The TrueTear Intranasal Tear Neurostimulator provides a temporary increase in tear production during neurostimulation to improve dry eye symptoms in adult patients with severe dry eye symptoms.
Device Story
TrueTear Intranasal Tear Neurostimulator; non-implantable electrostimulation device. Device inserts into nasal cavity; delivers electrical stimulation to nasal mucosa to trigger lacrimal functional unit; increases tear production. Used by adult patients with severe dry eye symptoms. Device includes patient-contacting tips; requires training for proper use. Output parameters (waveforms, frequency, pulse duration) controlled to ensure safety. Clinical benefit: temporary improvement in dry eye symptoms. Physician/patient labeling provides instructions for use, maintenance, and cleaning.
Clinical Evidence
Two pivotal trials (OCUN-009, OCUN-010) and usability study (OCUN-012). OCUN-010 enrolled 97 participants; 77 with severe dry eye symptoms. Primary endpoint: change in Ocular Surface Disease Index (OSDI) from baseline. At day 30, 25-39% of severe dry eye patients showed clinically meaningful improvement (based on MCID thresholds), while 7-12% showed worsening. Safety profile acceptable; no serious device-related adverse events reported.
Technological Characteristics
Intranasal electrostimulation device. Patient-contacting components must be biocompatible. Electrical output specifications include waveforms, output modes, voltage, current, pulse duration, frequency, net charge per pulse, phase charge, current density, and power density. Requires EMC testing, electrical/thermal/mechanical safety testing, and software verification/validation.
Indications for Use
Indicated for adult patients with severe dry eye symptoms to provide temporary increase in tear production via neurostimulation.
Regulatory Classification
Identification
An intranasal electrostimulation device for dry eye symptoms is a prescription non-implantable, electrostimulation device intended to increase tear production for improvement in dry eye symptoms.
Special Controls
In combination with the general controls of the FD&C Act, the intranasal electrostimulation device for dry eye symptoms is subject to the following special controls:
- (1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
- (2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
- (3) Patient-contacting components of the device must be demonstrated to be biocompatible.
- (4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
- (5) Software verification, validation and hazard analysis must be performed.
- (6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
- a. Summaries of electrical stimulation parameters.
- b. Instructions on how to correctly use and maintain the device.
- c. Instructions and explanations of all user-interface components.
- d. Information related to electromagnetic compatibility classification.
- Instructions on how to clean the device. e.
- Summaries of clinical performance testing demonstrating safety and effectiveness. f.
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
(2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
(3) Patient-contacting components of the device must be demonstrated to be biocompatible.
(4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
(5) Software verification, validation, and hazard analysis must be performed.
(6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
(i) Summaries of electrical stimulation parameters;
(ii) Instructions on how to correctly use and maintain the device;
(iii) Instructions and explanations of all user-interface components;
(iv) Information related to electromagnetic compatibility classification;
(v) Instructions on how to clean the device; and
(vi) Summaries of clinical performance testing demonstrating safety and effectiveness.
Submission Summary (Full Text)
{0}------------------------------------------------
#### DE NOVO CLASSIFICATION REQUEST FOR TRUETEAR INTRANASAL TEAR NEUROSTIMULATOR
#### REGULATORY INFORMATION
FDA identifies this generic type of device as:
Intranasal electrostimulation device for dry eye symptoms. An intranasal electrostimulation device for dry eye symptoms is a prescription non-implantable, electrostimulation device intended to increase tear production for improvement in dry eye symptoms
NEW REGULATION NUMBER: 21 CFR 886.5310
CLASSIFICATION: Class II
PRODUCT CODE: QBR
#### BACKGROUND
DEVICE NAME: TrueTear Intranasal Tear Neurostimulator
SUBMISSION NUMBER: DEN170086
DATE DE NOVO RECEIVED: October 23, 2017
CONTACT: Allergan 2525 Dupont Drive Irvine, California 92612
#### INDICATIONS FOR USE
The TrueTear Intranasal Tear Neurostimulator provides a temporary increase in tear production during neurostimulation to improve dry eye symptoms in adult patients with severe dry eye symptoms.
#### LIMITATIONS
The sale, distribution, and use of the TrueTear Intranasal Tear Neurostimulator are restricted to prescription use in accordance with 21 CFR 801.109.
