instructions and explanations of all user-interface components.
- d. Information related to electromagnetic compatibility classification.
- Instructions on how to clean the device. e.
- Summaries of clinical performance testing demonstrating safety and effectiveness. f.
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
(2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(3) Patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
(i) Summaries of electrical stimulation parameters;
(ii) Instructions on how to correctly use and maintain the device;
(iii) Instructions and explanations of all user-interface components;
(iv) Information related to electromagnetic compatibility classification;
(v) Instructions on how to clean the device; and
(vi) Summaries of clinical performance testing demonstrating safety and effectiveness.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 31, 2020
Allergan Barbara Simon Director, Regulatory Affairs 2525 Dupont Drive Irvine, CA 92612
Re: K193589
Trade/Device Name: TrueTear Intranasal Tear Neurostimulator Regulation Number: 21 CFR 886.5310 Regulation Name: Intranasal Electrostimulation Device For Dry Eye Symptoms Regulatory Class: Class II Product Code: QBR Dated: July 29, 2020 Received: July 30, 2020
Dear Barbara Simon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
J. Angelo Green, Ph.D. Assistant Director DHT1A: Division of Ophthalmic Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193589
### Device Name
TrueTear Intranasal Tear Neurostimulator
Indications for Use (Describe)
The TrueTear Intranasal Tear Neurostimulator provides a tear production during neurostimulation to improve dry eye symptoms in adult patients with severe dry eye symptoms.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
| I. APPLICANT: | Allergan, Inc. |
|------------------------|--------------------------------------------------------------------------------|
| | 2525 Dupont Drive |
| | Irvine, CA 92612 |
| | Phone: (714) 246-4500 |
| | Fax: (714) 796-3015 |
| Contact: | Barbara Simon |
| | Director, Regulatory Affairs |
| | Allergan, Inc. |
| | (714) 246-4393 |
| Date Summary Prepared: | April 17, 2020 |
| II. DEVICE | |
| Trade Name: | TrueTear Intranasal Tear Neurostimulator |
| Classification Name: | Intranasal electrostimulation device for dry eye symptoms<br>(21 CFR 886.5310) |
| Device Class: | Class II |
| Device Product Code: | QBR |
#### III. PREDICATE DEVICE
Allergan TrueTear Intranasal Tear Neurostimulator (DEN170086)
#### IV. DEVICE DESCRIPTION
The TrueTear device is a prescription only, non-implantable, electrostimulation device intended to increase tear production for improvement in dry eye symptoms. The device's technological principle is the application of low-level electrical stimulation to sensory neurons located in the nose to acutely increase tear production and improve dry eye symptoms. The device consists of three distinct non-sterile subassemblies - a base unit which controls and produces electrical stimulation waveform and enables the patient to control the neurostimulation, a disposable tip that is inserted into the nose and provides the contact surface for the stimulation to the target tissue located in the nose, and a charger case that protects the device and replenishes the sealed battery inside the base unit between uses.
{4}------------------------------------------------
The disposable tip connects to the base unit and provides the contact for conducting lowlevel electrical stimulation current. which is produced by the base unit, to the target site on the inside of the nose.
The device includes a wireless communications module and is capable of providing basic communication over a Bluetooth Low Energy (BLE) channel and a radio frequency identification (RFID) transceiver. The wireless module does not allow control or modification of the base unit that produces the stimulation. The BLE channel is a one-way, outbound-only channel that allows patients to view summary device use data on an optional mobile app. The RFID channel is used to authenticate that the disposable tip is genuine, untampered and allows the base unit to indicate when the tip is due for disposal.
#### V. INDICATIONS FOR USE
The TrueTear Intranasal Tear Neurostimulator provides a temporary increase in tear production during neurostimulation to improve dry eve symptoms in adult patients with severe dry eye symptoms.