Patient training is required on the proper use of the device before home use.
The TrueTear device is limited only to the improvement in dye eye symptoms as the safety and effectiveness in the treatment of dry eye disease has not been established.
{1}------------------------------------------------
The device increases tear production during neurostimulation, i.e., tearing was assessed only during stimulation.
The clinical study was not designed to evaluate any changes in nerve sensitivity.
Clinical study results demonstrate a trend of decreased effectiveness (tear production) over time. The mechanism for this decrease has not been identified and was not analyzed as part of the study.
# PLEASE REFER TO THE LABELING FOR A COMPLETE LIST OF WARNINGS. PRECAUTIONS AND CONTRAINDICATIONS.
### DEVICE DESCRIPTION
The TrueTear Intranasal Tear Neurostimulator is a non-surgical, non-implantable device intended for the application of low level electrical stimulation to sensory neurons of the nasal cavities to increase tear production to improve dry eye symptoms in adult patients with severe dry eye symptoms.
The design has not been significantly modified from the device granted under DEN160030. The following are the technical changes from DEN160030:
- 1) Enabling Bluetooth functionality within the device, to enable one-way (outbound only) wireless communication
- from the (b) (4) 2) Removal of (b) (4) material used in the disposable tip
- 3) Disabling of the tip detection feature in the Base Unit
- 4) Addition of an optional mobile application (TrueTear App) for reading and summarizing device usage data.
The device consists of four distinct non-sterile subassemblies, as listed below:
- 1) Disposable Tips that insert up to 28 mm into the nasal cavity and stimulate the target tissue of the intranasal skin and mucosa.
- 2) A reusable Base Unit which produces the electrical stimulation waveform.
- 3) A reusable Charging Station which recharges the sealed battery inside the Base.
- 4) A reusable Cover to protect the Disposable Tips.
Figure 1 provides figures of the device components; additional details for each component are discussed below.
{2}------------------------------------------------
Image /page/2/Figure/0 description: The image shows a base unit with a disposable tip, a cover, and a charger. The base unit has a control panel with buttons and indicators. The disposable tip is attached to the base unit and has a hydrogel pad on top. The cover is a protective sleeve that fits over the base unit and disposable tip. The charger is used to recharge the base unit.
# FIGURE 1. THE TRUETEAR INTRANASAL TEAR NEUROSTIMULATOR COMPONENTS
The device (Base) automatically turns off after one minute of stimulation. Alternatively, the device may also be turned off by holding down the "-" button for 2 seconds. The device will vibrate and the Light-Emitting Diodes (LED) will turn off to indicate that the power has been switched off.
Additionally, the device (Base) has a "Daily device usage limit" of thirty (30) minutes. If an orange LED flashes 2 times, the daily use limit (30 minutes) has been reached, and the device will no longer deliver stimulation.
Figure 2 is a schematic showing correct use of the device.
Image /page/2/Picture/5 description: In the image, there are two side-by-side images of a woman using a nasal rinse. The woman is fair-skinned with gray hair and is wearing a blue shirt. She is holding a blue and white nasal rinse bottle to her nose. The background is a solid blue color.
# FIGURE 2. USE OF THE TRUETEAR INTRANASAL TEAR NEUROSTIMULATOR. (L) STARTING POSITION AND (R) CORRECT TREATMENT POSITION
# DISPOSABLE TIPS
The Disposable Tips connect to the Base and incorporates a silicone hydrogel that touches the inside of the nose, at a depth of up to 28 mm from the nasal columella (the skin separating the two nostrils), to provide stimulation. The Disposable Tips are removed and replaced daily; a separate Cover can be used to protect the Disposable Tips between uses.
{3}------------------------------------------------
# BASE
The hand-held, battery-powered Base is the portion of the device which produces the required electrical output for stimulation. The Base produces a train of charge-balanced pulses that is patterned by modulating pulse amplitude, pulse width and pulse shape. The intensity of the stimulation is adjustable by using the (+) or (-) control buttons, which cycle the Base through six different levels (including zero, a non-stimulation level). The Base is held in the palm of the hand, allowing the patient to press one of two buttons to increase or decrease the intensity of stimulation. The base includes the wireless Bluetooth Low Energy communication channel for one-way (outbound) communication with the TrueTear mobile application. A series of LEDs on the Base illuminate to indicate powering the device on/off, the intensity level during use, and the charging of the Base's battery when connected to the Charging Station.
# CHARGING STATION
The Charging Station provides a dock for recharging the battery located inside the Base. With the Disposable Tips removed, the Base can be inverted and placed onto the Charging Station.
# COVER
The Cover is made of polycarbonate material and fits over the tips. The Cover can be used to protect the Disposable Tips between uses.
# TrueTear App
The TrueTear application is intended to be used on a mobile device (e.g., a smart phone) and uses the phone's Bluetooth functionality to connect to the TrueTear device, downloading device usage data. The TrueTear application, when connected to the TrueTear device, displays usage statistics for the TrueTear device and a summary of the TrueTear device's Instructions for Use. The usage statistics consist of summaries of the user's stimulation duration and frequency tabulated and/or displayed on a periodic (e.g., daily) basis. The mobile application does not provide any diagnostic and treatment recommendations and does not control or modify the TrueTear device.
# SUMMARY OF NONCLINICAL/BENCH STUDIES
# BIOCOMPATIBILITY/MATERIALS
The Disposable Tips of the TrueTear Intranasal Tear Neurostimulator were tested for cytotoxicity, sensitization, irritation, acute systemic toxicity per ISO 10993-1 "Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process." All results demonstrated acceptable performance.
#### SHELF LIFE/STERILITY
The TrueTear Intranasal Tear Neurostimulator is non-sterile. The Disposable Tips are disposable and are meant to be replaced daily. The Cover is reusable. Cleaning and maintenance instructions of the stimulator components of the device are included in the labeling.
{4}------------------------------------------------
# ELECTROMAGNETIC CAPABILITY (EMC) AND ELECTRICAL SAFETY
The device conformed to the following electromagnetic compatibility, electrical, mechanical, and thermal safety standards:
- . The TrueTear Intranasal Tear Neurostimulator was tested per the requirements of IEC 60601-1: 2005 + CORR. 1:2006 + CORR. 2:2007 + AM1:2012. "Medical Electrical Equipment Part 1: General Requirements for Basic Safety and Essential Performance." The Base and Charging Station are compliant to these standards.
- . The device was also tested per the requirements of IEC 60601-1-11:2015 "Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment." The device is in conformance with this standard.
- . The TrueTear Intranasal Tear Neurostimulator was tested specifically per the requirements of IEC 60601-2-10, Edition 2.0 Issue: 2012/06 "Medical Electrical Equipment Part 2-10: Particular Requirements for the Safety of Nerve and Muscle Stimulators - Includes Amendment A1: 2001." The device is in conformance with this standard.
- The TrueTear Intranasal Tear Neurostimulator was tested according to the IEC ● 60601-1-2. Issue: 2007/03/01 Ed:3.0 (Equivalent to AAMI/ANSI/IEC 60601-1-2:2007/(R)2012) "Medical Electrical Equipment - Part 1-2:General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests." Results demonstrated that the Base and Charging Station are compliant to this standard.
#### SOFTWARE
The following device functions of the TrueTear Intranasal Tear Neurostimulator are controlled by the software:
- · The allowance of multiple stimulation sessions of various intensity levels.
- · The control of stimulation intensity levels.
- · The notification to the user when the charging is complete.
- · The connection to mobile applications to download or clear historical stimulation data.
The software will mitigate safety risks as follows:
- The software limits the output current to a maximum of 5 mA. .
{5}------------------------------------------------
- . The software disables stimulation if the tip expiration time is exceeded, if the stimulation mode exceeds the maximum usage time, or exceeds the maximum usage time within the rolling usage time window.
A failure or latent flaw in the software for the TrueTear Intranasal Tear Neurostimulator could indirectly result in patient injury; therefore, the software of this device is considered to have a "Moderate" level of concern. The submission contained all the elements of software documentation corresponding to the "Moderate" level of concern, as outlined in the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." Adequate documentation describing the software/firmware, software specifications, architecture design, software development environment, traceability, revision level history, unresolved anomalies provide the foundation that the software will operate in a manner as described in the specifications. Hazard analysis was performed to characterize software risks including device malfunction, measurement related errors and cybersecurity. The submission included verification and validation (V&V) testing to address the potential hazards with satisfactory results.