The Indications for Use statement is identical to the predicate device.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The predicate device is the previous version of the TrueTear Intranasal Tear Neurostimulator that was described in de novo DEN170086. The predicate device and the subject device have the same technological principle: the use of low-level electrical stimulation to sensory neurons located in the nose to acutely increase tear production and improve dry eye symptoms. At a high level, the subject and predicate devices are based on the following same technological elements:
- A base unit that controls and produces electrical stimulation waveform and enables . the patient to control the neurostimulation
- A disposable tip that is inserted into the nose to reach target tissue located inside the . nose and provides the contact surface for the stimulation
- A charger that replenishes the sealed battery inside the base unit •
- . A wireless BLE communication module that allows one-way summary device use data to be viewed by the patient on an optional mobile app
The following technological differences exist between the subject and predicate devices:
- Addition of RFID transceiver to the base unit .
- . Addition of RFID tag embedded in the tip body, a non-tissue contact portion of the tip and not visible to the patient
{5}------------------------------------------------
- . Use of stainless steel 316 instead of a proprietary hydrogel as the stimulation contact surface of the tip
- . Consolidation of the charger and cover into a charging case; and updated base unit and charger dimensions to provide a more modern look and feel
- . Labeling modifications to reflect updated technological changes and the material change for the surface contact in the disposable tip; updated instructions to dispose of used tips within 28 days of first use versus disposal of used tips daily
#### VII. PERFORMANCE DATA
The following performance data are provided in support of the substantial equivalence determination.
## Electrical Output Specifications
Within the demonstration of substantial equivalence, performance testing for assessing the electrical outputs was conducted. The testing evaluated the output waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density. The performance results met the intended electrical output design specifications, which remain unchanged from the predicate device.
## Biocompatibility
Biocompatibility evaluation was performed according to ISO 10993-1:2018, Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. The disposable tip is classified per ISO 10993-1 as a surface device with long-term mucosal membrane contact. The stainless steel 316 electrode used in the subject device conforms to ASTM MIM-316L. The subject device meets the biocompatibility requirements and complies with ISO 10993-1:2018, ISO 10993-3:2014, ISO 10993-5:2009, ISO 10993-10:2010, and ISO 10993-18:2005.
## Electrical Safety Electromagnetic Compatibility (EMC)
Performance testing for electrical, thermal, and mechanical safety and EMC for the subject device and intended use environment (home use) was performed. The subject device complies with IEC 60601-1-2 (Edition 4.0):2014, IEC 60601-1:2005, COR1:2006, COR2:2007, AMD1:2012, IEC 60601-1-11:2015 (Second Edition) for use in conjunction with IEC 60601-1:2012 (Third Edition) + A1:2012, and IEC 60601-2-10: 2012 (Second Edition) for use in conjunction with IEC 60601-1:2005 (Third Edition)+A1:2012.
{6}------------------------------------------------
# Software
Software verification and validation testing was performed and complies with IEC 62304:2006 + A1:2015 (2015-06). The subject device software has been assigned a "moderate" level of concern, which is unchanged from the predicate device.
# Wireless Communications
Cybersecurity and coexistence evaluations were performed for the wireless communication module modifications. The subject device has appropriate mitigations for cybersecurity and coexistence testing confirmed that the device's essential performance is not affected by near range devices and data can be transmitted in the presence of active RF energy sources.
# Other Performance Data
Electrical and mechanical performance bench testing was evaluated for the disposable tip for the intended use duration of 28 days to confirm the subject device disposable tip met specified design criteria. The performance test results met the design specifications. Additionally, performance testing was conducted for the disposable tips to evaluate worst-case conditions for electrical stimulation durability and for durability during insertion and cleaning to demonstrate that the tips maintained electrical and mechanical integrity throughout their maximum usage period. The worst-case performance test results met the design specifications.
## Animal Studies
The design modifications did not require animal studies to demonstrate substantial equivalence.
# Clinical Studies
The design modifications did not require clinical testing to demonstrate substantial equivalence.
# VII. CONCLUSION
The objective evidence presented in the de novo submission for the predicate device demonstrated the clinical safety and effectiveness of the TrueTear Intranasal Tear Neurostimulator. The modified device relies on the same technological principle as the predicate device, and the modifications do not raise different questions of safety and effectiveness. The predicate and subject devices have the same intended use. The non-clinical performance evaluations demonstrate that the subject device performs comparably to the predicate device in the same use conditions. The changes that have been made to the device's overall design, hardware and software do not affect the intended use or risk profile of the device. Therefore the subject device described in this substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.