# SUMMARY OF CLINICAL INFORMATION
The applicant conducted two pivotal clinical trials, OCUN-009 and OCUN-010, and a human factors usability study, OCUN-012, which demonstrated that patients can use the TrueTear Intranasal Tear Neurostimulator properly. The safety and effectiveness of the device was evaluated based upon the outcomes of trials OCUN-009 and OCUN-010. The two pivotal trials were found to be appropriate to support the action of the device effect of increased tear production during neurostimulation in DEN160030. Please see FDA's summary of the De Novo Classification Request for DEN160030 for summaries of OCUN-009, OCUN-010, and OCUN-012. The current De Novo request was submitted for the new indication of improving dry eye symptoms. New effectiveness information was submitted from OCUN-10 to support the new indication. A brief summary of the symptoms information is provided below.
Participants meeting the enrollment criteria at a Screening Visit were enrolled at day 0 (baseline), provided patient training on the correct use of the device, and provided with a TrueTear device for home use.
Patient training for each subiect in OCUN-010 covered all content found in the proposed Instructions For Use (IFU) and Patient Quick Start Guide. A demonstration device was used with each subject as a visual aid. All training was provided by site personnel. This training took up to 30 minutes.
Participants were instructed to perform intranasal neurostimulation at least 2 times a day and as often as 10 times per day, as needed, and no more than 3 minutes per use. Participants were followed for 180 days and were seen for follow-up exams at days 7, 30, 90, and 180. Symptoms were assessed with the Ocular Surface Disease Index (OSDI) at the Screening Visit and days 0, 7 and 30. Dry eye symptom severity subgroups were determined by the Screening Visit OSDI total score where moderate dry eye was defined as an OSDI total score of 13 to 22 and a severe
{6}------------------------------------------------
dry eye was defined as an OSDI total score of 33 or more. Within participant change in OSDI from baseline was analyzed. The minimal clinically important difference (MCID) thresholds for the analyses were based on those characterized by Miller et al (Arch Ophthalmol. 2010;128(1):94-101).
The proportion of participants with a clinically important change in OSDI at follow-up days 7 and 30 for all available participants stratified by dry eye symptom severity subgroup and by the upper and lower limit of the MCID was analyzed. Of the 97 participants that were enrolled, 77 had severe dry eye symptoms at the Screening Visit and were seen following treatment. As shown in Table 4 below, of these participants seen at day 30, approximately 25-39% had a clinically meaningful improvement in their symptoms and approximately 7-12% had clinically meaningful worsening of their symptoms from baseline. The sample size of the moderate dry eye group was too small to make meaningful inferences regarding their results.
# TABLE 1: PROPORTION OF SUBJECTS WITH SEVERE DRY EYE SYMPTOMS AT SCREENING WITH A CLINICALLY IMPORTANT CHANGE IN THE OSDI* FROM BASELINE AT DAYS 7 AND 30
| Cohort | MCID** | Change | Day 7 | Day 30 |
|----------------------------|--------|----------|----------------|----------------|
| Severe Dry Eye<br>Subgroup | 7.3 | Improved | 42.86% (33/77) | 38.67% (29/75) |
| | 7.3 | Worsened | 12.99% (10/77) | 12.00% (9/75) |
| Severe Dry Eye<br>Subgroup | 13.4 | Improved | 23.38% (18/77) | 25.33% (19/75) |
| | 13.4 | Worsened | 5.19% (4/77) | 6.67% (5/75) |
*OSDI – Ocular Surface Disease Index
**MCID - Minimal Clinically Important Difference
#### Pediatric Extrapolation
In this De Novo request, existing clinical data were not leveraged to support the use of the device in a pediatric patient population.
# LABELING
The professional and patient labeling are adequate and meet the requirements of 21 CFR 801.109. Both guides contain requirements for use by prescription only and proper patient training. Indications for Use, contraindications, device description, warnings, precautions, potential complications, instructions for use, recommended stimulation schedule, instructions for device maintenance/cleaning, summary of clinical trials, information related to electromagnetic compatibility, expected service life, disposal & replacement, environmental operating conditions, electrical specifications, and symbols & markings. The patient labeling also includes additional information such as a glossary, "facts about dry eye," and potential benefits.
{7}------------------------------------------------
### RISKS TO HEALTH
The table below identifies the risks to health that may be associated with use of the intranasal electrostimulation device for dry eye symptoms and the measures necessary to mitigate these risks.
| Identified Risks | Mitigation Measures |
|--------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Tissue damage due to<br>overstimulation/understimulation<br>or mechanical injury (ex: tips too<br>long), device breakage | Non-clinical performance testing<br>Software verification, validation, and hazard analysis<br>Electrical, thermal, and mechanical safety testing<br>Labeling |
| Adverse tissue reaction | Biocompatibility evaluation<br>Labeling |
| Infection | Labeling |
| Electrical shock or burn | Electrical, thermal, and mechanical safety testing<br>Software verification, validation, and hazard analysis<br>Labeling |
| Interference with other devices | Electromagnetic compatibility (EMC) testing<br>Software verification, validation, and hazard analysis<br>Labeling |
| Pain, headache, or discomfort | Clinical performance testing<br>Non-clinical performance testing<br>Electrical, thermal, and mechanical safety testing<br>Labeling |
| Failure to mitigate dry eye<br>symptoms | Clinical performance testing<br>Training<br>Labeling |
Table 2 - Identified Risks to Health and Mitigation Measures
#### SPECIAL CONTROLS
In combination with the general controls of the FD&C Act, the intranasal electrostimulation device for dry eye symptoms is subject to the following special controls:
- (1) Clinical performance testing must evaluate improvement of dry eye symptoms under anticipated conditions of use.
- (2) Non-clinical performance testing must assess the following electrical output specifications: waveforms, output modes, maximum output voltage, maximum output current, pulse duration, frequency, net charge per pulse, maximum phase charge at 500 ohms, maximum current density, maximum average current, and maximum average power density.
- (3) Patient-contacting components of the device must be demonstrated to be biocompatible.
- (4) Performance testing must demonstrate the electrical, thermal, and mechanical safety along with electromagnetic compatibility (EMC) of the device in the intended use environment.
{8}------------------------------------------------
- (5) Software verification, validation and hazard analysis must be performed.
- (6) Training for the proper use of the device must be provided.
(7) Physician and patient labeling must include:
- a. Summaries of electrical stimulation parameters.
- b. Instructions on how to correctly use and maintain the device.
- c. Instructions and explanations of all user-interface components.
- d. Information related to electromagnetic compatibility classification.
- Instructions on how to clean the device. e.
- Summaries of clinical performance testing demonstrating safety and effectiveness. f.
#### BENEFIT-RISK DETERMINATION
The risks of the device are based on non-clinical laboratory data, as well as data collected in clinical trials described above and in FDA's summary of the De Novo Classification Request for DEN160030. The device exhibited an acceptable safety profile in the clinical studies which were conducted. No device-related serious adverse events were observed. Non-serious device-related adverse events were few in number, all self-limited, predominantly nasal in nature and decreased in incidence over the course of the study.
The probable benefits of the device are also based on nonclinical laboratory data, as well as data collected in clinical trials. There were more participants with severe dry eve symptoms that had a meaningful improvement in symptoms as measured with the Ocular Surface Disease Index (OSDI) than the number with clinically significant worsening of symptoms at day 7 and at day 30.
#### Patient Perspectives
Patient perspectives were considered for the TrueTear Intranasal Tear Neurostimulator during the review. Dry eye symptoms were assessed using the OSDI patient-reported outcome measure.
#### Benefit/Risk Conclusion
In conclusion, given the available information above, for the following indication statement:
The TrueTear Intranasal Tear Neurostimulator provides a temporary increase in tear production during neurostimulation to improve dry eye symptoms in adult patients with severe dry eve symptoms.
The probable benefits outweigh the probable risks for the TrueTear Intranasal Tear Neurostimulator. The device provides benefits and the risks can be mitigated by the use of general controls and the identified special controls.
{9}------------------------------------------------
### CONCLUSION
The De Novo request for the TrueTear Intranasal Tear Neurostimulator is granted and the device is classified as follows:
Product Code: QBR Device Type: Intranasal electrostimulation device for dry eye symptoms Class: II Regulation Number: 21 CFR 886.5310
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.